iHealth™ Wireless Blood Pressure Monitor (BP5) OWNER`S MANUAL
Contents
1. ANSI AAMI SP 10 2002 A1 2003 A2 2006 AAMI ANSI 80601 2 30 2009 IEC 80601 2 30 2009 Cor 2010 EN 80601 2 30 2010 Medical electrical equipment Part 2 30 Particular requirements for the basic safety and essential performance of automated non invasive sphygmomanometers ELECTROMAGNETIC COMPATIBILITY INFORMATION Table 1 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacture s declaration electromagnetic emissions The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the user of the MONITOR should assure that it is used in such an environment The Wireless Blood Pressure Monitor uses RF energy only for its internal function RE emissions Therefore its RF emissions are very low and are CSP re LI not likely to cause any interference in nearby electronic equipment RF emissions The MONITOR is suitable for use in all establishments including domestic Harmonic establishments and those directly connected to emissions Class A the public low voltage power supply network that TEC O 1L000 3 72 supplies buildings used for domestic purposes Voltage Fluctuations flicker Complies emissions IEC 61000 3 3 Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic immunity The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the us
2. Now Drcoverabha Now Diecoverable Monitor Status Bluetooth Indicator Waiting to connect Flashing blue light Connected and measuring Steady blue light Measurement completed and ready to Gradually extinguishing light disconnect MEASUREMENT PROCEDURES Blood pressure can be affected by the position of the cuff and your physiologic condition It 1S very important that the cuff should be placed at the same level as your heart Body Posture Sitting Comfortably During Measurement a Be seated with your feet flat on the floor without crossing your legs b Place your hand palm side up in front of you on a flat surface such as a desk or a table c The middle of the cuff should be at the level of the right atrium of your heart Lying Down During Measurement a G Lie on your back Place your arm straight along your side with your hand palm side up Cu The cuff should be placed at the same level as your heart Note Blood pressure can be affected by the position of the cuff and your PRVSTOLOGIC condition Apply the Cuff a Pull the cuff end through the metal loop positioning it outward away from your body b Place a bare arm through the cuff and position the cuff 1 2 1 2cm above the elbow joint C Tighten the cuff by pulling it towards your body securing it closed with the Velcro fastener d While seated place your hand palm side up in front of you on a flat surface such as a desk or table
3. 1Pad 3rd generation iPad 2 iPad 1Pad mini The 10S version of these devices should be V4 3 or higher SET UP PROCEDURES Download the Free iHealth MyVitals App Prior to first use download and install iHealth Myvitals from the APD OGCOrG Account Set Up and Registration Follow the on screen instructions to register and set up your personal account Charge Battery Before First Use Connect the monitor to a USB port using the charging cable provided until the green indicator light stabilizes Connect to iOS Device via Bluetooth e Apply the cuff or press the START STOP button the Bluetooth indicator will begin flashing e Turn Bluetooth On under the Settings menu on the iOS device e Wait until the model name printed on the monitor i e BP5 xxxxxx and Not Paired appear in the Bluetooth menu and select the model name BP5 XXXXXX to pair and connect The Bluetooth indicator will remain steady upon successful connection When using the monitor for the first time it may take up to 30 seconds for your 10S device to detect the Bluetooth Signal e Each subsequent time you use the monitor Not Connected will be displayed next to BP5 xxxxxx in the Bluetooth Menu e Please repeat these steps when you switch to another 10S device with the monitor No Service gt 16 49 100 fm Ho Service 16 49 100 Bluetooth on Bluetooth ON Devices Devices BP5 000041 Not Paired BPS 000041 Connected
4. 719 N Shoreline Blvd Mountain View CA 94043 USA tL 300 3 56 gt 705 www ihealthlabs com Lotus Global Co Ltd 15 Alexandra Road London UK NW8 ODP rel PO0044 20 75066010 Fax 0044 20 79006157 ANDON HEALTH CO LTD A No 3 Jinping Street Ya An Road Nankai District Tianjin 300190 China Tel 6 22 6052 6161 IMPORTANT INFORMATION REQUIRED BY THE FCC This device complies with Part 15 of the FCC Rules Its operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation Changes or modifications not expressly approved by iHealth Lab Inc would void the user s authority to operate the product Note This product has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference ina residential installation This product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this product does cause harmful interference to radio or television reception which can be determined by turning the equipment off and
5. Guide Measuring without an iOS Device Enable the Offline Measurement function on the App Device Setting gt Offline Measurement gt On Apply the cuff follow the Measurement Procedures and then press the START STOP button to begin measurement All offline measurements will be uploaded to the App automatically upon the next successful Bluetooth connection For detailed operating instructions please visit www ihealthlabs com Note Physical activity including eating drinking and smoking as well as excitement stress and many other factors influence blood pressure results ATTENTION You can stop the measurement process at any time by pressing and holding the START STOP button for 2 seconds SPECIFICATIONS 1 Product name Wireless Blood Pressure Monitor 2 Model BP5 3 Classification Internally powered Type BF applied part IPXO No AP or APG Continuous operation 4 Machine size approx 5 7 x 2 3 x1 2 145mm x 58mm x30mm Des CUTE cireimierence 62 6 L6 5 22Cm 42 0m 16 5 18 9 42cm 48cm XL size sold separately 6 Weight approx 5 167 145g including cuff approx 9 602 1609 ancluding XL curt 7 Measuring method Oscillometric method automatic inflation and 6 cs LA e 14 ike 16 ear LG LI measurement Memory volume 120 times with time and date stamp off line measurement only Power DC 5 0V 1 0A Battery 1 3 V L1 10n 400mAh Measurement r
6. blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit microwave oven etc 9 This product should not be used as a USB device 10 This product is verified by auscultatory method It 1s recommended that you check Annex B of ANSI AAMI SP 10 2002 A1 2003 A2 2006 for verification method details if needed 11 If the determined blood pressure systolic or diastolic is outside the rated range specified in part SPECIFICATIONS the app will immediately display a technical alarm on screen In this case consult a physician or ensure that proper measurement procedures are followed The technical alarm is preset in the factory and cannot be adjusted Or inactivated This technical alarm is assigned as low priority according to IEC 60601 1 8 The technical alarm is non latching and does not need to be reset 12 A medical AC adapter with an output of DC 5 0V and complies with IEC 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 is suitable for thismonitor such as ASP5 05010002J9U input 100 240V 50 60H2 Z00mMA output DC 5 0V 1 0A Please note that the monitor jack size is USB mini B This Wireless Blood Pressure Monitor is designed for adults and should never be used on infants young children pregnant or pre eclamptic patients Consult your physician before use on children This product might not meet its performance specifications if stored or used outside the specified temperature and hum
7. Position the monitor in the middle of your arm so that it is aligned with your middle finger e The cuff should fit comfortably yet snugly around your arm You should be able to insert one finger between your arm and the Curr Remember to i Make sure that the appropriate cuff size is used refer to the cuff Circumference range in the Specifications section of this manual Measure on the same arm each time Stay still during measurement Do not move your arm body or the monitor Stay still and calm for one to one and half minutes before taking a blood pressure measurement Prolonged over inflation of the bladder may cause ecchymoma of your arm Keep the cuff clean Cleaning the cuff after every 200 times of usage 1s recommended If the cuff becomes dirty clean it with a moistened cloth Do not rinse the monitor or cuff with running water During measurement press the START STOP button to stop measurement Press the START STOP button for 2 seconds to turn off the monitor manually Auto Connect Option Auto connect option allows the monitor to find the last used 10S device and re establish the connection automatically Auto connect option can be enabled in the App Device Setting gt Auto Connect gt On Taking Measurements with Multiple 10S Devices Turn off the Bluetooth of the last used 10S device if the Auto Connect option is enabled in your App then follow the Set Up Instructions in the Quick Start
8. YMBOLS cerir reeeo rere Eneee cao oi q ADS ata DARE Cena SA RA 12 IMPORTANT INFORMATION REQUIRED BY THE FCC ccccsssssssssssssssssssssssssscssssccccssosssooees 13 OTHER STANDARDS AND COMPLIANCEG cccccssssssssscssccccccccsssscccsscccsccccsssccosscccsscccssscccsscoossees 13 ELECTROMAGNETIC COMPATIBILITY INFORMAT TON ccccccccccccccccccccccccccccccccccccccccecs 14 INTRODUCTION Thank you for selecting the iHealth Wireless Blood Pressure Monitor The iHealth Wireless Blood Pressure Monitor is a fully automatic arm cuff blood pressure monitor that uses the oscillometric principle to measure your blood pressure and pulse rate The monitor works with your mobile device to test track and share vital blood pressure data PACKAGE CONTENTS o iHealth Wireless Blood Pressure Monitor Owner s Manual Quick Start Guide Charging Cable PrP RP PPB Travel Bag INTENDED USE The iHealth Wireless Blood Pressure Monitor Electronic Sphygmomanometer is intended for use ina professional setting or at home and 1S a non invasive blood pressure measurement system It is designed to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the upper arm The measurement range of the cuff circumference is 8 6 to 168 9 22cm 48cm BLOOD PRESSURE CLASSIFICATION FOR ADULTS The World Health Organization WHO has created the followi
9. ange Cuff pressure 0 300 mmHg Systolic 60 260 mmHg Diastolic 40 199 mmHg Pulse rate 40 180 beats minute Accuracy Pressure 3 mmHg Pulse rate t5 Wireless communication Bluetooth V3 0 EDR Class 2 SPP Frequency Band 2 402 2 480 GHz Environmental temperature for operation 5C 40C 41 F 104F Environmental humidity for operation S90 RH Environmental temperature for storage and transport 20C 55 C 4F 131F Environmental humidity for storage and transport S90 RH Environmental pressure 80kPa l05kPa Battery life more than 80 measurements on a full charge The blood pressure measurement system includes accessories pump valve cuff and sensor Note These specifications are subject to change without notice GENERAL SAFETY AND PRECAUTIONS Read all of the information in the Owner s Manual and other provided instructions before operating the unit Consult your physician for any of the following situations a The application of the cuff over a wound or inflamed area b The application of the cuff on any limb with intravascular access or therapy or an arterio venous A V shunt c The application of the cuff on the arm on the side of a mastectomy d Simultaneous use with other medical monitoring equipment on the same limb e The blood circulation of the user needs to be checked Do not use this product in a moving vehicle as this may result in inaccurate measurements Blood pressure me
10. asurements determined by this product are equivalent to those obtained by professional healthcare practitioners using the cuff stethoscope auscultation method within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometer When a call comes in during the measurement the measurement process will be terminated automatically It is recommended that the 10S device be set in Airplane mode during measurement to avoid strong magnetism interference 6 If Irregular Heartbeat IHB is detected during the measurement procedure the IHB symbol will be displayed Under this condition the Wireless Blood Pressure Monitor can keep functioning but the results may be inaccurate Please consult your physician for accurate assessment There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pulse period gt 25 2 The difference of adjacent pulse period 20 14s and the number of such pulse takes more than 53 percent of the total number of pulses 7 Please do not use any other cuff other than that supplied by the manufacturer as this may result in measurement errors and a biocompatible hazard 8 Information regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see ELECTROMAGNETIC COMPATIBILITY INFORMATION It is suggested that the
11. ccording to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol o NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the MONITOR is used exceeds the applicable RF compliance level above the MONITOR should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the MONITOR b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Recommended separation distances betwee
12. er of the Wireless Blood Pressure Monitor should assure that it 1S used in such an environment IMMUNITY IEC 6060ltest Electromagnetic environment Compliance level test level guidance Electrostat 6 kV contact tO KV CONTACL Floors should be wood ic discharge ESD LEC 61L000 4 2 Electrical fast transient b Urst TEC 61000 4 4 urge TEC 61000 4 5 Voltage dips short interruptio ns and voltage variations on power supply input lines LEG 61000 4 11 Power frequency 50760 Hz magnetic field LEC 6LO00 4 6 NOTE Ur is the a c 2 kV for power supply lines 1 kV line s to line s 2 kV line s to earth 299 O for 0 5 cycle 40 Ur 60 for 5 cycles TO 30 for 25 cycles dip in Ur dip in Ur dip in Ur S95 5 for 5 B dip in Ur 2 kV for power supply lines 1 kV line s to line s 2 kV line s to earth gt 90 3 for Uso Cycle 40 Ur 60 3 for 5 cycles TO 30 for 25 cycles dip in Ur dip in Ur dip an Ur gt 95 TORO iS dip in Ur Table 3 concrete or ceramic tile If floors are covered with the relative humidity should be at least 30 synthetic material Mains power quality should be that of atypical commercial or hospital environment Mains power quality should be that of atypical commercial or hospital environment Mains power quality should be that of atypical commercial o
13. iHealth Wireless Blood Pressure Monitor BP5 OWNER S MANUAL Table of Contents INTRODUCT LON cenione aa e ai a a a a a e 2 PACKAGE CONTENTS ooreis a a Eai 2 INTENDED USE au das e E E EN e A E A Aa DO RR a 2 BLOOD PRESSURE CLASSIFICATION FOR ADULTG ccccccccsssssssssssssssssssssssssssssssssssssscsscees 2 CONTRAINDICATION wiicsssssitictasstaacessscacusssecatdassnidectasacassaassexecunsacassusesidecdasseassesenedecunsasasssassedesdasteassens 3 PARTS AND DISPLAY INDICATORS ssssicsesssccebtvacedealsssdedeaiassbebewsdacebounsedealuessedeniassbebevadsesbtenteteaiene 3 SET UP REQUIREMENTO cuece ae E E R A a 3 SET UP PROCEDURE Sssesiessscuccdustvasicscdevedivcisatedscccucssastuaccesscesstavasucsedscadscssustwaosssscesssgusssoiuescaedcstustvars 3 MEASUREMENT PROCEDURES vreor aia aa E AEE A 4 SPECIFICATIONS ieioea ea aea e ae Aa 6 GENERAL SAFETY AND PRECAUTIONS ssssssscecccccssossssssossssssscceescssossosoossssssssseseeeeesssssssssssss 7 BATTERY HANDLING AND USAGE s sccccosccectecccusscusscosssdcaavsesensssasteceuwsusdsscosendeaassuscasasesguceesuatsscesees 9 TROUBLE SHOOT ENG oniani a eiedaucsuasdiuncacueuseumecdousssaeedecsuuadsancagveuseusueseuversens 9 CARE AND MAINTENANCE ocsissecccdaclssdsietascdsecsiecasssassccddelscnecsedescdsecsescasesanseoddelcuecsesesedsecsesvasesesseeds 11 WARRANTY INFORMATION scdcssciecaceasccbaciscasdenecauasndeesdesoceancebacisansdeposauasedesedusoeuanesbadseansdewesnuncadeeedeses 11 EXPLANATION OF S
14. idity ranges A Please do not share the cuff with any infectious person to avoid Ccross In eciLon BATTERY HANDLING AND USAGE e When the monitor is connected to the 10S device the battery volume will be displayed on the i0S device If the power is less than 25 please charge the battery The monitor will not work until the battery has enough power e When you charge the monitor the LED will display with different colors indicating the charging status See the table below for details e When charging is needed please connect the monitor to a power source The monitor can work normally while charging o It is suggested that you charge the battery when the battery is less chan 255 Monitor Status Status Indicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light for a few seconds Abnormal state Steady red light Do not change the battery If the battery can no longer be charged please contact Customer Service Ay Overcharging the battery may reduce its lifetime Lithium battery replacement by inadequately trained personnel could result ina hazard such as a fire or explosion A Do not plug or unplug the power cord into the electrical outlet with wet hands If the AC adapter is abnormal please change the adapter Do not pull out the adapter when you are using the monitor Do not use any other type of AC adapter as it may harm the monitor e The m
15. n 16 portable and mobile RF communications equipment and the Wireless Blood Pressure Monitor The Wireless Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Wireless Blood Pressure Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the MONITOR as recommended below according to the maximum output power of the communications equipment Rated maximum Separation distance according to frequency of transmitter output m power of ESO KHZ Co GOMEZ Wee MHZ TO SOO MZ S00 MHZ TO Z75 5 GHZ transmitter For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people
16. ng guide for assessing high blood pressure without regard to age or gender Please note that other factors e g diabetes obesity smoking etc also need to be considered Consult with your physician for accurate assessment Classification of blood pressure for adults Tor BLOOD PRESSURE SBP DBP COLOR CLASSIFICATION mmHg mmHg INDICATOR Optimal lt 120 lt 80 GREEN Normal 120 129 80 84 GREEN Mild Hypertension High norrnal 130 139 85 89 GREEN Grade 1 Hypertension 140 159 90 99 YELLOW Normal BP Grade 2 Hypertension 160 179 100 109 ORANGE Ea rd Grade 3 Hypertension 2180 2100 RED WHO I5H Definitions and Classification of Blood Pressure Levels Note This chart is not intended to provide a basis for any type of emergency condition or diagnosis based on the color scheme this chart only depicts different Classifications of blood pressure Consult your physician for proper interpretation of blood pressure results CONTRAINDICATION Att 1s not recommended for people with serious arrhythmia to use this Wireless Blood Pressure Monitor PARTS AND DISPLAY INDICATORS Bluetooth indicator gt Oe Status indicator START STOP Cuff SET UP REQUIREMENTS The iHealth Wireless Blood Pressure Monitor is designed to be used with the following iPod touch iPhone and iPad models 1Pod touch Sth generation 1Pod touch 4th generation 1Pod touch 3rd generation 1Phone 5 1Phone 45 iPhone 4 1Phone 3GS iPad 4th generation
17. ol for THE OPERATION GUIDE MUST BE READ The sign background color blue The sign graphical symbol white Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local authority or retailer for recycling advice Symbol for KEEP DRY Symbol for WARNING symbol for MANUFACTURER Cw Symbol for SERIAL NUMBER EC REP symbol for EUROPEAN REPRESENTATIVE Ce 0197 Symbol for COMPILES WITH MDD93 42 EEC REQUIREMENTS iHealth is a trademark of iHealth Lab Inc Made for iPod Made for iPhone and Made for iPad mean that an electronic accessory has been designed to connect specifically to iPod iPhone or iPad respectively and has been certified by the developer to meet Apple performance standards Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards Please note that the use of this accessory with iPod iPhone or iPad may affect wireless performance iPad iPhone and iPod touch are trademarks of Apple Inc registeredintheU S andother countries The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by iHealth Lab Inc is under license Other trademarks and trade names are those of their respective owners Manufactured for iHealth Lab Inc
18. on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This product complies with Industry Canada IC RSS 210 IC NOTICE This device complies with Industry Canada licence exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device This product is approved in accordance to R amp TTE directive transmitter OTHER STANDARDS AND COMPLIANCES The Wireless Blood Pressure Monitor corresponds to the following Standards IEC 60601 1 2006 Medical electrical equipment Part 1 General requirements for safety IEC 60601 1 2 2007 Medical electrical equipment Part 1 General requirements for safety Collateral Standard Electromagnetic compatibility Requirements and tests EN 1060 1 1995 Al 2002 A2 2009 Non invasive sphygmomanometers Part 1 General requirements EN 1060 3 1997 Ale 2005 AZ 2009 Non invasive sphyamomanometers Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems
19. onitor cable battery and cuff must be disposed of according to local regulations at the end of their usage Note Battery life and charge cycles vary by use and settings TROUBLESHOOTING PROBLEM FOSSIBLE CAUSE SOLUTION Battery is less than Low Battery 258 Charge the battery Blood pressure is Retest make sure your blood outside of pressure is within Display reads ERROR measurement range measurement range Arm or monitor was Retest make sure not to move moved during test your arm or the monitor Display reads an abnormal result Bluetooth Connection unstable No response The cuff does not inflate properly or pressure falls Gugekly during Test Irregular heartbeat arrhythmia The cuff was not properly applied The cuff position was Not correct or it was not properly tightened Body posture was not correct during testing Speaking moving arm or body being angry excited or nervous during test Bluetooth connection AS oOceEss Tu monitor is abnormal Or SEeErong electromagnetic interference is present Incorrect operation OE SEtono electromagnetic interference Review the cuff application instructions and retest It is inappropriate for people with serious arrhythmia to use this monitor Check with physician Review the cuff application instructions and retest Review the cuff application Aste ructions and terest lt Review body posture instruction
20. r If the user of the Wireless Blood hospital environment Pressure Monitor requires continued operation during power mains interruptions it 1s recommended that the Wireless Blood Pressure Monitor be powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical Location in a typical commercial or hospital environment mains voltage prior to application of the test level For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING Guidance and manufacturer s declaration electromagnetic immunity The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the user of the Wireless Blood Pressure Monitor should assure that it vs used in such an envaronment IEC 60601ltest level Compliance level Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the MONITOR including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF LEC gt Vrms 50 d 12NP 61000 4 6 kHz to 80 MHz Radiated RF 3 V m 80 MHz d 1 2 P 80 MHz to 800 MHz IEC to 2 5 GHz 61000 4 3 d 2 3NP 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W a
21. s and r test Retest when calm avoid speaking or movement during the test Reset 10S device Reset monitor by pressing the START STOP button about 10s Make sure the monitor and 10S device are away from other electrical equipment Please see GENERAL SAFETY AND PRECAUTIONS Press the START STOP button about 10 seconds to reset the device relaunch app and reconnect the TOS device tO the monitor CARE AND MAINTENANCE 1 If this monitor is stored near freezing temperatures allow it to acclimate to room temperature before use 2 If the monitor is not used for a long time please sure to fully charge it every month 3 It 1s recommended that product performance be checked every 2 years or after each repair Please contact the Customer Service 4 No monitor component needs to be maintained by the user The circuit diagrams component part lists descriptions Cali Dracion instructions or other information which will assist the user s appropriately qualified technical personnel to repair those parts of the equipment which are designated for repair can be supplied 5 Clean the monitor with a dry soft cloth or a moistened and well wrung soft cloth using water diluted disinfectant alcohol or diluted detergent 6 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years of usage and the cuff integrity 1S maintained after 1 000 open close cycles of the clos
22. ure 7 The battery can maintain the performance characteristics for a minimum of 300 charge cycles 8 It is recommended that if the cuff is used for example in a hospital or a clinic it be disinfected twice a week Wipe the inner side the side that contacts skin of the cuff with a soft cloth lightly moistened with Ethyl alcohol 75 90 Then air dry the cuff Do not drop this monitor or subject it to strong impact A Avoid high temperature and direct sunlight Do not immerse the monitor in water as this will result in damage to the monitor A Do not attempt to disassemble this monitor A Battery replacement should only be performed by a qualified iHealth technician To do otherwise will void your warranty and possibly damage your unit A Cuff replacement should only be performed by a qualified iHealth technician To do otherwise will possibly damage your unit WARRANTY INFORMATION The iHealth Wireless Blood Pressure Monitor is warranted to be free from defects in materials and workmanship within one year from the date of purchase when used in accordance with the provided instructions The warranty extends only to the end user We will at our option repair or replace without charge the iHealth Wireless Blood Pressure Monitor covered by the warranty Repair or replacement is our only responsibility and your only remedy under the warranty EXPLANATION OF SYMBOLS oymbol tor TYPE BE APPLIED PARTS Curr only AS symb
Download Pdf Manuals
Related Search
Related Contents
Olympus E-PL3 + EZ-M1442 II R JVC LYT1116-001A User's Manual 140217_Pavadentro F.indd Bedienungsanleitung - IMA-Box ENCORE ENHWI-3GN3 router Melange Arena II- III (“La evolución de la perfección” Yeastar NeoGate TG100 Tetrode Assembly Station User Manual Mode d`emploi Bras d`adaptation E3V « Pour une Europe sociale, apprenons la MOC Copyright © All rights reserved.