Home

Detector multigaz pentru patru gaze

image

Contents

1. Detector is operating properly Use caution in suspect areas 34 GasAlertQuattro Troubleshooting Problem Detector does not respond to pushbutton Possible Cause Battery is in critical low battery state or is completely depleted Solution Replace alkaline batteries Refer to Replacing the Alkaline Batteries on page 21 Refer to Charging the Rechargeable Battery Pack on page 20 Detector is performing operations that do not require Pushbutton operation restores automatically when the operation ends user input Sensor s requires calibration Refer to Calibration on page 9 Detector does not accurately Detect ris cold r h tterthan Allow the detector to attain ambient temperature before use measure gas gas temperature Sensor filter is blocked Refer to Replacing the Sensor Filter on page 26 Alarm setpoint s are set Refer to Sample Gas Alarm Setpoints in the GasAlertQuattro a P Technical Reference Guide Define the alarm setpoints in Fleet incorrectly Manager II Detector does not enter alarm Refer to Sample Gas Alarm Setpoints in the GasAlertQuattro Alarm setpoint s set to zero Technical Reference Guide Define the alarm setpoints in Fleet Manager II Detector is in calibration mode Complete the calibration procedure 35 GasAlertQuattro Operator s Manual Problem Detector intermittently enters alarm without reason Featur
2. oxygen on the instrument display Oxygen measurement performance as tested by BAM Oxygen response time tg 15 seconds for oxygen deficiency Oxygen response time tgp 15 seconds for oxygen enrichment Measurement stabilization time 120 seconds Instrument warmup time 32 seconds 31 GasAlertQuattro Operator s Manual Special conditions for safe use LEL measurement Performance approval The EC type examination certificate for LEL measurement applies only to the measurement of methane from 0 to 100 of the Lower Explosion Limit The LEL of methane is equal to 4 4 v v methane in air Additional tests of a notified body would be required in addition to this cer tificate for ATEX performance approval of the GasAlertQuattro with respect to other combustible gases Effect of other toxic gases on the LEL sensor If substances e g sensor poisons that could interfere with and affect the sensitivity of the sensing device are to be expected in the atmo sphere to be monitored which may cause a rapid change of sensitivity the calibration interval shall be reduced Measuring gases of other installed measuring channels of the GasAlertQuattro e g hydrogen sulphide may decrease the sensitivity of the LEL sensor The calibration interval should be reviewed taking into account any degradation in performance Methane LEL measurement performance as tested by BAM Methane response time tgo 15 seconds Stabilization time for methane se
3. Charge the battery immediately when the detector emits a low battery alarm Charge the lithium battery pack using the BW supplied charger and charger adapter only The charging adapter is specific to your region Use of the charging adapter outside your region will damage the charger and the detector Failure to adhere to this caution can lead to fire and or explosion Charge the lithium battery after each workday Ensure the charging connector surface is free of debris and moisture 1 Press and hold to deactivate the detector then plug the charger into an AC outlet Note The time required to charge will increase if the detector is activated 2 Connect the charging adapter to the detector IR receptacle Refer to the following illustration IR receptacle Charging adaptor 3 The lithium battery may require 6 hours to reach full capacity GasAlertQuattro Maintenance Replacing the Alkaline Batteries A Warning To avoid personal injury and or damage to the detector use only BW recommended alkaline batteries Refer to Specifications Change the alkaline batteries only in safe area that is free of hazardous gas 1 Press and hold to deactivate the detector 2 If using the retaining screw loosen the retaining screw 1 2 turns Remove the alkaline battery pack Refer to Replacing the Battery Pack 3 Unhook the ejector bar from the release clasp Move the ejector bar towards the top of the
4. If the IR Lock option is enabled the following screen displays to indicate calibration can only be performed using an IR device MicroDock II or IR Link 4 When the following screen displays attach the calibration cap and apply calibration gas at a flow rate of 250 500 ml min Refer to Connecting the Gas Cylinder to the Detector If a sensor is not yet due for calibration its box will have a greyed out checkmark 5 The detector initially tests for gas When a sufficient amount of gas is detected J displays beside each gas that is detected 10 Calibration IR lock enabled amp Connect to MicroDock or Fleet Manager Press to end waiting 24 Apply Calibration gas now Bra H s co oD uO Press to end Apply Calibration H S co fm o tev Press to end waiting 0 6 The detector then begins calibrating the sensors The following activities occur during the span calibrating displays at the bot tom of the screen Gas values adjust during the span Target gas values that are defined in Fleet Manager II dis play above or below the adjust ing gas value To abort calibration after the sensors have been zeroed press 7 When the following screen displays close the valve on the gas cylinder and remove the calibration cap from the detector A check mark displays beside each sensor that has calibrated s
5. FAILED Fy as cof oM aM waiting 272 Problem Last Calibration Failed Displays when the last calibration failed If the Force Calibration option is enabled the sensors must be calibrated Solution Press Q and calibrate the sensor s immediately Refer to Calibration on page 9 If Cal IR Lock is enabled an IR device IR Link or MicroDock Il must be used to calibrate Sensor self test ERROR A ns co oM aM Press to accept waiting 272 Self test Failed Sensor s fails the self test during startup Press to accept the failed sensor s The Sensor self test error accepted screen displays Replace the sensor when startup is complete Refer to Replacing the Sensors on page 25 Force Calibration enabled Hold to off or continue to i hold to Cal waiting 272 Force Calibration If the Force Calibration option is enabled the sensors must be calibrated to enter normal operation Press and hold to calibrate the sensors or press and release to deactivate the detector Refer to Calibration on page 9 If Cal IR Lock is enabled an IR device IR Link or MicroDock Il must be used to calibrate 37 GasAlertQuattro Operator s Manual Table 1 Startup Troubleshooting Error Screen All sensors FAILED self test aS Powering off Problem All Sensors Fail If the Lockou
6. GA NS 1 Neck strap w safety release 1 GA PA 1 GasAlertQuattro multi uni GA LY 1 Short strap 6 in 15 2 cm 1 MCS power adapter GA ES 1 Extension strap 4 ft 1 2 m 1 GasAlertQuattro multi uni QT C01 MC5 GA ARM 1 Arm band 1 cradle charger GA VPA 1 Vehicle power adapter GACK ERS 1 Hard sided carrying case for GPAs Replacement power adapter SPAK CC1 GasAlertQuattro and or 1 IR Devices motorized Sampling Pump GA USB1 IR roe IR connectivity Replacement Parts QT SCREW Replacement screw kit 40 1 DOCK2 2 1 J K1 screws and screwdriver C1N 00 N GasAlertQuattro docking module MicroDock Il station w charging cable Add one of the following applicable suffixes to the end of the order number to ensure power adapter is correct for region DOCK2 0 GasAlertQuattro docking 1 3 1C1N 00 N module w charging cable UK for United Kingdom Accessories EU for Europe GA BQT Concussion proof boot 1 AU for Australia China GA HQT Carrying holster A 50104941 023 D6445 2 English 2012 BW Technologies by Honeywell All rights reserved
7. R01 may present a risk of fire or chemical burn hazard if misused Do not disassemble heat above 212 F 100 C or incinerate Warning Do not use any other lithium batteries with the GasAlertQuattro detector Use of any other cell can cause fire and or explosion To order and replace the QT BAT R01 lithium battery contact BW Technologies by Honeywell Warning Lithium polymer cells exposed to heat at 266 F 130 C for 10 minutes can cause fire and or explosion OS Warning This instrument contains a lithium polymer battery Dispose of used lithium cells immediately Do not disassemble and do not dispose of in fire Do not mix with the solid waste stream Spent batteries must be disposed of by a qualified recycler or hazardous materials handler Keep lithium cells away from children Deactivating the detector by removing the battery pack may cause improper operation and harm the detector GasAlertQuattro Safety Information Read First GasAlertQuattro Operator s Manual Parts of the GasAlertQuattro tem Description tem Description tem Description tem Description 1 IntelliFlash 4 7 Liquid crystal display 10 7 green LED Pushbutton LCD Alligator clip 2 e indicator 5 oe LEL 8 Audible alarm 11 Battery pack 3 Hydrogen sulfide 6 Carbon monoxide 9 42 Charging connector H2S sensor CO sensor Oxygen Oz sensor and IR interface GasAlertQuattro Screen Elements Sc
8. Self Test Error is enabled and a sensor s has failed the startup self test Refer to Replacing the Sensorson page 25 and Lockout on Self Test Error in the GasAlertQuattro Technical Reference Guide Sensor s requires calibration Refer to Calibration on page 9 33 GasAlertQuattro Operator s Manual Problem The detector enters alarm immediately when activated Possible Cause Sensor needs to stabilize Solution Used sensor Wait 60 seconds New sensor Wait 5 minutes Low battery or critical low battery alarm Replace alkaline batteries Refer to Replacing the Alkaline Batteries on page 21 Refer to Charging the Rechargeable Battery Pack on page 20 Hazardous environment Leave the area immediately Deactivate and reactivate the detector in a safe area that is free of hazardous gas in an atmosphere of 20 9 oxygen A new sensor has been inserted Calibrate the sensor The activation startup self test fails General fault Contact BW Technologies by Honeywell Sensor error Refer to Startup Troubleshooting If required refer to Replacing the Sensors on page 25 Detector Operation Detector does not display normal gas reading after startup sequence Sensors not stabilized Used sensor Wait 60 seconds New sensor Wait 5 minutes Sensor s requires calibration Refer to Calibration on page 9 Target gas is present
9. are overdue for a bump test and the Force Bump test option is disabled Apply gas directly or use the MicroDock II other wise press _ to enter normal operation with the bump test overdue Refer to Bump Test on page 12 38 GasAlertQuattro Troubleshooting Table 1 Startup Troubleshooting Error Screen Problem Solution Error Screen Problem Solution Free tires Force Bump Apply gas ven oa Test immediately or use Caution If the Force the MicroDock II A ms co Bump Test station otherwise BW Technologies by Honeywell recommends to bump aO wo option is enabled press to deactivate test the sensors before each day s use to confirm their e ooie seio the detector ability to respond to ane by exposing the porn toa as concentration that exceeds the alarm setpoints enter normal Refer to Bump Test g Pp operation on page 12 39 GasAlertQuattro Operator s Manual Calibration Troubleshooting Table 2 Error Screen Problem Detector deactivates while trying to calibrate Detector does not enter calibration Displays OFF and then deactivates Solution Calibrate the sensor Error Screen Calibration IR lock enabled Connect to MicroDock or Fleet Manager Press to end Problem Cal IR Lock Enabled IR lock enabled screen displays Solution An IR device must be used to calibrate IR Link or MicroDock II For manual calibration refer to Calib
10. battery pack until it is aligned horizontally over the batteries 21 GasAlertQuattro Operator s Manual 4 Using the tab pull on the ejector bar 5 To the left of the tab pull up on the ejector bar 22 Remove the spent batteries Reset the ejector bar to its original flat position Ensure the ejector bar engages the release clasp 7 Insert the new batteries Position the positive end of the battery at a 30 angle and insert into the battery pack before pushing the negative end down Ensure the batteries are not inserted over the tab Note Ensure all three batteries are inserted with the positive end pointing toward the top of the battery pack 8 Before replacing the battery pack ensure the seal on the instrument and battery pack is free of debris and moisture 9 Replace the battery pack by inserting the bottom first then lower top into place Ensure the tab is tucked in before replacing the battery pack Press until the release tab engages If required tighten the retaining screw using 3 4 in lbs torque GasAlertQuattro WEEE Directive and Battery Directive WEEE Directive and Battery Directive Failure to comply with the following battery removal and dis posal instructions may result in battery shorting battery leak age and or other damage Ensure a qualified technician completes the following procedures Removal and Disposal of the Alkaline Batte
11. of sale or packing slip to establish that the product is within the warranty period THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BW SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT TORT OR RELIANCE OR ANY OTHER THEORY Since some countries or states do not allow limitation of the term of an implied warranty or exclusion or limitation of incidental or consequential damages the limitations and exclusions of this warranty may not apply to every buyer If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction such holding will not affect the validity or enforceability of any other provision Contacting BW Technologies by Honeywell USA 1 888 749 8878 Canada 1 800 663 4164 Europe 44 0 1295 700300 Other countries 1 403 248 9226 Email us at info gasmonitors com Visit BW Technologies by Honeywell s website at www gasmonitors com Introduction The operator s manual provides basic information to operate the GasAlertQuattro gas detector For complete operating instructions refer to the GasAlertQuattro Technical Reference Guide provided on the CD ROM The GasAlertQuattro ga
12. used to lock the battery pack on all European and IECEx scheme detectors and on all Canadian and U S Zone Certified detectors The screwdriver included with the detector has a double ended driver Loosen the brass nut to switch between a Phillips head and a hex head A hex tool is required to tighten and loosen the retaining screw Tighten the screw 1 2 turns using 3 4 in lbs of torque Do not overtighten the screw 19 GasAlertQuattro Operator s Manual 4 Before replacing the battery pack ensure the seal on Hex tool 7 Retaining Battery release latch Replacing the Battery Pack The alkaline and rechargeable battery packs can be changed in hazardous locations 1 Press and hold to deactivate the detector 2 If using the retaining screw loosen it 1 2 turns Push the battery release latch toward the top of the detec tor to release the battery pack 3 From the top of the battery pack lift upward to remove 20 the instrument and battery pack is free of debris and moisture 5 Insert a new battery pack Insert the bottom of the battery pack first then lower the top into place Press until the release tab engages Tighten the retaining screw if required Charging the Rechargeable Battery Pack A Warning To avoid personal injury and or damage to the detector adhere to following Charge only in a safe area that is free of hazardous gas within temperatures of 32 F to 104 F 0 C to 40 C
13. A 1 2 3 Uattro and 4 Gas Detector Operator s Manual BW cic Table of Contents Limited Warranty and Limitation Liability Contacting BW Technologies by Honeywell Introduction 224 scaled acs et oe godt Ait EA coe 1 Zeroing the Sensors 00 0 e p eee eee eee 1 Safety Information Read First 0000 1 Parts of the GasAlertQuattro 0 0 00 ee eee 4 Screen Elements cece eee eee eee eee 5 PUSIMDULOIN aas eE Gases E AER dor ene R 6 Sensor Poisons and Contaminants 7 Connecting the Gas Cylinder to the Detector 8 Calibration 2 2 ects eats ned teers ca ets ard OSE ees 9 Buo TAST e d ct3s 03bs aa bee didt lrg e back 12 PaE aii DEEE E Son pees Beene sees 13 User Options and Sensor Configuration 16 Maintenance ra tre dg as aoe eth ha AE Wer eee atl ah 19 WEEE Directive and Battery Directive 23 Removal and Disposal of the Coin Cell 24 Specifications oa aen Ea sa adeapeans E Paral ele ay Aa 27 ATEX Performance Approval 2 0 00005 30 Troubleshooung gx oeren iera 8 ie Ae DETARE TENE Mi 33 Replacement Parts and Accessories 41 Limited Warranty and Limitation Liability BW Technologies LP BW warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years beginning on the da
14. Quattro according to ATEX performance approvals for LEL and oxygen measurements are shown in Specifications on page 27 These specify the climatic range within which the instrument can be used in conformity to the performance certifications General use For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20 9 v v Oxy gen deficient atmospheres lt 10 v v may suppress some sensor outputs 30 Daily bump test To comply with ATEX performance certifica tion requirements a bump test must be completed before each day s use Calibration To minimize measuring errors the ambient condi tions of temperature humidity and pressure during calibration should be as close as possible to the actual environmental con ditions in which the instrument is intended to be used Calibration interval If the detector is to be used in atmo spheres which may contain compounds known to interfere with inhibit or poison the sensors the calibration intervals must be specified to take into account the possibility of a rapid loss in measurement sensitivity See page 7 Sensor Poisons and Contaminants Sensor impairment Some types and concentrations of dust in the monitored atmosphere may impair the measuring function of the gas detector Cross sensitivities described in the sensor manufacturer s data sheet must be considered Response time verification Before use confirm that the gas detector s res
15. TEL is the maximum permissible gas concentration a worker can be safely exposed to for short periods of time 5 15 minutes maximum Correction Factor LEL The Correction Factor option defines compensation factors for hydrocarbons other than methane The correction factor is only applicable to LEL and can only be applied if the LEL sensor has been cali brated with methane Detector operation using LEL correc tion factors has not been tested by BAM STEL Interval Define the short term exposure limit STEL from 5 15 minutes toxic sensors only TWA Period hours Define the time weighted average TWA from 4 16 minutes toxic sensors only TWA Method Select either the US Occupational Safety and Health Administration OSHA or the American Con ference of Governmental Industrial Hygienists ACGIH TWA calculating method 50 LEL CH4 Enter a percentage value to display the LEL reading as vol assuming a methane environ ment LEL only Auto Zero on Startup When enabled the sensors auto matically zero during the startup sequence The Auto Zero on Startup option is available for the CO H2S LEL and O3 sensors each sensor is enabled individually LEL by Volume CH4 If enabled the detector displays the LEL value as vol assuming a methane environment Note If changing the measurement unit from LEL to Vol or from Vol to LEL a calibration must be completed and the alarm setpoints changed For cali
16. an alarm event occurs Latching Alarms If enabled during an alarm condition he Latching Alarms option causes the low and high gas alarms audible visual and vibrator to persist until the alarm is acknowledged and the gas concentration is below he low alarm setpoint The LCD displays the peak con centration until the alarm no longer exists Local regula ions in your region may require the Latching Alarms option be enabled Force Calibration If enabled during startup if a sensor s is past due for calibration the sensor s must be calibrated o continue and enter normal operation A value must be entered in the Cal Interval days field before enabling Force Calibration Force Bump If enabled during startup if the sensor s is past due for a bump test a bump test should be performed and the overdue sensor s must enter into alarm A value must be entered in the Bump Interval days field before enabling Force Bump Cal IR Lock If enabled the sensor s can only be cali brated using an IR device IR Link or the MicroDock II sta tion Note If the Cal IR Lock option is enabled and a manual calibration is attempted the sensor s will auto zero but they will not be calibrated Flip Display The detector can display screens at 0 upright or 180 upside down depending upon how the detector is worn by the worker If the Flip Display option is enabled the LCD is viewed at 180 upside down Stealth When
17. bration information refer to Calibration on page 9 and for alarm setpoint information refer to Gas Alarm Setpoints in the GasAlertQuattro Technical Reference Guide 10 LEL of reading Over span If enabled the detector automatically over spans the LEL sensor by 10 of the span concentration Enable 10 LEL of reading Over Span to ensure the detector is in compliance with CAN CSA C22 2 No 152 20 8 Base Reading If enabled the detector is configured to detect 20 8 O2 as ambient air When disabled the detector is configured to detect 20 9 O as ambient air Low Alarm Acknowledge If enabled the audible alarm can be temporarily disabled during a low alarm by press ing The vibrator alarm LEDs and LCD remain opera tional toxic and LEL only Maintenance To maintain the detector in good operating condition perform the following basic maintenance as required Calibrate bump test and inspect the detector ona regular schedule e Maintain an operations log of all maintenance bump tests calibrations and alarm events Clean the exterior with a soft damp cloth Do not use solvents soaps or polishes Refer to Sensor Poisons and Contaminants GasAlertQuattro Maintenance Rechargeable Battery Capacity A rechargeable battery s runtime decreases approximately 20 over a two year period of typical use Battery Pack Retaining Screw The retaining screw QAQD 20x provided with the detector must be
18. compliance beep and IntelliFlash auto matically deactivate during a low battery alarm calibration fail bump test fail self test fail and in an alarm condition Critical Battery Alarm Fifteen minutes after low battery alarm activates sequence of 10 rapid sirens and alternating flashes with 1 second of silence in between sequence reactivates seven times e Vibrator alarm pulses Low Battery Powering Off dis plays and the detector deactivates Low battery Coo aN Powering off Note If the Low Alarm Acknowledge option is enabled the audible alarm can be disabled during a low alarm condition The LED and visual alarm indicators remain active until the alarm condition changes or the detector deactivates Press to acknowledge the low alarm and deactivate the audible alarm If the alarm escalates to a high TWA or STEL alarm the audible alarm reactivates 15 GasAlertQuattro Operator s Manual User Options and Sensor Configuration In order to modify user options and detector configuration the following items are required Detector IR Link adapter or MicroDock II Fleet Manager II software The following section describes some of the configuration options available in the detector Refer to the GasAlertQuattro Technical Reference Guide and Fleet Manager II Operator s Manual for complete information Device Configuration The Device Configuration section dis
19. duct Duracell MN1500 20 C lt Ta lt 50 C T4 129 9 C Energizer E91VP 20 C lt Ta lt 50 C T3C 135 3 C AA alkaline battery operating time 14 hours at 20 C 68 F Battery charger Charging adapter First time charge 6 hours Normal charge 6 hours Warranty 2 years including sensors EC Declaration of Conformity http www gasmonitors com Declarations_of_Conformity GasAlertQuattro Specifications Approvals Approved by CSA to both U S and Canadian Standards CAN CSA C22 2 No 157 and C22 2 152 ANS UL 913 and ANSI ISA 12 13 Part 1 CSA Class I Division 1 Group A B C D ATEX CE 0539 amp II 1 G Ex ia IIC Ga T4 for Zone 0 Group IIC KEMA 09 ATEX 0137 EN 60079 0 EN 60079 11 EN 60079 26 Ex ia IIC T4Ga IECEx CSA 09 0006 IEC 60079 0 IEC 60079 11 IEC 60079 26 BAM BAM 11 ATEX 1102 X EN 60079 29 1 for O up to 100 LEL methane BAM ZBF 006 11 EN 50104 for 0 up to 25 v v oxygen Release version GAQF_03_000 IECEx Firmware This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules and ICES 003 Canadian EMI requirements These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accor dance with the instructions may cause harmful interference to radio c
20. enabled the following features are dis abled backlight audible alarms visual alarms IntelliF lash and confidence compliance beep Only the vibrator and the LCD readings activate during an alarm condition Note Should the user wish to comply with ATEX performance certifications stealth mode must be disabled Datalog Interval The Datalog Interval seconds field defines how often the detector records a datalog every 1 120 seconds Enter the desired value GasAlertQuattro User Options and Sensor Configuration The total number of 8 hour days datalogs that can be recorded is assuming 90 of the day has no gas concen trations When the memory is full the detector replaces the oldest datalogs with the most recent datalogs IntelliFlash Interval The IntelliFlash Interval seconds field defines how often every 1 120 seconds the IntelliF lash occurs Confidence Compliance Beep Interval Define how often every 1 120 seconds the confidence compliance beep occurs Language The Language field provides a drop down menu that includes the following language options Eng lish Fran ais French Deutsch German Espa ol Spanish Portugu s Portuguese Select the language from the drop down menu in Fleet Manager II Sensor Configuration Sensor Disabled Enables disables the selected sensor A Warning Use extreme caution when disabling a sensor The disabled sensor cannot detect and alarm against
21. er II To acknowledge any of the due today messages calibration and bump test press If enabled the force calibration and force bump features cannot be bypassed Sensor Poisons and Contaminants Several cleaners solvents and lubricants can contaminate and cause permanent damage to sensors Before using cleaners solvents and lubricants in close proximity to the detector sen sors read the following caution and table A Caution Use only the following BW Technologies by Honeywell recommended products and procedures Use water based cleaners e Use non alcohol based cleaners Clean the exterior with a soft damp cloth Do not use soaps polishes or solvents GasAlertQuattro Sensor Poisons and Contaminants The following table lists common products to avoid using around sensors Cleaners and Lubricants Silicones Aerosols Silicone cleaners Bug Brake cleaners repellents and protectants and sprays Silicone based Lubricants adhesives sealants Lubricants and gels Hand body and Rust Rust inhibitors medicinal creams that Soe Pa inhibitors contain silicone Window and glass Tissues containing Window cleaners silicone cleaners Dishsoaps Mold releasing agents Citrus based Polishes cleaners Alcohol based cleaners Hand sanitizer Anionic detergents Methanol fuels and antifreezes GasAlertQuattro Operator s Manual Connecting
22. es and options are not operating as expected Possible Cause Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas Solution Detector is operating normally Use caution in suspect areas Check peak gas exposure reading Alarm setpoints are set incorrectly Refer to Sample Gas Alarm Setpoints in the GasAlertQuattro Technical Reference Guide Define the alarm setpoints in Fleet Manager II Sensor s requires calibration Refer to Calibration on page 9 Missing or faulty sensor s Refer to Replacing the Sensors on page 25 Changes in Fleet Manager II Verify that the settings in Fleet Manager II are correct Charging Battery has been charging for 6 hours Charging indicator on LCD shows the battery is still charging Battery is trickle charging Battery is fully charged and is ready for operation Battery indicator does not display when charging Battery is depleted below normal levels Charge the battery for 8 hours If the battery indicator does not light after charging contact BW Technologies by Honeywell 36 Startup Troubleshooting Table 1 Startup Troubleshooting GasAlertQuattro Troubleshooting Error Screen Auto Zero ERROR A ns col oM aM Problem Auto zero Error Sensor s failed to auto zero Solution Calibrate the sensor Error Screen Last Calibration
23. he Sensors Rear shell O sensor H2S sensor Sensor filter with gasket Front shell OO Machine screws 6 LEL sensor Side facing front shell Now 9 od gt O L Battery pack CO sensor QO TAS Side facing sensors J PEN 5 aras SSN _ N y NA al al GasAlertQuattro Operator s Manual A Warning To avoid personal injury and or property damage only use sensors that are specifically designed for the detector Replace the sensors in a non hazardous area Note Detectors that are configured for 1 2 or 3 gases may contain a dummy sensor in one of the four sensor locations To replace a sensor or sensor filter refer to the illustration Replacing the Sensors and the following steps 1 8 1 Press and hold to deactivate the detector Press the release latch and remove the battery pack Remove the six machine screws from the rear shell Remove the front shell Remove the spent sensor s Ensure no damage occurs to the LCD Insert the new sensor s Before reassembling the detector ensure the sealing surfaces on the front and rear shells is free of debris and moisture 7 Reassemble the detector Press the front and rear shells together firmly to ensure a proper seal Ensure the front and rear shells have a tight uniform 1 16 in 1 5 mm seal on all sides of the detector 26 8 Replace the six machine screws using 3 4 in Ibs torque Do not ove
24. high gas alarms audible visual and vibrator to persist until the alarm is acknowledged by pressing and the gas concentration is below the low alarm setpoint The peak concentration values display continually until the alarm no longer exists Enable disable Latching Alarms in Fleet Manager II Local regulations may require the Latching Alarms option be enabled The Latching Alarms option must be enabled if the detector is to be used in accordance with ATEX gas performance certification requirements BAM 14 GasAlertQuattro Alarms Alarm Low Battery Alarm Sequence of 10 rapid sirens and alternating flashes with 7 seconds of silence in between continues for 15 minutes Ess flashes e Vibrator alarm pulses After 15 minutes of the low battery alarm sequence the detector enters critical battery alarm see Critical Battery Alarm below Screen H2S ppm 0 0 20 9 OQ vi gt bv CO ppm y LEL Alarm Confidence compliance Beep One beep every 1 120 seconds beep frequency is defined with Confidence compliance Beep Interval option IntelliFlash default one flash every 1 second One flash every 1 120 seconds flash frequency is defined with IntelliFlash Interval option Heartbeat O d pulses once every second to verify detector is operating correctl Screen HS ppm 00 0 Kd CO ppm 209 0 O LEL v ap bv Note Confidence
25. larm conditions TWA alarm STEL alarm low alarm high alarm multi alarm over limit OL alarm low battery alarm crit ical low battery alarm sensor failure alarm IntelliFlash confidence compliance beep Audible alarm 95 dB at 30 cm 12 in variable pulsed beeper Visual alarm Red light emitting diodes LEDs IntelliFlash Green light emitting diode Flash frequency is user defined with IntelliFlash interval option Confidence compliance beep Audible beep from variable pulsed beeper Beep frequency is user defined with confidence compliance beep interval option 28 Display Alphanumeric liquid crystal display LCD with flip display 0 or 180 capability user defined in Fleet Manager II Backlight Activates upon startup and deactivates when self test is complete Activates when the pushbutton is pressed and deactivates after 10 seconds Also activates during an alarm condition and remains lit until alarm ceases Internal vibrator Vibrates during activation deactivation and all alarms Self test Initiated during activation self test runs continuously on the battery and electrochemical sensors H2S and CO while detector is operational Calibration Zero and automatic span User options Startup message lockout on self test error safe mode IntelliFlash confidence compliance beep latching alarms force calibration force bump calibration IR lock flip display stealth mode datalog interval IntelliFlash in
26. ly before operating or servicing GasAlertQuattro Operator s Manual If using the detector near its upper or lower operating temperature BW Technologies by Honeywell recommends zeroing or activating the detector in that environment Charge the detector before first time use BW Technologies by Honeywell recommends the detector be charged after every workday Calibrate the detector before first time use and then ona regular schedule depending on use and sensor exposure to poisons and contaminants BW recommends that the sensors must be calibrated regularly and at least once every 180 days 6 months Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20 9 oxygen The combustible sensor is factory calibrated to 50 LEL methane If monitoring a different combustible gas in the LEL range calibrate the sensor using the appropriate gas Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International BW recommends that the combustible sensor be checked with a known concentration of calibration gas after any exposure to contaminants poisons such as sulfur compounds silicon vapors halogenated compounds etc BW recommends to bump test the sensors before each day s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints Manually verify that the audible visual and vibrato
27. mp test the sensors before each day s use to confirm their ability to respond to gas by exposing the sensors to a gas concentration that exceeds alarm setpoints To operate the detector in accordance with ATEX performance certification requirements the user must complete a bump test before each day s use Per EN 60079 29 1 and EN 60079 29 2 1 Connect the calibration hose to the 0 5 l min regulator on the gas cylinder Refer to Connecting the Gas Cylinder to the Detector To bump test using the MicroDock II station refer to the MicroDock II User Manual 2 Connect the calibration hose to the intake inlet on the calibration cap Arrows on the calibration cap indicate the direction of gas flow 12 3 Attach and tighten the calibration cap onto the detector and apply gas Verify the visual audible and vibrator alarms activate 4 Close the regulator and remove the calibration cap The detector temporarily remains in alarm until the gas clears from the sensors Note In normal operating mode it is possible to display measured values with calibration gas applied to determine any measurement errors Alarms GasAlertQuattro Alarms Refer to the following table for information about alarms and corresponding screens For more information about alarms refer to the GasAlertQuattro Technical Reference Guide Alarm Low Alarm Slow siren upward tone Slow flash Black box around gas flashes Vibra
28. neywell CG Q34 4 Oo 18 0 H2S 25 ppm CO 4 100 ppm bal Nz 34 I Table 3 Replacement Parts and Accessories Dual gas cylinder 50 LEL age CH 2 5 O2 20 9 bal N Model No Description Qty CG4 a 08 ARENE 1 Sensors Single gas cylinder H2S 25 SR W04 75C Combustible LEL sensor 1 G0042 H25 ppm bal No 58 1 SR X10 C1 Oxygen O2 sensor 1 CG2 M 200 Single gas cylinder CO 200 SR M04 SC Carbon monoxide CO sensor 1 103 ppm bal N2 103 I SR H04 SC Hydrogen sulfide H2S sensor 1 Bump alarm gas aerosol CH4 9 9 i Sensor Filters CG BUMP1 2 5 02 10 H2S 40 ppm 1 QT SS Sensor filters Kit of 2 1 S229 RATY x te it SSOP ren aie Quad calibration kit with QT SS K1 Sensor filters kit of 10 1 ease regulator quad gas cylinder A Regulators CG Q34 4 hose and carrying REG DF 1 Demand flow regulator 1 a E 7 i uad calibration kit wit REG 0 5 0 5 I min regulator 1 Kons regulator quad gas cylinder i CG Q58 4 hose and carrying case 41 Model No Description Model No Description Qty QT TC 1 Calibration cap QT AF K1 perp lees w LCD protector 1 Battery Packs wit ilter i QT SS AF Replacement auxiliary filters kit QT BAT RO1 Selig oa lithium battery K1 of 5 filters 1 FOTE Alkaline battery pack batteries QT VMB 1 Vehicle mount attachment 1 included XT AG 1 Alligator clip stainless steel 1 Chargers and Power Options
29. nsor 2 120 seconds Warmup time 32 seconds LEL correction factors Detector operation using LEL correc tion factors has not been tested by BAM 32 Changing measurement range from LEL to Vol If changing the measurement unit from LEL to Vol or from Vol to LEL a calibration must be completed and the alarm setpoints changed For calibration information refer to Calibra tion on page 9 and for alarm setpoint information refer to Gas Alarm Setpoints in the GasAlertQuattro Technical Reference Guide Measuring values from 6 LEL to 3 LEL are indicated as 0 LEL in the measuring mode Measuring values in the 3 LEL ranges of the configured span gas concentration are indi cated as the configured span gas concentration GasAlertQuattro Troubleshooting Troubleshooting f the problem persists contact BW Technologies by Honeywell Problem Startup Possible Cause Solution The detector does not activate Depleted batteries Replace alkaline batteries Refer to Replacing the Alkaline Batteries on page 21 Refer to Charging the Rechargeable Battery Pack on page 20 Damaged detector Contact BW Technologies by Honeywell Detector automatically deactivates Automatic deactivation due to critical low battery Replace alkaline batteries Refer to Replacing the Alkaline Batteries on page 21 Refer to Charging the Rechargeable Battery Pack on page 20 Lockout on
30. ommunica tions However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help 29 GasAlertQuattro Operator s Manual ATEX Performance Approval Special conditions for safe use To comply with ATEX performance approval requirements the detector must be operated in the following manner Operating manual Read and understand the operating man ual It is essential that the instructions for correct use are fol lowed Instrument power up The instrument must be powered up in a safe area that is free of hazardous gas and in an atmosphere of 20 9 oxygen The detector must be turned on and verified to be operating in the normal measurement mode before it can be brought into a hazardous area Low battery alarm In the event of a low battery alarm the user must leave the hazardous area immediately Operating environment The operating environment ranges that apply for use of the GasAlert
31. plays data about the detector allows for a startup message to be entered and defines and enables disables settings for the detector Serial Number Field This field displays the serial number e g QA111 001000 of the detector Firmware Version This field displays the current firmware version that displays on the detector LCD during the startup sequence If new firmware is uploaded to the detector the Firmware Version field automatically updates Hardware Version This field displays the current hard ware version of the detector Startup Message Enter text to display on the detector LCD during startup 50 characters maximum Enter infor mation such as employee name plant area emergency number s etc 16 Lockout on Self Test Error If Lockout on Self Test Error is enabled and a failure occurs during the self test the screen displays Sensor Self Test Error Lockout Enabled and the detector deactivates Safe Mode If enabled SAFE displays continuously on the LCD unless an alarm condition occurs Confidence Compliance Beep If enabled the Confi dence Compliance Beep provides continuous audible con firmation that the detector is operating correctly Frequency of the beep is defined with the Confidence Compliance Beep Interval option every 1 120 seconds Note Confidence Compliance Beep automatically disables during a low battery alarm self test fail calibration fail bump test fail and when
32. ponse time is fast enough to trigger alarms so that unsafe situations are avoided If necessary set the alarm levels below the standard safety related limit values to allow enough time for protective measures to be carried out Instrument configuration FleetManager II version 2 6 0 or higher must be used to make configuration changes to the GasAlertQuattro To operate the instrument in accordance with ATEX gas perfor mance configuration requirements Latching alarms must be enabled See page 16 Stealth mode must be disabled See page 17 Force Bump must be enabled See page 12 Bump Interval days must be set to 1 for all sensors IntelliFlash must be enabled See page 17 LEL Low Alarm and High Alarm values cannot be 0 GasAlertQuattro ATEX Performance Approval Special conditions for safe use oxygen measurement Performance approval The EC type examination certificate applies to the measurement of oxygen up to 25 v v Baseline and span zone blanking Please note that oxygen measurements in the range from 20 5 v v up to 21 3 v v are indicated as 20 9 oxygen on the instrument display Measurement values in the range within 0 2 v v of the span gas concentration are displayed as the span gas concentration For example if the span gas concentration used for oxygen bump testing is configured as 18 v v measured values in the range of 17 8 v v to 18 2 v v will be indicated as 18 0
33. r alarms are activated Calibrate if the readings are not within the specified limits For an additional bump test caution relating to the ATEX performance certification refer to Bump Test on page 12 Caution High off scale LEL readings may indicate an explosive concentration Any rapid upscaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit which can be hazardous For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20 9 v v Oxygen deficient atmospheres lt 10 v v may suppress some sensor outputs Extended exposure of the GasAlertQuattro to certain concentrations of combustible gases and air may stress the detector element that can seriously affect its performance If an alarm occurs due to a high concentration of combustible gases a calibration should be performed or if needed the sensor replaced Before using common products around sensors refer to Sensor Poisons and Contaminants _on page 7 High concentrations of certain toxic gases for example H2S may have an adverse effect on the LEL sensor This effect known as inhibition is usually temporary but in extreme circumstances can impair the sensitivity of the LEL sensor After any gas exposure that causes an alarm in the toxic gas sensors the LEL sensor should be verified with a bump test and recalibrated if necessary Warning The lithium battery QT BAT
34. rage temperature 40 C to 60 C 40 F to 140 F Operating humidity 10 to 100 relative humidity non condensing Operating environment specifications for use according to ATEX performance approvals Oxygen and methane measurements Operating temperature range certified by BAM 20 C to 50 C Operating humidity as tested by BAM 5 r H to 95 r H Extended Range of Use for Temperature and Humidity as Compared To EN 50104 Oxygen performance and EN 67009 29 1 LEL performance Storage temperature range tested by BAM 25 C to 60 C Operating pressure as tested by BAM 80 kPA to 120 kPa Storage duration Two years from purchase date 27 GasAlertQuattro Operator s Manual Dust and moisture ingress P66 67 with screw engaged Alarm setpoints May vary by region and are user defined Detection range H3S 0 200 ppm 0 1 ppm increments from 0 0 to 39 9 ppm 1 ppm increments above 40 ppm CO 0 1000 ppm 1 ppm increments Oz 0 30 0 vol 0 1 vol increments Combustible LEL 0 100 1 LEL increments or 0 5 0 v v methane Sensor type H2S CO Os Single plug in electrochemical cell Combustibles Plug in catalytic bead O measuring principle Capillary controlled concentration sensor Bump test specified limits BW recommends using a gas cyl inder that will ensure the combustible sensor has an accuracy of 0 to 20 of actual reading reference CAN CSA C22 2 No 152 A
35. rating Using an IR Device in the GasAlertQuattro Technical Reference Guide For automated calibration refer to the MicroDock II User Manual Self test Fail If a sensor s fails auto zero an error message displays showing which sensor s failed H S ppm CO ppm v 208 x LEL vik m Gv Replace the sensor or contact BW Technologies by Honeywell Refer to Replacing the Sensors on page 25 Calibration gas not detected x Press to skip Calibration Error Insufficient amount of gas detected Verify the span gas values on the cylinder match the span gas values set for the detector Ensure gas is applied at a flow rate of 250 500 ml min Ensure cylinder is not empty or expired Replace immediately if required Replace the regulator if required Bump Test Troubleshooting For bump test troubleshooting recommendations refer to Table 1 Startup Troubleshooting 40 Replacement Parts and Accessories A Warning GasAlertQuattro Replacement Parts and Accessories Model No Description Gas Cylinders and Kits Qty Quad gas cylinder CH4 2 5 To avoid personal injury and or damage to the detector CG Q58 4 O5 18 0 H2S 25 ppm CO 1 use only the specified replacement parts 100 ppm bal N 58 I To order parts or accessories contact BW Technologies by Quad gas cylinder CH4 2 5 Ho
36. reen Elements iis Calibration gas cylinder Display during startup to indicate audio and visual alarm pass or fail during a MicroDock bump test Displays if calibration is initiated and the Cal IR Lock option is enabled Bump test gas cylinder Kf Displays when the Stealth option is enabled Displays during calibration and when startup is complete Indicates pass for startup sensors calibrations and bump tests Displays when the detector is in alarm not applicable to TWA and STEL Battery full charge Indicates fail for startup sensors calibrations and bump tests A Displays when there is a warning failure error or low battery Battery half charge Pushbutton displays when screen provides an option to end or skip kd Heartbeat pulses continually during normal operation to verify the detector is operating correctly Low battery warning 20 8 02 Reading displays with white back ground during normal operation Displays for STEL alarms and setpoints Displays when the detector is connected to an IR Lin Reading displays with alternating black background when the sensor is in alarm Displays for TWA alarms and setpoints Displays when the detector is communicating with Fleet Manager II Grey check box displays during bump tests or calibration when a gas is not due M Displays during peak gas exposure informa
37. rtighten the screws Replace the battery pack 9 New sensors must be calibrated prior to use Calibrate the new sensor s immediately Refer to Calibration Replacing the Sensor Filter To replace the filter refer to the illustration Replacing the Sen sors on page 25 and the following steps 1 6 1 Press and hold to deactivate the detector Press the release latch and remove the battery pack 2 Remove the six machine screws from the rear shell Remove the front shell Remove the sensor filter 4 Refer to the following illustration before inserting the new filter Ensure the filter is laying flat and that the holes are correctly aligned over the filter posts Facing front shell Facing sensors Before reassembling the detector ensure the sealing surfaces on the front and rear shells is free of debris and moisture Replace the front shell Press the front and rear shells together firmly to ensure a proper seal Ensure the front and rear shells have a tight uniform 1 16 in 1 6 mm seal on all sides of the detector Replace the six machine screws using 3 4 in lbs torque Do not overtighten the screws Replace the battery pack GasAlertQuattro Specifications Specifications Instrument dimensions 13 x 8 1 x 4 7 cm 5 1 x 3 2 x 1 5 in Weight 316 g 11 15 oz with rechargeable battery pack 338 g 11 92 oz with alkaline battery pack Operating temperature 20 C to 50 C 4 F to 122 F Sto
38. ry Pack Only a qualified technician should complete the following proce dures To remove the alkaline batteries refer to steps 1 to 6 in Replacing the Alkaline Batteries on pages 21 and 22 Removal and Disposal of the Rechargeable Battery Pack To remove the rechargeable battery pack refer to steps 1 to 3 in Replacing the Battery Pack on page 20 Dispose of the battery pack according local laws 23 GasAlertQuattro Operator s Manual Removal and Disposal of the Coin Cell The detector contains a coin cell to power the real time clock Only a qualified technician should complete the following proce dure Item Description 1 Rear machine screws 6 2 Rear shell 3 Front shell and PCB 4 Coin cell 24 DD Ors Be 208 7 Press and hold to deactivate the detector If the battery pack has not yet been removed refer to Removal and Disposal of the Alkaline Battery Pack on page 23 or Removal and Disposal of the Rechargeable Battery Pack on page 23 Remove the six machine screws on the rear shell Remove the two screws on the main PCB Remove the main board The coin cell is connected to the board by four leads Clip the four leads individually to remove the coin cell A Caution Do not touch two or more leads when disconnecting the battery cell 8 Dispose of the coin cell according to local laws GasAlertQuattro Removal and Disposal of the Coin Cell Replacing t
39. s detector the detector is designed to warn of hazardous gas levels above user defined alarm setpoints The detector is a personal safety device It is your responsibility to respond properly to the alarm Note The detector is shipped with English as the default displayed language Additional languages provided are French German Portuguese and Spanish The screens for the additional languages are displayed on the detector and in the corresponding operator s manual Zeroing the Sensors To zero the sensors refer to steps 1 3 in Calibration on page 9 GasAlertQuattro Safety Information Read First Use the detector only as specified in this operator s manual and the technical reference guide otherwise protection provided by the detector may be impaired Read the following Cautions before using the detector A Cautions Warning Substitution of components may impair Intrinsic Safety Before using the detector refer to Sensor Poisons and Contaminants Protect the combustible sensor from exposure to lead compounds silicones and chlorinated hydrocarbons Although certain organic vapors such as leaded gasoline and halogenated hydrocarbons may temporarily inhibit sensor performance in most cases the sensor will recover after calibration Caution For safety reasons this equipment must be operated and serviced by qualified personnel only Read and understand the technical reference guide complete
40. ses to gas This calibration procedure is written as the procedure is intended If an error or alarm screen displays refer to Calibration Trouble shooting in the GasAlertQuattro Technical Reference Guide A Caution Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20 9 oxygen If performing single gas calibration calibrate O first Note The maximum hose length for calibration is 3 ft 1 m The following steps are written for use with a standard quad gas cylinder Calibration can only be aborted after the sensors have been zeroed If is pressed to abort CALIBRATION cancelled displays 1 Press and hold as the detector performs the Powering off countdown OFF Poweringloffias Continue holding when OFF displays and the detector briefly deactivates 2 The detector activates again and performs the calibration countdown Continue holding until Starting Calibration displays 3 The detector enters the zero function zeroing displays while the detector zeros all of the sensors A Caution If a sensor fails to zero it cannot be calibrated Refer to Startup Self Test Troubleshooting in the GasAlertQuattro Technical Reference Guide GasAlertQuattro Calibration OFF Starting Calibration Continue holding E to calibrate or zero zeroing H2S ppm CO ppm oo 0 4 0 LEL GasAlertQuattro Operator s Manual
41. t on Self test Error option is enabled and all sensors fail the detector automatically deactivates Solution A failure of all sensors can be caused by poisoning alcohol and silicon Allow sensors to recover for 1 hour If sensors fail startup again refer to Replacing the Sensors on page 25 Error Screen H S days CO days CAL 3q__DUE 180 x 0 days Let Press to skip waiting 272 Problem Calibration Overdue Displays when calibration is past due If the Force Calibration option is enabled the sensor s must be calibrated to enter normal operation Solution Press to continue and calibrate the sensor s immediately Refer to Calibration on page 9 If Cal IR Lock is enabled an IR device IR Link or MicroDock Il must be used to calibrate Last Bump Test FAILED D ns col oM eM Press to accept waiting 272 Last Bump Test Failed If the last bump test failed and the Force Bump Test option is enabled a bump test must be performed Use the MicroDock II station to perform a bump test otherwise press _ to deactivate the detector If you do not have a MicroDock ll change the bump interval in Fleet Manager Il refer to Bump Test on page 12 Bump Test due Use MicroDock or apply gas now ent oO O Press to ignore waiting 272 Bump Test Overdue Screen displays when sensors
42. te of shipment to the buyer This warranty extends only to the sale of new and unused products to the original buyer BW s warranty obligation is limited at BW s option to refund of the purchase price repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period In no event shall BW s liability hereunder exceed the purchase price actually paid by the buyer for the Product This warranty does not include a fuses disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use b any product which in BW s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation handling or use c any damage or defects attributable to repair of the product by any person other than an authorized dealer or the installation of unapproved parts on the product or The obligations set forth in this warranty are conditional on a proper storage installation calibration use maintenance and compliance with the product manual instructions and any other applicable recommendations of BW b the buyer promptly notifying BW of any defect and if required promptly making the product available for correction No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW and c the right of BW to require that the buyer provide proof of purchase such as the original invoice bill
43. terval con fidence compliance beep interval and language selection Sensor options Sensor enable disable calibration gas values calibration interval bump test interval alarm setpoints low high TWA STEL STEL interval TWA period auto zero at startup enable disable LEL correction factor 10 of reading over span low alarm acknowledge O measurement LEL gas measurement vol methane measurement Year of manufacture The detector s year of manufacture is determined from the serial number The second and third num ber after the second letter determines the year of manufacture E g QA111 001000 2011 year of manufacture Approved lithium battery for GasAlertQuattro product Lithium ion polymer QT BAT RO1 as per standards UL913 EN60079 11 EN60079 0 IEC 60079 0 IEC 60079 11 EN 60079 29 1 EN 50104 and C22 2 No 157 Rechargeable battery QT BAT R01 Temperature code Lithium polymer 20 C lt Ta lt 50 C T4 Lithium battery operating time One rechargeable lithium polymer battery provides the following operating runtimes 20 hours at 20 C 68 F 18 hours at 20 C 4 F Lithium Battery operating time ATEX performance approval 26 hours tested according to EN 60079 29 1 2007 and EN 50104 2010 Approved alkaline battery pack for GasAlertQuattro QT BAT A01 as per standards UL913 EN60079 11 EN60079 0 EC 60079 0 IEC 60079 11 C22 2 No 157 Approved alkaline batteries for GasAlertQuattro pro
44. the applicable gas Calibration Gas ppm Define the span gas concentra tion for each sensor The span gas concentration must match the span value on the gas cylinder e Calibration Interval Define how often a sensor should be calibrated 0 365 days in the Calibration Interval days 17 GasAlertQuattro Operator s Manual field A different calibration interval can be defined for each sensor A Caution BW recommends that the sensors be calibrated at least once every 180 days 6 months 18 Bump Interval Define how often a bump test should be performed for each sensor 0 365 days in the Bump Interval days field A different bump interval can be defined for each sensor Low Alarm Define the low alarm setpoints for each sen sor Refer to Sample Gas Alarm Setpoints for factory defined alarm setpoints in the GasAlertQuattro Technical Reference Guide High Alarm Define the high alarm setpoints for each sen sor Refer to Sample Gas Alarm Setpoints for factory defined alarm setpoints in the GasAlertQuattro Technical Reference Guide TWA Alarm The time weighted average TWA is a safety measure used to calculate accumulated averages of gases Using the US Occupational Safety and Health Administration OSHA method or the American Confer ence of Governmental Hygienists ACGIH method an average is calculated to ensure the detector alarms when the TWA has accumulated STEL Alarm The short term exposure limit S
45. the Gas Cylinder to the Detector Gas Cylinder Guidelines Fy e To ensure accurate calibration use a premium grade calibration gas Use gases approved by the National Institute of Standards and Technology e Ifa certified calibration is required contact BW Technologies by Honeywell e Do not use a gas cylinder past its expiration date Gas Cylinder Connection Note Only use the calibration cap during the calibration process Read the following steps 1 5 prior to initiating calibration 1 Verify the calibration gas being used matches the span concentration value s that are set for the detector 2 Connect the calibration hose to the 0 5 l min regulator on the gas cylinder For use with the MicroDock II use a demand flow regulator and refer to the MicroDock lI User Manual 3 Connect the calibration hose to the intake inlet on the calibration cap Arrows on the calibration cap indicate the direction of gas flow 4 Begin the calibration procedures Do not attach the calibration cap until instructed to apply gas When instructed place the calibration cap on the detector and tighten the knob NOTE Ensure the cap is securely fastened before applying gas 5 When calibration is complete disconnect the hose from the calibration cap and the regulator Remove the calibration cap from the detector Calibration Calibration is performed to adjust the sensitivity levels of sensors to ensure accurate respon
46. tion screens Displays when the detector s firmware is being updated Displays when the most recent calibration or bump test failed but a previous calibration or bump test is still valid within the due date Also displays during auto zero a Displays during an operation such as auto zeroing Displays when gas should no longer be applied after a bump fest or calibration GasAlertQuattro Operator s Manual Pushbutton Pushbutton Description To activate the detector press and hold in a safe area that is free of hazardous gas and in an atmosphere of 20 9 oxygen To deactivate the detector press and hold O during the powering off countdown Release when OFF displays To view the date time current battery power calibration due date bump test due date TWA STEL and peak readings press twice rapidly To clear the TWA STEL and peak readings press and hold O when the LCD displays Hold to reset peaks TWA STEL To initiate calibration press and hold while the detector performs the OFF countdown Continue holding while the LCD briefly deactivates and then reactivates to begin the calibration countdown Release when Calibration started displays To activate the backlight press and release To acknowledge latching alarms press To acknowledge a low alarm and disable the audible alarm press The Low Alarm Acknowledge option must be enabled in Fleet Manag
47. tor alarm activates Screen HS ppm CO ppm Alarm TWA Alarm Fast siren downward tone Fast flash Black box around gas flashes Vibrator alarm activates Screen High Alarm Fast siren downward tone e Fast flash Black box around gas flashes e Vibrator alarm activates H2S ppm CO ppm 00 Pal STEL Alarm Fast siren downward tone Fast flash Black box around gas flashes Vibrator alarm activates H2S ppm CO ppm 13 GasAlertQuattro Operator s Manual Alarm Multi Alarm Alternating low and high alarm siren and flash Black box around gas flashes Type of alarm alternates Vibrator alarm activates H2S ppm CO ppm 200 02 LEL v m v Over Limit OL Alarm Fast siren downward tone Fast flash Black box around gas flashes e Vibrator alarm activates Note LCD may also display an under limit reading OL Screen HS ppm CO ppm 0 0 0 209 Bll O LEL v m Gv Sensor Failure Alarm displays H2S ppm VALJU 209 O CO ppm 0 x LEL v ap v Normal Deactivation Sequence of alternating beeps and alternating flashes e Vibrator alarm activates Countdown initiates OFF displays OFF Note If enabled during an alarm condition the Latching Alarms option causes the low and
48. uccessfully H S ppm 20 CO ppm 250 100 0 18 0 4 50 0 O LEL Press to end calibrating Turn gas off G26 Hs OM wt Press to skip 8 When calibration is complete the following screen displays Note The calibration due date cannot be reset for a sensor that fails calibration If a sensor fails or an error screen displays refer to Calibration Troubleshooting in the GasAlertQuattro Technical Reference Guide 9 All successfully calibrated sensors automatically reset to the number of days defined in the Cal Interval field in Fleet Manager II The calibration due dates can be changed in Fleet Manager II 10 The detector now enters normal operation Calibration complete ik Press to skip HS days CO days O NEXT 34 CAL O days LEL days Press to skip HS ppm 00 209 0 v aD by CO ppm LEL GasAlertQuattro Calibration 11 GasAlertQuattro Operator s Manual Bump Test A bump test applies test gas to force the detector into alarm A bump test should be performed regularly to confirm the sensors are responding correctly to gas and that the audible visual and vibrator alarms activate during an alarm condition The detector can also prompt for a bump test during startup when the Bump Test Interval is defined Refer to the GasAlert Quattro Technical Reference Guide A Caution BW recommends to bu

Download Pdf Manuals

image

Related Search

Related Contents

Konica Minolta VECTISS100 Film Camera User Manual    Mode d`emploi Version complète>  Bosch WAE24369GB washing machine  Rust-Oleum Automotive CC1000 Use and Care Manual  Samsung KRS180PTDA User Manual  Pro-Ject USB Box S+  Jewel Series Manual  Entreprise  un produit dustbane pour l`entretien de tapis  

Copyright © All rights reserved.
DMCA: DMCA_mwitty#outlook.com.