1. DX 200 DIVER HELD MAGNETOMETER User Manual 1e SFIN scan International Limited DX 200 PROFESSIONAL KIT DX 200 Professional Kit Includes Optional Shipping Carrying Storage case for DX 200 Control Unit Sensor Probe D Ring Waist Belt Charger Bonephone Underwater Headphones and Armsaver 18 DX 200 DIVER HELD MAGNETOMETER User Manual Manufacturer Contact Information If you should need to contact Aquascan International Limited for advice on your purchase to order further equipment or to arrange a repair please use the following contact information Mailing Address Aquascan International Limited Aquascan House Hill Street Newport South Wales NP20 1LZ United Kingdom Tel 44 0 1633 841117 Fax 44 0 1633 254829 E mail Sales Enquiries sales aquascan co uk General Inquiries info aquascan co uk Technical Support support aquascan co uk Website www aquascan co uk Aquascan International Ltd DX 200 DIVER HELD MAGNETOMETER User Manual EC Declaration of Conformity We hereby declare that the following equipment complies with the essential requirements at the Electronic Compatibility Directive 89 336 91 26 3 and 92 31 This equipment should not be modified without our approval as this declaration will lose its authority Equipment description Diver Held Magnetometer Model DX 200 Manufacturer Aquascan International Ltd Aquascan
2. by users of boat towed magnetometers is a diver held device to help pinpoint magnetometer targets located with a conventional towed magnetometer The DX 200 fills this need extremely well Example of appropriate use of the DX 200 The following is an example of an application where the DX 200 might be utilised with great success and great savings in time and effort A magnetometer target is found with a boat towed system in 35 feet of water The visibility is minimal and upon examination by a diver the bottom is found to be flat sand and no target could be detected within the range of a metal detector The DX 200 is taken to the bottom by the diver after the unit is switched on and set to Hi sensitivity the diver scans the identified area in a series of straight lines radiating out from the initial seabed marker and notices an increase in the audio frequency in one direction Moving slowly in the direction where the audio output of the DX 200 increased the diver notes where the peak response occurs using the range switches to 3 DX 200 DIVER HELD MAGNETOMETER User Manual desensitise as necessary to note the point where the peak response occurs The diver next determines if a larger peak response can be on a line that intersects at 90 degrees to the original search direction this will assist in determine the exact centre of the target and to an extent the size and shape of the target The above described technique using steady straigh
3. in an arc and notice where the highest audio output occurs which should be as the probe is pointed directly at the target Using this technique you can use a sweeping motion of the probe to actually walk yourself to the target 8 As you get closer and closer to the target the DX 200 will begin to have constant audio output so you will need to lower the sensitivity to be able to actually pinpoint the target Do this by changing the SENSITIVITY setting to MED And then as you get even closer change the setting to LO 9 If possible experiment with several different targets of different sizes to get an idea of how the unit behaves under different circumstances Having performed the above checks on land you are now ready to use the Aquascan DX 200 in the water Underwater Operation NOTE As explained above in the Land Test you must be free of objects containing ferrous materials Steel dive tanks are probably out of the question Most Stainless Steel does not actually contain any iron so it usually won t interfere with the operation of the DX 200 But check yourself carefully something as small as a set of keys can affect the operation of the unit and give you false readings Also don t try to operate the unit in the vicinity of a large steel vessel 1 The control unit of the DX 200 can be worn or carried in a number of ways the choice will be influenced by the conditions and diver preference The most obvious choice is worn clip
4. this self discharge causes the battery to slowly discharge itself over a period of months To maintain maximum performance it is important to charge your DX 200 before use Checking Battery Pack Performance Note For routine maintenance this should be carried out every three months The recommended procedure is as follows 1 0 1 1 1 2 1 3 1 4 1 5 1 6 Connect the Sensor Probe to your DX 200 module Power on your DX 200 and leave it on overnight to fully discharge the battery Power off the DX 200 and disconnect the Sensor Probe Charge the DX 200 for at least 14 hours using the provided charger Once the DX 200 is fully charged disconnect the charger Connect the Sensor Probe and Bonephone Headphone to the DX 200 unit and switch it on Place the Sensor Probe away from any ferrous target and adjust the SET AUDIO control to get a steady slow ticking sound if possible This may be difficult if you are inside a building due to the electronic noise and many metal objects Make a note of the time DX 200 DIVER HELD MAGNETOMETER User Manual 1 7 Confirm the functionality of the unit every 30 minutes re adjusting the tick over as necessary 1 8 When it becomes impossible to set the threshold or the audio ceases altogether this can be considered to be the end of charge status for the battery This should be in the range of 10 to 12 hours for a healthy battery pack 1 9 If the performance becomes unacceptab
5. D and the LO as you get closer 6 If you are just doing a general search of an area you will probably want to establish some sort of grid and move in straight lines scanning back and forth in an arc to insure you cover the entire area to be searched Note Different environmental conditions may make a small difference to the threshold setting as you perform your search which can be readjusted by the SET AUDIO control knob Care and Maintenance Proper care of your Aquascan DX 200 will be repaid by a long and trouble free life and attention should be given to the following points 1 After use and before removing plugs make sure that the unit is switched off and wash down with clean fresh water NEVER use chemicals to clean your DX 200 as these will damage o rings and seals which can cause them to fail and allow the incursion of water into the control unit 2 Remove the Sensor Probe and Bonephone Headphones plugs from the bulkhead sockets and make sure the Rubber Seal is free from sand or other debris remove and clean if necessary Dry off the plug pins and blow any moisture out of the sockets You can use a Q Tip to clean and dry the sockets Ikelite does not recommend using silicone grease or any other material on the connectors The rubber seal provides adequate isolation of the connectors from the water Use of these type products tends to accumulate sand and other debris which can actually interfere with the seal Never utili
6. DX 200 DIVER HELD MAGNETOMETER User Manual AQUASCAN INTERNATIONAL LTD AQUASCAN HOUSE HILL STREET NEWPORT SOUTH WALES UK NP20 1LZ Aquascan International Limited Tel 44 1633 841117 Fax 44 1633 254829 UNDERSEA DETECTION EQUIPMENT Email info aquascan co uk Website www aquascan co uk DX 200 DIVER HELD MAGNETOMETER Operating Instructions This manual is copyright 2009 Aquascan Int Ltd all rights reserved This document must not be replicated in whole or part without the prior express permission of Aquascan International Ltd Rev 5 18 Jan 2012 DX 200 DIVER HELD MAGNETOMETER User Manual Page Contents L Front 2 Page of Contents 3 Introduction Applications Example of appropriate use of the DX 200 4 General comments The DX 200 Equipment 6 Initial Testing of your DX 200 Charging the batteries 8 Underwater Operation 9 Care and Maintenance 10 Battery Care amp Checking Battery Performance Status 12 Troubleshooting Guide No Response from the DX 200 Module Erratic Performance from the DX 200 Module 14 DX 200 Specifications 15 DX200 Accessories 17 DX 200 Optional Accessories 18 DX 200 Professional Kit 19 Manufacturer Contact Information 20 EC Declaration of Conformity Product Guarantee loose insert DX 200 DIVER HELD MAGNETOMETER User Manual Introduction The AQUASCAN name has been established for more than two decades and conveys the message
7. House Hill Street Newport NP20 1LZ S Wales Tel 44 0 1633 841117 255645 Fax 44 0 1633 254829 Email info aquascan co uk Applicable directives EN 55011 General Emission Standard Class A Group 1 EN 50082 Generic Immunity Standard Part 2 A technical construction file for this equipment is retained at the manufacturing base Ist January 1996 Signed gt l i Date im nssiesisssersoissrsossssursssssosrosesss D DINTE a J R WILLIAMS Managing Director a Name Position fa Wessidicdsesbcscossscs sasssbedeessed densesbeses EREE EAEE 20
8. UG THE CHARGER FROM THE MAINS AND PLUG IT BACK IN TO INITIATE ANOTHER REGULAR 6 HOUR CHARGE CYCLE The batteries will not be damaged by being left on charge for a longer period than necessary due to the controlled charge current 14hr rate If the batteries become fully discharged due to natural usage for the maximum operating period or through the unit accidentally being left in the on state the unit should be fully recharged as soon as possible to avoid irreparable damage to the battery cells see above for charging from a discharged state DO NOT USE ANY OTHER BATTERY CHARGER THAN THE ONE SUPPLIED AS DAMAGE TO THE BATTERY PACK MAY RESULT DX 200 DIVER HELD MAGNETOMETER User Manual INTERNATIONAL When the charge period has elapsed disconnect the AC supply to the charger first and then unplug the connector from the socket on the detector 2 For initial testing on land you will have to find a location where you can position yourself at least 30 metres from any large ferrous objects in all directions Get rid of anything containing iron on your body or in your clothing and leave your cell phone at home An empty farm field or park or sports field is ideal for testing Park your car at the edge of the field at least 20 metres from where you will be testing or find a location where there is some large ferrous target in one direction only 3 To prepare the DX 200 for use plug the connector on the Sensor Probe into t
9. e dive equipment currents etc So please take the time to perform this Initial Testing of your unit before venturing into the water with it 1 N B N B N B Charging the batteries The DX 200 is supplied only partially charged and will require a full 14 hours charging duration to bring the unit to a fully charged status Make sure the DX 200 is in the OFF position Plug the output lead from the charger into the AUDIO amp CHARGE socket on the control bulkhead Please note that this can only be inserted one way round due to the locating polarising pin Screw down the retaining collar This should be good finger tightness only DO NOT ON ANY ACCOUNT PLUG THE CHARGER INTO THE PROBE SOCKET AS DAMAGE MAY BE DONE TO THE ELECTRONICS Ensure the charger is correctly connected to the Charge Audio socket before plugging the charger into the mains supply 100 250vAC After running for an hour or so the charger may get slightly warm to the touch This is quite normal THE TYPE OF CHARGERS SUPPLIED WITH THE DX 200 CHARGE AT A REGULAR RATE FOR APPROXIMATELY 14 HOU HOURS AND THEN GO INTO A TRICKLE CHARGE MAINTENANCE MODE WHILE CHARGING AT THE REGULAR RATE THE GREEN LIGHT WILL BE ON ONCE THE CHARGER DROPS INTO TRICKLE CHARGE MODE THE LIGHT WILL GO OUT THIS SHOULD NORMALLY TOP OFF THE CHARGE IN THE BATTERY PACK AFTER A DAY OF WORK BUT IF YOUR DX 200 IS FULLY DISCHARGED YOU SHOULD WAIT UNTIL THE LIGHT GOES OUT THEN UNPL
10. ence of a buried wreck site Large ferrous masses will be detected at greater distances than smaller masses and will be noted to have a much slower build up to the peak audible response The DX 200 Equipment A standard AQUASCAN DX 200 kit consists of the following items Sealed electronic unit Bone conductor earphone Bonephone Sensor Probe with cable and plug Mains Battery charger 100 250v with AC charging lead D Ring mounting system Operating instructions oe gt DX 200 DIVER HELD MAGNETOMETER User Manual The Aquascan DX 200 Diver Held Magnetometer is a precision built instrument using the latest Fluxgate magnetic sensors and has been designed for both ease and simplicity of use in the underwater environment featuring three sensitivity settings for use in pinpointing targets The unit is ruggedly built to withstand normal diving usage but to get the best trouble free operation and long life certain simple operation and maintenance procedures must be followed Read the instructions carefully before diving with your DX 200 DX 200 DIVER HELD MAGNETOMETER User Manual Initial Testing of your DX 200 Note It is highly recommended that you familiarize yourself with the DX 200 on land before taking it into the water This section explains the basic functionality of the DX 200 and how to use the controls It is much easier to learn to operate the unit without the challenges of limited visibility restrictiv
11. eseseesseessseeserrserrsssrrssee GRP Probe Dia cats cates ncutecmicncc testi ewecenaaees 25mm Probe Leng aioe cerespierates cssiacadescceveancevers 750mm CABLE Construction Twin core shield Diamet t siossissrinra riii 4 5mm io ile Ad eee eee nee oe etn meee ene eae es Polyurethane TS BUD snd ses ceccstanasensasnevasadeetencsauheven 2m approx 6ft POWER Power Internal Rechargeable NiMH battery BoC een ee ee Universal 100 250v AC Charge time from empty s2sasasscodiecsussscceoseeas 14Hrs CONTROLS Power On Off Incorporated with Sensitivity Sensitivity Switch we Off Low Med High Set AMG IG rerne Multi turn Pot WEIGHTS Control Tit eseese e 1Kg Probe amp AIG cecin tesco acetrasvoietngenta aceasta 0 7Kg C3Smm i Dimensions Case amp Sensor 1Q0rmm E ee Sn 1 Basic outline of Case amp Sensor Stem DX 200 DIVER HELD MAGNETOMETER User Manual 4quascan International Limited DX 200 ACCESSORIES AQ 4 061 DX 200 Bonephone Fully waterproof headphone that utilises an Ikelite underwater connector to connect to the DX 200 Control Module The low profile module is normally placed against the skull below or behind the ear whilst underwater enabling the diver to clearly hear the change in the Detector tone when metal is sensed AQ 4 062 DX 200 Bonephone with extending flex cable This version is in all other respects the same as the AQ 4 061 The option o
12. f a flexible cable that will extend and retract is preferred by some divers particularly in low visibility conditions to minimise the risk of the cable fouling any obstructions AQ 4 030 DX 200 Land Headphones Rugged headphones optional accessory for use above water enabling The DX 200 to be utilised as a powerful ferrous metal land detector The Headphones feature an adjustable volume control and a comfortable fit for the user AQ 4 031 DX 200 Underwater Headphones Rugged headphones for use underwater giving an alternative to the Bonephone These phones are particularly useful in warm waters where a diving hood is not normally worn They also provide a more powerful sound level that can enhance the ability to sense even the most subtle changes in the audible output DX 200 DIVER HELD MAGNETOMETER User Manual AQ 4 010 Battery Charger Switched mode unit with 100v to 250v AC 50 60HZ input voltage range this unit is used to recharge the internal battery pack of the DX 200 Control Module The charger can be supplied with option of US UK or European type Power Connector in the form of a push fit shoe or as a conventional power lead DX 200 Control unit with standard D Ring harness DX 200 DIVER HELD MAGNETOMETER User Manual x ESk nigi j Optional DX 200 Armsaver probe extended DX 200 Control Unit mounted in optional Armsaver Close up of DX 200 Armsaver Probe connection 17
13. he SENSOR socket and screw down the retaining collar This should be good finger tightness only N B DO NOT EVER USE PLIERS OR OTHER TOOLS TO TIGHTEN THE PLUG COLLARS THESE ARE DESIGNED TO BE USED FINGER TIGHT WITH THE BLACK RUBBER BULKHEAD SEAL IN PLACE ON THE PLUG PINS OVER TIGHTENING OR INSTALLING THE PLUG WITHOUT THE RUBBER SEAL CAN PERMANENTLY DAMAGE THE SOCKET WHICH MAY CAUSE THE CONTROL UNIT TO FLOOD Note This plug can only be inserted one way round due to the polarising pin 4 Plug the connector on the Bonephone or headphone lead into the PHONE amp CHARGE socket and screw down the collar as above NB For testing on land the headphones if available are much easier to work with 5 Locate yourself to the centre of the testing area hold the Sensor Probe at full arm s length from your body pointing down at about 45 degrees and switch the unit on by rotating the POWER SENSITIVITY control to the LO position If a continuous sound is heard immediately the SET AUDIO control should be rotated one way or the other until the audio threshold is reached this is indicated by a low frequency ticking sound in the Bonephone Headphones There is a centre point where you will achieve the low frequency ticking sound and turning the SET AUDIO control either way from this point will result in an increase in the audio output Note The SET AUDIO control is a 10 turn potentiometer Do not turn it past its stop point in eit
14. her direction or permanent damage will result The audio threshold should be somewhere in the middle of the 10 turns so just start turning one way until you either reach the audio threshold or the stop point of the control If you reach the stop point without finding the audio threshold start turning the other way until you do Once you have found the audio threshold the first time you test the unit the SET AUDIO control should be in the near vicinity of the correct location for future uses so you will normally just need to tune it back and forth a few turns unless someone has turned it away from this setting DX 200 DIVER HELD MAGNETOMETER User Manual 6 Once you achieve the audio threshold adjustment turn the POWER SENSITIVITY control to HI You will probably have to make minor adjustments to the threshold setting when you switch to HI Once you have the threshold set to the low frequency ticking you are ready to begin testing Test the operation of the DX 200 by turning slowly in a circle with the Sensor Probe pointed down at about 45 degrees As the Sensor Probe is pointing in the direction of a target you should hear an increase in the audio output If you make a full circle without noticing any change in the audio output move 5 metres closer to your target and complete the circle test again 7 Once you are able to detect an increase in the audio signal frequency or pitch you should be able to move the Sensor Probe back and forth
15. le after less than what you the operator determine to be an acceptable minimal operational period then the DX 200 Battery Pack needs to be replaced Note If your battery needs to be replaced please contact our technical support dept to arrange for an installation of a replacement battery or in the case of an approved service agent please refer to the servicing data provided Please consider NiMH batteries as hazardous material and dispose of your old battery pack in line with local guidelines DX 200 DIVER HELD MAGNETOMETER User Manual Troubleshooting Guide Symptom No Response from DX 200 Module Check the following l Check that the battery has been charged as the voltage drops below a certain point erratic behaviour will result Further decrease in voltage will result in total lack of functionality gt If not place DX 200 on charge and then re test 2 Check that both Sensor Probe and Bonephone are correctly connected 3 Try using an alternative Bonephone AQ1B Headphones 4 Try using an alternative Sensor Probe if available 5 Check that both the Sensor Probe and Bonephone are in good condition Check by substitution if possible or by monitoring the nominal resistance of Bonephone Headphones Connecting a good quality multi meter across the pins of the plug can check the resistance NB It is good policy to firstly check the multi meter s reading with the leads shorted together this provides a measure of what
16. of underwater detection excellence using Proton Magnetometers and Pulse Induction Metal Detectors Aquapulse The newest addition to the AQUASCAN line of underwater detection equipment is the AQUASCAN DX 200 DIVER HELD MAGNETOMETER The DX 200 is a hand held gradient magnetometer utilizing fluxgate technology It is ideally suited for an array of search applications both on land or underwater Applications Wreckage The DX 200 is the ideal tool for the location of buried ferrous wreck material There are unlimited applications in all types of archaeological wreck investigation It is the ideal diver operated tool to locate items buried deep in sediment It allows accurate mapping of all buried ferrous material prior to seabed disturbance It can also detect the presence of some forms of ballast material which in many cases gives a magnetic response Pipes and Cables The DX 200 is also ideal for locating and mapping pipes and armoured cables buried in sediment It can detect at burial depths of between 1 metre and 7 metres dependent on the diameter of the pipe or cable E O D amp Security There are unlimited applications in all types of underwater investigation for the presence of ordinance weapons etc Salvage The DX 200 is a useful tool for the successful recovery of lost items either in low visibility or where lost in soft sediment It is the ideal tool to complement surface operated searching The tool most requested
17. ped to any suitable location on the diver with the D Ring mounting system provided with the unit Other options include the quick release belt harness strapping it to the upper arm or mounting it to the DX 200 Armsaver optional accessory see section DX 200 Optional Accessories 2 The Bonephone should be placed under the hood If no hood is worn it can be placed under the strap of the facemask or under a suitable neoprene headband Position the Bonephone so that it is on a bony part of the head close to the ear N B NEVER PLACE THE PHONE DIRECTLY OVER THE EAR AS THIS COULD INTERFERE WITH PRESSURE EQUALISATION ON THE EARDRUM DX 200 DIVER HELD MAGNETOMETER User Manual 3 Dive to the search area to be searched 4 Switch the unit on and rotate the SENSITIVITY control to HI Hold the Sensor Probe at full arm s length in front of you and pointing down at about 45 degrees Rotate the SET AUDIO control to achieve the audio threshold point keeping the Sensor Probe at arms length and angled down at about 45 degrees 5 If you are trying to pinpoint a hit from a towed magnetometer the standard procedure is to start scanning around in a complete circle If you detect even the smallest change in audio in any given direction move in that direction and repeat the scan Just as in the Land Test above you should be able to guide yourself right to the target by following the Sensor Probe and adjusting the SENSITIVITY control down to ME
18. residual reading to take into account gt See DX 200 Resistance Chart below item Resistance ohms Bonephone 5 4 Land Headphones 15 0 Minimum Volume U W Headphones 10 5 6 There is no user test for the Sensor Probe other than visual mechanical inspection Inspect the connector and cable looking for any signs of damage that could allow ingress of water into the cable or connector If damage is found contact your Aquascan agent for servicing replacement Symptom Erratic Performance from DX 200 Module 1 First check as above 2 Check that the contacts of the bulkhead connectors are clean 3 Check that the connectors of the Sensor Probe and Bonephone are clean If necessary reduce the contact resistance by slightly prising the gap in the male contacts this will increase the firmness of the connector on insertion 12 DX 200 DIVER HELD MAGNETOMETER User Manual 4 Check for visible damage to Sensor Probe cable and connector which could have caused water ingress 5 Check that the male connector seals are in place and clean 6 If you are unable to determine the cause of the erratic behaviour contact your Aquascan agent for assistance DX 200 DIVER HELD MAGNETOMETER User Manual DX 200 SPECIFICATIONS MATERIALS amp DIMENSIONS INSTRUMENT CASE Housing material eee Machined uPVC Housing Di r ceisio tsian 100mm Oyerall length ssnin 187mm MAGNETOMETER PROBE Probe housing esees
19. t line searching and use of the range switches to reduce the sensitivity where necessary to determine the true centre of the response will pay dividends in terms of diver bottom time and in focussing any following excavation in the most likely productive area General comments The DX 200 is NOT a metal detector It detects localised disturbances in the earth s magnetic field that are caused by the presence of ferrous material The user should note that other things can cause magnetic disturbances that will affect the DX 200 operation These include electromagnetic fields produced by electrical and electronic devices Working around large industrial areas or under transformers and high tension power lines will generally give the user problems with false readings In the water it is just important to be clear of any influence that may be generated by a large steel vessel and of course equipment on the diver such as a steel dive cylinder If you are searching for small items of debris such as a cannonball a pistol a hand tool you will probably be more successful utilising a metal detector such as the Aquascan AQIB The DX 200 is more suited towards larger items which will cause a larger magnetic deviation The DX 200 is ideal for commercially searching for buried UXO pipes cable and large ferrous debris For archaeologists and wreck hunters it is ideal for locating cannons anchors chains and other large ferrous items that can confirm the pres
20. ze petroleum products such as Vaseline since these will harden the rubber seal making it non functional The rubber seal should be replaced any time it becomes stiff as this will interfere with its function Note A dirty or corroded connection is the most common cause of poor performance of underwater detection equipment Keep your connectors clean and dry between uses DX 200 DIVER HELD MAGNETOMETER User Manual Dry the DX 200 paying particular attention to the area around the plugs Clear any excess water present on the bulkhead by blowing or wiping with a soft cloth Recharge the batteries if the total hours used is approaching 10 Never leave your DX 200 lying in the hot sun as this could raise the internal temperature to the point where it may distort the case and harm the electronics This warning also applies to the Sensor Probe and headphones Bonephone The electronic unit is sealed and pressure tested in the factory no attempt should be made to access the electronics If any servicing is required contact your supplier who will advise you where to send the unit Battery Care amp Checking Battery Performance Status The DX 200 internal battery pack is an 8 cell series NiMH AA size pack giving a capacity of approximately 2800mA Hrs This capacity relates to an operational period of between 10 and 12 hours from full charge Note self discharge occurs naturally in a rechargeable battery During inactivity
Download Pdf Manuals
Related Search
Related Contents
DeLOCK Notebook Car Power Supply User`s Manual 08.97 8-Bit CMOS Microcontroller http://www.siemens 型式:TR-406HH 取扱説明書 スタビリンカーブッシュツール Service Manual_4001351 Rev B MSM1036 Ma 1507-29.indd White Rodgers 1F82-0261 Thermostat User Manual Manual del Usuario BBQ Pro Vertical Lp Gas Smoker-- GENERAL INFO User Manual for EGO VIII - Max-Planck Copyright © All rights reserved.