Home
        Eclipse Manual 7x10
         Contents
1.       Oxygen Concentration    90      3  for all flow settings       Oxygen Concentration Indicator    Green Light   Normal Operation  Yellow Light   Warning or Caution  less than 85      3   Red Light Flashing   Abnormal Operation  less than 70      5        Maximum System Pressure    15 psig  103 5 kPa        Oxygen Outlet Pressure    5 0 psig  34 5 kPa  nominal       Electrical Power   AC Operation  DC Operation  Power Cartridge  AC Power Supply  DC Power Supply    100   240 VAC  50 60 Hz  12V   195 W Hrs   200 Watts Max   150 Watts Max       Operating Temperature    50   F to 104   F  10   C to 40   C        Operating Humidity    10    95  at 82 4   F  28   C  dew point       Nominal Sound Level    48 dBA at 3 0 LPM in Continuous Flow Mode  40 dBA at 3 0 LPM in Pulse Flow Mode          Dimensions  H x W x D  19 3 inches x 12 3 inches x 7 1 inches   49 0 cm x 31 2 cm x 18 0 cm   Weight 17 9 pounds  8 1 kg  with Power Cartridge    14 5 pounds  6 6 kg  without Power Cartridge       Continuous Flow Indication    Expressed in liters per minute  LPM        Audible Alarm Indicators    Back Up Alarm Power    Loss of power Hot Power Cartridge  Low Battery Warm Power Cartridge  Low Oxygen Output   O  flow outside normal limits   No inspiration detected in Pulse Mode  Unit Malfunction    9V Internal Battery       Filters    Device Classification    Air Inlet  HEPA  IEC Class    Type B Applied Part  IPXO       Transport Storage Requirements    Temperature   4   F   20  C  to  140   F
2.      Contact your dealer provider  Check tubing or replace    Clean Air Inlet Filter       Humidifier    Check humidifier attachment  and tubing       ECLIPSE not ON    Turn ECLIPSE ON       Tubing or cannula is not properly  connected or kinked       Power Cartridge is discharged  or no external power is present    Check tubing  cannula and  connections    Install charged Power Cartridge  or attach external power       Malfunction    Restriction in tubing or humidifier    Contact your dealer provider    Repair or replace tubing  or humidifier          Filter restricted    Malfunction    Restriction in humidifier or tubing    Clean or replace air inlet filter  Contact your dealer provider    Repair or replace humidifier  or tubing       Filter Blocked    Clean or replace air inlet filter       Malfunction    Tubing cannula longer than  7 feet  2 1m     Contact your dealer provider    Attach 7 foot  2 1m   tubing cannula       Humidifier attached    Remove humidifier          No inspiration detected    Power Cartridge needs  calibration    Contact your dealer provider    Fully discharge the Power  Cartridge and recharge    ECLIPSE    Model 1000 User Manual   31    OTER    Accessories          Use of certain oxygen administration accessories not specified for  use with this oxygen system may impair the performance  ONLY use  recommended oxygen delivery accessories           ECLIPSE    Model 1000  Cart           7  lt           DC Power Supply  Model 6105    AC Power Supply    M
3.    60   C   Humidity  up to 95  Non Condensing       Back pressure Effect on Flow Indicator    None       Altitude       0  13 123 feet  0   4 000 meters        38   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual          Pulse Flow Mode       Principle of Operation    Pulse Flow Mode is designed to deliver 16 ml boluses of oxygen  per incremental flow setting for each patient inspiration detected   Bolus volume is held constant regardless of breath rate within   a given setting  Delivery of the bolus is triggered by detection   of the end of inspiration followed by the start of a subsequent  inspiration  Sensitivity is dealer provider adjustable  Based on  this inspiration trigger signal  a    pulse    of oxygen is released to  the patient by an electronically controlled flow valve  There is   no mixing of oxygen with room air        Range of Operation     Settings    Increments of 1  6 times  x  the bolus of 16 ml   16 to 96 ml per breath        Accuracy of Delivered Bolus      15        Breathing Frequency range    6 to 25 breaths per minute average  sustainable       Minimum time between breaths    1 25 seconds  maximum 3 consecutive breaths        Response to missing breaths       ECLIPSE switches to Continuous Flow Mode after 30 seconds  without breath detection          Specifications are subject to change without notice     ECLIPSE    Model 1000 User Manual   39    OTER    Summary of All Warnings and Cautions    FEDERAL LAW RESTRICTS THIS DEVICE TO SALE B
4.    turn the system back ON to resume oxygen production     24   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Operating Your ECLIPSE    STEP 1  LOCATING YOUR ECLIPSE  m Place the ECLIPSE in a well ventilated place so that the air inlet and exhaust  have clear access     m Position the ECLIPSE so that any alarm can be heard       Check to see if the air inlet filter is in place before using  If the air inlet filter needs  to be cleaned  wash with soapy water and allow drying before using  If necessary   replace with a new air inlet filter     m Plug unit into AC Power  DC Power  or make sure there is a charged Power  Cartridge installed        ECLIPSE    Model 1000 User Manual   25    OTER    Operating Instructions  continued        STEP 2  TURN ON THE UNIT AND ALLOW TO WARM UP    m Press and hold the    ON OFF    Button for one  1  second to turn ON      your ECLIPSE  A brief self test will happen  Verify that all the indicators  are lit and a 3 second buzzer sounds  indicating the system is in  working order       Warm up Period    Oxygen Monitoring   The ECLIPSE Oxygen System has an oxygen   monitoring system with a yellow  alert  and red  alarm  visual indicator and  buzzer  The buzzer will activate in the event of an internal malfunction or power  outage  external power loss with no Power Cartridge installed   The red   yellow  and green  all OK  indicators will initially light up and then the red  and yellow indicators sequentially go out when the oxy
5.  ALERT  Yellow  Indicator   Low and Medium Priority Alerts     When steadily illuminated  this indicates a low priority awareness  J condition or Caution  Continue to use your system and refer to the    ECLIPSE    Model 1000 User Manual   13    OTER    Understanding the Controls and Front Panel  continued        Troubleshooting section for the proper response  A flashing yellow  indicates a medium priority alert  Prompt response is needed to deal  with a condition  Refer to the Troubleshooting section     ALARM  Red  Indicator   High Priority Alarms    Indicates a high priority alarm condition  Immediate response is  necessary to deal with the condition  Refer to the Troubleshooting  section and contact your dealer provider     Buzzer  An audible alarm  or buzzer  is used to alert you of the  condition of the unit  either a warning or failure  and to confirm a valid  key press     Flow Rate Setting Indicator  Green   This is the main focus on your  control panel  Your dealer provider will set up your prescribed flow  rates for your Continuous Flow Mode  LPM  and your Pulse Flow  settings  if pulse flow has been prescribed  DO NOT change these  settings unless it is under the order of your physician  Each time you  turn the unit on  the last flow mode and flow setting will be used     Power Cartridge Status Gauge  This indicator displays the charge  remaining in the Power Cartridge  When the Power Cartridge is fully  charged all 10 bars are illuminated  Each bar represents app
6.  ECLIPSE     m Store the ECLIPSE unit in a cool  dry location     ECLIPSE    Model 1000 User Manual   29       Alarm Conditions and Troubleshooting       The table shows possible alarm conditions and suggests appropriate responses  If any other alarms occur   call your authorized SeQual dealer provider                             Green Yellow Red Audible  Condition Indicator Indicator Indicator Alarm What to do  uw      Warming Up  Please Wait    On None Wait  The system typically takes  a 5 3 5 minutes to reach specified   lt 2  GE performance   aA    5    2 0     or flow setting  Off None Nothing  The system is operating  o properly at the specified flow rate      Low or Warm Power Blinking   Off Beeps Plug into external power or replace  Cartridge    with a charged Power Cartridge      O2 Concentration  lt  85     Blinking   Off None Continue Use  Check or clean air  intake filter  If condition persists   contact dealer provider      O  Concentration  lt  70     Blinking   Beeps   Continue Use  Check or clean air  intake filter  If condition persists   z       contact dealer provider   E  Z      Flow Rate Error  Blinking   Off Beeps   Check tubing and or humidifier    Blocked Flow    for obstruction  Check clean  F air intake filter  If condition  a persists past 10 minutes   H contact dealer provider        No Inspiration Detected    Off Beeps Make sure tubing does not   gt     in Pulse Flow Mode only  exceed 7 ft  2 1m  when using  Pulse Flow Mode  Audible alarm  will be
7.  System and its accessories  Thoroughly  read all the information in this manual in its  entirety before operating your ECLIPSE   Contact your dealer provider for proper  training in the use of this device     YOUR PHYSICIAN HAS PRESCRIBED  SUPPLEMENTAL OXYGEN AS PART  OF A TREATMENT PLAN  LONG TERM  OXYGEN THERAPY IS NOW WIDELY  ACCEPTED AS THE STANDARD  TREATMENT FOR CHRONIC HYPOXIA  CAUSED BY CHRONIC OBSTRUCTIVE  PULMONARY DISEASE  OXYGEN  THERAPY HAS BECOME MORE  VERSATILE IN RECENT YEARS DUE  TO THE DEVELOPMENT OF OXYGEN CONCENTRATORS  THE ECLIPSE MAKES  IT EASIER FOR PHYSICIANS TO SELECT A MODE OF OXYGEN THERAPY THAT  SUITS THE PATIENT   S LIFESTYLE        THE ECLIPSE MAY BE USED IN HOME  INSTITUTION  AND MOTOR VEHICLES     USE OF THE ECLIPSE IN HOSPITALS SHOULD INCLUDE THE USE OFA     HOSPITAL GRADE    POWER CORD     YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE  SETTING TO MEET YOUR INDIVIDUAL NEEDS  THIS FLOW RATE IS DETERMINED  BY THE AMOUNT OF OXYGEN YOU NEED  RECOMMENDED OXYGEN FLOW  RATES SHOULD BE ADJUSTED ONLY UNDER THE ADVICE OF A PHYSICIAN     10   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Understanding the Controls and Front Panel       Description of the ECLIPSE    The key features of the ECLIPSE Oxygen System and control panel are important  items to become familiar with     Control Panel    Oxygen Outlet Port    Service Port    Simplified User  Not for Patient Use   Instructions   Under Base  Air Inlet Filter    Externa
8.  and  make sure the air inlet and exhaust airflow are not blocked  This may  impact performance             WARNING   DO NOT leave the ECLIPSE or DC Power Supply plugged into the  vehicle without the engine running or attempt to start the vehicle  while the DC Power Supply is connected to the vehicle  This may  drain the vehicle s battery              A buzzer in the DC Power Supply is intended to warn user when low vehicle power is  present  Immediately disconnect the DC Power Supply from the vehicle for low voltage  alarm conditions        If the vehicle power source  vehicle accessory connector  drops below  11 5 Volts  brownout  the ECLIPSE will revert to Power Cartridge  operation  if present   The External Power Present Indicator may remain on           TYPICAL FLOW RATES AVAILABLE AT EACH FUSE SETTING                                     10 AMP 15 AMP   Setting Continuous Pulse  0 5    1 0 Yes  1 5 Yes  2 0 Yes  2 5 Yes  3 0 Yes  3 5 Yes  4 0 Yes  4 5 Yes  5 0 Yes  5 5 Yes  6 0 Yes          Actual performance may vary dependent upon usage conditions and vehicle     20   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       When operating in the 10 AMP switch position  flow rate settings above 2 0 LPM  continuous flow will automatically switch to 2 0 LPM continuous flow when the DC  Power Supply is connected      3  RECHARGEABLE POWER CARTRIDGE  The ECLIPSE system can also be  powered by a rechargeable Power Cartridge     Installing the Power Cartridge  Align
9.  and eyes  Keep the Power Cartridge away from children           ECLIPSE    Model 1000 User Manual   41    OTE         WARNING   Use of a damaged Power Cartridge may cause personal injury       WARNING   Exposing the Power Cartridge to water or other liquids may cause  personal injury                 ONJ Protect the ECLIPSE and AC DC Power Supplies from fluid spills or  drips to avoid possible shock hazards     DO NOT use any cord if damaged  Improper use of the Power Cord  may result in a fire or electrical shock hazard     DO NOT touch the Power Cartridge cavity connectors with the Power  Cartridge removed while the unit is connected to an external power  source  This may result in electrical shock                         WARNING   Secure the ECLIPSE and the DC Power Supply in your vehicle and  make sure the air inlet and exhaust airflow are not blocked  This may  impact performance              DO NOT leave the ECLIPSE or DC Power Supply plugged into the  vehicle without the engine running or attempt to start the vehicle while  the DC Power Supply is connected to the vehicle  This may drain the  vehicle   s battery           If the vehicle power source  vehicle accessory connector  drops below  11 5 Volts  brownout  the ECLIPSE will revert to Power Cartridge  operation  if present   The External Power Present Indicator may  remain on             WARNING   When operating in Continuous Flow Mode  to ensure adequate flow   DO NOT use a tubing cannula extension more than 50   
10.  in a small  enclosed space  such as a  bathroom or closet  with cannula leading out of the enclosed space                 8   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Power Cartridge Safety      WARNING   This Power Cartridge may explode and cause potential injury  if exposed to or disposed of in a fire              DO NOT short circuit the Power Cartridge   s metal contacts with  metallic objects  such as keys or coins  It may cause sparks or  excessive heat              DO NOT disassemble  puncture  or crush the Power Cartridge  Power  Cartridge electrolytes may be toxic if swallowed and can be harmful  to skin and eyes  Keep the Power Cartridge away from children           Use of a damaged Power Cartridge may cause personal injury             WARNING   Exposing the Power Cartridge to water or other liquids may cause  personal injury           CAUTION   DO NOT expose the Power Cartridge to temperatures above  140  F  60  C   such as in a vehicle parked in the sun or on a hot day              CAUTION   Charging the Power Cartridge below 50 F    10   C  or above  104   F  40   C  may degrade the performance              CAUTION   DO NOT drop the Power Cartridge or expose it to mechanical shock                 CAUTION   Only use the Power Cartridge for its intended purpose        ECLIPSE    Model 1000 User Manual   9    E    Introduction to Your ECLIPSE Oxygen System       This User Manual will inform you about the  use and care of the ECLIPSE Oxygen 
11.  quilt  etc       DONOT overfill the humidifier  if used  Fill with water only to the level shown  by the manufacturer     m When using ECLIPSE in a vehicle  make sure it is properly secured or restrained     ECLIPSE    Model 1000 User Manual   7    OTER    Safety Guidelines  continued        Operating Safety  continued       WARNING   DO NOT remove the cover  no serviceable parts inside  Qualified  service personnel should only remove the cover                WARNING   DO NOT operate device or accessories on wet surfaces or in standing  water and DO NOT submerse or expose to water  The ECLIPSE  enclosure does not provide protection against the harmful effects of  liquid ingress  Electrical shock or damage to the unit may result             WARNING   Protect electrical cords from sharp edges  Electrical shock and serious  physical injury may result             WARNING   Use only the SeQual AC Power Supply and DC Power Supply supplied  with your ECLIPSE Oxygen System  Use of a non SeQual Power    Supply may be hazardous  cause damage to your oxygen system  and will void the warranty           CAUTION   Availability of an alternate source of oxygen is recommended in  case of power outage or mechanical failure              CAUTION   DONOTexpose unit to temperature outside of specified operating  or storage temperatures as this may damage the unit              CAUTION   Locate oxygen tubing and Power Supply cords to prevent tripping  hazards           CAUTION DO NOT place ECLIPSE
12.  the Power Cartridge so that it will slide into the  Power Compartment  Push the Power Cartridge into the Power Compartment until it  is secure and flush with the back of the system  When the Power Cartridge is properly  inserted and the ECLIPSE is ON  the Power Cartridge Status Gauge will appear on  the Control Panel   If you DO NOT have a Power Cartridge installed  the Status Gauge  will not light up      To remove the Power Cartridge  push up on the latch on the Power Cartridge   The Power Cartridge will slide out so that you can grasp it for removal     Power Cartridge Capacity  The capacity of the Power Cartridge is lit up on the control  panel  A variety of factors  such as flow rate and Pulse or Continuous Flow Mode  operation determine the duration of operating time  The following table provides  estimates of amount of time that the ECLIPSE can operate  on a fully charged Power Cartridge given certain  flow rates and operating conditions          Power Cartridge    ECLIPSE    Model 1000 User Manual   21    OTE    Operating Instructions  continued        TYPICAL NEW POWER CARTRIDGE OPERATION TIME                         Flow Rate Setting Continuous Flow Mode  LPM  Pulse Flow Mode  Setting   0 5 4 4 hours     1 0 3 7 hours 4 4 hours   2 0 2 4 hours 3 6 hours   3 0 1 3 hours 3 0 hours   4 0 a 2 6 hours   5 0   2 3 hours   6 0    2 1 hours          Operating time will degrade with Power Cartridge use and age        CAUTION   Store your Power Cartridge in a cool  dry locat
13.  you operate and maintain the unit  If you have any further questions or problems  call your SeQual dealer provider or authorized service representative     Important  DO NOT attempt to operate the ECLIPSE System without first reading  the    Safety Guidelines    section of this manual  Follow all operating  instructions  Observe all warnings on the product and in the operating  instructions  To reduce the risk of fire  bodily injuries  and damage to  the oxygen system  observe all safety precautions        CON Protect the ECLIPSE and AC DC Power Supplies from fluid spills or  drips to avoid possible shock hazards              CAUTION   Always check to see that the Air Inlet and the Exhaust Vent are not  blocked and the air inlet filter is clean before using your system                     Air Inlet and    Exhaust Vent Air Inlet Filter       ECLIPSE    Model 1000 User Manual   17    Ls    Operating Instructions  continued        Powering Your ECLIPSE System    The ECLIPSE is a lightweight  portable oxygen system capable of being operated  directly from three different power sources   1  an AC Power Supply   2  a DC Power  Supply  or  3  a rechargeable Power Cartridge     3  z     3  RECHARGEABLE  POWER CARTRIDGE        1  AC POWER SUPPLY     2  DC POWER SUPPLY    The ECLIPSE charges the optional Power Cartridge when external power  AC or DC   is present and when the Power Cartridge temperature is below a safe charging  temperature  When external power is disconnected  the sys
14. 15 2m  long              42   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual            WARNING   Operating your ECLIPSE outside the recommended operating  environment may impact performance  cause damage and will void  the warranty           Pulse Flow settings should be determined for each patient individually   Settings from Continuous Flow applications may not be applicable to  Pulse Flow Mode           When operating in Pulse Flow Mode  to ensure adequate breath  detection  the length of the tubing cannula must not exceed 7 feet   2 1m        WARNING   As with all conserving devices  the ECLIPSE may not be able to detect    some respiratory efforts                WARNING   DO NOT change from prescribed flow rate setting without consulting  with a qualified clinician              AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS  RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL  FAILURE  CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK UP  SYSTEM REQUIRED  DO NOT IGNORE ALARMS                WARNING   Use of certain oxygen administration accessories not specified for  use with this oxygen system may impair the performance  ONLY use  recommended oxygen delivery accessories                 EZOM The tube fitting may be tight  DO NOT use oil or grease to lubricate        ECLIPSE    Model 1000 User Manual   43    OTE       If tubing becomes disconnected during operation  ECLIPSE does  not alarm  Check flow at the cannula to verify oxygen delivery   DO NOT tamper with  
15. ESENCE OF FLAMMABLE ANESTHETICS           CON AS WITH ANY ELECTRICALLY POWERED DEVICE  THE USER MAY  EXPERIENCE PERIODS OF NON OPERATION AS THE RESULT OF  ELECTRICAL POWER INTERRUPTION  OR THE NEED TO HAVE THE  OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN  AN OXYGEN  CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO  WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE  RESULT OF SUCH TEMPORARY INTERRUPTION     ECLIPSE    Model 1000 User Manual   5             OTER    Safety Guidelines       This section of the manual provides warnings and cautions that apply to hazards or  unsafe practices that could result in personal injury or property damage     Oxygen Therapy Safety       Read Instruction Manual     DO NOT operate this oxygen system   without first reading and understanding this manual  If you are unable  to understand the warnings and instructions  contact your authorized  SeQual dealer provider before attempting to use this oxygen system     be hazardous  therefore  it is advisable to seek medical advice before  using the oxygen system     i    Caution or Attention     In certain circumstances  oxygen therapy can    DO NOT deviate from prescribed flow rates  Adjust the flow only to your prescribed  flow rate  unless otherwise instructed by your physician          WARNING      No Smoking or Naked Flames     The use of oxygen therapy requires  that special care be taken to reduce the risk of fire  Oxygen enhances  S the combustion of flammable material  For safety conce
16. N   DO NOT connect the ECLIPSE to an extension cord or electrical  outlet controlled by a switch           CAUTION   Store your Power Cartridge in a cool  dry location  This will help to  assure the longevity of your Power Cartridge              CAUTION   DOT UN Regulations require the removal of the Power Cartridge  for international travel              CAUTION   There are no serviceable parts inside the Power Cartridge   DO NOT open              CAUTION   Replace the Power Cartridge with only SeQual approved Power  Cartridge                 CAUTION   The Power Cartridge used in this device may present a risk of fire or  chemical burn if mistreated  DO NOT disassemble  incinerate  or heat  above 140   F  60   C                  CAUTION   Return your Power Cartridge to your dealer provider for proper  disposal                 46   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual          CAUTION   Keep your Power Cartridge away from children           CAUTION   Store the Power Cartridge in a cool  dry place when not in use           CAUTION   DONOT leave your ECLIPSE or your Power Cartridge in your vehicle  or in a trunk during a hot day              CAUTION   The ECLIPSE system can only work with a SeQual Power Cartridge   Use of another Power Cartridge or Battery may damage the unit   present a risk of fire or explosion and will void the warranty                    CAUTION   To ensure there is adequate flow to deliver oxygen  the length of the  tubing must not ex
17. SE outside the recommended operating    environment may impact performance  cause damage and will void  the warranty           CAUTION   Locate the unit in a well ventilated room that provides adequate air  flow and check if air is being prevented from entering or exiting the  system by furniture  draperies  bedding  clothing or pets                    ECLIPSE    Model 1000 User Manual   15     as    Recommended Operating Environments  continued        The Proper Location    Select a location that avoids fumes and pollutants and allows the oxygen system to  draw in room air through the air inlet filter on the back of the cabinet and exhaust air  out the vent at the bottom of the case without being restricted     Locate the unit such that the alarms can be heard   Route oxygen tubing such that it does not kink or is subject to occlusion     DO NOT cover unit with a blanket  towel  quilt  or other covering  as the unit may  overheat     DO NOT locate near flammable materials or cleaning agents or in the direct path of  any heat source  such as a heat register or car heater     Keep the oxygen system at least five  5  feet  1 5 m  away from hot  sparking objects  or open sources of flame        16   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual    Operating Instructions       Before Operating    Your SeQual dealer provider or authorized service representative will demonstrate  how to operate the oxygen system  This manual serves as a supplemental reference to  help
18. SS SEQuaL    The Smart Choice        Model 1000 User Manual       Y  e  S 4  9  be        Contents    TEOL ETETEN                                                                                     Important Safety Instructions 2  Symbols Used in User Manual and on Oxygen System 2  Indications For Use 5  Contraindications 5  Safety Guidelines 6  Oxygen Therapy Safety 6  Operating Safety 7  Power Cartridge Safety 9  Introduction To Your ECLIPSE Oxygen System 10  Understanding the Controls and Front Panel 11  Description of the ECLIPSE 11  Control Panel Features 12   User Controls 13   Front Panel Icons 13   System Status Indicators 13  Recommended Operating Environments 15  The Proper Location 16  Operating Instructions 17  Before Operating 17  Powering Your ECLIPSE System 18  Operating Your ECLIPSE System 25  Alarm Conditions and Troubleshooting 30  Accessories 32  Cleaning  Care and Routine Maintenance 35  Technical Data 38  Summary of All Warnings and Cautions 40  Electromagnetic Compatibility 48  Disposal of Equipment and Accessories 49       OTER    Important Safety Instructions       Important Safety Instructions are Shown  Under the Following Headings     SeQual   Technologies Inc   SeQual  recommends that this User Manual be given to  each patient receiving an ECLIPSE    Oxygen System  This User Manual describes all  the necessary information in order for a patient or caregiver to operate the ECLIPSE          WARNING   Important safety information for hazards that migh
19. Supply or the DC Power Supply  if adequate power is available  The Power  Cartridge will take longer to recharge when connected to the DC Power Supply and  will depend on the flow setting and the amperage switch setting     Power Compartment  The Power Compartment is available when Power Cartridge  operation is not required  When using the Power Compartment  your ECLIPSE will    ONLY operate on the AC and DC Power Supplies   ECLIPSE    Model 1000 User Manual   33    OTER    Accessories  continued        AC Power Supply  Model 6104  Your AC Power Supply enables the ECLIPSE to be  connected to a grounded AC electrical outlet with the country specific power cord   Please contact your dealer provider for the country specific power cord     DC Power Supply  Model 6105  Your DC Power Supply enables the ECLIPSE to be  operated in an automobile or other vehicle with an accessory outlet  Contact your  dealer provider to set up the DC Power Supply for proper use in your own automobile     The indicator on the AC and DC Power Supplies will be either green or red  Green  indicates normal operation and red indicates abnormal operation     DO NOT expose the AC and DC Power Supplies to water  Electrical  shock or serious injury may result     Cart  Your ECLIPSE Oxygen System can be easily transported while attached to the  Cart  The cart has wheels and a telescoping handle for easy mobility              DO NOT drip  spill or expose the ECLIPSE to water during portable  use  The ECLIPSE encl
20. Y OR ON THE  ORDER OF A PHYSICIAN     THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE  SUSTAINING APPLICATIONS  NOR DOES IT PROVIDE ANY PATIENT  MONITORING CAPABILITIES                    IN CERTAIN CIRCUMSTANCES  THE USE OF NON PRESCRIBED  OXYGEN CAN BE HAZARDOUS  THIS DEVICE SHOULD ONLY  BE USED WHEN PRESCRIBED BY A PHYSICIAN     FOR ADULT USE ONLY  NOT QUALIFIED FOR USE ON  PEDIATRIC PATIENTS       WARNING   NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS                       AS WITH ANY ELECTRICALLY POWERED DEVICE  THE USER MAY  EXPERIENCE PERIODS OF NON OPERATION AS THE RESULT OF  ELECTRICAL POWER INTERRUPTION  OR THE NEED TO HAVE THE  OXYGEN CONCENTRATOR SERVICED BY A TECHNICIAN  AN OXYGEN  CONCENTRATOR IS NOT APPROPRIATE FOR ANY PATIENT WHO  WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE  RESULT OF SUCH TEMPORARY INTERRUPTION                 No Smoking or Naked Flames     The use of oxygen therapy requires  that special care be taken to reduce the risk of fire  Oxygen enhances  the combustion of flammable material  For safety concerns  all possible  sources of ignition must be kept away from the oxygen system and  preferably out of the room in which it is being used  Smoking in the  proximity of an operating oxygen concentrator is dangerous  and   can permanently damage the device and void the warranty              40   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual            WARNING      Use No Oil or Grease     A spontaneous and viole
21. ae will appear on the control panel        CAUTION   The AC Power Cord used with the ECLIPSE Oxygen System must  meet the requirements of the country where the product is used           CAUTION   Use with grounded  3 pin receptacle  DO NOT remove ground pin from  cord  Use only with SeQual supplied cords                 CAUTION DO NOT connect the ECLIPSE to an extension cord or electrical  outlet controlled by a switch                  2  DC POWER SUPPLY  A DC Power Supply allows the system to operate from  DC accessory outlets  such as those found in motor vehicles  You need to  determine the Fuse rating in your motor vehicle  either 15 AMP or 20 AMP   Consult your vehicle   s Operator s Manual or look at your fuse box in the vehicle   Set the switch on the DC Power Supply to the appropriate one  Use the 10 AMP  position on the DC Power Supply for a 15 AMP vehicle fuse and use the 15 AMP  position for a 20 AMP or greater vehicle fuse  Start your vehicle  Insert the DC  cord in the proper receptacle on the side of the ECLIPSE  Next  insert the DC  Power Supply plug into the DC accessory outlet in your vehicle  When the system  is properly plugged in and receiving power from the DC source  a green indicator  light on the Power Supply  and the External Power Present indicator  a  on the control panel will appear     ECLIPSE    Model 1000 User Manual   19    OTER    Operating Instructions  continued             WARNING   Secure the ECLIPSE and the DC Power Supply in your vehicle
22. ceed 50 feet  15 2m  when using Continuous Flow  Mode  and must not exceed 7 feet  2 1m  when operating in Pulse  Flow Mode for adequate breath detection              CAUTION   When operating the ECLIPSE on the cart  DO NOT bang  drop or  bump the unit while in transport                 CAUTION   The filter should be free of liquid water and dry to the touch before  reinstallation  as excess moisture may impair the proper operation  of the oxygen system              CAUTION   When the filter is removed  the Service Port is exposed  DO NOT  attach any cables to this Service Port  It is for Service ONLY                    CAUTION  DO NOT poke objects through the vent holes        ECLIPSE    Model 1000 User Manual   47       Electromagnetic Compatibility       This equipment has been tested and found to comply with the limits for medical  devices to the IEC60601 1 2  2001 09 Electromagnetic Compatibility standard   These limits are designed to provide reasonable protection against harmful  interference in a typical medical installation  This equipment generates  uses  and can  radiate radio frequency energy and  if not installed according with the instructions   may cause harmful interference to other devices in the vicinity  However  there is no  guarantee that interference will not occur in a particular installation  If this equipment  does cause harmful interference to other devices  which can be determined by turning  the equipment off and on  the user is encouraged to try t
23. d  clean it at least once a week     1  Remove the filter     2  Wash the filter in warm water  using a mild detergent solution     3  Rinse the filter thoroughly and  squeeze out the excess water     4  Allow the filter to air dry        5  Reinsert the filter in the cabinet     ECLIPSE    Model 1000 User Manual   35    OTE          CAUTION   The filter should be free of liquid water and dry to the touch before  reinstallation  as excess moisture may impair the proper operation of  the oxygen system                 CAUTION   When the filter is removed  the Service Port is exposed  DO NOT  attach any cables to this Service Port  It is for Service ONLY                 CAUTION   DO NOT poke objects through the vent holes              The air inlet filter should be replaced as needed  If the oxygen system is used in a  dusty environment  the filter may need to be replaced more often  You should not  operate the ECLIPSE without a filter installed for longer than 30 minutes     Cabinet  AC Power Supply  DC Power Supply and Power Cord  Turn OFF the  ECLIPSE and disconnect from AC or DC power before any cleaning or disinfection  activity  Use a damp  not soaking wet  cloth or sponge and a mild detergent solution  to clean the cabinet and power supplies  To disinfect the ECLIPSE use Lysol   Brand II  disinfectant  Proceed as directed by the manufacturer        Unplug Power Cords  AC or DC Power Supplies before cleaning the  exterior cabinet  DO NOT use denatured alcohol or apply liqu
24. disassemble  crush or heat the Power Cartridge    above 140   F  60   C   The Power Cartridge may present a risk of fire  or explosion and will void the warranty                   WARNING   DO NOT expose the AC and DC Power Supplies to water  Electrical  shock or serious injury may result                WARNING   DO NOT disassemble the ECLIPSE or attempt to perform any routine  maintenance other than what is described in the Patient Cleaning and  Care section below  Disassembling the oxygen system may create a  hazard of electrical shock and void your warranty  Contact your  dealer provider for any maintenance required              Unplug Power Cords  AC or DC Power Supplies before cleaning  the exterior cabinet  DO NOT use denatured alcohol or apply liquid  spray or aerosol cleaners             WARNING   DO NOT tamper with or try to repair the Power Cartridge  No  serviceable parts inside       WARNING   Exposing the Power Cartridge to water or other liquids may cause    personal injury                   WARNING   DO NOT drip  spill or expose the ECLIPSE to water during portable  use  The ECLIPSE enclosure does not protect against the harmful  effects of liquid ingress  Electrical shock or damage to the unit  may result              44   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       CAUTION   Availability of an alternate source of oxygen is recommended in case  of power outage or mechanical failure              CAUTION   DONOT expose unit to temperature o
25. ep after 15 seconds if no  inspiration is detected  If no  inspiration detected 15 seconds  after that  the ECLIPSE will  automatically switch to  Continuous Flow Mode   u      Power Cartridge Shutdown   Off Off Blinking   On Seek external power source   5 or hot Power Cartridge     with If Power Cartridge malfunction  E  no external power present  persists  contact dealer  provider   a  A    Loss of External Power    Off Off Blinking   On Seek an alternative oxygen        without charged Power supply until external power is  z 3 r i       Cartridge installed  restored or install a charged  2 Power Cartridge   5    System Malfunction    Off Off On On Contact dealer provider  Use  N                back up oxygen supply           AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS  RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL  FAILURE  CONSULT YOUR PHYSICIAN FOR THE TYPE OF  BACK UP SYSTEM REQUIRED  DO NOT IGNORE ALARMS           30   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual          Symptom    ECLIPSE does not power on  when ON OFF button is pressed    No Oxygen    Low Oxygen Concentration    Low Oxygen Flow    No Oxygen delivered in  Pulse Mode    Power Cartridge Status Gauge  never indicates fully charged    Probable Cause    No Power Cartridge installed    Remedy    Install charged Power Cartridge  or plug into external power       Power Cartridge is discharged    Plug into external power       Malfunction    Blocked or kinked tubing    Filter Blocked  
26. gen purity is within  specification     After initially turning ON the unit  allow up to five  5  minutes for the unit  to reach its specified performance  The control panel will indicate when  the unit is done warming up  When the green ON indicator is lit  the system  is ready for use     Green light Yellow light Red light   ready   alert   alarm                                   UUWUUWUUU0  4                26   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       STEP 3  CONNECT THE TUBING TO THE OXYGEN OUTLET AND THE CANNULA  PER THE INSTRUCTIONS FROM YOUR DEALER PROVIDER     m Replace the cannula and tubing on a regular basis  Consult your dealer provider  for cannula and tubing cleaning and replacement information        STEP 4  SELECT FLOW MODE  m Press the Flow Mode button to select the flow delivery mode       Continuous Flow Mode Operation  e When operating in Continuous Flow Mode  a continuous supply of oxygen  will flow through your tubing and nasal cannula     When operating in Continuous Flow Mode  to ensure adequate flow   DO NOT use a tubing cannula extension more than 50     15 2m  long              ECLIPSE    Model 1000 User Manual   27    OTE         Pulse Flow Mode Operation  e Pulse Flow Mode should only be used under the direction of a qualified  clinician or physician     e When operating in Pulse Flow Mode  an alarm will beep after 15 seconds  if inspiration is not detected  If another 15 seconds elapses and no  inspiration is detected  t
27. he unit will switch to Continuous Flow Mode  at the last Continuous Flow Mode rate setting used        Pulse Flow settings should be determined for each patient individually   Settings from Continuous Flow applications may not be applicable to  Pulse Flow Mode           When operating in Pulse Flow Mode  to ensure adequate breath  detection  the length of the tubing cannula must not exceed  7 feet  2 1m        WARNING   As with all conserving devices  the ECLIPSE may not be able to detect  some respiratory efforts   STEP 5  ADJUST FLOW RATE TO PRESCRIBED LEVEL    Using the Flow Rate Setting buttons  adjust the flow rate to  the prescribed setting supplied by your physician           DO NOT change from prescribed flow rate setting without consulting  with a qualified clinician or physician                       28   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       STEP 6  BEGIN USE  m Breathe normally through the cannula   m You will get longer operating time on your Power Cartridge if you operate your    ECLIPSE in Pulse Flow     STEP 7  CART OPERATION    m The Cart is designed for use on pedestrian surfaces   m Align the locator holes on the bottom of the unit to the tabs on the cart     m Align the screw to the insert on the back of the ECLIPSE  and hand tighten the knob on the cart       Push the button on the cart handle  to adjust the height        STEP 8  TURN OFF THE UNIT    m Press and hold the    ON OFF    Button for two  2  seconds to turn  OFF your
28. id spray  or aerosol cleaners           Humidifier  If used  clean your humidifier daily to reduce possibility of contamination   Replace humidifier monthly or as recommended by the manufacturer or your  dealer provider     Cart  Before cleaning the cart  remove the ECLIPSE  Use a damp  not soaking wet   cloth or sponge and mild detergent solution to clean the Cart     36   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Power Cartridge Cleaning  Care and Maintenance    The Power Cartridge in the ECLIPSE Oxygen System requires special care to  assure a longer life and higher performance  The SeQual Power Cartridge is the  only approved Power Cartridge to be used     Once a month  completely discharge your Power Cartridge and fully recharge  to calibrate the Power Cartridge     Use a damp cloth or sponge and mild detergent to clean the Power Cartridge  case and latch     DO NOT tamper with or try to repair the Power Cartridge   No serviceable parts inside     Power Cartridge Disposal  Your Power Cartridge is rechargeable and can  be recycled  Always return it to your dealer provider for proper disposal              Power Cartridge Storage  Your Power Cartridge should be stored in a cool   dry location     Exposing the Power Cartridge to water or other liquids may cause  personal injury              ECLIPSE    Model 1000 User Manual   37    Technical Data           Flow Rate Settings    1 to 6 setting in Pulse Flow Mode  0 5 to 3 0 LPM in Continuous Flow Mode 
29. ion  This will help to  assure the longevity of your Power Cartridge           CAUTION   DOT UN Regulations require the removal of the Power Cartridge  for international travel              CAUTION   There are no serviceable parts inside the Power Cartridge   DO NOT open              CAUTION   Replace the Power Cartridge with only SeQual approved Power  Cartridge                 CAUTION   The Power Cartridge used in this device may present a risk of fire  or chemical burn if mistreated  DO NOT disassemble  incinerate   or heat above 140   F  60   C               CAUTION   Return your Power Cartridge to your dealer provider for proper  disposal                    CAUTION   Keep your Power Cartridge away from children        22   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Typical Power Cartridge Recharge Time    The typical time to recharge your Power Cartridge to achieve 80  capacity from  a fully discharged Power Cartridge is 1 4 hours to 5 0 hours  dependent upon the  flow setting     If the Power Cartridge is too warm  charging will not begin until it sufficiently cools     The Power Cartridge may not charge or the charge time may increase when operating  from the DC Power Supply set in the 10 AMP position     INITIAL POWER CARTRIDGE OPERATION     The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully  charged  Before using your ECLIPSE Oxygen System for the first time  you must fully  discharge the Power Cartridge     1  Turn y
30. is dangerous  and can permanently damage the  device and void the warranty        Use No Oil or Grease     Indicates important information about the      SS possibility of spontaneous and violent ignition that may occur if oil   grease or other petroleum based substances come into contact with  oxygen under pressure       PXO    Drip Proof Equipment IPX0     The ECLIPSE enclosure does not  provide protection against the harmful effects of liquid ingress   IPXO   per IEC 60529       PX1    Drip Proof Equipment IPX1     The ECLIPSE AC and DC Power  Supplies provide protection against the harmful effects of liquid  ingress   IPX1  per IEC 60529     a    IEC Icon for DC Power  Input         Type B Equipment     Equipment providing a particular degree of  IK protection against electric shock  particularly regarding       Allowable LEAKAGE CURRENT     Reliability of the protective earth connection     A    Amperes       ECLIPSE    Model 1000 User Manual   3    OTE    Symbols  continued        Off   e QqDo Bb        Alternating Current          Audible Alarm     An audible indicator is provided to announce a  reduction in oxygen concentration or when the power has been  interrupted or lost        ON OFF     Turns the unit ON and OFF but does not directly  disconnect the mains power        Increase Flow Setting     Increases the flow setting in 0 5  increments each time the button is pushed        Decrease Flow Setting     Decreases the flow setting in 0 5  increments each time the b
31. l Power Cart Insert Screw          Receptacle  Exhaust Vent  Power Cartridge  Latch  Main Rating Label  FRONT  Under Base   Power Cartridge  BACK   INPUT OUTPUT CONNECTIONS   Oxygen Outlet Port Extension tubing or a cannula is attached to this port    Air Inlet Filter Ambient air is drawn into the unit through the air inlet located on the    top back of the unit  This filter prevents dust and debris from entering  the ECLIPSE and should be cleaned regularly     Service Port This connection allows service personnel to diagnose and monitor  ECLIPSE performance  It is not for patient use        External Power Receptacle Either the ECLIPSE AC or DC Power Supply output cords may be  connected to this receptacle        Exhaust Vent Exhaust air from the ECLIPSE is expelled from this vent        ECLIPSE    Model 1000 User Manual   11       Understanding the Controls and Front Panel  continued        Control Panel Features    The ECLIPSE displays important operating information using the icons  numbers and  visual indicators on the control panel  This section will help you understand these features     No Smoking Icon  ON OFF Button   ON OFF Indicator  Green   Alert Indicator  Yellow     Flow Rate Setting  2 Digits    Green     Alarm Indicator  Red     Increase Flow Rate Button                                                                                                                                           Decrease Flow Rate Button    Power Cartridge Status  Gauge  Green     Ex
32. nt ignition may  occur if oil  grease or other petroleum substances come into contact   SS  with oxygen under pressure  Keep these substances away from the  oxygen system  tubing and connections and any other oxygen source   DO NOT use any petroleum based or other lubricants             WARNING   DO NOT remove the cover  no serviceable parts inside  Qualified  service personnel should only remove the cover             WARNING   DO NOT operate device or accessories on wet surfaces or in standing  water and DO NOT submerse or expose to water  The ECLIPSE  enclosure does not provide protection against the harmful effects   of liquid ingress  Electrical shock or damage to the unit may result             WARNING   Protect electrical cords from sharp edges  Electrical shock and  serious physical injury may result             WARNING   Use only the SeQual AC Power Supply and DC Power Supply supplied  with your ECLIPSE Oxygen System  Use of a non SeQual Power    Supply may be hazardous  cause damage to your oxygen system  and will void the warranty             WARNING   This Power Cartridge may explode and cause potential injury if  exposed to or disposed of in a fire             WARNING   DO NOT short circuit the Power Cartridge   s metal contacts with  metallic objects  such as keys or coins  It may cause sparks or  excessive heat           DO NOT disassemble  puncture  or crush the Power Cartridge  Power  Cartridge electrolytes may be toxic if swallowed and can be harmful  to skin
33. o correct the interference by  one or more of the following measures     m Reorient or relocate the receiving device    E Increase the separation distance between the equipment    m Connect the equipment into an outlet on a circuit different from that which  the other device s  are connected    m Consult the manufacturer or service technician for help     Guidance and Manufacturer   s declaration   electromagnetic emissions   The SeQual Technologies ECLIPSE Oxygen System is intended for use in the  electromagnetic environment specified below  The customer or the user of the  SeQual Technologies ECLIPSE Oxygen System should assure that it is used in  such an environment     Emissions Test Compliance Electromagnetic Environment   Guidance       RF emissions CISPR 11 Group 1 The SeQual Technologies ECLIPSE Oxygen  System uses RF energy only for its internal  function  Therefore  its RF emissions are  very low and are not likely to cause any  interference in nearby electronic equipment        RF emissions CISPR 11 Class B The SeQual Technologies ECLIPSE Oxygen  System is suitable for use in all establishments   including domestic establishments and those  directly connected to the public low voltage  power supply network that supplies buildings  Voltage fluctuations  Complies used for domestic purposes    flicker emissions IEC 61000 3 3    Harmonic emissions Class A  IEC 61000 3 2             48   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual    Disposal of Equipmen
34. odel 6104 Power Cartridge    Model 2400    There are many different types of oxygen tubing  cannula  and humidifiers  The  following accessories are recommended to be used with your ECLIPSE Oxygen  System  Refer to the    Maintenance    section of this manual for accessory  maintenance instructions     Humidifier   Salter Labs   Bubble Humidifier  Part Number 7600 or equivalent    If your physician has prescribed a humidifier  follow the manufacturer s instructions  for use  Attach the humidifier to the oxygen outlet port of the oxygen system  The  O  symbol marks the oxygen outlet port on the concentrator  Use of humidifiers not  specified for this oxygen system may impair performance and void the warranty     DO NOT use a humidifier in Pulse Flow Mode  The ECLIPSE will not detect an  inspiration and will default to Continuous Flow Mode     Humidifier Adapter   SeQual Part Number 3509  If your physician has prescribed a  humidifier  you may need to use the SeQual Humidifier Adapter  Follow the instructions  for use  Attach the Humidifier Adapter to the oxygen outlet port of the oxygen system  and then to the humidifier  Attach cannula or extension tubing to humidifier outlet     32   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Oxygen Tubing   Salter Labs   Series 2000 or equivalent tubing  internal diameter  of the tubing should be no less than 3 16     0 48 cm   Connect the oxygen tubing to  the outlet port of the humidifier or directly to the oxygen o
35. osure does not protect against the harmful  effects of liquid ingress  Electrical shock or damage to the unit  may result           CAUTION   When operating the ECLIPSE on the cart  DO NOT bang  drop or  bump the unit while in transport              34   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual    Cleaning  Care and Routine Maintenance       Routine Maintenance  A trained technician will carry out all routine maintenance at least once a year or    as needed     Your dealer provider will perform inspections and maintenance of the compressor   filters  alarms  9V battery  Power Cartridge and other internal parts as needed        DO NOT disassemble the ECLIPSE or attempt to perform any routine  maintenance other than what is described in the Patient Cleaning and  Care section below  Disassembling the oxygen system may create a  hazard of electrical shock and void your warranty  Contact your  dealer provider for any maintenance required           Patient Cleaning and Care    Cannula Replacement  Replace your tubing and cannula on a regular basis as  recommended by your dealer provider  Your physician or dealer provider will provide  you with cleaning  disinfection and replacement information     Air Inlet Filter  Ambient air is drawn into the unit through the air inlet located on the  top back of the unit  Cleaning the air inlet filter is the most important activity you will  perform to keep your oxygen system performing properly  Check this air filter daily an
36. our ECLIPSE ON and allow the Power Cartridge to completely discharge   Unit will sound alow Power Cartridge alert until the unit completely shuts down   This could continue for up to approximately 20 minutes     2  Connect the AC Power Supply and plug in the AC Power Cord to completely  charge the Power Cartridge  This may require up to five hours of uninterrupted  charging  This completes the calibration of the Power Cartridge        ECLIPSE    Model 1000 User Manual   23    OTER    Operating Instructions  continued             WARNING   DO NOT tamper with  disassemble  crush or heat the Power Cartridge  above 140  F  60  C   The Power Cartridge may present a risk of fire  or explosion and will void the warranty           CAUTION   Store the Power Cartridge in a cool  dry place when not in use        CAUTION   DO NOT leave your ECLIPSE or your Power Cartridge in your vehicle  or in the trunk during a hot day                 CAUTION   The ECLIPSE system can only work with a SeQual Power Cartridge   Use of another Power Cartridge or Battery may damage the unit   present a risk of fire or explosion and will void the warranty                 In the event of a power interruption  the ECLIPSE Oxygen System will automatically  switch to the Power Cartridge operation  When power is restored  the Power Cartridge  will automatically start recharging  If the Power Cartridge is not present during a power  interruption the ECLIPSE Oxygen System will stop operation  When power is restored
37. rns  all possible  sources of ignition must be kept away from the oxygen system and  preferably out of the room in which it is being used  Smoking in the  WARNING      proximity of an operating oxygen concentrator is dangerous  and  can permanently damage the device and void the warranty                 Use No Oil or Grease     A spontaneous and violent ignition may  occur if oil  grease or other petroleum substances come into contact  with oxygen under pressure  Keep these substances away from the  oxygen system  tubing and connections and any other oxygen source   DO NOT use any petroleum based or other lubricants              6   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Operating Safety     m Keep oxygen system and cord away from hot surfaces       DO NOT use the oxygen system if it has a damaged cord or plug  contact your  dealer provider     m Your oxygen system should be located in a well ventilated area to allow for  adequate air intake and to avoid the intake of airborne pollutants and fumes       ONLY use this device with accessories designed or recommended by SeQual  Technologies Inc     m  f your oxygen system has been dropped  damaged or exposed to water  contact  your dealer provider for examination or repair       DONOT drop or insert any object into any opening       DONOT block the air inlet or the exhaust vent of the ECLIPSE when it is on  a soft surface  such as a bed or couch or car seat       DONOT cover unit with a blanket  towel 
38. roximately  10  of the total Power Cartridge charge  When the Power Cartridge is  being charged  the charge indicator bars will blink in a waterfall fashion     If the Power Cartridge is not installed  or if it is improperly installed  the  Status Gauge will not light     External Power Present Indicator  When the ECLIPSE system is  properly plugged in using the AC or DC Power Supply  this indicator  will appear on the Control Panel     14   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       Recommended Operating Environments       The following information recommends certain boundaries that will allow the  ECLIPSE Oxygen System to operate at its full potential           Operating Temperature 50   F to 104   F  10   C to 40   C   Operating Humidity 10    95  at an 82 4  F  28  C  dew point  Transport Storage Temperature  4   F to 140   F   20   C to 60   C     Humidity  Up to 95  Non Condensing       Electrical Use no extension cords or electrical  outlets controlled by a switch        Altitude O     13 123 feet  0   4000 meters        Placement Do Not block air inlet or bottom exhaust vent   keep at least 3 inches away from walls   draperies  furniture  etc         Environment Smoke  pollutant  and fume free        Operating Time Up to 24 hours per day when connected  to external power        When moving the ECLIPSE from an extreme environment  allow time for your  ECLIPSE to acclimate to the recommend operating environment       WARNING   Operating your ECLIP
39. t and Accessories       Follow local governing ordinances and recycling plans regarding disposal of device  components  DO NOT dispose of the Power Cartridge  Power Cartridges should be  returned to your dealer provider     MANUFACTURED BY  EUROPEAN UNION AUTHORIZED AGENT   SeQual Technologies Inc  MeddiQuest Limited   11436 Sorrento Valley Road Business  amp  Technology Centre   San Diego  CA 92121 USA Bessemer Drive    858  202 3100 Stevenage SG1 2DX    800  826 4610 United Kingdom   www sequal com Tel    44 1763 222505    Fax    44 1763 222049       DEALER PROVIDER NAME AND ADDRESS              PHYSICIAN NAME AND ADDRESS              SeQual   and ECLIPSE    are trademarks of SeQual Technologies Inc   San Diego  CA 92121  Salter Labs   is a registered mark of Salter Labs  Arvin  CA 92303    Lysol   is a registered trademark of Reckitt Benckiser  UK     ECLIPSE    Model 1000 User Manual   49    ECLIPSE       Model 1000 User Manual  PN 6112 03 06       
40. t cause  serious injury           CAUTION   Important information for preventing damage to the ECLIPSE             Oxygen System   DO NOT Information to which you should pay special attention   Important  Clarifying information  specific instructions  commentary  sidelights     or interesting points     Symbols Used in User Manual and on Oxygen System    Refer to this section for an explanation of the symbols and the warnings that  accompany them        Read Instructions     This symbol  when used alone or in conjunction   Ci  with any of the following symbols  indicates the need to consult the  operating instructions provided with the product  A potential risk  exists if the operating instructions are not followed     symbol indicates important information that requires the user to take    f    Attention or Caution     Consult accompanying documents  This  special precautions when certain conditions are present     2   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual          No Smoking or Naked Flames     The use of oxygen therapy   requires that special care be taken to reduce the risk of fire  Materials  that burn in air  and even some materials that will not burn in air  are  easily ignited and burn rapidly in high concentrations of oxygen  For  safety concerns  all possible sources of ignition must be kept away  from the oxygen concentrator and preferably out of the room in which  itis being used  Smoking in the proximity of an operating oxygen  concentrator 
41. tem will automatically  switch over to the Power Cartridge  if installed  When external power is restored  by  plugging into AC or DC outlets  the system will accept power from the external source  and replenish the Power Cartridge while operating  if adequate power is available     Connect the AC or DC Power Supply by aligning the arrow on the plug locking ring   with the white dot on the receptacle of the unit  Push until the locking ring twists   slightly counter clockwise  then snaps back into alignment with the dot on the   receptacle  Disconnect by grasping the plug  twist counter clockwise and pull   Connect the AC Power Cord to a grounded power outlet     18   ECLIPSE    Model 1000 User Manual       Model 1000 User Manual       DO NOT use any cord if damaged  Improper use of the Power Cord  may result in a fire or electrical shock hazard   DO NOT touch the Power Cartridge cavity connectors with the Power    Cartridge removed while the unit is connected to an external power  source  This may result in electrical shock                  1  AC POWER SUPPLY  The ECLIPSE System includes a universal AC Power Supply  for use at home or where standard AC power is available  To plug into AC  make  sure the AC Power Supply cord is securely inserted in the power receptacle on the  side of the unit  and plug the AC Power Supply into an AC outlet  When the unit is  plugged in properly  a green indicator on the AC Power Supply will light and the  External Power Present Indicator    
42. ternal Power Present  Indicator  Green     Pulse Flow Mode  Indicator  Green     Flow Mode Button    12   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual       User Controls    This section describes the control buttons and function for your ECLIPSE Oxygen System   FRONT PANEL ICONS    ON OFF Button and Indicator  This button turns the system ON or  OFF  The Green Indicator is lit when the unit is ON     Increase or Decrease Flow Rate Button  Use these buttons to set the  flow rate to your prescribed level   Flow Mode Button and Indicator  The ECLIPSE contains a button to  toggle between Continuous Flow and Pulse Flow mode  Pulse Flow  mode allows a significant increase in the amount of time the system  can operate from the Power Cartridge  When the Pulse Flow Mode is  activated  the green Flow Mode indicator goes on and a pulse of  oxygen is delivered upon inhalation  When the pulse of oxygen is  delivered  the indicator briefly goes off  Only change your flow settings  at the recommendation of your physician     SYSTEM STATUS INDICATORS    The ECLIPSE control panel displays important operating information  This section will  help you understand these features     Green  Yellow  and Red indicators on the front panel indicate the operating condition  of the system  For more information on the indicators  see the Alarm Conditions and  Troubleshooting table      ly  OA ON OFF  Green  Indicator   System OK     Indicates that the ECLIPSE is ON and operating properly  
43. utlet port if you do not use a  humidifier  Connect the other end of the tube to the nasal cannula  if not already attached   Tubing not specified for use with this oxygen system may impair the performance        BGR The tube fitting may be tight  DO NOT use oil or grease to lubricate             WARNING   lf tubing becomes disconnected during operation  ECLIPSE does not  alarm  Check flow at the cannula to verify oxygen delivery              CAUTION   To ensure there is adequate flow to deliver oxygen  the length of the  tubing must not exceed 50 feet  15 2m  when using Continuous Flow  Mode  and must not exceed 7 feet  2 1m  when operating in Pulse  Flow Mode for adequate breath detection                 Cannula   Salter Labs   Part Number 1600 Series or equivalent  Your physician will  have prescribed a cannula to deliver oxygen  In most cases they are already attached  to the oxygen tubing  If not  follow the instructions included with the cannula to attach  it to the oxygen tubing  Use of cannula not specified for use with this oxygen system  may impair the performance     Power Options  You have various options to provide power to your ECLIPSE   A Power Cartridge  AC Power Supply and DC Power Supply     Power Cartridge  Model 2400  The rechargeable Power Cartridge supplied with  your ECLIPSE is capable of providing you with battery operation depending on your  prescribed flow rate     Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC  Power 
44. utside of specified operating or  storage temperatures as this may damage the unit              CAUTION   Locate oxygen tubing and Power Supply cords to prevent tripping  hazards              CAUTION   DONOT place ECLIPSE in a small enclosed space  such as a  bathroom or closet  with cannula leading out of the enclosed space              CAUTION   DO NOT expose the Power Cartridge to temperatures above 140   F   60   C   such as in a vehicle parked in the sun or on a hot day              CAUTION   Charging the Power Cartridge below 50 F    10   C  or above 104   F   40   C  may degrade the performance           CAUTION   DO NOT drop the Power Cartridge or expose it to mechanical shock           CAUTION   Only use the Power Cartridge for its intended purpose              CAUTION   Locate the unit in a well ventilated room that provides adequate air  flow and check if air is being prevented from entering or exiting the  system by furniture  draperies  bedding  clothing or pets           CAUTION   Always check to see that the Air Inlet and the Exhaust Vent are not  blocked and the air inlet filter is clean before using your system                 ECLIPSE    Model 1000 User Manual   45    OTET       CAUTION   The AC power cord used with the ECLIPSE Oxygen System must  meet the requirements of the country where the product is used              CAUTION   Use with grounded  3 pin receptacle  DO NOT remove ground pin  from cord  Use only with SeQual supplied cords              CAUTIO
45. utton is pushed        Flow Mode     Activates either Continuous Mode or Pulse Flow  Mode Operation        External Power Present Indicator     Indicates the presence of  external power to operate the concentrator        Power Cartridge Status     Shows the amount of charge remaining in  the Power Cartridge        Power Cartridge Symbol     The symbol on the Power Cartridge  pack        Oxygen Output     The port in which the oxygen flows out and where  the tubing is attached     4   ECLIPSE    Model 1000 User Manual    Model 1000 User Manual    Indications for Use       THE ECLIPSE OXYGEN SYSTEM IS INDICATED FOR THE ADMINISTRATION OF  SUPPLEMENTAL OXYGEN  THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT  NOR DOES IT PROVIDE ANY PATIENT MONITORING CAPABILITIES     YOUR PHYSICIAN HAS PRESCRIBED A SPECIFIC OXYGEN FLOW RATE  SETTING TO MEET YOUR INDIVIDUAL NEEDS     RECOMMENDED OXYGEN FLOW RATES SHOULD BE ADJUSTED ONLY  UNDER THE ADVICE OF A PHYSICIAN     FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE  ORDER OF A PHYSICIAN     Contraindications    THE ECLIPSE IS NOT INTENDED FOR LIFE SUPPORTING OR LIFE  SUSTAINING APPLICATIONS  NOR DOES IT PROVIDE ANY PATIENT  MONITORING CAPABILITIES                    IN CERTAIN CIRCUMSTANCES  THE USE OF NON PRESCRIBED  OXYGEN CAN BE HAZARDOUS  THIS DEVICE SHOULD ONLY BE  USED WHEN PRESCRIBED BY A PHYSICIAN                WARNING   FOR ADULT USE ONLY  NOT QUALIFIED FOR USE ON PEDIATRIC  PATIENTS                WARNING   NOT FOR USE IN THE PR
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
CLH 7482 BLH 7502 / BLH 7602 / BLH 7603 BPH 7502  Printing Barcodes User Guide  Manuel d`utilisation Générateur de courant, Essence - BC  Omega Manual - Thompson Center Arms  SAS to SAS/SATA RAID Subsystem User Manual  Sony MZ-E40 User's Manual  Sharkoon Side Panel closed  Informações Essenciais para Aerospray Hematology" Slide Stainer    Coby CVE196 headphone    Copyright © All rights reserved. 