1. A Graham Field Brand Wallmax Professional Aneroid Sphygmomanometer Gaps Model V222B User Manual Read this manual before operating the Wallmax Professional Aneroid Spohygmomanometer Save this manual for future reference The latest version of this manual can be found at www grahamfield com V222B INS LAB RevB11 CONTENTS SAFETY GUIDELINES PLEASE READ BEFORE USE ccccssssssssssseesseeneeeeees 2 INTENDED US Beicccscessccsesevcerccseeseyessesedeseteyestants eteestesstnecetevebevessutetelevescsstsetevesectearsutevets 2 INSTALLATION ioe oc sce cece tads caus ov eccelseavstcevalesanzstes her scvexstessserorsunsvouevacessuseieveuacecuessst deer 3 BLOOD PRESSURE wscisessesccceiscesctecccceisioabeceedezczonthscvsdaectencbacuades seucdlesuevhcos aad sce dee csseive 5 MAINTENANGE eis corcsesescsessecescarscsecescerccbicsscscccnczsicaieasd sassstcesersaee couivitessdes satbetsacereiciusaes 7 WARRANTY i vssccesesicasecscttetscdecssnacasestsasassactncesiasasanctacsddesanrastntesdosssancenendissansusbatsdiaasszanin 7 Tools needed for installation Drill with 3 16 drill bit Info A stethoscope is necessary to perform the auditory component of blood pressure readings SAFETY GUIDELINES PLEASE READ BEFORE USE Z mportant Read and understand these instructions before installing or using the Labtron Wallmax Aneroid Sphygmomanometer If you do not understand any part of these instructions contact your medical profess
2. does not include any labor charges incurred in replacement part s installation or any associated freight or shipping charges to GF Health Products Inc The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 7 Graham Field and Labtron are registered trademarks and Wallmax is a trademark of GF Health Products Inc Packaging warranties products and specifications are subject to change without notice GF Health Products Inc is not responsible for typographical errors USA Corporate Headquarters GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 telephone 770 368 4700 fax 770 368 2386 GRAHAM FIELD www grahamfield com V222B INS LAB RevB11 2011 GF Health Products Inc
3. g the air release valve clockwise Squeeze the inflation bulb at a steady rate until the gauge s needle points at approximately 30mmHg above the individual s normal systolic pressure value If the individual s normal blood pressure is not known it is recommended to inflate to 200mmHg Systolic blood pressure reading Open the air release valve slowly by turning it counter clockwise while holding the diaphragm of the stethoscope over the brachial artery Proper deflation rate is vital for an accurate reading The recommended deflation rate is 2 3mmHg per second or a drop of one to two marks on the pressure gauge with each heartbeat Do not keep the cuff inflated any longer than necessary As the cuff begins to deflate listen carefully with the stethoscope Note the reading on the gauge as soon as a faint rhythmic tapping or thumping sound is heard The first sound is the systolic pressure reading Always check with your health care provider to ensure readings are performed correctly Diastolic blood pressure reading Allow the pressure to continue dropping at the same deflation rate Note the reading on the gauge when the last audible thumping swishing or blowing sound is heard this is the diastolic blood pressure reading After a few seconds have passed and no audible thumping swishing or blowing sounds are heard deflate the cuff using the air release valve completely Remove the cuff and stethoscope from the arm Record the
4. ional or authorized Graham Field distributor for direction in the use of this product Z WARNING If components are damaged or missing contact your Graham Field dealer immediately DO NOT use substitute parts Z WARNING GF Health Products Inc assumes no responsibility for any damage or injury caused by improper installation or use of this product INTENDED USE The Labtron Aneroid Sphygmomanometer shown on the next page is intended to measure arterial blood pressure Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 2 KU N 140 0 WUO My 160 480 Vy 200 V222B Wallmax Professional Aneroid Sphygmomanometer INSTALLATION The Wallmax Aneroid Sphygmomanometer wall mount is packaged complete with gauge swivel bracket and mounting screws Info Mount the gauge at the intended users eye level to facilitate easy measurement reading Z WARNING The wall bracket mounts on two screws which are installed in the wall You must mount these screws in a wall stud otherwise the sphygmomanometer could fall resulting in damage to the sphygmomanometer and or personal injury Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 1 Mount the bracket using the two center holes Swivel the sphygmomanometer on its bracket to expose the mounting holes as shown below manometer mounting position on wall 4 screw f bracket SPE Installation V222B Wallmax Professional Aneroid Sphyg
5. momanometer 2 Locate the wall stud where you will mount the bracket Z DANGER risk of electrocution Ensure that there is no electrical wiring at the location where you will mount the bracket 3 Hold the bracket up to the wall at the desired mounting height Place the two center mounting holes directly over the wall stud Ensure that the manometer s face is correctly oriented and that the bracket is level Mark on the wall in the center of both holes Remove the bracket 4 Drill a 3 16 hole approximately 7 8 deep at each of the two marked locations 5 Hold the bracket up to the wall at the mounting location ensuring that the manometer s face is correctly oriented and that the bracket is level 6 Install the two mounting screws through the bracket center mounting holes into the wall stud Do not overtighten Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 4 BLOOD PRESSURE Blood Pressure is a measure of the blood s pressure in the circulatory system which changes constantly during the course of the cardiac cycle Blood pressure readings report two values The higher reading systolic pressure shows the highest pressure in the arteries occurring when the heart contracts The lower reading diastolic pressure shows the lowest pressure in the arteries which occurs right before the heart contracts Blood pressure readings are written with the highest value first then the lowest value Readings of 120 80 a
6. re considered to be normal with high blood pressure being defined as a systolic pressure which is 140mmHg or more at rest and a diastolic pressure which is 90mmHg or more at rest Only a patient s physician is qualified to determine whether the readings obtained are normal for that person Measuring blood pressure Have the patient while relaxed and in a sitting position extend the arm from which the blood pressure will be taken to the front or laterally with the palm of the hand up Attach the cuff and diaphragm Hold the end of the cuff containing the inflation bag firmly against the inside of the bare upper arm with the artery symbol positioned over the brachial artery Never place the cuff over clothing Pull the opposite end of the cuff snugly around the arm and secure the hook and loop fastener strips The range lines indicated by the arrows show the correct cuff size when the cuff is placed on the arm If the index line on the end of the cuff falls between the two range lines the cuff is the proper size If the index line falls outside the range lines a larger or smaller cuff should be used The cuff should be snug but not too tight If one or two fingers can fit between the cuff and the arm the cuff is properly secured Place the diaphragm of the stethoscope over the brachial artery inferior to the cuff Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 5 Inflating the cuff Close the bulb s air valve by turnin
7. systolic and diastolic readings Repeat the measurement two or more times to ensure accuracy Only a patient s physician is qualified to analyze blood pressure Labtron V222B Wallmax Professional Aneroid Sphygmomanometer 6 MAINTENANCE A NOTICE To prevent product damage please follow these recommended care and maintenance instructions Do not drop or pull excessively on the sphygmomanometer components Never inflate beyond 300mmHg Do not expose the cuff to direct sunlight Do not put the sphygmomanometer in contact with sharp objects which could pierce the material and cause damage Do not dismantle or disassemble Cleaning Wipe off manometer and bulb with a clean damp cloth Wash cuff with soap and cold water rinse and air dry Do not iron cuff Sterilization Do not use steam heat or liquid disinfectants to sterilize cuff inflation system or manometer Gas sterilization may be used on cuff only if necessary Always deflate the cuff completely before storage WARRANTY GF Health Products Inc Graham Field warrants the Labtron Wallmax Professional Aneroid Sshygmomanometer Model V222B for a period of one year for defects in workmanship and materials If a product is deemed to be under warranty GF Health Products Inc shall provide at its option 1 replacement of any defective part or product or 2 a credit of the original selling price made to GF Health Products Inc s initial customer The warranty

Download Pdf Manuals

image

Related Search

Related Contents

The MC@NLO 3.3 Event Generator  User Manual  Feit Electric 73992/6 Instructions / Assembly  情報きょういく通信第4号 - 福島県立あぶくま養護学校  Mode d`emploi - Andermatt Biocontrol AG  Samsung 204BW, 205BW Computer Monitor User Manual  “Brocantes, vides-greniers, puces, braderies, Leboncoin … une  KB-7441  Mode d`emploi (français)  Funcionamiento  