Home
Druck DPI 821/822 - GE Measurement & Control
Contents
1. G IDOS 30 g G Dos 4 4 30 g 30 g 5 03 15 03 15 88 16 11 18 15 74 16 11 35 0 14 _276 min Settings Start Settings a Pressure b Leak test Figure 7 Example configuration Pressure measurement with a UPM When the connections are complete make the necessary IDOS selections Table 2 3 If you re attach a UPM the instrument uses the same measurement units that you used before The instrument keeps a record for the last 10 modules UPM Measure the pressure To measure the pressure Figure 7 dk 2 Select the applicable pressure task from Select Task Table 2 3 and if necessary adjust the Set Up Table 1 and the Settings Table 4 5 If necessary do a zero correction Table 4 To measure pressure with another operation Figure 8 use the same procedure K346 Issue 2 W 9 CJ a iDos af MEA OSO Ti Ge los 28 exe Ge CKT 02243 87 9 G 30 5 03 Figure 8 Example configuration To measure pressure and temperature UPM Leak test 1 1 o do a leak test on a pressure system Figure 7 Select an applicable leak test from Select Task Table 2 3 and if necessary adjust the Set Up Table 1 and the Settings Table 4 2 Set the period for the leak test Table 4 If necessary do a zero correction Table 4 To st
2. 212 O 100 3 4 9 S EC 584 212 3216 100 1769 2 5 4 E EC 584 454 238 270 150 6 0 9 E EC 584 238 1832 150 1000 0 7 0 4 N EC 584 454 4 270 20 8 0 N EC 584 4 2372 20 1300 M 0 6 L DIN 43710 328 1652 200 900 0 6 0 3 U DIN 43710 328 212 200 100 0 9 0 5 U DIN 43710 212 1112 100 600 0 6 0 3 32 2732 0 1500 8 1 0 C 2732 3632 1500 2000 2 5 14 3632 4208 2000 2320 3 4 19 D 32 3092 0 1700 8 1 0 D 3092 3992 1700 2200 29 1 6 D 3992 4523 2200 2495 6 5 3 6 Cold Junction CJ error Maximum Range 50 86 F 10 30 C 0 4 F 0 2 C 14 50 F 86 122 F Add 0 01 CJ error ambient temperature change for ranges 10 10 C 30 50 C Specification mV range Specification Electrical connectors A2 Range mV Impedance 10 100 020 Range Measure Oto 55 mA 12 EN English Accuracy 0 02 of reading 3 counts Temperature coefficient 14 SO F 86 122 F 0 0011 FS F 10 10 C 30 50 C 0 002 FS C Switch detection Open and closed 2 mA current Loop power output 24V 1096 HART resistor 250 Q menu selection Connectors A2 Three 0 16 in 4 mm sockets K346 Issue 2 Customer Service Visit our web site www gesensing com
3. 100 s ues for the Step output I Auto Repeat Optional dr 0 Additional data Table 6 Select Settings IB M r Step To select and set up values for the Defined Step output Example output cycle 100 4 s I Auto Repeat Optional dr 0 Additional data Table 6 Select Settings i IM To select and set up values for the Ramp output Example output cycle 100 sd 0 et Additional data Table 6 Select Settings I IM Auto Repeat Optional Table 6 Additional data for Settings Output Item Value Span Check Low 096 Set the 096 value High 10096 Setthe 100 value Dwell d Set the period Hours Minutes Seconds between each change in value 96 Step Low 0 High 100 Dwell d As above Step Size s Set the change in value for each step as a 96 percentage of the full scale range High Low Defined Step Low 096 High 10096 Dwell d As above Step Size s Set the change in value for each step as a temperature or mV value Ramp Low 0 High 100 Dwell d As above Travel t Set the period Hours Minutes Seconds to go from the Low 096 value to the High 10096 value Auto Repeat continuously f applicable select this item to repeat a cycle EN English 5 Operation This section gi
4. No Yes Other functions Hold Maximum Minimum Average Filter Tare Scaled values Backlight Alarm Refer to Specification data Optional item Safety Before you use the instrument make sure that you read and understand all the related data This includes all local safety procedures the instructions for the UMM if applicable and this publication WARNING e Itis dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its normal condition Use the applicable protection and obey all safety precautions Donotusethe instrument in locations with explosive gas vapor or dust There is a risk of an explosion Continued EN English 1 Safety Continued To prevent electrical shocks or damage to the instrument do not connect more than 30V between the terminals or between the terminals and the ground earth UPM only To prevent a dangerous release of pressure isolate and bleed the system before you disconnect a pressure connection Before you start an operation or procedure in this publication make sure that you have the necessary skills if necessary with qualifications from an approved training establishment Follow good engineering practice at all times Safety Marks and symbols on the instrument Complies with European Warning refer to the C Union di
5. or use the A W buttons e See the example below Select Start Stop i IM the A W butt wor TF elect Start Stop liil ll or use the uttons to make the step changes manually Va Select Start Stop IB IM Example procedure Nudge output Display 3 T1 output 1 Edit 2 3 Edit 4 an 0100 0 Sk1 Edit Sk2 Settings Display T1 output 5 6 A A v 101 0 v 102 0 Increment 1 Table 5 K346 Issue 2 Operation Simulate thermocouple values To simulate the temperature or mV values of a thermocouple 1 Connect the instrument Figure 2 and if necessary adjust the Set Up Table 1 2 Select the T1 or mV output task from Select Task Table 2 3 and if necessary adjust the Settings Table 5 3 Supply the output values to the system Table 7 Bannan Inn on Q mv oT om a Gb CK Cjc243 Gb 100mV e 94 0 e 2 90 Edit Settings f Edit Settings a Temperature b mv Figure 2 Example configuration To simulate the temperature or mV values of a thermocouple Operation Transmitter calibration DPI 822 only To calibrate a transmitter 1 Connect the instrument Figure 3 4 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 5 3 Supply the output values to the system Ta
6. Sensing Druck DPI 821 822 Thermocouple calibrator and Thermocouple loop calibrator User manual K346 e ee gt s m DPI 822 T M 27 A 123ABC_ 1234567890 QWERTYUIOP ASDFGHJKL z Xv BNM 6x Jen es CEC K346 Issue 2 Table of Contents Safety Safety Marks and symbols on the instrument Tostart nee A W A O AC 2 To start Location of items nennen 2 To start Items on the display To start Prepare the instrument To start Power on or off s 3 To start Set up the basic operation To start Select a task Measure and or simulate To start Set up the settings Toxstart Edit THN CHONG asesinato m ania 5 Operation Operation Thermocouple connections Operation Communications port connections 6 Operation Measure thermocouple values E Operation Change the output values 6 Operation Simulate thermocouple values m Operation Transmitter calibration sod Operation mA measurements 8 Operation Switch test acd Operation UPM Pressure measurements 9 Operation Error indications 9 Maintenance 10 Maintenance Clean the unit 0 Maintenance Replace the batteries 10 Calibration 10 Calibration Before you sta
7. Task Settings selection 2 3 T1 mA If applicable Sk1 Start Stop Sk2 Settings Menu Settings Settings AA 19 BA A om v Table 4 yt Menu Settings AB DB a o og Sk2 Table 5 ejt If there is additional data for a menu option select Settings B Il to see the values that are set up If necessary adjust the values Refer to Edit functions 4 EN English AY o include maximum minimum and average values for the measurement task o select and set up a tare value for the measurement task a specified value or the reading on the display he instrument subtracts a positive tare value and adds a negative tare value Additional data Select Settings Wi Wi xy o select and set up a scale of values One local scale for each measurement task Maximum 5 Additional data Example 1 2 Select Settings WI M e o select and set up the filter values to give a smoother output for the measurement task Band as a 96 of full scale FS The filter compares each new value with the previous value If the new value is outside the band it is not filtered V Low pass filter time constant in seconds Increase the value to increase damping factor Additional data Select Settings Wi Wi To select and set up the alarm values for the measurement task maximum and minimum Additional data Select Settings Wi Wl UPM only Gage sensors or sensors with differential opera
8. art the leak test select Start lil I When the test is finished the instrument calculates the leak rate in the applicable units minute Operation Error indications If the display shows lt lt lt lt or Make sure that the range is correct Make sure that all the related equipment and connections are serviceable EN English 9 Maintenance This section gives procedures to maintain the unit in a good condition Return the instrument to the supplier for all repairs Maintenance Clean the unit Clean the case with a moist lint free cloth and a weak detergent Do not use solvents or abrasive materials Maintenance Replace the batteries To replace the batteries refer to B1 Then re attach the cover Make sure that the time and date are correct The calibration facility uses the date to give service and calibration messages All the other configuration options stay in memory Calibration Note GE can provide a calibration service that is traceable to international standards We recommend that you return the instrument to the manufacturer or an approved service agent for calibration If you use an alternative calibration facility make sure that it uses these standards Calibration Before you start To do an accurate calibration you must have thecalibration equipment specified in Table 8 astable temperature environment 70 2 F 21 1 C Table 8 Calibration equipment Calibration e
9. ata Manual Select Settings I M To select and set up the backlight facility timer Additional data Select Settings Wi Ii To select and set up the power off facility timer Additional data Select Settings Wi Wi To show the battery level 96 Scale o gt 9 To set the display contrast 96 A Increases 96 W decreases 96 K346 Issue 2 Table 1 Part of table Menu options Set Up Options If applicable Description To set the time date The calibration facility uses the date to give service and calibration messages To set the language option To calibrate the instrument Additional data Refer to Calibration To select and show the applicable status data Software Build Calibration Due date Serial Number IDOS Information To start Select a task Measure and or simulate When the instrumentis set up Table 1 use the Select Task menu to select the applicable task Menu Select Task Display 1 T1 output 2 3 Sk1 Edit Sk2 Settings Table 2 3 In Table 2 3 IDOS is a Universal Measurement Module UMM If you attach a UMM to the communications port A1 item 9 the Select Task menu shows the applicable IDOS options Table 2 Menu options Select Task Options If applicable Description G Tlor An input measurement task mv Measure thermocouple temperature OR MV Measure thermocoup
10. ble 7 K346 Issue 2 Maximum 30V ec M sur Hasana Kuna i GCK 02243 Gb 100mV e 100 0 s 3 15 Ge 55m Ge 55mA 4 000 4 000 Edit f Settings Start Settings a Temperature b mv Figure 3 Example configuration Transmitter calibration with external loop power O T1 3e mA 24V OP mv Cie mAQAv OPIER Ge mara m 24V 24V e O CK 032243 G 100mV e e 1000 Z 3 15 m F n 4 000 4 000 f Edit Settings Start Settings a Temperature b mv Figure 4 Example configuration Transmitter calibration with internal loop power EN English 7 Operation mA measurements DPI 822 only To measure a current 1 Connect the instrument Figure 5 and if necessary Example Ramp output a Set High and Low values that are applicable to the switch value Table 6 Then to get an accurate Switch value set a long Travel period b Use Start Stop il Bl to start and stop the Ramp cycle 4 If necessary supply the output values in the opposite direction until the switch changes condition again The display shows the applicable values to open and close the switch 5 To do the test again press ESC to reset the values adjust the Set Up Table 1 2 Se
11. ecification data All accuracy statements are for one year Specification General Languages English Default Operating 14 122 F 10 50 C temperature Storage 4 158 F 20 70 C temperature Humidity 0 to 90 without condensation Def Stan 66 31 8 6 cat III Shock Vibration BS EN 61010 2001 Def Stan 66 31 8 4 cat III EMC BS EN 61326 1 1998 A2 2001 Safety Electrical BS EN 61010 2001 CE Marked Size L W H 7 1x 3 3 x 2 0 in 180 x 85 x 50 mm Weight 14 oz 400 g Power supply 3 x AA alkaline batteries Duration T1 mV 70 hours Measure mA 25 hours mA 11 hours 24 V Source at 12 mA Duration T1 mV 70 hours Supply EN English 11 Specification Temperature ranges Thermocouple type Standard Range F Range C Accuracy F Accuracy C K EC 584 454 328 270 200 3 6 2 0 K EC 584 328 2498 200 1370 JT 0 6 J EC 584 346 2192 210 1200 0 9 0 5 T EC 584 454 292 270 180 2 5 14 T EC 584 180 0 9 0 5 T EC 584 70 0 6 0 3 B EC 584 50 7 2 4 0 B EC 584 m 500 3 6 2 0 B EC 584 2192 3308 1200 8 0 R EC 584 58 32 50 O 5 4 3 0 R EC 584 32 572 0 300 3 6 2 0 R EC 584 572 3216 300 1769 8 0 S EC 584 558 sss 32 750 5 0 4 5 2 5 S EC 584 32
12. for the instrument Calibration Procedures mV input or output 1 Connect the instrument to the calibration equipment e mVinput Figure 1 e mV output Figure 2 2 Letthe equipment get to a stable temperature minimum 5 minutes since the last power on 3 Usethe calibration menu Table 9 to do the calibration e mV input three point calibration FS Zero and FS e mV output two point calibration Zero and FS The display shows the applicable instructions to complete the calibration K346 Issue 2 To make sure that the calibration is correct select the applicable mV input or output task Table 2 and apply the necessary values mV input 10 5 O short circuit Then mv 5 10 20 50 100 mV output 10 5 0 5 10 20 50 100 Make sure that the error is in the specified limits Table 10 Table 10 mV input or output error limits Input Calibrator error Permitted DPI 821 822 or output mv error mV mv a mV mV Ge mV mV 10 0 0005 0 00018 0 010 0 010 5 0 0004 0 00011 0 009 0 009 0 0 00005 0 008 0 008 20 0 0006 0 00030 0 011 0 011 50 0 0008 0 00068 0 016 0 016 100 0 0012 0 00130 0 024 0 024 Calibration Procedures CJ input 1 2 3 4 Connect the instrument to the calibration equipment Figure 1 Let the equipment get to a stable temperature minimum 5 minutes since the last power on Use the calibration menu Table 9 to do a
13. he loop power supply is on Refer to Select Task Table 2 3 18 The measured value satisfies one of the alarm conditions Refer to Settings Table 4 19 The data on the display is on hold To continue press the HOLD button again 20 T z Shows the battery level 0 100 21 2 Identifies the type of data and the measurement range Ge Input Ge IDOSinput Refer to Select Task Table 2 3 Gp Output 22 23 24 22 amp w The settings applied to the input or output The units or a specified scale x y Table 4 5 The thermocouple type IK J T Table 4 5 The cold junction temperature Table 1 a P Output operation Table 5 Filter aximum x Average inimum A T Table 4 Tare v 25 A soft key function To select an available function press the soft key below it Example gt Move left ove right 26 The measured value or values applicable to the task selection 27 The Edit display to set up text xy Scaling Table 4 OK Accept the new text label Shift Change the keys 123ABC or _ abc C Add a space BS Back space Delete character labels 6 characters To start Prepare the instrument Before you use the instrument for the irst time Make sure that there is no damage to the instrument and that there are n
14. le mV Gb Tlor An output task mv Simulate thermocouple temperature OR mV Simulate thermocouple mV Gr mA PI 822 only A mA measurement task D G mA 24V DPI 822 only A mA measurement task the loop power supply is on PN DPI 822 only A switch test Ge IDOS UMM only An IDOS measurement task 4 UPM only A leak test 4 o set up the way the instrument works Additional data Refer to Set Up Table 1 Table 3 shows all the one and two function operations that are available If you attach a UMM you can only use the options that include IDOS EN English 3 Table 3 Permitted 1 and 2 function operations Table 4 Menu options Settings Input Options If applicable Units Description o select the temperature units C or F UPM only Pressure Units if you select an IDOS task Table 2 3 Select one of the fixed units of measurement psi mbar type o select an applicable thermocouple type IK J T Function Gr O Gr gt qe T1 Ti mV mV IDOS 1 1 1 1 1 mA 1 x 2 x 2 2 mA 24V 1 x 2 x 2 2 mA x x 2 x 2 2 xx x x x 2 Ge DOS 1 2 2 2 2 x DPI 822 only To start Set up the settings When the task is set up Table 2 3 use the Settings menu to adjust the input and or output operation Display
15. lect the applicable mA input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 Maximum 30V o fur Ge ma Ge mA 24V O80 24V Oe oss 0 083 Setinge Settings Qn Gm V a with oma loop b oka loop Figure 5 Example configuration mA measurement E m ws Operation Switch test DPI 822 only To do tests on a thermocouple switch 1 Connect the instrument Figure 6 and if necessary adjust the Set Up Table 1 2 Select the applicable switch test from Select Task Table 2 3 and if necessary adjust the Settings Table 5 The display shows the switch condition open or closed in the top right hand corner 3 Supply the output values to the system Table 7 1000 345 1020 335 100 0 345 Edit Settings Start Settings a Temperature b mv Figure 6 Example configuration Switch test Example Nudge output a Use Edit I Ml to set a value less than the switch value b Use the A W buttons to change the value in small increments 8 EN English K346 Issue 2 Operation UPM Pressure measurements Read all the instructions supplied with the UPM and then use the specified procedures to connect it Figure 7 8 A
16. o missing items Remove the plastic film that protec s the display Use the tag B in the top right hand corner Install the batteries refer to B1 Then re attach the Cover K346 Issue 2 To start Power on or off To turn the instrument on or off press O A1 item 1 The instrument does a self test and then shows the applicable data When the power is off the last set of configuration options stays in memory Refer to Maintenance To start Set up the basic operation Use the Set Up menu to set up the basic operation of the instrument Menu Menu 1 2 3 Set Up 4 5 Select Task Table 2 Table 1 Uy If there is additional data for a menu option select Settings WI Il to see the values that are set up If necessary adjust the values Table 1 Part of table Menu options Set Up Options If applicable Description DPI 822 only To add a series resistor into the mA circuit You can then use this instrument together with a HART communicator to set up and calibrate HART devices HARTI To select the applicable international temperature scale IPTS 68 or ITS 90 CJ To select the type of cold junction CJ compensation Automatic The instrument monitors the CJ temperature and applies the necessary CJ compensation Manual Measure the CJ temperature and set the applicable value The instrument uses this value to apply the necessary CJ compensation Additional d
17. one point calibration FS The display shows the applicable instructions to complete the calibration To make sure that the calibration is correct select the applicable T1 input task Table 2 ake sure that the DPI 821 822 gives a probe emperature that agrees with the temperature on he digital thermometer 0 2 F 0 1 C Calibration Procedures mA input M DPI 822 only Connect the instrument to the calibration equipment Figure 5 Let the equipment get to a stable temperature minimum 5 minutes since the last power on Usethe calibration menu Table 9 to do a two point calibration FS Zero and FS The display shows the applicable instructions to complete the calibration K346 Issue 2 4 To make sure that the calibration is correct select the applicable mA input task Table 2 and apply these values mA 55 40 24 18 12 6 0 short circuit Then mA 0 6 12 18 24 40 55 5 Make sure that the error is in the specified limits Table 11 Table 11 mA input error limits Applied Calibrator Permitted mA error DPI 821 822 error mA mA 55 0 0022 0 005 40 0 0018 0 004 24 0 0014 0 003 18 0 0004 0 003 12 0 0003 0 002 6 0 0002 0 002 0 Short circuit 0 001 Calibration Procedures IDOS UMM Refer to the user manual for the IDOS UMM When the calibration is complete the instrument automatically sets a new calibration date in the UMM Sp
18. quipment mV calibrator Accuracy Refer to Table 10 CJ Standard RTD probe Accuracy 50 mk for 23 82 4 F 5 28 C Digital thermometer Accuracy 10 mK Function mV Pressure UPM only Refer to the user manual for the IDOS UPM mA mA calibrator Accuracy Refer to Table 11 Before you start the calibration make sure that the time and date on the instrument are correct Table 1 10 EN English Selection sequence gt Select Task Table 2 gt Set Up Table 1 gt Calibration gt Menu e 3 Select Channel B 3 Display Enter Calibration PIN M L L Factory PIN 2 4321 Table 9 Table 9 Calibration options Options Description a mV o calibrate the input mV o mv o calibrate the output mV CJ o calibrate the cold junction channel qx IDOS UMM only To calibrate the specified IDOS UMM Refer to the user manual for the IDOS UMM a mA DPI 822 only To calibrate the mA input Calibration Due To set the date of the next P calibration for the instrument After the specified calibration date there is a warning message There is a selection box to stop the warning To change the calibration PIN Personal 8 Identification Number when you select a channel the display shows the applicable instructions to complete the calibration when the calibration is complete select Calibration Due and set the new calibration date
19. rectives A manual na Read the manual t Battery Ground Earth ON OFF T1 Identifies the thermocouple connector To start To start Location of items ESN Item Description 1 O Onor off button 2 mu Left hand soft key Selects the function above it on the display Item 25 Example Edit 3 Moves back one menu level ESC Leaves a menu option Cancels the changes to a value 4 A Increases or decreases a value vw Highlights a different item 5 HOLD Holds the data on the display To continue press the HOLD button again 6 Shows the Select Task menu MENU Selects or accepts an item or value Selects v or cancels a selection 7 nu Right hand soft key Selects the function above it on the display Item 25 Example Settings 8 Display Refer to A3 9 sensor Communications port Use to connect a Universal PC Measurement Module UMM or a RS232 cable 10 Thermocouple connector Refer to Operation 11 Connection point for some of the optional accessories Refer to the datasheet 12 Battery compartment Refer to B1 13 14 15 DPI 822 only Terminals to measure current to supply 24V source and to do switch tests 2 EN English To start Items on the display Item Description DPI 822 only Task indication for the switch test switch closed oo e switch open UPM only Task indication for the leak test Refer to Select Task Table 2 3 17 24V DPI 822 only T
20. rt 10 Calibration Procedures mV input or output 0 Calibration Procedures CJ input 1 Calibration Procedures mA input Calibration Procedures IDOS UMM Specification data cete ertt ee Specification General T Specification Temperature ranges 12 Specification mV range 2 Specification Electrical connectors A2 12 Customer service sse Back cover 2005 General Electric Company All rights reserved Trademarks All product names are trademarks of their respective companies K346 Issue 2 Introduction The DPI 821 Thermocouple Calibrator and DPI 822 Thermocouple Loop Calibrator are part of the Druck DPI 800 series of hand held instruments The DPI 800 series uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Universal Pressure Module UPM The DPI 821 822 include these functions Function DPI 821 DPI 822 Measure simulate Yes thermocouple temperature or mV Cold Junction CJ Automatic Manual compensation Step Ramp functions Automatic Manual Communications port IDOS or RS232 Language selection Yes Measure pressure Leak test External IDOS UPM Snapshot Up to 1000 displays with a date time stamp Measure mA No 0 55mA HART resistor No Yes V dc output No 24V Switch test
21. tion A zero correction that makes the instrument read zero at local pressure Leak Test only To set an applicable period for the leak test Hours Minutes Seconds K346 Issue 2 To start Edit functions Example 1 Set up a label for x y Scaling 96 Settings xy mA GRE 504 000 04 000 20 009 20 009 a mj 1 Edit 2 Edit 3 A xw Sk1 Edit Example 2 Se Settings 4 mm L A3 item 27 up values for x y Scaling 0 to 100 Settings 2 Edit 3 Xy 0 mm E20 000 100 00 Sk1 Edit a Flow scaling mA pressure only Table 5 Part of table Menu options Settings Output A 00000 vw an Options Description If applicable Units To select the temperature units C or F type To select an applicable thermocouple type IK J T A To select and set up a value for the Nudge output M Example 10 C increments Additional data Select Settings Wi Wi wor To select and set up values for the Span Check 0 output Example output cycle idk 100 args This cycle repeats automatically 0 Additional data Table 6 Select Settings i Wl K346 Issue 2 Table 5 Part of table Menu options Settings Output Options If applicable r Step Description To select and set up va Example output cycle i
22. ves examples of how to connect and use the instrument Before you start Read and understand the Safety section e Do not use a damaged instrument Operation Thermocouple connections Attach the thermocouple wires to the applicable thermocouple mini connector Figure 1 The wider blade is the negative Then attach the connector to the instrument Operation Communications port connections Use the communications port A1 item 9 to attach an IDOS Universal Measurement Module UMM When you attach the cable from a UMM Figure 7 8 the instrument automatically changes the menus to give you all the applicable options Table 2 3 Operation Measure thermocouple values To measure the temperature or mV values of a thermocouple 1 Connect the instrument Figure 1 and if necessary adjust the Set Up Table 1 2 Selecta T1 or mV input task from Select Task Table 2 3 and if necessary adjust the Settings Table 4 M un en Ge mv OH CKT1 2243 Gr 100mv T1 87 9 2 56 b mV a Temperature Figure 1 Example configuration To measure the temperature or mV values of a thermocouple 6 EN English Operation Change the output values When the output operation is set up Table 5 use one of these procedures to change the output values Table 7 Procedures to change the output Output Procedure A Select Edit lil BI and
Download Pdf Manuals
Related Search
Related Contents
Ask Proxima Campus 300 User's Manual Galaxy Metal Gear 87040 User's Manual Metro 100XT 41 RCH Metro 100XT Tunnel 41 RCH Massive Wall light 17131/42/10 Massive Wall light 17218/47/10 Bedienungsanleitung Schichtdickenmessgerät PCE 第 49 回 新春読書感想文コ THE PUG TONEARM Bedienungsanleitung Unité de contrôle OptiControl CM04 - liste des pièces Copyright © All rights reserved.
Failed to retrieve file