1. D D o o C D JET Insert N ozzle blocks repeatedly 1 Powder is moisture contaminated lumpy Discard powder 2 Air filter is contaminated Clean or replace filter N o bleed air 1 Blocked JET nozzle Clean JET N ozzle using supplied tool 2 Blocked filter Clean or replace filter N o or poor cleaning action 1 Very low powder level or empty Powder chamber Fill Powder Chamber 2 Blocked JET Nozzle Clean JET N ozzle using supplied tool 3 Fill Cap is loose Tighten powder fill cap and turn System ON Continuous powder agitation Fill cap not securely sealed Shut System OFF Removethe O ring seal from the fill cap and clean residual powder from the cap Be careful not to scratch or otherwise damage the plastic cap W ipe off the O ring and place it in the fill cap and turn System ON 10 2 Technical Support and Repairs For technical support and assistance call 1 800 347 7412 M onday through Friday 8 00 AM to 5 00 PM Eastern T ime For other areas contact your local D EN T SPLY representative Section 11 Specifications Cavitron JET Combination System 100 115 VAC Systems Electrical Voltage 100V Gen 120J 115V Gen 120 Current 9 8 amps 100v 115v Frequency 50 60 Hertz Water Pressure 25 60 psig Air Pressure 65 100 psig Flow Rate Minimum setting CCW 0 15 ml min Maximum setting CW gt 55 ml min Weight 12 Ib INSIS SYSTEMON OFF Dimensions Height 4 in Width 7 3 8 in D
2. 4 Insert connector of ejector tube into operator saliva ejector Figure 2 g _ AV 1 Insert air polishing nozzle F 2 Seat shoulder flush on nozzle 3 Incorrect positioning 5 Place ejector tube clip around ejector tube and handpiece cable such that cables drape together for patient and clinician comfort 6 O ptional Place handpiece clip along handpiece close to handpiece cable connection such that clip is holding the JetShield evacuation tube in place along handpiece 7 Verify that slurry exits from the JetShield when air polishing unit is activated Adjust saliva ejector as necessary 8 Proceed with routine air polishing applying cup to tooth so cup flares slightly Pivot nozzle inside cup adapting to mesial and distal surfaces Apply 2 seconds of spray for each segment of tooth releasing foot control each time If spray escapes readjust cup to maintain flat contact with tooth surface adjusting angle by pivoting nozzle inside of handpiece Section 8 Techniques for Use cont d After completion of each procedure remove J etShield assembly and dispose of JetShield according to applicable regulations W ipe disinfect all other components before next use Figure 3 8 6 PerformingJET Air Polishing Procedures with J etShield NOTE Refer to the Infection Control Card supplied with your System for general procedures to be followed at the beginning of each day and between patients e Install J
3. A manual shut off valve on the office air supply line should be used so that the air line can be completely shut off and the line pressure relieved when the office is unoccupied e Your System must be supplied with clean dry air T o help prevent water condensation from forming in the air supply line which may cause the system to malfunction it is strongly recommended that an air dryer be used on the compressor line supplying the System 5 4 Electrical Requirements Refer to Section 11 Specifications 5 5 Unpacking the System Carefully unpack your Cavitron JET Combination System and verify that all components and accessories are included 1 Cavitron JET Combination System with water Supply line air supply line handpiece and cable connector assembly factory installed Air filter assembly Detachable AC Power Cord set Cavitron JET Insert N ozzle with cleaning tool D ual position Foot Control Assembly Cavitron U Itrasonic Inserts quantity optional Efficiency Indicator for C avitron Inserts JeShield Aerosol Reduction D evice Literature Packet S OQ Uu Row 5 6 ON OFF rocker switch Section 5 Installation Instructions cont d Power C ord Power C onnection Verify the Power Control ON OFF rocker switch located at the left front underside of the System is set to the OFF position before proceeding Plug the detachable AC cord into the back of the System Plug the 3 prong grounded plug
4. For Air Polishing set the power level between medium mid and maximum Handpiece Holds all Cavitron 30K Ultrasonic Inserts and transmits power and lavage from the System to the insert It also holds AC Power Indicator Light Lighted when System is switched ON the Cavitron JET insert nozzle which automatically selects air polishing mode Power Control ON OFF switch located at the left front underside of the System Boost Indicator Light The two position Foot Control provides the capability to increase the front panel Power dial setting When lighted indicates Foot Control is fully depressed and System is functioning in Boost mode Lavage Flow Adjustment Turn knob to select Lavage Flow level during System operation Clockwise increases flow at insert tip counter clockwise decreases flow The rate of flow through the Handpiece determines the temperature of the lavage Low flow rates produce warmer temperatures higher flow rates produce cooler temperatures ENGLISH 6 Section 6 Cavitron Jet Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System cont d 6 2 Handpiece T he handpiece accepts all Cavitron 30K Ultrasonic Inserts and the JET air polishing insert nozzle T he open end of the handpiece accepts all Cavitron 30K inserts Handpiece Powder Flow Window Air Polishing Powder Delivery Port Cable Insert Port 6 3 Cavitron 30K Ultrasonic nserts T he many
5. may cause caking of the air polishing cleaning powder and clogging the lines at the air polishing tip e Cavitron Prophy J et Cleaning Powder is a water soluble sodium bicarbonate powder specially formulated for use in 2 2 Warnings Persons fitted with cardiac pacemakers have been cautioned that some types of electronic equipment might interfere with the operation of a pacemaker Although no instance of pacemaker interference has ever been reported to DENT SPLY we recommend that the handpiece and cables be kept at least 6 to 9 inches 15 to 23 cm away from any pacemaker and pacemaker leads during use Do not direct the air polishing stream at soft tissue or into the sulcus T issue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulcus Cavitron Air Polishing Systems D o not attempt to use any other cleaning materials in the air polishing powder reservoir Empty the air polishing powder reservoir at the end of the day to prevent caking of the powder and clogging the lines and air polishing tip 3 2 Precautionsfor U Itrasonic Prophylaxis Procedures Like a toothbrush ultrasonic inserts wear out with use Inserts with just 2 mm of wear lose about 5096 of their scaling efficiency In general it is recommended that ultrasonic inserts be discarded and replaced after 2 mm of wear to maintain optimal efficiency and avoid breakage f excessive w
6. the work site before proceeding to the next 2 3 teeth If desired the bleed air foot control released passing through the air polishing insert tip can be used to dry the work site during inspection Flushing the patient s tongue with water only foot control depressed to the first position helps reduce the saline taste Maintain a 3 to 4mm insert tip to tooth operating distance e n the posterior regions use the cheek to help contain aerosol spray T he patient s lip should be held between the thumb and index finger of your free hand and pulled out and up or down to form a cup Do not retract the lips When air polishing the facial surfaces of the anteriors center the spray on the middle third of the tooth and use a constant circular motion T he edge of the spray will clean the teeth to the e U se a saliva ejector or HVE gingiva e Tilt the patient s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Do not aim directly at the soft tissue Avoid prolonged use on polished metal surfaces as a matte finish may result n the upper anterior region use the upper lip and your hand as a shield to help contain the aerosol spray 8 8 Proper Angulation of the Air Polishing Insert without JetShield X LII A Recommended angulation on the Recommended angulation on the buccal and Recommended angulation anterior teeth is 60 with the tip aime
7. without all filters in place Section 10 Troubleshooting Although service and repair of the Cavitron JET Combination System should be performed by D EN T SPLY personnel the following are some basic trouble shooting procedures that will help avoid unnecessary service calls Generally check all lines and connections to and from the System aloose plug or connection will often create problems Check the settings on the System s knobs 10 1 Troubleshooting G uide System will not operate Power Indicator Light is not lighted 1 Check that the Power switch isin the ON position and that detachable Power Cord is fully seated in the receptacle on back of System 2 Check that the System s three prong plug is fully seated in an appropriate AC receptacle and that AC current is present Power Indicator Light is lighted 1 Check that the Foot Control Connector is fully seated in the Foot Control Receptacle on the back of the System System operates N o lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s dental office water supply are open 3 If connected to D ualSelect Dispensing System check that fluid level in the selected bottle is sufficient M ake sure valves are open when using external water source ENGLISH 14 Pointer Fill Cap View Window Filter depressurizing System Powder Bowl Air Inlet C c 3 Ke
8. Any other sub stances or additives may clog the U nit and will void the warranty For your convenience the cleaning powder is supplied in sealed packets Keep packets dry and stored in alocation that does not exceed 95 F A special container is provided with your System for powder filling and storage The opening in the container s fill cap permits easy filling of the powder chamber An additional cap without a hole is provided for storage e t is strongly recommended that the Powder Chamber be emptied at the end of each day T his will reduce moisture absorption and minimize clogging T o fill or refill the powder chamber Turn the Unit OFF Unscrew the Powder Fill Cap Transfer the powder from the packet into the fill container and screw on fill cap Place finger over hole and shake the powder container vigorously to break up any lumps that may have formed from settling Carefully pour powder into the chamber until the level reaches the top of the center tube Replace the Chamber Powder Fill Cap and tighten e Turn the Unit ON NOTE Use only Cavitron Prophy Jet Cleaning Powder in the unit Prophy Jet Cleaning Powder should be kept dry and stored in a location that does not exceed 95 F LL eT ENGLISH 10 T o adjust the flow of powder Adjust the powder flow rate by positioning the control at H 12 O clock M 9 O clock or L 6 O clock For heavy stain removal set the control on
9. D EN T SPLY D ealer If your dealer is unable to assist you call 1 800 989 8826 M onday through Friday 8 00 AM to 5 00 PM Eastern T ime For other areas contact your local D EN T SPLY representative Caution U S Federal Law restricts this device to sale by or on the order of a dentist ENGLISH 1 Table of Contents Section Number 1 10 11 12 Section Title Page Description of Contents Number Halo crs C10 S RETRO THRONO TRUE PRE EEE 191 REUNIR RN MN 1 1 Ultrasonic procedures 1 2 Air Polishing procedures Contrandicatlons and MOERS seeded sees oblatae a uet MR Fw E I Du ft V id edu 2 1 Contraindications 2 2 Warnings iso MIO RIED DU OT 3 1 Precautions for All Systems 3 2 Precautions for U Itrasonic Prophylaxis Procedures 3 3 Precautions for Airpolishing Procedures Ratis ener Eae TN ECL Maines ne E HP HH nmn EN 4 1 Infection Control Information Reference C ard 4 2 General Infection Control Recommendations 4 3 Water Supply Recommendations hii Stall abt oi LnStrU CtloNS coord erani ote eRe LO A M Erie aun s Een DS EU E 5 1 General Information 5 2 Water Line Requirements 5 3 Air Line Requirements and Recommendations 5 4 Electrical Requirements 5 5 Unpacking the System 5 6 Power Cord Power Connection 5 7 Foot Control Cable Assembly Connection 5 8 Incoming W ater Supply Line Connection 5 9 Air Supply Line Connection Cavitron JET Combination System U Itrasonic Scaler and Air Polishing Prophy
10. DENSPLY PREVENTIVE CARE Cavitron JET Combination System U Itrasonic Scaler and Air Polishing Prophylaxis System Directions For Use Systeme combineC avitron JET D artreur ultrasonique et systeme de prophylaxie par polissage l air M ode d emploi Sistema combinado C avitron JET Escarificador ultras nico y unidad profil ctica de pulido neumatico Instrucciones para el uso Please read carefully and completely before operating unit Veuillez lire soigneusement et completement avant d utiliser le systeme Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad DENSPLY PREVENTIVE CARE Introduction Cavitron JET Combination System U Itrasonic Scaler and Air Polishing Prophylaxis System DENT SPLY Preventive Care is an ISO 9001 and EN 46000 certified company The Cavitron JET Combination System is classified by U nderwriters Laboratories nc with respect to electric shock fire mechanical hazards only in accordance UL 2601 1 and Can CSA C22 2 NO 601 assigned control 13VA The Cavitron JET dental prophylaxis system is a precision engineered and manufactured instrument It contains controls and components for ultrasonic scaling and air polishing modes T he System operates by converting ordinary AC house current into high frequency current T he ultrasonic system consists of two parts an insert and the SPS Sustained Performance System electronic system The S
11. H For light stain removal set the control on L T he control can be set at any position between H and od id The view window at the center of the pointer lets you observe the powder flow small white circle of powder during operation If no flow is seen check for clogging or add cleaning powder 8 5 Air Polishing using theJetShield Accessory Aerosol reduction device for use with air polishing systems JetShield assembly and accessories see Fig 1 The disposable JetShield assembly is designed for single use all other compo nents are reusable and are compatible with standard clinic equipment Figure 1 A Ejector tube B JetShield assembly we F Connector UI E Ejector tube clip C Connector D Handpiece clip Precautions TheJetShield itself is designed for single use only U se of chemical or heat sterilization will lead to poor performance and malfunction of product T he reusable components are wipe disinfectable using clean cloths and appropriate non immersion type disinfectant W ipe once with disinfectant to remove bioburden then use fresh cloth and more disinfectant to clean surface Dry with clean paper towel or air Step by Step Instructions 1 Remove entire assembly from package 2 Insert air polishing nozzle through JetShield cup and seat shoulder flush on nozzle See Figure 2 3 Insert connector on JetShield assembly into the flexible ejector tube
12. PS system incorporates two closed loops One provides automatic tuning operating frequency is adjusted to be at resonance for each insert the second loop automatically controls the tip stroke over different working conditions The Cavitron JET Combination System produces 30 000 microscopically small strokes per second at the insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional operator and patient comfort ENGLISH Cavitron JET Combination System U Itrasonic Scaler and Air Polishing Prophylaxis System Directions For Use Please read carefully and completely before operating unit With the addition of the Air Polishing function your Cavitron JET becomes a complete prophylaxis system M ode selection is automatic when you place the Airpolishing insert nozzle or one of the interchangeable 30K ultrasonic inserts into the handpiece Technical Support T here are no serviceable parts in this device Refer all serviceto a qualified service repair facility For technical support and repair assistance in the U S call 1 800 347 7412 M onday through Friday 8 00 AM to 5 00 PM Eastern T ime For other areas contact your local D EN T SPLY representative Supplies amp Replacement Parts T o order supplies or replacement parts for your Cavitron JET Combination System contact your local
13. Service Centers Centres de service apres vente mondiaux Centros de servicio mundiales U S A DENTSPLY Preventive Care Technical Service and Repair Department 1301 Smile Way York PA 17404 Phone 800 347 7412 Deutschland DENTSPLY DeTrey GmbH De Trey Strasse 1 D 78467 Konstanz Germany Phone 7531 5830 United Kingdom DENTSPLY Ltd Hamm Moor Lane Addlestone Weybridge Surrey KT15 2SE England Phone 0 1932 853422 Swiss Representative France DENTSPLY DeTrey 17 Michael FARADAY 78380 Montigny Le Bretonneux France Phone 1 3014 77 77 Italia DENTSPLY DeTrey Italia s r l Via A Cavaglieri 26 l 00173 Roma ltalia Phone 06 723 3626 DENTSPLY DeTrey GmbH Konstanz Zweigniederlassung Baar Oberdorfstrasse 11 CH 6342 Baar DENSPLY PREVENTIVE CARE Manufactured by DENTSPLY Preventive Care DENTSPLY International 1301 Smile Way York PA 17404 CC 0086 Authorized EU Representative DENTSPLY DeTrey GmbH D 78467 Konstanz Germany Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Phone 905 851 6060 Printed in U S A Form 80684 Rev C 11 98 Distributed by DENTSPLY Canada Woodbridge Ontario L4L 4A3
14. binet Power Cord H andpiece Connector and cable assembly Foot Control and cable assembly by applying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufacturer T o clean System generously spray disinfectant solution on a clean towel and wipe all surfaces D iscard used towel T o disinfect system generously spray disinfectant on a clean towel and wipe all surfaces Allow disinfectant solution to air dry 5 f using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut D own Procedures at the end of the day Turn theUnit OFF 2 Remove Cavitron U Itrasonic Insert or Cavitron JET Insert N ozzle clean using the wire cleaning tool Sterilize 3 U nscrew the powder chamber cap 4 Remove the powder chamber from the cabinet Pour the powder into the powder storage jar and secure the lid 5 H old the open end of the powder chamber away from you Activate the System for 15 seconds to clear the powder chamber A high volume evacuator can be used to remove any residual powder 6 Secure the lid on the powder chamber 7 Clean and disinfect the surfaces of the cabinet Power Cord H andpiece Connector and cable assembly Foot Control and cable assembly by applying an approved non immersion type disinfectant solution carefully following the instructions provided by the disinfectant solution manufactur
15. ce Place the heater rod into the insert port of the handpiece Align the powder delivery tube with the powder delivery port and gently push into handpiece until fully seated ENGLISH 7 Section 6 Cavitron Jet Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System cont d N 6 5 FootControllnformation amp O peration Z T he Foot Control is a two position momentary switch which activates both ultrasonic energy and Lavage Flow at the insert tip 6 6 Operation for U Itrasonic Prophylaxis Procedures e gt Foot control Foot control Foot control depressed all the released S L depressed half way ta Ultrasonics and lavage flow No ultrasonic energy or lavage flow l are activated A CUTU 6 7 Operation for Air Polishing Procedures Foot control Foot control Foot control released lt gt depressed E depressed SS Za half way all the way X Bleed air exits insert rinsing function Air Polishing Lavage flow nozzle tip kaivaa Function is activated ENGLISH 8 Section 7 Accessories 7 1 Accessories List AC Power Cord Set Cavitron Ultrasonic Inserts Cavitron Jet Insert N ozzle D ual position Foot Switch Assembly D ualSelect M edicament D ispensing System JeShield Aerosol Reduction D evice cU BP WN H For detailed information contact your local D ENT SPLY Representative Section 8 Techniques for Use 8 1 PatientPositioning For optimal access to bo
16. d lingual surfaces of posterior teeth is 80 with to occlusal surfaces is 90 at the middle third of the tooth surface the tip aimed slightly distally ENGLISH 12 Section 9 System Care 9 1 DailyM aintenance Start Up Procedures at the beginning of the day 1 O pen the manual shut off valve on the dental water supply System 2 With the system OFF unscrew the powder chamber cap Verify the powder bowl is empty Turn the Unit ON for 15 seconds to eliminate residual powder or moisture in the lines T urn the system OFF 3 Shake the powder in the storage jar to create an even consis tency of powder mixture 4 Pour the powder into the chamber until it reaches the top of the center tube 5 Secure the lid on the powder chamber 6 Turn the system ON Hold the H andpiece without an insert installed over a sink or drain activate the Foot Control and flush the water line at maximum water flow for at least 2 minutes 7 Place a sterilized Cavitron Ultrasonic Insert or Cavitron JET Insert N ozzle into the H andpiece and adjust the Power and Lavage Flow as required For theJET mode the powder flow can be adjusted Between Patients 1 Remove Cavitron U Itrasonic Insert or Cavitron JET Insert N ozzle clean using the wire cleaning tool Sterilize 2 Set Water Control to highest setting Activate the System and flush at least 30 seconds 3 Turn the System OFF 4 Clean and disinfect the surfaces of the ca
17. eShield on Cavitron Air Polishing N ozzle as indicated on instructions provided Select the appropriate amount of powder using the dial on the Cavitron JET or PROPHY JET 30 air polishing system Verify that slurry exits from the cup when held outside the mouth adjust saliva ejector as necessary e Divide the tooth visually into segments to polish incisal middle and gingival Apply cup to middle third with light pressure so cup flares slightly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes readjust cup to maintain flat contact with tooth surface Adjust angle by pivoting nozzle inside the J etShield cup T he suction from theJetShield can be used to remove excess saliva from the patient s mouth After completing each procedure dispose of clear JetShield assembly according to Federal State and local regulations Properly disinfect all other J etShield accessories prior to reusing them on future patients ENGLISH 11 Section 8 Techniques for Use cont d 8 7 Performing Air Polishing procedures without JetShidd The recommended normal procedure is to clean 2 3 teeth with the air polishing spray foot control depressed to the second position and then rinse the area with water only foot control depressed to the first position to inspect
18. ear is noted or the insert has been bent reshaped or otherwise damaged discard the insert immediately e Ultrasonic insert tips that have been bent damaged or reshaped are susceptible to in use breakage and should be discarded and replaced immediately e Retract the lips cheeks and tongue to prevent contact with the insert tip whenever it is placed in the patient s mouth ENGLISH 3 Section 3 Precautions cont d 3 3 Precautions for Air Polishing Procedures Only set the air polishing powder flow control to the maximum position H when it is necessary to remove particularly difficult stains Return the powder flow control to the medium position upon the completion of the procedure Avoid prolonged or excessive use on restorative dental materials Section 4 Infection Control 4 1 Infection Control Information Reference C ard For your convenience an Infection Control Information reference card printed on durable stock has been included with your Cavitron JET Combination System If your card is missing or misplaced you can request a replacement in the U S by calling Customer Service at 1 800 989 8826 M onday through Friday 8 00 AM to 5 00 PM Eastern T ime For other areas contact your local D EN T SPLY representative 4 2 Generallnfection C ontrol Recommendations As with all dental procedures use standard personal protection equipment i e wear face mask eyewear or face shield gloves and protective
19. epth 91 2 in Handpiece Cable 81 in Foot Control Cable 96 in Power Cord 84 in s Water Supply Line 96 in R TYPE B EQUIPMENT Air Supply Line 98 in Cavitron JET Combination System 200 230 VAC Systems Electrical Voltage 200V Gen 120D 230V Gen 120U N Current 5 4 amps 200v 230v A FOOTSWITCH Frequency 50 60 Hertz Water Pressure 172 414 kPa Air Pressure 448 690 kPa Flow Rate Minimum setting CCW 0 15 ml min Maximum setting CW gt 55 ml min Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for operating theatres Protection Class 1PX0 Operating Environment Temperature 15 to 40 Deg Celsius Relative Humidity 30 to 75 non condensing Transport and Storage Conditions Temperature O to 70 Deg Celsius Relative Humidity 10 to 95 non condensing Atmospheric Pressure 500 to 1060 hPa Section 12 Classification Type of protection against electric shock Class 1 Degree of protection against electric shock Type B Degree of protection against the harmful ingress of water Ordinary Mode of operation Continuous Degree of safety of application in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide Equipment not suitable for use in the presence of flammable anesthetics or oxygen ENGLISH 15 Worldwide
20. er T o clean System generously spray disinfectant solution on a clean towel and wipe all surfaces D iscard used towel T o disinfect system generously spray disinfectant on a clean towel and wipe all surfaces Allow disinfectant solution to air dry N OT E Water based disinfection solutions are preferred Some alcohol based disinfectant solutions may be harmful and may discolor plastic materials If the System is connected to a D ualSelect M edicament D ispensing System follow the recommended procedures listed in the D ualSelect manual If powder fluffing is observed when Foot Control is not in use this would indicate an air leak T o correct turn the Unit OFF remove the Fill Cap clean any residual powder from the O ring seal and threads replace the Fill Cap tighten and turn theU nit back ON 9 2 Periodic M aintenance The two repairable filters should be disassembled and inspected every 90 days Repairable Filters Duckbill A d Ring ENGLISH 13 Section 9 System Care cont d Turn System OFF Remove the Powder Chamber e Remove filter from holder and remove the tubing from one end of the filter Disassemble the filter as shown e Inspect the filter element Replace if necessary Reassemble the filter and reconnect the tubing and replace filter in holder e nspect second filter in a similar manner Replace the Powder Chamber Turn the System ON NOTE N ever operate the System
21. gown Avoid prolonged use on cementum or dentin Avoid prolonged use on polished metal surfaces as a matte finish may result Patients who are on a restricted sodium diet or have severe respiratory illness should consult their physician before undergoing air polishing prophylaxis procedures Patients wearing contact lenses should remove them prior to air polishing treatment As with high speed handpieces and other dental devices the combination of water and ultrasonic vibration from your Cavitron JET Combination System will create aerosols W ith proper technique much of the Cavitron JET Combination System s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patient safety carefully follow the Infection Control Information procedures detailed on the reference card accompanying your System 4 3 Water Supply Recommendations It is highly recommended that all dental water supply systems conform to applicable C D C U S Centers for D isease Control and Prevention and AD A American D ental Association standards and that all recommendations be followed in terms of flushing chemical flushing and general infection control procedures See sections 4 2 and 9 Section 5 InstallationInstructions 5 1 Generallnformation If theinstallation of your Cavitron JET Combination Syste
22. into an approved outlet 5 7 FootControl C able Assembly C onnection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated a ENGLISH 5 Section 5 Installation Instructions cont d 5 8 Water Supply LineC onnection 5 9 Air Supply Line Connection Connect the free end of the system s water supply line to e Connect the free end of the system s air supply line to office water supply line or a Cavitron D ualSelect office air supply line or a Cavitron D ualSelect D ispensing System dispensing System e Inspect all connections to make certain there are no leaks e Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System Description 6 1 System Controls Powder Fill Cap Removable cap for filling and emptying the powder chamber An adjustable pointer increases or decreases powder flow rates Top Grooves Provide a convenient temporary place ment for the handpiece during treatment procedures Always place handpiece in holder when not being used Handpiece Holder Securely holds the System s Handpiece when system is not being used Power Adjustment Turn knob to select ultrasonic power level for operation clockwise increases system power counter clockwise decreases system power NOTE
23. laxis System 6 1 System Controls 6 2 H andpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishing Insert N ozzle 6 5 Foot Control Information and O peration 6 6 O peration for Ultrasonic Prophylaxis Prodedures 6 7 O peration for Air Polishing Prodedures ACCOSSOPIES mesh a Te s cS cM ed Pe a C I S i c E MEE M eR Ae aM EE EE 7 1 Accessories List RE enini TON biz s M 9 8 1 General Information on Patient Positioning 8 2 Performing U Itrasonic Scaling Procedures 0 3 Patient Comfort Considerations 8 4 Air Polishing Powder Chamber 8 5 Air Polishing using the etShield Accessory 8 6 Performing JET Air Polishing Procedures with J etShield 8 7 Performing Air Polishing Procedures without J eShield 8 8 Proper Angulation of the Air Polishing Insert without J etShield TT SS Cases eee I M H M PET 13 9 1 Daily M aintenance Start U p Procedures Between Patients Shut D own Procedures 9 2 Periodic M aintenance TOUTS UO EIEN coco miri at D Sac ay ata fede iuba ume ov boten ta aute tenants taa etos ua deu a 10 1 Troubleshooting 10 2 T echnical Support and Repairs sese fece q L 2 OE AGES T 3 A0 441420 NEM RE TOM ERU 215074 2175 000115 E ERES ENGLISH 2 Section 1 Indications 1 1 Ultrasonic procedures All general supra and subgingival scaling applications Periodontal debridement for all types of period
24. m is performed by someone other than trained D EN T SPLY D ealer personnel care should be taken to observe the following require ments and recommendations 5 2 Water LineRequirements T he System s water supply line is factory installed Do not disconnect from the System Incoming water supply line pressure to the System must be 25 psi 172 kPa to 60 psi 414kPa If your dental water system s supply line pressure is above 60 psi 414 kPa install a water pressure regulator on the supply lineto your Cavitron JET Combination System ENGLISH 4 e A manual shut off valve on the dental water system supply line should be used so that the water can be completely shut off when the office is unoccupied A filter in the dental water system supply line is recom mended so that any particulates in the water supply will be trapped before reaching the System After the above installations are completed on the dental water supply system the dental office water line should be thoroughly flushed prior to connection to the System 5 3 Air Line Requirements amp Recommendations e An air supply line is factory installed on Cavitron JET Combination System D o not disconnect it Incoming air supply line pressure to the System must be 65 psig 448kPa to 100 psig 690kPa If your office air line pressure is above 100 psig 690kPa install an air pressure regulator on the supply line to your Cavitron JET Combination System
25. ncreased finger pressure to function adequately T he use of a saliva ejector or H VE is recommended during all procedures 8 3 Patient Comfort C onsiderations Procedures Reasons for sensitivity NOTE Refer to the Infection Control Information card Form no 80722 supplied with your Unit for general procedures to be followed at the beginning of each day and between patients T he edges of Cavitron U Itrasonic Inserts are intention ally rounded so thereis little danger of tissue laceration with proper ultrasonic scaling technique W henever the insert tip is placed within the patient s mouth the lips cheek and tongue should be retracted to prevent acciden tal prolonged contact with the activated tip H old the empty H andpiece in an upright position and activate the Foot Control until fluid exits 1 Incorrect tip placement Point should be directed away from root surfaces 2 Notkeepingtip in motion on tooth D o not allow the insert to remain in a static position on any one area of the tooth Change the insert s path of motion 3 Applying pressure U se extremely light grasp and pressure especially on exposed cementum 4 f sensitivity persists decrease power setting and or move from the sensitive tooth to another and then return ENGLISH 9 Section 8 Techniques for Use cont d 8 4 Air Polishing Powder C hamber Use only Cavitron Prophy Jet Cleaning Powder in your Cavitron Jet Combination System
26. ontal diseases For patients with a history of sensitivity to ultrasonics Endodontic procedures 1 2 Air polishing procedures e Removal of a variety of extrinsic stains e g tobacco coffee amp tea chlorhexidine Prophylaxis of orthodontic patients Preparing tooth surfaces prior to bonding and sealant procedures Section 2 Contraindications and Warnings A 2 1 Contraindications Ultrasonic Systems should not be used for restorative dental procedures involving the condensation of amalgam For best performance use only ultrasonic inserts manufactured by D entsply Preventive Care e Use only Cavitron Prophy J et Cleaning Powder Section 3 Precautions 3 1 Precautionsfor All Systems U Itrasonics Do not placethe System on or next to a radiator or other heat source Excessive heat may damage the System s electronics Placethe System where air is free to circulate on all sides and beneath it T he System is portable but must be handled with care when moving Equipment flushing and dental water supply system maintenance are strongly recommended See W ater Supply under Section 9 System Care Close the coolant water shut off valve to the System every night before leaving the office T he use of an in line water filter is recommended Air Polishing T heuse of an air dryer on the compressor line supplying the System will prevent condensation from forming in the air line which in turn
27. styles of Cavitron 30K Ultrasonic Inserts are easily interchangeable for various procedures and applications See enclosed booklet for specific information O Ring Connecting Body Provides seal for handpiece Transfers and amplifies mechanical coolant motion of stack to insert tip Insert Tip L Finger Grip Magnetostrictive Stack Shape and size of tip determines access and Converts energy provided by the hand adaptation Preheated Lavage directed to Insert Marking piece into mechanical oscillations tip Manufacturer Date YYMM used to activate the insert tip Frequency Type E g DENTSPLY 9608 30K FSI SLI 10S H old the handpiece in an upright position Activate to bleed any air bubbles that might be trapped inside the handpiece Lubricate the rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE 6 4 Cavitron JET Air Polishing Insert nozzle T he Air Polishing Insert N ozzle automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs Insert Heater Rod Heats the Air Polishing Insert Tip air powder flow to insert tip delivered water for patient Tube in a tube design comfort deliveres precise air water powder mixture at point of delivery O Ring Seals water when Insert Nozzle is fully seated in Handpiece Lubricate the rubber O ring on the insert with water before placing it into the handpie
28. th the upper and lower arches the backrest of the chair should be adjusted as for other dental procedures T his assures patient comfort and clinician visibility H ave the patient turn their head to the right or left Also position their chin up or down depending upon the quadrant and surface being treated Evacuate irrigant using either a saliva ejector or H VE 8 2 PerformingUltrasonic Scaling Lubricate the rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE e Set the U nit s Power Adjustment knob to the lowest power setting for the application and the selected insert Activatethe Foot Control and hold the H andpiece over a sink or drain Check spray temperature to verify fluid is reaching the working end of the insert tip Adjust the Lavage flow knob to ensure adequate flow for the selected Power setting Greater flow settings provide cooler irrigant t may be necessary to adjust lavage with the System in Boost mode Foot Control fully depressed so adequate fluid will be available to cool tip tooth interface n general it is suggested that a feather light touch be used both supra and subgingivally The motion of the activated tip and acoustic effects in the irrigating fluid in most cases is adequate to remove even the most tenacious calculus e U se Cavitron efficiency indicator if an insert seems to require i