User Manual - PM265 / 268
Contents
1. 1 Main Power Switch 2 Power Cord Female Receptacle 3 Fuses i a 4 Air Filter 4 5 Mounting Brackets 2 3 FRONT PANEL 1 Comfort Adjust Levels The Comfort keys adjusts the pump pressure output to the mattress overlay When pressing the output pressure increases making the mattress overlay firmer Pressing decreases the air pressure making the mattress overlay softer When the pump is started it always begins in the inflation mode and then changes to Alternate Once Alternate is illuminated the carer sets the pressure to the appropriate weight of the client and adjusts afterwards for client comfort and clearance 2 Therapy XD Press the Therapy button to change the therapy mode A Inflate Mode This is the default mode each time the power switch is turned on This ensures the pump is able to reach its maximum operating pressure Once the max pressure level is reached the pump will automatically switch into alternating mode The user can also activate this function to assist with easier transfer on and off the mattress overlay surface B Alternate Mode The normal operating mode of the system The mattress overlay system will alternate every 10Mins C Static Mode The static mode inflates all mattress overlay cells equally and stops all alternation The mode can be activated or suspended by pushing on the static button The advantage of operating in a static mode is it allows caregivers to perform therapy on
2. 2 The specification are suitable for all areas operating with same power supply 3 The manufacturer reserves the right to modify the specification without notice SAVE THESE INSTRUCTIONS GROUNDING INSTRUCTIONS This product should be grounded In the event of an electrical short circuit grounding reduces the risk of electric shock by providing an escape wire for the electric current This product is equipped with a cord having a grounding wire with a grounding plug The plug must be plugged into an outlet that is properly installed and grounded DANGER Improper use of the grounding plug can result in a risk of electric shock If repair or replacement of the cord or plug is necessary do not connect the grounding wire to either flat blade terminal The wire with insulation having an outer surface that is green with or without yellow stripes is the grounding wire NOTE If the repair or replacement of the cord is necessary please contact a qualified electrician or serviceman To reduce the risk of electric shock do not modify the cord or plug in any way Check with a qualified electrician or serviceman if the grounding instructions are not completely understood or if in doubt as to whether the product is properly grounded 11 AUSTRALIAN AGENT GMS REHABILITATION 154 PRINCES HWY TRARALGON 3844 VICTORIA AUSTRALIA PH 1800 060 919 FAX 1300 734 998 EMAIL sales gmsrehab com au APEX MEDICAL S L Alameda de Recalde 50 Despac
3. a level surface Also some clients may initially find alternation prevents them from going to sleep so this mode can assist The Static automatically reverts back to alternation after 20 minutes It is not recommended or therapeutic to extend the static mode beyond 20 minutes D Sitting Mode The seat inflation feature provides additional pressure support to the mattress overlay and is usually activated during bed backrest elevation The extra pressure reduces the potential of the client bottoming out whilst sitting upright lt can be activated to work alone or in combination with the static mode Lock Unlock Panel Lock TU Should the panel remain untouched for 3 minutes the panel lock feature will lock all panel keys with green LED light on to prevent accidental changing of the setting during normal operation To unlock simply press and hold the Panel lock button for 3 seconds Alarm Mute Press alarm mute button to temporarily suspend the low pressure alarm LED light and buzzer Should the situation not be resolved within 5 minutes the alarm shall resume A PFA Power Failure Alarm During power failure situation the Power Failure LED light will light up with the buzzer sounding Pressing the mute button will disable both buzzer and LED and turn off the pump The pump will restart when power supply resumes but will need the operator carer to reset Do not have the client on the mattress overlay during power failure or low pressure
4. comfort setting can be adjusted to suit the individual o client It is essential to test for clearance of at least 50mm A between the clients sacral bottom region and the base of the mattress overlay NOTE Unzip cover and place hand between the deflated cell and clients sacrum bottom There should be 2 5cm gap when the cell is deflated If not increase the comfort level and retest If unable to achieve ring your local dealer for advice STANDARD MATTRESS 4 2 CPR ang NG When there is an emergency that requires CPR to be ARN performed on the client quickly pull the CPR valves to release air from mattress The CPR valves are located at the head end right hand side of the mattress The quick connector found on the pump unit can be disconnected for even faster deflation 4 3 PRESSURE SET UP Users can adjust pressure of air mattress to a desired softness by adjusting the comfort keys Please consult your physician allied health professional for a suitable setting Clients can lie on the mattress after the pump is running under the alternating mode this indicates the pressure has reached normal operating levels NNOTE Check to see if the suitable pressure is selected by sliding one hand between the deflated air cells and the patient to feel patient s buttocks Users should be able to feel the minimum contact 4 4 LOW PRESSURE INDICATION FUNCTION When an abnormal low pressure occurs the Low Pressure indicator yel
5. contact our local dealers 7 2 FUSE REPLACEMENT 1 Disconnect the plug from mains power when a blown fuse is suspected gt 2 Remove the cover of the fuse holder by means of a small screwdriver 3 Insert a new fuse of the correct rating and replace the cover of the fuse holder The fuse should be rated as T1A 250V type and VDE approved 7 3 AIR FILTER REPLACEMENT 1 Loosen the screws from the air filter plate located at the back of pump KA 2 The filter is reusable and can be washed gently with a mild detergent and water Dry the filter before use 3 Check and replace air filter regularly if environment is dirty 8 TECHNICAL DESCRIPTION Specifications Item Specification Power Supply Note See rating nc 230 240V 50 60 Hz 0 07A for 230V system label on the product Operation 10 C to 40 C 50 F to 104 F Temperature Storage 15 C to 50 C 5 F to 122 F SEE Shipping 15 C to 70 C 5 F to 156 F Operation 10 to 90 non condensing Storage 10 to 90 non condensing Shipping 10 to 90 non condensing Type BF IPKO Applied Part Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture No AP or APG protection Mattress Specification 8 Mattress Dimension L x W x H 200 x 90 x 20 3 cm or 78 7 x 35 4 x 8 Weight 7 5 kg or 16 5 lb Classification 10 NNOTE 1 Consult the local agent or call GMS Rehabilitation for further technical documents
6. electronic equipment For more detailed information about the recycling of this product please contact your local city office household waste disposal service or the retail store where you purchased this product 2 iad 1 D A yal 1 INTRODUCTION This manual should be used for initial set up of the system and for reference purposes 1 1 GENERAL INFORMATION The system is a high quality and affordable mattress system suitable for stage II III ischemic ulcer treatment and prevention of pressure related complications The system has been tested and successfully approved with the following standards EN 60601 1 EN 60601 1 2 EN 55011 Class B IEC61000 3 2 IEC 61000 3 3 1 2 INTENDED USE This product is intended to help and reduce the incidence of pressure ischemic ulcers whilst optimizing patient comfort Recommended for e An Individual in a home care setting and long term care who suffer from pressure ischemic ulcer s or risk of ischemic ulcer formation e Pain management as prescribed by physician health care professional A NOTE Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide 2 PRODUCT DESCRIPTION 2 1 PUMP AND MATTRESS SYSTEM a a FA p p f 1 CPR 2 Pump unit 3 Mattress System 4 Quick Connectors 2 2 PUMP UNIT s gt 7 AG tA Front tae E ig E 1 Quick Connectors h 2 Front Panel NOT Rear Dr i
7. position Grounding Before any connection to the output connectors is made the unit shall be connected to a protective earth conductor via the three core main cable the mains plug shall be inserted only into a socket outlet provided with a protective earth contact The protective action shall not be negated by the user of an extension cord without protective conductor Use only the cord supplied with the unit Several installation tips are listed below After installation the extra length of the power cord if any should be neatly arranged to avoid any tripping accidents The pump should be firmly placed in a position where users carers can access easily 4 OPERATION A NOTE Always read the operating instructions before use 4 1 GENERAL OPERATION 1 Switch on the main power switch found at the side of the pump GREEN HE 2 Press the ON OFF the pump unit delivers air to the mattress and takes approximately 20 min to fully inflate p 3 The low pressure indicator yellow LED will light up with si an alarm sounding if normal operating pressure is not reached from power on around 40mins When the appropriate pressure is reached the low pressure indicator yellow LED will go off and pump will automatically revert to the status of last power off e g enter the alternating mode GMS 4 Once fully inflated and in alternation mode with no alarms q i gt sounding the client can be placed on the mattress overlay The
8. Dynamic Alternating Therapy Pump Overlay and Replacement Mattress System PM265 PM268 Instruction Manual A GMS REHABILITATION PRODUCT Model No 9P 052 series PLEASE READ ALL INSTRUCTIONS BEFORE USE Contents Important Safeguards iia o O A AA e non KANAN 3 Product Description stemmer 3 Pump Unit AAA IAA ee eee 4 Therapy Modes amp Alarm tt ttttinnnnnnnnnnnsannnnnnnnannnninniiicnnni 5 RE A AN 6 Operation uses gt Cleaning seeren 3 Storage vassere eres 9 Meier EE cele 9 Technical Description mie 10 Grounding Instructions tienen 11 IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USING DANGER To reduce the risk of electrocution NO of WD gt Always unplug this product immediately after use Do not use whilst bathing or in wet areas Do not place or store product where it can fall or be pulled into a tub or sink Do not place in or drop into water or other liquid Do not reach for a product that has fallen into water Unplug immediately Do not open the pump casing return to service centre for repair Opening the casing pump housing will void warranty WARNING To reduce the risk of burns electrocution fire or injury to persons 1 2 Close supervision is necessary when this product is used by on or near children or invalids Use this product only for its intended use as described in this manual Do not use attachments not recommended by the manufacturer Ne
9. ho 304 48008 Bilbao Vizcaya SPAIN APEX MEDICAL CORP 9 Min Sheng St Tu Cheng Taipei County 236 Taiwan
10. low LED will light up along with the buzzer Check to ensure all connections are connected properly and that they are correctly installed as per installation instructions ANOTE If the pressure level is consistently low check for any leakage tubes or connecting hoses If necessary replace any damaged tubes or hoses Contact local qualified dealer for repair Do not continue to use the product in a low pressure condition The Low Pressure indicator will light up until the low pressure fault condition is resolved 4 5 STATIC MODE Press the STATIC button to suspend the alternating mode of mattress By pressing this button again the mattress will switch back to alternating therapy mode This will auto revert to alternating mode after 20 minutes 4 6 ALARM MUTE When a low pressure situation has occurred both the LED light and buzzer will sound off to warn the client carer Press the button to temporarily mute the buzzer so the care giver can check for possible air leaks Should the situation not be resolved within 5 minutes the alarm will resume beeping 5 CLEANING It is important to follow the cleaning procedures before using the machine on human bodies to reduce the risk of infection Wipe the pump unit with a damp cloth pre soaked with a mild detergent and keep it away from dust If other detergent is used choose one that will have no chemical effects on the surface of the plastic case of the pump unit CAUTION Do not imme
11. rse or soak pump unit Wipe the mattress unit with a damp cloth pre soaked with warm water not to exceed 65 containing a mild detergent and keep it away from dust The cover may also be cleaned using sodium hypochlorite diluted in water 1 concentration All parts should be air dried thoroughly before use CAUTION Do not use phenolic based product for cleaning CAUTION After cleaning dry the mattress without direct exposure to sunlight The carrying bag if available should be turned inside out and completely wiped down using the disinfectant solutions Allow it to air dry thoroughly Once the inside is dry turn it back and wipe down the outside of the bag with disinfectant solutions 6 STORAGE 1 To store disconnect air hoses and remove CPR plugs 2 Lay the mattress out flat and upside down 3 Roll from the head end towards the foot end 4 Foot end strap can then be stretched around the rolled mattress to prevent unrolling A Note Do not fold crease or stack the mattresses 7 MAINTENANCE 7 1 GENERAL 1 Check main power cord and plug to ensure there are no abrasions or excessive wear 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are held together correctly 3 Check airflow from the air hose connector The airflow should alternate between each connector every half cycle time if it s in alternating mode 4 Check the air hoses for any kinks or breaks For replacement please
12. situations B Low Pressure Indicator When low pressure LED lights up the pressure inside of air mattress is below normal Please refer to troubleshooting 5 Service LED This service LED will illuminate when the unit fails due to a mechanical fault Please contact your local service technician or call GMS Rehabilitation on 1800 060 919 DO NOT CONTINUE TO USE THE PRODUCT remove the client from this surface 3 INSTALLATION Unpack the box to inspect for damage which may have occurred during shipment lf there is any damage please contact your dealer immediately 3 1 PUMP 8 MATTRESS INSTALLATION ON Place the mattress overlay on top of the bed frame Please note the foot end Overlay to go on top of existing foam mattress Mattress replacements on to bed base Hang the pump onto bed rail foot end and adjust hangers or place the pump on the floor in safe position Connect air hose connectors from air mattress to the pump unit When a click sound is felt or heard the connection is completed and secured NNOTE Check and ensure the air hoses are not kinked or tucked under mattress Plug the power cord into electrical outlet and pump unit N NOTE 1 Make sure the pump unit is suitable for the local power voltage CAUTION The pump can only be used with the mattress recommended by the manufacturer Do not use it for any other purpose 5 Then turn the main Green power switch to the ON
13. ver operate this product if it has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water Return the product to GMS Rehabilitation or authorized service centre for examination and repair Keep the cord away from heated surfaces Never block the air openings of this product or place it on a soft surface such as a bed or couch where the openings may be blocked Keep the air opening free of lint hair and other similar particles Never drop or insert any object into any opening or hose Connect this product to a properly grounded outlet only See Grounding Instructions NOTE CAUTION AND WARNING STATEMENTS NOTE Indicates some tips CAUTION Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the equipment or other property WARNING Calls attention to a potential danger that requires correct procedures or practices in order to prevent personal injury Mm lt W O r wn Authorised representative in the European community Manufacturer BF symbol indicates this product is manufactured according to the degree of protection against electric shock for type BF equipment Grounding terminal for 230V system Attention read the instructions Disposal of Electrical amp Electronic Equipment WEEE This product should be handed over to an applicable collection point for the recycling of electrical and