1. CAUTION Maintain cleanliness of electrodes between main unit and mesh cartridge If Medication or saliva on the outside of the mesh surface will block nebulization absorb with gauze Disinfecting Disinfect the medication bottle mesh cartridge mask adapter mouth piece and mask by boiling at 100 C for 2 minutes Don t boil the mesh cartridge for more than 2 minutes or damage may occur When disinfecting with disinfectant soak parts in a 0 196 solution of Benzalkonium Chloride for 10 minutes Don t soak in disinfectant for more than 10 minutes Other types of disinfection agents may cause rust within the unit You must rinse off disinfectant with clean water after disinfection and dry with sterile gauze Don t dry with cotton or cloths as they may be con taminated Dry completely Reassemble Make sure you do not lose the silicone ring which pro vides a proper seal when the cover is closed If the silicone ring sepa rated from the cover put it back on its place CAUTION DO NOT BOIL MAIN UNIT BATTERY AC ADAPTOR amp USB CHARGER OR BATTERY COVER DO NOT BOIL THESE PARTS Main unit Battery AC adaptor amp USB charger Storage Keep unit and accessories dry Keep unit out of the reach of small children Always keep unit unplugged when not in use Do not store battery in direct sunlight high temperatures or in humid conditions Do not place battery
2. This symbol shall accompany the name and address ofthe European Community EU representative authorized representative WARRANTY This device is guaranteed for 1 year from the date of purchase The guar antee does not extend to consumable parts such as mesh cartridge mask mouthpiece The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming date of purchase and the receipt Should the unit be damaged due to user misuse or negligence the manufacturer or authorized dealers shall not be responsible for any damage or loss to the unit All assembly extension adjustment or repair of this unit must be performed only by the authorized service centre of distributor 21 22 in your country For repairs or to purchase parts please call authorized service centre of distributor in your country SPECIFICATIONS MODEL NE SM1 Model NE SM1 Power amp Consumption Lithium ion Battery 3 6 4 2V DC 1 4W Dimensions W x D x H 52 mmx53 mmx109 mm Weight Main unit 88 0 mask 50 9 Nebulization rate Approximately 0 25 ml min 23 Particle size Average 5 um Volume of medication bottles 6 ml Operating time with full charge 5 hours 30 minutes Charging life cycle of battery Over 500 cycles battery life differs according to usage and charging conditions Ultrasound frequency Approximately 108 KHz Ope
3. on metal or wet surfaces or inside metal boxes Such placement could short circuit the battery s positive amp negative contacts and cause the battery to heat or explode Do not place unit where it could be exposed to excessive impact or vibration Do not expose unit to chemical or explosive gases Maintenance Repairs must be performed by authorized personnel Disassembling unit or battery may short circuit the battery causing accidents Should the main unit become contaminated wipe with sterile gauze Don t immerse main unit or battery in water as this may cause damage and electrical shock Don t touch mesh vibrator or insides of medication bottle as this may cause breakdown or contamination The unit must be washed after use to prevent contamination with germs If medication or saliva sticks to the mesh cartridge it may be broken If medication or saliva sticks to the mesh cartridge you can remove contaminants by boiling the mesh cartridge The mesh cartridge is 17 consumable If there is residual medication in the bottle after use liquid can leak through the mesh Wipe with sterile gauze Replacing Battery 1 Open battery cover with a 2 Insert the provided lithium ion battery locker or coin battery into the main unit in the proper polarity direction e 3 When inserting the battery cover arrow marks must be aligned Close the battery cover with a battery locker or use a coin while pushing and turning
4. APARAT DE AEROSOLI PORTABIL SUPER MESH USER MANUAL Model NEPLUS SM1 reo DEAR NePlus SM1 SUPER MESH NEBULIZER OWNER Thank you for choosing the Ultrasonic Nebulizer model NE SM1 We are sure that having appraised worthily the high quality and reliability of this device you will become a regular user of the our products Please read this manual carefully before operating the unit For any ques tions arising during unit operation please contact service center TABLE OF CONTENTS INTRODUCTION What Is Nebulizer and How Does It Work What Is Nebulizing Therapy Used Advantages of Nebulizing Therapy Mesh Technology Advantages of Nebulizer NE SM1 BEFORE STARTING TO USE THE UNIT Essential Safety Rules Important Information Product Components Product Descriptions Use Procedure HOW TO USE THE UNIT Assembly How to Remove the Mesh Cartridge Charging the Internal Battery Filling with Medication Basic Operation HOW TO MAINTAIN THE UNIT Gl ahirit ent nets Disinfecting Storage Maintenance Replacing Battery Replacing Mesh Cartridge TROUBLESHOOTING SYMBOL DESCRIPTION WARRANTY SPECIFICATIONS APPENDIX QUICK USE GUIDE INTRODUCTION What Is Nebulizer and How Does It Work The nebulizer system is a device that turns a liquid d
5. bacterial infections disinfect and wash medication Electrical connections must comply with relevant regulations bottle mouthpiece mask adapter and mask before use Disinfect and wash after purchase and after long periods of storage Follow your physician prescribed regimen of drug therapy or symptoms may worsen Thoroughly disinfect wash and dry medication bottle mesh cartridge mouthpiece mask adapter and especially the mask after each use 6 Rinse thoroughly with water after disinfection Symptoms may worsen if disinfectant remains on surfaces 7 Don t turn on power without medication as this may cause damage to the mesh cartridge 8 Don t immerse the main unit in water This may cause damage and or electrical shock 9 This unit has small parts Keep these parts away from infants and small children to prevent swallowing 10 This unit is a medical device intended for use with a single patient KEEP THESE INSTRUCTIONS IN A SAFE PLACE FOR FU TURE REFERENCE Important Information Use only factory packed batteries with the unit Using or charging un specified batteries will cause loss of warranty endanger the user and may cause injury from fire Do not use battery with equipment other than the NE SM1 nebulizer or attempt to modify battery for other uses To save battery power and avoid accidental operation always turn the Switch to off when not in use Don t use water for inhalation Don t hit dro
6. children elderly and infirm patients Mesh Technology Advantages of Nebulizer NE SM1 1 Nebulizer NE SM1 belongs to new generation of ultrasonic nebu lizers Membrane mesh nebulizers with Vibrating MESH Technology VMT operate on an advanced mechanism vibrating membrane with many microscopic holes mesh through which liquid drug passes and turns into an aerosol Ultrasonic vibrating mesh releases medication via 4 my of fine holes Vibrations of ultrasonic mesh do not damage the large molecules as it happens with compressor nebulizers Membrane nebulizer NE SM1 is featured in very small residual volume of medication bottle approximately 0 1 ml The unit provides high lung deposition of medications carries out nebulizing of solutions and sus pensions also medications with large molecules proteins peptides insulin glucocorticosteroids and antibiotics Portable device of pocket size operates on built in battery delivers medication to upper and lower respiratory tracks The device allows to nebulize at different angles It can be used to nebulize in lying position children infants Totally noiseless technology allows to nebulize sleeping child The device can be used both for adults and for children The device is simple in operation only one button BEFORE STARTING TO USE THE UNIT Essential Safety Rules 1 Operate the unit only as directed by this manual 2 3 To avoid viral or
7. clockwise 4 Close battery cover until a click is heard 18 NOTE Use only batteries supplied by the manufacturer Replace battery pack after 500 cycles of discharge charging Batteries are recyclable At the end of their useful life various state and local laws may prohibit battery disposal into municipal waste streams Replacing Mesh cartridge 2 Install new mesh cartridge into medication bottle as shown in diagrams 1 Remove mesh cartridge NOTE Use only Mesh cartridges supplied by the manufacturer At the end of one year replace Mesh cartridge with a new one or sooner if need arises To replace the mesh cartridge contact service center of TM Dr Frei At the end of its useful life state and local laws may prohibit disposal of this battery into municipal waste streams 19 20 TROUBLESHOOTING In case of problems please refer to the table shown below If problems persist please contact service centre TM Dr Frei in your country Problem Cause and Solution Power on green lamp does not come on Battery inserted in wrong direction Correct battery direction see page 20 The battery is exhausted Charge battery see page 14 The unitis charging The unit doesn t work and yellow battery indicator lamp is on indicating charging Disconnect charger before use Power lamp is on but unit does not nebulize The unit is tilted too much Straighten the unit as nebu
8. ee page 16 NOTE Clean and disinfect nebulizer parts and applicable inhalation attachments before initial use or after long periods of storage M J N STEP 2 HOW TO INSERT amp CHARGE BATTERY See pages 14 amp 20 NOTES Insert battery after purchase Charge battery if yellow battery indicator lamp blinks Jl C STEP 3 HOW TO FILL WITH MEDICATION See page 15 NOTES Follow a physican prescribed regimen of drug therapy Do not drop medication or liquid onto the main unit P Y gt STEP 4 HOW TO INSTALL THE INHALATION MASK See page 12 NOTES Nebulizing will not occur if the units is tilted at too great an angle Be sure to clean washable parts after each use v H STEP 5 HOW TO OPERATE THE NEBULIZER See page 15 NOTES Do not turn on power without medication Do not turn on power while battery is charging Nebulizing will not occur when the unit is tilted at too great an angle HOW TO USE THE UNIT Assembly 1 Insert the medication bottle into the main unit until a click is heard or the unit may not work correctly 2 Insert the mesh cartridge at a right angle to the medication bottle then turn downwards and push until a click is heard 3 Insert the mesh cartridge and push down as seen in the diagram or unit may not work correctly 4 Insert the mask adapter into the main unit until a click i
9. lization will only occur when medication touches the mesh cartridge see page 15 edication or saliva sticks to mesh cartridge gt Wash according to cleaning instructions If unit does not nebulize even after washing replace m cartridge with a new one see pages 13 amp 1 The mesh is broken Replace mesh cartridge with a new one It is replaceable see page 21 The electrodes between main unit and mesh cartridge are dirty 3 Wipe lightly with moist gauze see page 18 The medication bottle isn t properly connected to the main unit 3 Insert medication bottle tightly into the main unit until a click is heard see page 12 Nebulizing rate decreases Mesh cartridge is dirty Wash according to cleaning instructions If washing does not improve nebulization replace mesh cartridge with a new one see pages 13 amp 21 The yellow low battery lamp is blinking gt Charge battery see page 14 SYMBOL DESCRIPTION N This symbol shall accompany the manufacturer s serial number Serial Number Ll Consult enclosed documents Operating instructions e A This symbol shall accompany equipment Applied part classification of type B IP22 This symbol shows degree of P rotection against dust and water When sprayed vertically at a 15 tilt for 1 minute IP degree t This symbol shall accompany manufacturers name and address Manufacturer
10. ocker if yellow indicator ora coin lamp blinks 6 Charging is complete when yellow LED turns to green LED Charger 3 Close medication cover until a click is heard How to use the unit 1 Install the mask adapter 2 Push the power 3 Inhale amp inhalation mask button to start stop Green LED Caution during use Inhalation is impossible Do not power the unit If medication leaks before at angles over 75 without medication use wipe with gauze and use unit promptly How to Clean after use A Insert the medication Boil all washable parts Dono se with until tightly into the main unit for 2 minutes at least 9 a click is heard once a week Do not dry with Do not touch mesh with For cleaning wipe parts with microwave oven pins or other objects gauze not cotton or cloths
11. p or cause impact to this instrument Don t drop medications or liquids on the main unit particularly the charging inlet as this may cause damage and or electric shock Don t disassemble modify or repair the main unit as this may cause fire and or electrical shock Don t touch mesh with fingers or any objects The unit can t be used while charging the battery This unit is approved by IEC60601 1 please use in areas free from potential electromagnetic waves Inhalation masks are applied part of type B any other parts are ac cessible part Please separate the battery from unit when not in use for long periods Please dispose of properly at the end of useful life following relevant state and local laws Battery disposal may be illegal in municipal wmm waste streams Product Components Mesh cartridge NE SM1 1 Medication bottle NE SM1 2 Mask adapter NE SM1 3 Small inhalation mask NE SM1 4 Large inhalation mask optional NE SM1 5 Battery locker NE SM1 6 Rechargeable lithium battery NE SM1 7 Mouthpiece NE SM1 8 Product descriptions CD Medication bottle and bottle cap ON OFF switch amp The main unit ON lamp green and Battery indicator lamp yellow amp Mesh cartridge amp Mask adapter Inhalation mask Battery Battery cover Use procedure 7 STEP 1 HOW TO CLEAN THE NEBULIZER S
12. rating environment Temperature 10 40 C humidity 30 8595 Pressure 70 106 kPa Battery charging environment 0 45 C Storage environment Temperature 20 60 C humidity 10 9596 R H Pressure 70 106 KPa Expected service life Main unit 5 years from purchase Mesh cartridge 1 year from purchase Instrument warranty 1 year from purchase Mesh car tridge is consumable part Contents Nebulizer unit mask mouthpiece AC adaptor amp USB charger user s manual carrying bag Battery Fairman 1ICR19 66 3 7V 2600 mAh Li ion Recommended charger IEC 60950 1 approved travel adapter input 100 240 V 50 60 Hz output DC5V 500 1 000 mA 1 Particle size was measured in accordance with European standard EN 13544 1 2 The device NE SM1 meets the requirements of European standard EN 13544 1 TECHNICAL DATA DISTRIBUTION OF AEROSOL CLOUD MASS ACROSS PARTICLE SIZES 100 2 15 00 Cumulative volume Es lume 96 10 00 i 0 00 10 T Particle size u 23 24 APPENDIX QUICK USE GUIDE How to insert the battery 1 Open the battery cover 2 Insertthe battery with battery locker or a coin 3 Put battery cover align with 2 arrow marks 4 Push and close the battery 5 Charge battery cover with battery l
13. rug into a fine aerosol under the influence of ultrasound An aerosol contains larger and smaller particles When breathing the gravity makes the larger particles accumu late on the walls of the upper respiratory passages while the smaller par ticles have enough time to reach the lower bronchial branches Therefore the drug comes to all parts of the patient s bronhial tree including the smallest bronchi and alveole and gets into the blood vessels on the walls of the alveole in big therapeutic dozes with tiny or none side effects What Is Nebulizing Therapy Used for Nebulizer systems are used by the patients with acute respiratory diseases obstructive pulmonary disease chronic inflammatory processes in respiratory passages such as chronic bronchitis bronchial asthma incl chronic or unstable ones exacerbation of asthma chronic pharyngitis respiratory passages infections abnormalities in secretion quality and secreting in bronchi different types of coughing Advantages of Nebulizing Therapy Nebulizing therapy gives a possibility to quickly effect the inflammatory zone by big drug dozes which increases the influence of the treatment and reduces the risk of side effects to other organs Nebulizing therapy is the easiest way of inhalation treatment and it is also an economical means of treatments from the point of view of purposeful use of drug Aerosol treat ment with the help of nebulizing systems can be used for
14. s heard 5 Install mask adapter or mouth piece mask to the unit mask for adult is optional How to remove the Mesh cartridge 1 Remove the medication bottle by pushing buttons on both sides of the bottle 2 Push the Battery locker or a coin into the edge of the medication bottle in the direction of the arrow as seen in the diagram The mesh cartridge will pop out 11 Charging the internal battery 1 Charge battery after use if the yellow battery indicator lamp blinks during use Use a USB power cable for PCs or a micro 5pin USB smart phone charge adapter individual purchase If you use the unit without charging the unit will stop automatically 2 Connect the charge adapter to the charge receptacle in the main unit after inserting the specified battery into the main unit charging will begin Over 90 will charge within 3 hours Low battery 1000 mA charger adapter Charging is complete when the yellow LED changes over to the green LED Remove the USB charger from the unit CAUTION Don t use the unit when the yellow battery indicator lamp is on during charging Charge battery after emptying residual medication from the bottle Charging time required approximately 3 hours Charging life cycle of battery over 500 cycles NOTE Use a Micro 5pin USB smart phone charge adapter IEC 60950 1 approved travel adapter input 100 240V 50 60 Hz output DC5V 500 1000 mA as sho
15. wn in the following diagram Filling with medication 1 Following a physician pre Use the unit as soon as the Push up the hook on the medication bottle as seen in diagram The bottle cover will open Make sure there is a silicone ring on the cover which provides a proper seal when the cover is closed scribed regimen pour medi cation into the medication bottle and close the cover until a click is heard medication bottle has filled Basic operation 1 Inhale gas while holding down the sides of the unit Push the on off switch in front of the unit nebulizing will begin as seen in diagram In cases where the inhalation angle is tilted more than 75 nebulizing will not occur because mesh must be soaked with medica tion Model 1 Nebulizing stops if the on off y switch is pushed again Model 2 If the on off switch is pressed for more than 1 5 seconds the unit nebulizes only while the switch is pushed In this way you can synchronize nebulizing and inhalation so that medication can be used efficiently 13 14 4 The unit will stop automatically after 20 minutes The unit may also stop if running out of medication CAUTION The rate of nebulization may be greatly reduced if used with a high viscosity medication or a medica tion at low temperature If the unit isn t used within 10 minutes after filling with medication liquid may leak If so wipe with gau
16. ze or turn on power Nebulizing will begin after 2 seconds of absorption Do not leave children alone while using the nebulizer HOW TO MAINTAIN THE UNIT Cleaning 1 Dispose of residual medication after use Don t store medication or carry the unit while filled with medication Medication may leak and or the unit may become damaged After filling with water nebulize for 60 seconds in order to clean mesh 2 Wash and dry all the following parts completely after every use medication bottle mesh cartridge mask adapter mouthpiece and mask 3 Washable parts medication bottle mesh cartridge mask adapter mouthpiece and mask WASHABLE PARTS Mesh cartridge Medication Mask adapter Mouthpiece T 4 There is no need to remove the mesh cartridge from the medication bottle for cleaning Put clean water in a container Gently shake parts in water after disassembly Do not clean parts under running water 5 Dry parts with sterile gauze after washing 5 Main body Clean with a soft cloth moistened with water and a non abrasive cleanse 7 Keep all parts dry 8 Reassemble CAUTION Never immerse the main unit in water as this may damage the unit Do not clean parts under running water to avoid possible damage Gently shake parts in water Do not touch mesh with fingers Never dry in a microwave oven 15 16 Always unplug this product before cleaning
Download Pdf Manuals
Related Search
Related Contents
① RVMパンフレット - 株式会社エコスポット・ソリューションズ IDEAL Security SK110W Instructions / Assembly EPSON GP-700 詳細取扱説明書 Incidents reliés à l`eau potable Garmin 430 8T0 054 630 A MA PDC Audi A4 + A5 Sportback.fm Hotpoint DY46 User's Manual Article 17 Instant tips for right and effective approach to solve HPLC Copyright © All rights reserved.