Home

A microcontroller based pacemaker tester

image

Contents

1. interval T Then it puts outa SQR SSQ wave 40MS wide delayed by an interval D which is more than the pacing refractory period from the last pacer pulse The pacemaker senses this wave The delay time D is then decremented in steps during every cycle until the SQR SSQ waveform is not sensed by the pacemaker The maximum value of the interval D for which the pacemaker does not sense this wave is the pacing refractory period The sensed refractory period is measured in a similar manner except that the tester now generates two SQR SSQ waves instead of one The first wave is generated at a fixed delay from a pacer pulse which is greater than the paced refractory period The pacemaker always senses this wave The second one is generated at a delay D from the first wave Initial value of D is selected to be greater than the sensed refractory period Hence the first time on the pacemaker senses both the waves On subsequent cycles the delay D is reduced in steps until the pacemaker is unable to sense the second one The maximum value of D for which the pacemaker does not sense the second wave is the sensed refractory period For sensitivity measurement also different waveforms Square and Haversine are selectable with widths varying in steps from 10 to 200MS Once a waveform with a particular width is selected the tester first generates this waveform with an amplitude of 100mV irrespective of the sensitivity control setting and pacer rate cont
2. normal test operation and is also used as temporary storage during floating point calculations The pacemaker output directly feeds into an input selection logic network which selects the test load using a multiplexer MUX based on the manufacturer s specifications The output of this network feeds into an ADC for converting the analog signals into digital data for subsequent processing The digital data output of the ADC is saved in the RAM A digital to analog converter DAC is used to generate different test waveforms for the pacemaker An LCD display is used for menu display and output indications The keypad selects the particular test to be performed The power supply unit provides regulated 5V dc for the whole system and 2 5V dc to the reference inputs of ADC and DAC This power is derived from a 9V dc battery using a charge pump and a few voltage regulator chips 3 1 Microcontroller The microcontroller selected is a derivative of the well known 80C31 the Philips 80C52 1 This is an 8 bit microcontroller which contains an 8kx8 ROM a 256x8 RAM four 8 bit ports Po Pi P2 P3 three 16 bit counters timers a six source two priority level 406 Journal of Network and Computer Applications 1996 19 405 413 Power supply gt Microcontroller Pacer Selection Logic Figure 1 Block Diagram of Pacemaker Tester 10uF Address Decoder 3V4 High order 10k Address Chip Select Signals Control Signals ADC
3. wave The signal level at this point must be higher than that for point 2 Once the pacer is on the next step is to detect the termination of the pacer This is achieved simply by comparing the input signal level with the selected threshold value Once the signal level falls below this value the pacer is assumed to have ended Having detected the occurrence of a pacer the next step is to perform various measurements These tests are selected from the displayed menu using the keypad To measure the pacer rate a 1MS timer is started whenever the tester senses the beginning of a pacer pulse The timer is stopped at the beginning of the next consecutive pacer pulse Now a count of the number of 1MS timer interrupts is a good measure of the pacer rate One of the microcontroller timers is used for this purpose Pacer width is measured based on the total number of ADC samples while the pacer is on Hence pacer width accuracy depends on the sampling rate of ADC Pacer amplitude measurement is done by measuring the voltage developed across the precision load resistance with the pacemaker output current flowing through it This voltage when divided by the precision load resistance gives the amplitude A V delay is measured by counting the number of IMS timer interrupts between an atrial pacer and the subsequent ventricular pacer The measurement of paced refractory period uses a few cycles of the pacemaker output First the tester measures the pacer to pacer
4. Control Lines Data Bus Figure 2 Microcontroller Address Decoder and Control Signals nested interrupt structure a serial I O port and on chip oscillator and clock circuits The main program is written in assembly code and the compiled code is stored in the internal ROM The program code depends on the different measurements to be carried out The Ports P and Po are used as address and data busses respectively Port P is used for controlling the keypad and Port P3 is used for controlling the analog to digital converter ADC An 8 bit latch Ly is used for generating extra control signals for the LCD display digital to analog converter DAC and the multiplexer MUX A simple RC circuit is used for initializing 407 Journal of Network and Computer Applications 1996 19 405 413 the microcontroller on power up RST as shown in Figure 2 A good choice is R 10k and C 10uF A crystal of 11 0592 MHz is a good choice for clock frequency An 8 bit latch 74C373 L is used for latching the lower order address byte from Po bus The higher order 8 bits of the address is available directly from P2 port A 3X8 address decoder is used to generate the chip select CS signals for the different devices in the tester The CMOS decoder 74C138 is used for this purpose The 3 enable signals together with the 3 address lines in 138 can be carefully selected to generate the proper decoder outputs The RXD P3 and TXD P3 leads can be used to transf
5. Journal of Network and Computer Applications 1996 19 405 413 A microcontroller based pacemaker tester D Radhakrishnan Division of Computing Systems School of Applied Science Nanyang Technological University Nanyang Avenue Singapore 639798 The design of a microcontroller based pacemaker tester is presented in this paper Both invasive and non invasive transthoracic pacemakers can be tested using this device Two test loads are provided a 500 ohms load for testing invasive pacemakers and a 50 ohms load for testing non invasive pacemakers A two line LCD display is used to indicate the test parameters test results and the test in progress The test results can be conveniently transferred to a PC or to a printer through the RS232C serial port provided The unit can be packaged inside a small package and hence can be used as a handheld tester 1 Introduction Electrocardiogram ECG is one of the common tools used today for the diagnosis of a large number of diseases The number of heart patients is growing at an enormous rate each year The use of pacemakers external and implanted to correct any malfunctioning of the heart is a very common practice among cardiac patients Since a pacemaker is a very crucial device the importance of its testing and calibration need not be overemphasized The pacemaker tester described in this paper can be used to test a large variety of external pacemakers used today Since the range of parameters and co
6. e Generator Technical Reference Manual Medtronic 5880A External Demand Pacemaker Technical Reference Manual Medtronic 5330 AV Sequential Demand Pacemaker Technical Reference Manual Medtronic 5345 DDD Temporary Pacemaker Technical Reference Manual Physio Control LIFEPAK 9P Pacemaker Defibrillator Technical Reference Manual HP CodeMaster Defibrillator Pacemaker Technical Reference Manual MDE Escort 300A Defibrillator Pacemaker Technical Reference Manual Zoll PD100 Pacemaker Defibrillator Technical Reference Manual 10 Bio Tek Pacemaker Tester User Manual 11 Environics PaceAlyzer PFA 1A Technical Reference Manual OMANNDNMNBPWNK 412 Journal of Network and Computer Applications 1996 19 405 413 Damu Radhakrishnan received the B Sc degree in Electronics and Communication Engineering from the University of Kerala India in 1968 and M Tech and Ph D degrees from the Indian Institute of Technology Kanpur India and the University of Idaho Moscow Idaho USA both in Electrical Engineering in 1975 and 1983 respectively From 1983 to 1985 he was an Assistant Professor in the Department of Electrical Engineering Old Dominion University Norfolk Virginia and from 1985 to 1990 with the Department of Electrical Engineering University of Idaho Moscow Idaho From 1990 to 1994 he served as the Director of R amp D at Netech Corporation New York Presently he is with the School of Applied Science at Nanyang Technological Univ
7. ed by a differential amplifier gain 10 before feeding to the analog to digital converter 408 Journal of Network and Computer Applications 1996 19 405 413 3 3 Analog to digital Converter ADC An 8 bit parallel analog to digital converter ADC1205 is used for converting the pacer signal to digital data The ADC control leads WR RD INT STS and CLK are fed from the microcontroller The chip select CS line is controlled from the address decoder logic The ALE output of the microcontroller feeds the clock input CLK to ADC The frequency of this signal is 1 6 of the microcontroller clock frequency The data output lines are tied to the data bus on the microcontroller DAC AD7528 Data 20k SQ SIN SSQ Waveform Output 10k To 80C52 PORT 3 Figure 4 Waveform Generator using Digital to analog Converter 3 4 Digital to analog Converter DAC A dual channel 8 bit DAC AD7528 is used to generate the test waveforms Sine SIN Square SQR and Haversine SSQ as shown in Figure 4 Three separate look up tables are used to store these waveforms The SQR and SSQ waveforms are used to test the sensitivity of the pacemaker as well as its refractory period The SIN waveform is used for testing the noise immunity Of the two channels in the DAC one of them is used to control the amplitude of the generated waveform Channel A This allows 256 different amplitude selections for the waveform 4 Test Generation A co
8. er data to a PC through an RS232C interface SQ SIN SSQ Waveform from DAC Dual MUX ADC ADC1205 Ventricular Input To Microcontroller Figure 3 Input Selection Logic and Analog to digital Converter 3 2 Input Selection Logic The pacemaker output leads feed directly into the input selection logic For normal testing these outputs are connected to the Ventricular input leads of the tester as shown in Figure 3 The Atrial input leads are connected to the Atrial output leads on the pacemaker during A V delay measurement The input network is responsible for the selection of different loads to be connected to the pacemaker and also for amplifying the pacer pulses before feeding to the input of ADC Two loads are provided one a 50Q and another 500Q Invasive pacemakers are always tested using the 500Q load and non invasive ones using the 50Q load A slide switch S1 is used for selecting the load When S1 is open the load seen at the ventricular input leads is 500Q otherwise 50Q for transthoracic pacemakers In transthoracic pacemakers the current level could go to as high as 200mA Hence an input attenuation network is included to scale down the voltage This attenuation network feeds into a dual channel multiplexer DG409 The scale factors used are 1 10 and 1 100 at Xo and X inputs of the MUX Similar attenuation values are provided on the atrial pacer input leads for X2 and X3 respectively The MUX output is amplifi
9. ersity in Singapore His research interests include VLSI design fault tolerant design of digital systems microcomputer applications and switching theory 413
10. hese tests with the given default values or running one or more of them with user selectable values or skipping one or more of these tests If the user modifies a value this modified value can be saved so that it can be used for subsequent tests as long as the unit is not switched Off The user can also leave the modified value unsaved so that the original default values will be output during subsequent tests The test sequence continues until all the parameters are tested Once a test is run on the tester it stores the measured value for future reference At the completion of the test sequence the results can either be sent to a printer or to a PC through the RS232C interface At this time both the set and measured values can be compared and if there is a mismatch of more than a certain percentage then a comment to that effect can be printed 5 Conclusion A simple low cost pacemaker tester is presented here The design is based on the Philips 80C52 microcontroller All test results are based on three elements the crystal oscillator voltage reference source and the precision load resistance Time measurements use a IMS timer controlled by the crystal oscillator running at 11 0592MHz Voltage and current measurements use a precision load resistance and the reference voltage source The reference voltage source Vref is set at 2 5V dc References Philips 1994 80C51 Based 8 Bit Microcontrollers The Netherlands Medtronic 5375 Puls
11. mplexity of operation varies considerably among different manufacturers it is a very difficult task to integrate all these requirements into a single tester The design of pacemaker testers must be made based on easily available components The selection of a specific microcontroller plays a major role in the design Since the pacer widths of earlier pacemakers were below 1MS analog to digital converters ADC must be carefully selected to provide the required accuracy The overall cost of the system must also be a major consideration 2 Terminology The different parameters related to pacer pulses are given below e Pacer rate is the number of pacer pulses delivered per minute PPM by the pacemaker e Pacer Width is the duration of a pacer pulse MS e Pacer Amplitude is the current delivered by the pacemaker at a particular output setting e Refractory period is the time interval during which the pacemaker is insensitive to any external inputs If a QRS is detected during this period the pacemaker ignores it On the otherhand if a QRS is detected outside the refractory interval then the pacemaker resets its internal timer and the next pacer pulse is generated after a delay of one period from this QRS wave There are two refractory periods pacing and sensing 405 Journal of Network and Computer Applications 1996 19 405 413 e Pacing Refractory period is the maximum time interval after the generation of a pacer pulse during which
12. mprehensive set of tests for the pacemaker include the measurement of pacer amplitude pacer width pacer rate refractory periods sensitivity A V delay and noise immunity In fact the pacemaker must be tested for its full range of parameter values as specified by the manufacturer Hence for example pacer amplitude measurement must be done under varying test loads at various current settings refractory period and sensitivity measurements using signals of varying shapes pacer rate at various rate settings and so on To this end the individual characteristics and parameter ranges of different pacemakers are first tabulated 2 9 A comprehensive set of such 409 Journal of Network and Computer Applications 1996 19 405 413 requirements is given in Table 1 This table lists the pacemaker parameter ranges for both invasive and non invasive pacemakers Furthermore Square and Haversine waveforms are recommended by different manufacturers for testing the refractory periods as well as the demand sensitivity Based on the parameter ranges given in Table 1 and the test waveform requirements a comprehensive list of tester parameters with appropriate test ranges are identified This is given in Table 2 Table 1 Pacemaker Parameter Test Ranges Test Parameters Invasive Pacemakers Non invasive Pacemakers Rate PPM 30 180 30 180 Current mA 0 1 20 0 200 Load ohms 300 1200 Sensitivity mV 1100 A V Delay MS _ o 300 50 60 Hz Noise Immu
13. nity 100mV at max sensitivity Table 2 Parameter Specifications for the Pacemaker Tester Test Parameters Invasive Pacemaker Non invasive Pacemaker Test Load ohms Amplitude mA 0 1 25 1 200 Pacer Rate PPM 20 1000 20 1000 Pacer Width MS 0 8 80 0 8 80 Refractory Period 10 25 40 100 200MS 10 25 40 100 200MS Sensitivity SQ SSQ Waves SQ SSQ Waves 4 1 Algorithm The first requirement in generating an algorithm is to detect the pacer waveform in the presence of noise signals This depends on the signal level of the waveform As the signal level decreases detecting the waveform becomes more and more difficult A three point detection method is used to detect the presence of a pacer wave First a minimum signal threshold value is selected so that when the input level is below above this represents the absence presence of a pacer wave This is fixed based on the minimum pacer amplitude Once the threshold value is fixed the following criteria is used for the selection of the three points 1 The first point is selected to be lower than the specified threshold value 410 Journal of Network and Computer Applications 1996 19 405 413 2 The second point identifies the starting of a waveform The magnitude at this point must be greater than the selected threshold value 3 The third point is chosen so that the waveform can be separated from transient noise spikes and also to indicate the presence of a sustained pacer
14. rol setting on the pacemaker as long as the pacing rate is set below 150PPM This waveform is delayed from the pacer pulse so that it is outside the pacing refractory period The tester then checks whether or not this wave is sensed by the pacemaker If it is not sensed then a message exceeded is displayed which means that the pacemaker needs an amplitude more than 100mV for sensing at that setting If the wave is sensed the tester then reduces the amplitude in steps until it reaches the lowest value required for 411 Journal of Network and Computer Applications 1996 19 405 413 the pacemaker to sense it The algorithm used here converges to the lowest value in the least number of cycles This lowest value is the sensitivity Noise immunity measurements are done by feeding sine waveforms at 50 amp 60 Hz to the pacemaker These waveforms are generated by the waveform generator of Figure 4 During this test the amplitudes of these waveforms are varied until they exceed the rated noise immunity level 100mV The pacemaker sensitivity must be set to its maximum value during this test 4 2 Auto Test Mode This is a very powerful feature provided wherein the tests are performed in a semi automatic manner without much user intervention Test and documentation time can be greatly reduced using this feature For each parameter being tested a number of different test values are preprogrammed in the tester The user has the option of running all t
15. time the presence of a QRS wave is ignored by the pacemaker e Sensing Refractory period is the maximum time interval after a QRS wave is sensed by the pacemaker during which time the presence of a second QRS wave is ignored e Sensitivity is the minimum QRS amplitude required to cause the pacemaker to operate in the demand mode A QRS wave of this amplitude will be sensed by the pacemaker e A V delay is the interval between an atrial stimulus and the subsequent ventricular stimulus e Noise Immunity is the maximum amplitude of noise signal that will not affect the correct performance of the pacemaker A few pacemaker testers are available in the market today 10 11 These testers are equipped to measure the above mentioned parameters But none of them have the capability for automatic measurement of any of the above test parameters For example the measurement of pacing refractory period is carried out by manually selecting different time intervals on the tester and continuously monitoring the pacemaker until an appropriate response is received The last tester setting gives the pacing refractory period On the otherhand the tester described in this paper measures each test parameter automatically without any manual intervention 3 System Design A block diagram of the pacemaker tester is shown in Figure 1 The microcontroller houses the program and controls all peripherals The external static RAM stores all the data collected during the