Home
        Lucidea Universal ScoreCard Plate Format
         Contents
1.                                          Control  Cy3 Cy5  Expected Raw Normalized  Calib 1 1 00 0 49 0 91  Calib 2 1 00 0 54 1 00  Calib 3 1 00 0 63 1 16  Calib 4 1 00 0 60 1 12  Calib 5 1 00 0 50 0 93  Calib 6 1 00 0 68 1 27  Calib 7 1 00 0 63 1 16  Calib 8 1 00 1 06 1 96  Calib 9 1 00 0 88 1 63  Ratio 3 3 00 1 67 3 11  Ratio 4 0 33 0 16 0 30  Ratio 7 10 00 4 90 9 10  Ratio 8 0 10 0 06 0 12  Utility 1 0 20 0 08 0 15    Asia Pacific Tel   852 2811 8693 Fax   852 2811 5251 Australasia Tel   61 2 9899 0999 Fax   61 2 9899 7511 Austria Tel  01 576 0616 22 Fax  01 576 0616 27 Belgium  Tel  0800 73 888 Fax  03 272 1637 Canada Tel   1 800 463 5800 Fax   1 800 567 1008 Central  East  and Southeast Europe Tel   43 1 982 3826  Fax   43 1 985 8327 Denmark Tel  45 16 2400 Fax  45 16 2424 Finland  amp  Baltics Tel   358  0 9 512 39 40 Fax   358  0 9 512 17 10 France Tel  01 69 35 67 00  Fax  01 69 41 96 77 Germany Tel  0761 4903 291 Fax  0761 4903 405 Italy Tel  02 27322 1 Fax  02 27302 212 Japan Tel   81 3 5331 9336 Fax   81 3 5331 9370  Latin America Tel   55 11 3933 7300 Fax   55 11 3933 7315 Middle East and Africa Tel   30 210 96 00 687 Fax   30 210 96 00 693 Netherlands Tel  0165 580 410  Fax  0165 580 401 Norway Tel  2318 5800 Fax  2318 6800 Portugal Tel  21 417 70 35 Fax  21 417 31 84 Russia  amp  other C I S   amp  N I S  Tel   7095 232 0250  956 1137  Fax   7095 230 6377 Southeast Asia Tel   60 3 8024 2080 Fax   60 3 8024 2090 Spain Tel  93 594 49 50 Fax  93 594 49 55 Sweden Te
2.   O Neg 1    Neg 2       63 0050 26   p2    microarrays       Table 1  Spotting parameters for Lucidea Array Spotter  The  spotting parameters are user definable  For more information  please refer to the Lucidea Array Spotter User   s Guide  1                                                Parameters cDNA  Total slides  27  Total plates  8  Spotting mode  Normal  Replicates  2  Repeats  1  Complete last row  Yes  Use default pitch  No  Row pitch  250 um  Column pitch  250 um  Slide thickness  0 9 1 1 mm  Barcode enable  Vv  Output file enable  Vv  Humidity control  Vv  Humidity setting  65  RH  Rows per block  8  Columns per block  16  Spots in last row  16       Spotting complete    7 1  At the end of the spotting run  a message window will  appear stating    The spotting session is complete  Please  remove your sample plates and slides  If you click OK  the  humidifier will be turned off     Leave the humidifier on by not  pressing OK     7 2  Allow the slides to remain inside Lucidea Array Spotter for  30 60 min depending on the drying rate of the spotting solution  used  This step allows all the spots on the slides to dry under  the same humidity conditions     7 3  Then press OK and unload the plates and slides and  remove the pen set as instructed by the Lucidea Array Spotter  User   s Guide     The slides were fixed and hybridized following the slide manu   facturer   s protocol  The slides were manually hybridized with  17 pmol of Cy3 labelled human heart and 17 pmol
3.  of Cy5   labelled human spleen cDNA targets  Slides were imaged using  an GenePix    4000B Array Scanner  The slide images were  quantitated using ArrayVision    6 0 software     Results    From Figure 2  the spotting quality of Lucidea Universal  ScoreCard controls and Easy to Spot DNA Human 1 0 genes  using Lucidea Array Spotter is shown  The spots all have    63 0050 26   p3    good morphology and good reproducibility as seen when  visually comparing the spot intensities of duplicate spots  The  nine calibration controls span a range from 3 to 30 000 pg 2 pl  of spike  The four ratio controls have a targeted Cy3 CyS ratio  of 3 1 and 10 1 and the reverse ratio for both  The utility  control has a targeted Cy3 CyS ratio of 1 5        Fig 2  A section of the microarray image  Cy3 labelled cDNA  heart  shown in green  and Cy5 labelled cDNA spleen  shown  in red  target were mixed and hybridized onto a spotted slide  with a total of 2688 human genes printed in duplicate  Shown  here are two duplicate blocks out of a total of 48 blocks of spots  on the whole slide  Row 1 of each block contains the calibration   yellow boxed   negative  red boxed   ratio  green boxed   and  utility  blue boxed  controls from Lucidea Universal ScoreCard   Rows 2 through 7 contain a duplicate set of 112 human genes  from Easy to Spot DNA Human 1 0           The calibration controls can determine the dynamic range of  the slide  The concentrations of the calibration controls used  in this experime
4. e the data     From Table 2  it can be noted that  in general  the normalization  procedure brings the Cy3 CyS ratio much closer to the expected  Cy3 Cy5 ratio than when using the raw signal intensities   Calib 8 and Calib 9 have a low mRNA spike concentration of  10 and 3 pg 2 pl of spike  respectively  As a result  Calib 8 and  Calib 9 have spot signals close to background which causes  high signal variation and skewing of the normalization ratios     Conclusions    Lucidea Universal ScoreCard is an essential tool for microarray  analysis  Lucidea Universal ScoreCard controls make it possible  to validate normalization and generate calibration curves for  microarray data  Most importantly  having Lucidea Universal  ScoreCard controls on microarray slides allows the comparison  of data from multiple experiments  Together with Lucidea  Array Spotter  microarray data is generated with excellent  accuracy and quality     References    1  Lucidea Array Spotter User   s Manual  Amersham Biosciences   pp  375 752  2002      Table 2  Lucidea Universal ScoreCard ratio analysis  Cy3 Cy5    Calib stands for calibration  Expected ratios are determined from  the known mRNA spike concentrations  Raw ratios are calculated  from the observed Cy3 and Cy5 signal intensities  Normalized  ratios are calculated from the constant factor normalized Cy3  and Cy5 signal intensities  The ratios listed in the table are the  average of the Cy3 Cy5 ratios for each control     Ratio Analysis  Cy3 Cy5   
5. f Utility 1 spike to the human heart  mRNA labelling reaction  Add 1 ul of Utility 1 spike to the  human spleen mRNA labelling reaction     1 4  Proceed with direct labelling and purification of the targets  using CyScribe First Strand cDNA Labelling Kit     1 5  Quantitate the concentration of the target     Lucidea Universal ScoreCard spotting sample preparation       Lucidea Universal ScoreCard spotting samples are lyophilized  in individual tubes to allow users the flexibility to resuspend the  samples in their choice of spotting buffer and customize the  format of the spotting plate pattern         Amersham     Biosciences    microarrays       2 1  Resuspend the samples in 600 ul of 50  DMSO to a final  concentration of 20 ng ul  50  DMSO is the recommended  spotting solution for Lucidea Microarray Reflective Slides    For other spotting buffers and slide types  the required final  concentration of Lucidea Universal ScoreCard spotting samples  may be different        Spotting plate pattern    3 1  Transfer 25 ul of each sample to a well in the barcoded  384 well microplate following the format shown in Figure 1     3 2  Centrifuge the plate for 1 min at 2750 x g to spin down  the samples        Spotting parameters    4 1  Create or edit a spotting protocol to the spotting parameters  in Table 1        Spotter setup    5 1  Install the Lucidea Spotting Pens pen set  Be careful not to    damage the pens     5 2  Load the plates for spotting  Place the plate containing    L
6. l  018 612 1900  Fax  018 612 1910 Switzerland Tel  01 802 81 50 Fax  01 802 81 51 UK Tel  0800 616928 Fax  0800 616927 USA Tel   1 800 526 3593 Fax   1 877 295 8102    Amersham Biosciences UK Limited  Amersham Place  Little Chalfont  Buckinghamshire  England HP7 9NA    Amersham Biosciences AB  SE 751 84 Uppsala  Sweden    Amersham Biosciences Corp   800 Centennial Avenue  PO Box 1327  Piscataway  NJ 08855 USA    Amersham Biosciences GmbH  Munzinger Strasse 9  D 79111 Freiburg  Germany    Amersham Biosciences  SV  Corp   928 East Arques Avenue  Sunnyvale CA 94085 USA    63 0050 26   p4    Cy  CyDye  CyScribe  Lucidea  and ScoreCard are trademarks of the Amersham Biosciences Ltd  Amersham and Amersham Biosciences are trademarks of Amersham plc   ArrayVision is a trademark of Imaging Research Inc   Easy to Spot is a trademark of Incyte Genomics   GenePix is a trademark of Axon Instruments Inc       Amersham Biosciences Corp  2003   All rights reserved    All goods and services are sold subject to the terms and conditions of sale of the company  within the Amersham Biosciences group which supplies them  A copy of these terms and  conditions is available on request     www amershambiosciences com        Amersham     Biosciences    
7. lication note       microarrays    Lucidea Universal ScoreCard Plate Format       Lucidea Array Spotter    Key words  microarray  gene expression  Lucidea  cDNA    Gene expression profiling with microarrays is being widely used  to compare the expression levels of thousands of genes using a  single microarray slide  To properly evaluate microarray results   it is necessary to minimize the variations in the measurements so  that accurate comparisons can be made within an experiment  and across multiple experiments  One method to minimize  variations is the use of control samples to quantitatively relate  the experimental data to the control data  The control samples  also provide a means of validating normalization  Normalization  is necessary to factor out differences in dye incorporation   hybridization efficiency  scanner dye sensitivity  and other  experimental factors     Lucidea    Universal ScoreCard    is a set of 23 microarray  controls consisting of calibration  ratio  negative  and utility  controls  All the controls except the utility controls are artificial  genes that generate pre determined signal intensities and ratios  that do not change across samples or experiments  While the  utility controls are also artificial genes  their signal intensities  and ratios are user definable  Utility controls can be used to  troubleshoot and estimate sample recovery during the RNA  purification process as well  All the controls are used for  validating and normalizing microa
8. nt span over four orders of magnitude  The  average signal intensities from the calibration control spots of  the experiment show that this slide has a dynamic range of  approximately 3 3 5 orders of magnitude  as seen in Figure 3   For higher expressing genes  the DNA on the spot was  hybridized to saturation causing the fluorescence intensity to  plateau  For lower expressing genes  the signal of the spots  was not high enough above background to be quantitated                            accurately   Calibration Curve   10 000 000    cys    1000000     Cy3     3 100 000         o  Ty  s  S 10 000   lt    1000  1 10 100 1000 10 000 100 000  Amount of RNA  pg              Fig 3  Calibration curve from Lucidea Universal ScoreCard  calibration controls for the microarray slide  The average signal  was calculated for each calibration control  The signal intensities  for Cy3 and Cy5 have been background subtracted  Rfu is an  abbreviation for relative fluorescent unit and is a measure of  fluorescence intensity     microarrays       The accuracy of the normalization protocol can be determined  from Lucidea Universal ScoreCard controls  This can be done   by comparing the normalized Cy3 CyS ratios against the   expected Cy3 CyS ratio of the controls  For each CyDye     label  constant factor normalization was performed on the  quantitated data using the average signal from the spots as  the constant factor  The constant factor is divided into the    signal of each spot to normaliz
9. rray data  They can also  be used to generate a calibration curve for determining limits  of detection  linear range  and the amount of mRNA in each  spot  Using Lucidea Universal ScoreCard with Lucidea Array  Spotter ensures high quality and accurate microarray results   This application note describes how Lucidea Universal  ScoreCard can be used in association with Lucidea Array  Spotter to determine the quality of microarray data     Products used    Lucidea Array Spotter 63 0040 09  Lucidea Spotting Pens  24 pen set 63 0040 10  Lucidea Universal ScoreCard  1000 hybridizations 63 0042 86  Lucidea Microarray Reflective Slides inquire  CyScribe    First Strand cDNA Labelling Kit 25 8007 67  384 well barcoded microplate 25 0199 34    63 0050 26 Rev A 2003 02   p1    Other materials required  e Easy to Spot    DNA Human 1 0  Incyte Genomics   e 50  DMSO    Protocol    The protocol below represents an example for spotting cDNA  on Lucidea Microarray Reflective Slides  The protocol can be  modified by the user to accommodate the needs of the  experiment  It is recommended to use only 384 well barcoded  microplates from Amersham Biosciences for spotting     Lucidea Universal ScoreCard target labelling       1 1  Add 2 ul of reference spike mix to the Cy   3 target  labelling of human heart mRNA     1 2  Add 2 ul of test spike mix to the Cy5 target labelling  of human spleen mRNA     1 3  Create an additional 1 5 ratio control using Utility 1 control   Add 1 ul of a 1 5 dilution o
10. ucidea Universal ScoreCard in plate position 1     5 3  Load the slides onto the slide tray     5 4  Make sure that the plates and slides are seated properly in    the correct orientation        Start spotting    6 1  To start the spotting session  click Start in the Lucidea    Array Spotter Control window     6 2  If more than 12 plates are being spotted  the Change  Plates In timer will include when to change the plates  drain    the waste container  and fill the wash solutions and humidifier    if necessary           Fig 1  Spotting Plate Pattern setup for  the Lucidea Universal ScoreCard control  plate on the Lucidea Array Spotter  The  arrangement of the calibration  Calib    ratio  negative  Neg   and utility controls  are shown     OOOOCOOOCOOCOOCSSOOO8008800080  OOOO OOOOOO8ODODO0D0DO000000      COCOOCOCOOCOCOCO SOOO O808008    SOOOOOOOOOOOOOOOOCOOCOCOO     OOOO OOO OOOO OOOOO8OO8OOO di   DDD D DDD DDI DDD L ILILILILIL II I    OOOO OOOOOOOOOOOOOOOOOO      DQOOODOOODDOODSDOOOSOOOOSOO      01010101000000 D S LIL ILILILILILLL IJ    LLALL LLLLILILLLLILLLLLLLL LA Ii   01001000D DIDID D   LL LILALLLAL ALA A    LLALLA LLLLILLLLLLLLLLLL AA di  ODOOCOCCOCOCO0EOOO08OO0888    SOOOOOOO OOOO 1000000000 00O   DOOOSCCOOOCCOCESOOOOSOOOOO8       OOOO a a OOOO OOOOCOOOCOCOO         oor out ON      _  _                Calib 01      calib 02     Calib 03     Calib 04    calib 05    calib 06     Calin 07     Calib 08    Calin 09    O Ratio 3    Ratio 4    Ratio 7  O Ratio 8      Utility 1
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
User`s Manual for  Kenwood TK-840/940/941 Interface  Hannspree Hanns.G HE225ANB  Manuel d`utilisation  CMX224・CM225・CM184 取扱説明書  Nilfisk SH 460  I. Prélèvement d`échantillons  Importation de véhicules automobiles  Shelf Logic  Mises en garde au sujet des piles Remplacement des - Fisher    Copyright © All rights reserved. 