1. DO NOT store filled oxygen cylinders near a furnace hot water tank in the trunk of a car or other high temperature area Storage in areas such as this can result in bursting of the cylinder or fire The use of this device is limited to the oxygen patient Cylinders MUST be used ONLY by the oxygen patient and are not to be distributed to any other individual for any purpose DO NOT use outdoors The maximum patient flow while filling a cylinder with the IOH200 is 3 L min for Platinum 5 XL and Perfecto concentrators and 5 L min for Platinum 10 concentrators Patients with a higher prescribed flow should not fill a cylinder while receiving supplemental oxygen treatment from the Platinum Concentrator HomeFill Compressor 12 Part No 1100873 SECTION IMPORTANT SAFETY INSTRUCTIONS NEVER block the air openings of the product or place it on a soft surface such as a bed or couch where the air opening may be blocked Keep the openings free from lint hair and the like If the compressor has a damaged cord or plug if it is not working properly if it has been dropped or damaged or submersed in water call a qualified technician for examination and repair If any leakage of the oxygen cylinder is detected DO NOT attempt to use the cylinder Turn the flow selector knob on the regulator to Off If leakage persists place the cylinder outdoors and notify your Home Health Care Provider or Service Representative of this condition
2. Invacare Platinum 5 10 XL and Perfecto M HomeFill compatible oxygen concentrators The oxygen concentrator can be set to the following maximum flowrates when operating with HomeFill compressor Platinum 5 XL and Perfecto 3 L min or less Platinum 10 5 L min or less To determine if your oxygen concentrator is HomeFill compatible examine the lower right corner of the rear of the oxygen concentrator for an outlet fitting Refer to the illustration below TOP VIEW Outlet Fitting Power Cord Concentrator HomeFill Compressor 10 Part No 1100873 SECTION IMPORTANT SAFETY INSTRUCTIONS SECTION IMPORTANT SAFETY INSTRUCTIONS A WARNING This section contains important information for the safe operation and use of this product A DANGER Risk of electric shock DO NOT disassemble Refer servicing to qualified service personnel No user serviceable parts TO REDUCE THE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS NEVER use near any type of flame or flammable explosive substances vapors or atmosphere Operating Information DO NOT allow oil from your hands or other sources to come into contact with the compressor coupling or bottle fittings These substances may become flammable in the presence of oxygen and cause injury Avoid touching the bottle fitting whenever possible Example of items to avoid are hand creams or lotions c
3. 5 L min for Platinum 5 and 5 L min for Platinum 9 The HomeFill Oxygen system consists of IOH200 gas compressor connection tubing and the Invacare supplied portable gas cylinders with the integral flow regulator and interlocking fill fitting The IOH200 compressor consists of a multistage compressor internal reservoir control electronics with an integral oxygen concentration sensor interlocking fill fitting inlet fitting and drive motor The compressor operation is controlled by the electronics assembly The motor compressor is allowed to fill a cylinder only when the oxygen concentration of the inlet gas stream is greater than 90 The motor compressor fill operation is completed when the gas in a cylinder reaches the maximum fill pressure of 2000 PSI 13786 KPa The IOH200 compressor can fill cylinders for portable use by ambulatory patients The Invacare supplied portable gas cylinders have an interlocking fill fitting that mates only with the Invacare IOH200 compressor assembly The gas cylinder also contains an integral flow regulator with fixed selectable flow settings Part No 1100873 15 HomeFill Compressor SECTION 3 INITIAL SETUP SECTION 3 INITIAL SETUP NOTE Dealers Only Refer to Dealer Information on page 9 before performing this procedure Initial Set Up of Compressor Compressor Initial Setup Checklist Before operating the HomeFill Compressor for the first time complete the following checklist
4. FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS IS Invacare Corporation www invacare com INVACARE USA 2010 Invacare Corporation All One Invacare Way rights reserved Republication Yes you can Elyria Ohio USA duplication or modification in whole or 44036 2125 in part is prohibited without prior 800 333 6900 written permission from Invacare Trademarks are identified by and Technical Services All trademarks are owned by or 800 832 4707 licensed to Invacare Corporation or its subsidiaries unless otherwise noted 3 in l oil is a registered trademark of American Home Products Corporation WD 40 is a registered trademark of the WD 40 Company Part No 1100873 Rev 11 10
5. Q Choose a location for the compressor Q Transport the compressor to the desired location Connect the power cord to the compressor Q Connect the compressor to the concentrator Choosing a location The compressor should be at least three inches away from walls draperies or furniture to ensure sufficient air flow Avoid placing the compressor next to heaters radiators or hot air registers DO NOT use in a closet The compressor should sit on a hard flat stable surface such as a table or the Ready Rack A WARNING NEVER use near any type of flame or flammable explosive substances vapors or atmosphere HomeFill Compressor 16 Part No 1100873 SECTION 3 INITIAL SETUP Transporting the Compressor A WARNING NEVER transport a compressor with a cylinder connected to or stored on the compressor otherwise injury or damage can occur CAUTION When transporting the compressor be careful not to bump the compressor or the connectors into obstacles Otherwise damage to the compressor may occur NOTE For this procedure refer to FIGURE 3 1 1 There are two methods to safely transport the compressor e One Hand Grasp the hand grip at the bottom rear of the compressor Hand Grip Hand Grip Two Hands Grasp the hand grips on the sides Hand Grip of compressor FIGURE 3 1 Transporting the Compressor Installing the Power Cord NOTE For this p
6. The HomeFill is equipped with a high pressure relief valve to ensure the user s safety When activated this safety feature will make an extremely loud noise If this noise occurs turn the unit Off and contact your Home Health Care Provider or Service Representative Unplug unit from outlet when not in use and before cleaning Maintenance DO NOT disassemble The unit is provided with double insulation and has no serviceable parts If any service is required contact Invacare Technical Service Double Insulated Products In a double insulated product two systems of insulation are provided instead of grounding No grounding means is required on a double insulated product nor is a means for grounding to be added to the product A double insulated product is marked with the words DOUBLE INSULATION or DOUBLE INSULATED or the symbol Hl Part No 1100873 13 HomeFill Compressor SECTION IMPORTANT SAFETY INSTRUCTIONS DO NOT unplug the power cord by pulling on the cord To unplug grasp the plug not the cord otherwise damage may occur Disposal of Equipment and Accessories Follow local governing ordinance and recycling plans regarding disposal of device components Radio Frequency Interference This equipment has been tested and found to comply with EMC limits specified by IEC EN 60601 1 2 These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installat
7. con nection FULL light is On in lt 10 mins Cylinder filling stops Cannot fill another cylinder until compressor is turned Off and On again O flowing to user if concentrator is On ATTENTION RED Audible Alarm HomeFill Compressor Compressor is On Cylinder NOT filling Causes Full cylinder attached when unit turned On replace with empty cylinder or Internal failure contact dealer 30 Compressor does not begin to fill or stops filling Unit must be turned Off and On before filling can occur O flowing to user if concentrator is On Part No 1100873 SECTION 7 TROUBLESHOOTING SECTION 7 TROUBLESHOOTING TROUBLESHOOTING PROBLEM SOLUTION No indicator lights on when compressor turned on Turn compressor Off 2 Check that power cord is properly plugged into compressor and wall outlet 3 Turn the compressor On 4 Clean air filters Refer to Cleaning the Compressor Filter on page 33 5 If performing STEPS 1 4 does not cause any indicator lights to come on contact your home health care pro vider or Invacare for service RED Light ON Check the cylinder gauge to see if it is FULL Cylinder full Remove the cylinder and turn the unit Off Perform the Cylinder Prefill Inspection on page 20 with a cylinder that is NOT FULL Cylinder not full Proceed to STEP 2 2 Check that the cylinder is connected properly 3 Turn the compress
8. violent ignition See Owner s Manual before cleaning P N 1075220 Rev C 10 04 WARNING DO NOT REMOVE UNIT COVER This unit contains moving and high pressure components Manufacturer Service Only P N 1075221 Rev B 10 04 WARNING Keep all connections free of oil and grease See Owner s Manual before cleaning Never oil or lubricate coupling Cylinders filled on this unit MUST be labeled Contents of this cylinder are the product of an oxygen concentrator P N 1075238 Rev C 11 17 06 DEALER INFORMATION NOTE Invacare recommends leaving a full cylinder of oxygen with the patient after setting up the HomeFill compressor Before leaving a HomeFill Compressor with a user the following checklist MUST be completed Q Make sure the pressure test on cylinder has not expired Q Instruct the user on the safe operation of the concentrator and review all warnings the user Q Instruct the user on the safe operation of the HomeFill compressor and review all warnings Leave a full cylinder of oxygen with the user Leave a copy of all warnings and filling instructions with Part No 1100873 9 HomeFill Compressor HOMEFILL INFORMATION HOMEFILL INFORMATION HomeFill is an accessory device to the oxygen concentrator that enables the patient to refill custom portable oxygen cylinders for personal use NOTICE The HomeFill compressor is for use with specially designed
9. C IEC 60601 1 2 TYPICAL PRODUCT PARAMETERS COMPRESSOR Alternating C t Type BF equipment emm _ dp omm 0 oe Attention Consider Accompanying AN Documents DO NOT smoke 5 aso Protected Electrical NEM NEM 120 VAC 10 108 VAC 132 Lic NNNM 60 Hz Operating Temperature 50 to 95 F 10 to 35 C 20 to 60 non condensing humidity Storage Temperature 10 to 150 F 23 to 65 C 15 to 9576 humidity Part No 1100873 7 HomeFill Compressor TYPICAL PRODUCT PARAMETERS Input Pressure Required 14 21 psi Input Flow Required 2 L min Input O Required gt 90 O Not suitable for use in the presence of a flammable anaesthetic mixture Electrical No extension cords Placement No closer than 7 5 cm 3 in from any wall furniture draperies or furniture to assure sufficient air flow Avoid deep pile carpets and heaters radiators or hot air registers Floor location only 2 m 7 ft cannula with a maximum 15 m 50 ft of Crush Proof Tubing DO NOT pinch Time of Operation Recommended Storage and 29 C to 65 C 20 F to 150 F at Shipping Temperature 15 95 relative humidity Environment Smoke and soot free No confined spaces Example No closets Minimum Operating Time 30 Minutes Up to 24 hours per day HomeFill Compressor 8 Part No 1100873 LABEL LOCATION LABEL LOCATION WARNING Keep all connections free of oil and grease to avoid
10. CATION ce eeeeeeoececcceccccecceceeees 9 DEALER INFORMATION ccccccccccccccccccccccccccccces 9 HOMEFILL INFORMATION 10 SECTION IMPORTANT SAFETY INSTRUCTIONS 000000000000000000000000000000000000000000000 Operating Information 2 E soeia anea E ENEE EARR 13 Double Insulated Products eee 13 Disposal of Equipment and 14 Radio Frequency Interference esee 14 SECTION 2 TECHNICAL DESCRIPTION 15 SECTION 3 INITIAL SETUP 0000000000000000000000000000 16 Initial Set Up of Compressor serene l6 Compressor Initial Setup Checklist l6 1 tiere toii eie l6 Transporting the COMPressOM cscsssssssssssssssssesseeseesssseess 17 Installing the Power 17 Connecting the Compressor to the Concentrator 18 SECTION 4 CHECKLIST AND CYLINDER INSPEC TION 0000000000000000000000000000000000000000000000000000000000000 20 Compressor Operation Checklist sss 20 Cylinder Prefill Inspection eene 20 The Hydrostatic Testing Date ee 22 External Examination 22 SECTION 5 COMPRESSOR OPERATI
11. ON 23 Part No 1100873 3 HomeFill Compressor TABLE OF CONTENTS Connecting Disconnecting the Cylinder to from the COMPRESSOR concedido oda apa 23 Connecting the Cylinder to the Compressor 23 Disconnecting the Cylinder from the Compressor 26 Turning the Compressor On esent 28 SECTION 6 INDICATOR LIGHTS 00000000000000000000 30 SECTION 7 TROUBLESHOOTING 0000000000000000000 3 SECTION 8 MAINTENANCE 00000000000000000000000000 33 Cleaning the Compressor Filter eee 33 Cabinet RE 33 LIMITED WARRANTY 000000000000000000000000000000000000 35 HomeFill Compressor 4 Part No 1100873 SPECIAL NOTES SPECIAL NOTES Check all parts for shipping damage In case of damage DO NOT use Contact Invacare for further instruction Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the following table for definitions of the signal words SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoi
12. User Manual HomeFill Compressor User Before using the product read this manual and save for future reference Dealer This manual must be given to the end user For more information regarding Invacare products parts and services please visit www invacare com Yes you can WARNING A WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner s manuals service manuals or instruction sheets supplied with this product or optional equipment If you are unable to understand the warnings cautions or instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur NOTICE The information contained in this document is subject to change without notice A ACCESSORIES WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products NOTE Updated versions of this manual are available on www invacare com HomeFill Compressor 2 Part No 1100873 TABLE OF CONTENTS SPECIAL NOTES cccccccccccccccccccccccccccccccccccccccccce D TYPICAL PRODUCT PARAMETERS 7 LABEL LO
13. a Keep away from heat flame and spark Keep out of reach of children DO NOT drop Filling of this gas is performed by Invacare HomeFill oxygen system Only Cylinders that have been refinished and or exposed to elevated temperatures must be hydrostatically tested before filling Cylinder temperature should not exceed 125 F 52 C Aluminum cylinders subject to the action of fire or heated to temperatures in excess of 350 F must be withdrawn from service and condemned by trained personnel DO NOT use caustic paint strippers Valve and safety relief devices must be removed and replaced by trained and authorized personnel DO NOT alter or change valve assembly Invacare Corporation DO NOT REMOVE THIS PRODUCT LABEL 2101 E Lake Mary Blvd Sanford FL 32773 Not for Commercial Resale 1 800 832 4707 Part Number 1125090 Rev D FIGURE 4 1 Cylinder Prefill Inspection Part No 1100873 21 HomeFill Compressor SECTION 4 CHECKLIST AND CYLINDER INSPECTION The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years A WARNING DO NOT fill cylinders that have not been tested in the past five years Otherwise serious personal injury may result Contact your dealer for replacement External Examination 1 Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist Dents or dings e Arc burns Oil or grease e Any other signs of damage that migh
14. and will burn with great intensity in oxygen enriched air Caution Federal law restricts this device to sale by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use or order the use of this device OPERATING INSTRUCTIONS See Operator s Manual E eel Full um YELLOW light GREEN Light Power Switch FIGURE 5 5 Turning the Compressor On Part No 1100873 29 HomeFill Compressor SECTION 6 INDICATOR LIGHTS SECTION 6 INDICATOR LIGHTS INDICATOR LIGHT COLOR None COMPRESSOR STATUS Compressor is Off EFFECT flowing to user if concentrator is on O BELOW NORMAL YELLOW Compressor is On and warming up 0 3 minutes after power is On Cylinder filling not started O flowing to user if concentrator is On O from concentrator is lt 90 any time after three minute warmup Cylinder filling stops will restart when O gt 90 O flowing to user if concentrator is On FILLING GREEN Compressor is On and cylinder is filling from concentrator is gt 90 any time after three minute warmup Cylinder is filling O flowing to user if concentrator is On FULL GREEN Compressor is On Cyl inder is finished filling full remove cylinder FULL light is On in gt 10 mins or Cylinder is not attached properly check
15. ded may result in property damage or minor injury or both ACCESSORIES There are many different types of oxygen tubing cannulas and masks that can be used with this device You should contact your local home care provider for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance and cleaning CAUTION Caution Federal law restricts this device to sale or rental by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use or order the use of this device Part No 1100873 5 HomeFill Compressor SPECIAL NOTES A DANGER DO NOT SMOKE while using this device Keep all matches lit cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death HomeFill Compressor 6 Part No 1100873 TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Standard Listing Double Insulated Product ETL Certified as Complying with UL60601 First Edition ETL Certified Complying with CAN CSA 22 2 No 601 1 M 90 Canada EM
16. e from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty For Model Nos HF2PCL4 HF2PCL6 and HF2PC9 Invacare warrants the HomeFill regulator NOT including the cylinder to be free from defects in materials and workmanship for a period of two years from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty DO NOT OPEN OR ATTEMPT TO SERVICE UNLESS SERVICE IS PROVIDED BY AN INVACARE CERTIFIED TECHNICIAN otherwise this will void any and all warranties If within such warranty periods any such product shall be proven to Invacare s satisfaction to be defective such product shall be repaired or replaced at Invacare s option This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare dealers It does not apply to the labor performed by the purchaser or user This warranty does not include normal wear and tear failure to adhere to the product instructions or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items such as filters are excluded from this warranty For warranty service please contact Invacare s service department at the toll free number on the back page during normal working hours Upon rece
17. ector fillport and cylinder fillport NOTE When using the post valve type cylinder attach an oxygen regulator per manufacturers instructions 7 Examine the cylinder gauge to ensure cylinder is full needle pointing into GREEN area If the cylinder is not full perform the following A Repeat Connecting the Cylinder to the Compressor on page 23 B If cylinder is not full after second attempt to fill contact the provider Cylinder Fillport Cover Connector Fillport Cover FIGURE 5 4 Fillport Covers Part No 1100873 27 HomeFill Compressor SECTION 5 COMPRESSOR OPERATION Turning the Compressor On NOTE For this procedure refer to FIGURE 5 5 on page 29 1 Make sure the concentrator is On Refer to the concentrator Owner s Manual CAUTION The concentrator flow rate to the patient MUST be set at maximum 3 L min or less for Platinum 5 XL and Perfecto and 5 L min for Platinum 10 Otherwise the oxygen level to the compressor will be lt 90 and the compressor will not fill the cylinder 2 Set the concentrator flow rate to the patient to 3 L min or less for Platinum 5 XL and Perfecto2 and 5 L min for Platinum 10 Refer to the concentrator Owner s Manual Make sure the cylinder is connected to the compressor Refer to Connecting the Cylinder to the Compressor on page 23 Push the power switch on the control panel to the On 1 position Examine the control panel The following sequ
18. ence of events should occur A 0 3 minutes The O Below Normal YELLOW light will be lit while the compressor warms up B After three minutes The FILLING GREEN light will be on while the cylinder is filling C The FULL GREEN light will be on when the cylinder is finished filling Proceed to STEP 6 Remove the full cylinder Refer to Disconnecting the Cylinder from the Compressor on page 26 Push the power switch to the Off O position HomeFill Compressor 28 Part No 1100873 SECTION 5 COMPRESSOR OPERATION NOTE The O BELOW NORMAL YELLOW light may come on if the oxygen level from the concentrator has not yet reached or drops below 90 The compressor will begin or resume filling when the oxygen level reaches 90 If light does not go Off within 10 minutes refer to Troubleshooting on page 31 NOTE If the ATTENTION RED light and audible alarm are on Refer to Troubleshooting on page 31 Control Panel Attention RED light and se Alarm Filling O GREEN Light PS FILLING Z3 INVACARE HOMEFILL ATTENTION _ O2 BELOW NORMAL d DANGER RISK OF ELECTRICAL SHOCK DO NOT remove cover refer servicing to qualified service personnel WARNING RISK OF FIRE NO SMOKING Keep all sources of ignition out of the room in which this product is located and from areas where oxygen is being delivered Textiles oil and other combustiles are easily ignited
19. ion Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards To determine if the emissions from the Perfecto is causing the interference turn the Perfecto Off If the interference with the other device s stops then the Perfecto is causing the interference In such rare cases interference may be reduced or corrected by one of the following measures Reposition relocate or increase the separation between the equipment e Connect the equipment into an outlet on a circuit different from that to which the other device s is connected Avoid creation of any spark near medical oxygen equipment This includes sparks from static electricity created by any type of friction SAVE THESE INSTRUCTIONS HomeFill Compressor 14 Part No 1100873 SECTION 2 TECHNICAL DESCRIPTION SECTION 2 TECHNICAL DESCRIPTION The HomeFill Oxygen System IOH200 is an electro mechanical prescription system for use in the home Its intended function and use is to supply pressurized oxygen to fill Invacare proprietary gas cylinders for ambulatory use while the patient receives therapeutic oxygen from an accessory oxygen concentrator The IOH200 gas system is designed for use with an Invacare Platinum oxygen concentrator with an integral HF2 accessory port that can supply a gas concentration greater than 9076 O around 2 L min while supplying a patient therapy flow rate up to 2
20. iving notice of an alleged defect in a product Invacare will issue a serialized return authorization It shall be the responsibility of the purchaser at the purchaser s cost to return the unit pack the unit in a manner to avoid shipping damage and to ship the unit to either Invacare s plant or service center as specified by Invacare in advance Defective units must be returned for warranty inspection using the serial number as identification within thirty days of Part No 1100873 35 HomeFill Compressor return authorization date Do not return products to our factory without prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE SOOT OR SMOKE FILLED ENVIRONMENTS OR OTHER THAN NORMAL APPLICATION USE OR SERVICE OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION S CONTROL THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
21. l Compressor 20 Part No 1100873 SECTION 4 CHECKLIST AND CYLINDER INSPECTION DANGER HomeFill Improper filling or abuse of this cylinder or failure to follow these safety Cylinder instructions may cause serious bodily injury or death High pressure oxidizing gas vigorously accelerates combustion Label DO NOT alter or modify cylinder or related components DO NOT use oil or grease Keep cylinder cylinder valve all equipment and connections free of oil and grease to avoid violent ignition resulting in serious bodily injury or death A WARNING For medical applications Use only as directed by a licensed practitioner Uninterrupted use of high concentrations of oxygen over a long duration without monitoring its effect on oxygen content of arterial blood may be harmful Oxygen Cylinders filled by the Invacare HomeFill oxygen system shall be used for personal use only Not to be filled for resale or use by professional users OXYGEN This cylinder is to be used with the Invacare HomeFill 2 oxygen system This cylinder contains 93 3 oxygen produced by an oxygen concentrator using the pressure swing adsorption process DO NOT handle cylinder or use contents until you are professionally trained including emergency procedures Use in accordance with Invacare HomeFill oxygen system Operator s Manual Open valve slowly Close valve after each use and when empty Secure cylinder during storage and use No smoking in cylinder are
22. lth care provider or Invacare Fill times exceed maximum as recommended fill times OR GREEN filling light stays On 1 Ensure the cylinder regulator is set to Off 2 Check the recommended fill times shown in the individual cylinder s manual 3 Check that the cylinder is connected properly Turn the compressor Off Remove the cylinder and reinstall Turn the compressor ON 4 Check the compressor concentrator line for proper connection and or damage i e leaks or kinks 5 Check the concentrator liter flow setting is at three L min or less If necessary adjust the liter flow setting refer to the concentrator owner s manual If the GREEN filling light does not go Off the compressor and or concentrator may need service Contact your home health care provider or Invacare Excessive noise when compressor is On Call home health care provider or Invacare for service Hissing sound from Regulator when flow selector is Off Turn flow selector Off 2 If hissing sounds continue call home health care provider or Invacare Regulator is On but O is not being delivered Refill cylinder 2 Check that cannula is straight and not pinched If damaged replace 3 Call home health care provider or Invacare for service NOTE Refer to the cylinder s manual for proper fill times NOTE Filling times are approximate and may vary on environmental conditions HomeFill Compressor 32 Par
23. not fill 2 Set the regulator flow dial on the cylinder to Off Refer to the cylinder s manual w Remove the cylinder and connector fillport covers if present Part No 1100873 23 HomeFill Compressor SECTION 5 COMPRESSOR OPERATION 4 Momentarily push DOWN on the outer ring sleeve of the connector fillport until GREEN dots are visible to reset the connector FIGURE 5 1 NOTE If the outer ring sleeve is in the UP position GREEN dots not visible the connector fillport will not be able to accept the cylinder fillport Pushing DOWN momentarily will reset the connector fillport GREEN dots visible to accept the cylinder fillport PUSH Outer Ring Sleeve DOWN Connector Fillport GREEN Dots BEFORE N COUPLING 5 Ps CYLINDER Bav PUSH DOWN FP GREEN DOTS ARE VISIBLE FIGURE 5 1 Resetting Connector Fillport 5 Grasp the cylinder in the area behind the cylinder fillport FIGURE 5 2 6 Position the cylinder in the compressor cradle FIGURE 5 2 7 Align the cylinder fillport with the connector fillport FIGURE 5 2 8 Pull UP on the outer ring sleeve of the connector fillport while pushing DOWN on the cylinder to couple the cylinder fillport into the connector fillport FIGURE 5 2 NOTE The cylinder is properly connected when an audible click is heard HomeFill Compressor 24 Part No 1100873 SECTION 5 COMPRESSOR OPERATION NOTE Fillport covers no
24. omeFill Compressor 18 Part No 1100873 SECTION 3 INITIAL SETUP DETAIL DETAIL C Insert End of Interconnect Hose into Inlet Outlet Fitting Inlet Outlet Fitting Interconnect Hose CONCENTRATOR TOP VIEW DETAIL Outlet Fitting ower Cord Compressor m Inlet Fitting FIGURE 3 3 Connecting the Compressor to the Concentrator Part No 1100873 19 HomeFill Compressor SECTION 4 CHECKLIST AND CYLINDER INSPECTION SECTION 4 CHECKLIST AND CYLINDER INSPECTION Compressor Operation Checklist Each time the HomeFill Compressor is used to fill a cylinder complete the following checklist Ensure the concentrator has been On for at least thirty minutes Refer to the concentrator Owner s Manual Perform the prefill inspection on the cylinder Connect the cylinder to the compressor Push the compressor power switch to the On 1 position Examine the indicator lights on the control panel Disconnect and remove the full cylinder Push the compressor power switch to the Off O position If filling another cylinder repeat this checklist Cylinder Prefill Inspection A WARNING ONLY use cylinders that have the Invacare HomeFill connection and HomeFill label All cylinders MUST be inspected before attempting to fill Otherwise injury or damage may occur NOTE For this procedure refer to FIGURE 4 1 on page 21 HomeFil
25. ooking oils suntan oils or sunblocks and similar oily products If you use these types of products wash your hands prior to using the HomeFill unit If the bottle fitting should come in contact with such substances it should be cleaned off with a damp cloth prior to connecting it to the compressor Part No 1100873 HomeFill Compressor SECTION IMPORTANT SAFETY INSTRUCTIONS NEVER oil or lubricate the compressor coupler or bottle fittings NEVER use any penetrating oil or lubricant such as WD 408 or 3 in 1 oil If connection difficulty is experienced verify that the bottle fitting is being inserted straight down into the compressor coupler If continued difficulty is experienced contact your supplier for assistance Before moving or repositioning either the compressor or concentrator ALWAYS push the power switch to the Off O position disconnect the AC power cord and the interconnect hose between the compressor and concentrator Failure to do so may result in damage to the unit or personal injury Children should ALWAYS be supervised around the HomeFill unit When used with the HomeFill table or similar elevated support surfaces a child could pull over any such table causing the unit to fall Failure to observe this warning may result in personal injury or product damage Use extreme care when handling and filling an oxygen cylinder Full oxygen cylinders are under pressure and can become a projectile if dropped or mishandled
26. or Off 4 Remove the cylinder and reinstall 5 Turn the compressor On If cylinder is not full and is connected properly the compressor may have internal failure Contact your home health care provider or Invacare Compressor cycles ON and Off or O BELOW NORMAL YELLOW and FILLING GREEN lights are alternating or O BELOW NORMAL YELLOW light goes On and Off Part No 1100873 Check the concentrator liter flow setting is at 3 L min or less If necessary adjust the liter flow setting refer to the concentrator owner s manual Wait at least 20 min utes for the GREEN light to appear If it does not appear proceed to STEP 2 2 The concentrator may need service contact your home health care provider or Invacare 31 HomeFill Compressor SECTION 7 TROUBLESHOOTING TROUBLESHOOTING PROBLEM SOLUTION BELOW NORMAL YELLOW light stays On Ensure the concentrator has warmed up for at least 20 minutes If it has not been 20 minutes turn the compressor Off until the concentrator warms up If the YELLOW light still remains on proceed to STEP 2 2 Inspect the connection between the compressor and the concentrator for damage or pinching If damaged replace If pinched straighten tubing Turn the compressor Off for at least 10 seconds Turn the compressor On If the YELLOW light still remains on after 10 minutes proceed to STEP 3 3 The concentrator may need service contact your home hea
27. person or entity and is not transferable or assignable to any subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state Invacare warrants the HomeFill Compressor to be free from defects in materials and workmanship for a period of one year from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty Invacare warrants the HomeFill internal wear components piston seals and rings bearings drive belts check valves pressure relief components and electric motor for a period of three years from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty or 5 000 hours whichever comes first Invacare warrants the HomeFill control panel components electrical switches reset switch and indicator lights for a period of three years from the date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty or 5 000 hours whichever comes first For Model Nos IOH2MD IOH2M6 and 2 9 Invacare warrants the HomeFill regulator NOT including the cylinder to be free from defects in materials and workmanship for a period of two years from date of purchas
28. rocedure refer to FIGURE 3 2 on page 18 1 Choose a proper location Refer to Choosing a location on page 16 2 Make sure the power switch is in the Off O position 3 Plug the power cord into the electric connector in the back of the compressor 4 Plug the opposite end of the power cord into the wall outlet Part No 1100873 17 HomeFill Compressor SECTION 3 INITIAL SETUP Wall Compressor Outlet REAR Electrical Connection Power Cord E m TM NOTE This product may with a polarized plug one blade is wider than the other As a safety feature this plug will fit in a polarized outlet only one way If the plug does not fit fully in the outlet reverse the plug If it still does not fit contact a qualified electrician DO NOT attempt to defeat this safety feature FIGURE 3 2 Installing the Power Cord Connecting the Compressor to the Concentrator NOTE For this procedure refer to FIGURE 3 3 on page 19 1 Insert one end of the interconnect hose into the outlet fitting on the back of the concentrator Detail A 2 Insert the opposite end of the interconnect hose into the inlet fitting on the side of the compressor Detail B NOTE The metal tabs on the concentrator outlet fitting and the compressor inlet fitting will pop out with an audible click when the end of the interconnect hose is properly installed Detail C H
29. t No 1100873 SECTION 8 MAINTENANCE SECTION 8 MAINTENANCE Cleaning the Compressor Filter A WARNING Push the power switch to the Off O position and unplug the compressor before cleaning DO NOT operate the compressor without the filter installed CAUTION DO NOT unplug by pulling on cord To unplug grasp the plug not the cord otherwise damage may occur NOTE For this procedure refer to FIGURE 8 1 NOTE All compressors have one filter on the back of the cabinet NOTE Perform this procedure at least once a week 1 Unplug the compressor Grasp the plug not the cord 2 Remove the filter 3 Clean the filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly B Dry the filter thoroughly before reinstallation Filter at alone FIGURE 8 1 Cleaning the Compressor Filter Cabinet A WARNING To avoid electrical shock DO NOT remove cabinet 1 Clean the cabinet with a mild household cleaner and non abrasive cloth or sponge Part No 1100873 33 HomeFill Compressor NOTES NOTES HomeFill Compressor 34 Part No 1100873 LIMITED WARRANTY LIMITED WARRANTY NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer This warranty is not extended to any other
30. t cause a cylinder to be unacceptable or unsafe for use 2 Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blistering of the paint or other protective coating or heat sensitive indicator If fire or thermal damage is found replace the cylinder 3 Inspect the cylinder regulator assembly for the following and replace if found e Debris oil or grease Noticeable signs of damage e Signs of corrosion inside the valve e Signs of excessive heat or fire damage HomeFill Compressor 22 Part No 1100873 SECTION 5 COMPRESSOR OPERATION SECTION 5 COMPRESSOR OPERATION Connecting Disconnecting the Cylinder to from the Compressor A WARNING NEVER use tools of any kind to connect disconnect the cylinder and the compressor Otherwise severe injury and or damage may occur Connecting the Cylinder to the Compressor A WARNING DO NOT drop oxygen cylinders Use two hands when handling transporting oxygen cylinders Otherwise injury or damage may occur DO NOT use a liquid leak detector to test for leaks NOTE For this procedure refer to FIGURE 5 1 on page 24 and FIGURE 5 2 on page 25 1 Examine the cylinder gauge If the cylinder pressure is less than 1500 psig proceed to STEP 2 NOTE For location of the cylinder gauge refer to the cylinder s manual NOTE If the cylinder pressure is greater than 1500 psig DO NOT attempt to top off this cylinder It may
31. t shown for clarity Cylinder Fillport Power Switch Connector Fillport Compressor Outer Ring Cradle Sleeve Compressor Cylinder Fillport Connector Fillport Outer Ring Sleeve Cylinder FIGURE 5 2 Connecting the Cylinder to the Compressor Part No 1100873 25 HomeFill Compressor SECTION 5 COMPRESSOR OPERATION Disconnecting the Cylinder from the Compressor NOTE For this procedure refer to FIGURE 5 3 1 Press the compressor power switch to the Off O position 2 Grasp the cylinder in the area behind the cylinder fillport 3 With the other hand grasp the outer ring sleeve of the connector fillport and push DOWN 4 Lift up on the cylinder to remove from the connector fillport PUSH DOWN Outer Ring Cylinder Sleeve Compressor H d Connector Fillport FIGURE 5 3 Disconnecting the Cylinder from the Compressor 5 When the cylinder fillport is disconnected from the connector fillport release the outer ring sleeve of the connector fillport and use two hands to remove the cylinder from the compressor cradle HomeFill Compressor 26 Part No 1100873 SECTION 5 COMPRESSOR OPERATION A WARNING The fillport cover on the connector fillport and the cylinder fillport MUST be replaced after filling and whenever not in use If either of the fillport covers are missing contact your healthcare provider before using 6 Place the fillport covers onto the conn

Download Pdf Manuals

image

Related Search

Related Contents

AFG310 and AFG320 Service Manual  Brother MFC-822 User's Manual  DTN-V800地図データ収録について [PDF:118KB]  Winter 2010 - Engineering Dynamics Corporation  Scarica il PDF  UKMON V4 - UK Meteor Observation Network  Manual de instalación y uso de la lavadora Manuel d`installation et d  NetComm - Foxtel Community  Manual de instalación y mantenimiento de tuberías de refrigerante  USER'S MANUAL @ BEDIENUNGSANLEITUNG I MANUEL D  