Home
        HCV Real TM Quant 100 PCR NEW
         Contents
1.                          Chick on 5 cycle below to modiy it         This cycle repeats 42  tmels      Chek an one ol the siege below to noodiy      or press  gt  or to add and remove steps for thir cycle          BO dag for 40 sect    W    Press the button OK    4  Inthe window New Run Wizard click Calibrate  Gain Optimisation    for Rotor Gene 6000   In the new window  Channel Setting select channels Joe  Yellow  and Fam  Green   Indicate Min Reading 5  Max Reading 10 and  select function Perform calibration  Optimization  before 1     Acquisition  Click the Close button     Auto Gain Calibration Setup E alf x   Calibration os  different gain levels until it finds one at which the fluorescence levels are    acceptable  The range of fluorescence you are looking for depends on the  chemistry you are performing     9  Auto Giain Calibration will read the fluaresence on the inserted sample at     Set temperature to     degrees        Calibrate Al       Calibrate Acquiring          Perform  Calibration At BD D egrees AE Beginning Of Aun          Channel Settings                88    Edit       Remove      Remove   ll       Sat   Hamal    Close   Help    5  Select Next and click Start run    Program the position of the tubes in the carousel of the Rotor Gene 3000 6000 Q and enter the concentrations of the  Quantitative Standards  reported on the HCV TM RG Quant Data Card  in order to generate Standard curves     Data Analysis   IC amplification analysis  Cycling A Fam Green    Pre
2.                       um   9   um    3                                 In the above table the data show that almost all the results obtained from the clinical and control samples with HCV       Real TM Quant test differ by less than 0 5 Logio from the values obtained with Reference manufacturer s kit  The  figure 2 illustrates the graphic correlation between the HCV RNA viral load measured by two tests  The correlation    factor is 0 975  p   lt  0 01      29    Figure 2  Graphic comparison among the results obtained with    HCV Real TM Quant     Sacace  kit and reference    manufacturer   s kit     6 0    5 0                               D4                      4 0                                   j                   Joy 0160        3 0    3 5 4 5 6 5    em  O1    log10 Sacace    30    17  TROUBLESHOOTING    l     o    Weak  Ct    34  signal of the IC  Fam channel   retesting of the sample is required   e The PCR was inhibited       Make sure that you use a recommended RNA extraction method and follow the  manufacturer s instructions     gt  Re centrifuge all the tubes before pipetting the extracted RNA for 2 min at maximum  speed  12000 16000 g  and take carefully supernatant  Don t disturb the pellet  sorbent  inhibit reaction   e The reagents storage conditions didn t comply with the instructions       Check the storage conditions  e The PCR conditions didn t comply with the instructions       Check the PCR conditions and for the IC detection select the fluorescence cha
3.                  DefaultUser Dec 18  2004 01 18 PM   Default User Dec 18  2004 10 38 AM  HBVquan 171204 1 Done DefaultUser Dec 17  2004 11 31 AM  HCVquan 161204 1 Done DefaultUser Dec 16  2004 12 53 PM  Done DefaultUser Dec 15  2004 02 05 PM   Done DefaultUser Dec 14  2004 05 01 PM   Done Default User Nov 18  2004 01 01 PM   K Done DefaultUser Jul 19  2004 02 23 PM   95 62 65 cmw taq f 22 05 Done Default User May 23  2004 04 24 PM       eem 161204 1    User  Default User   Jeroan    standard  Started  Dec 16  2004 12 53PM       Finished  Dec 16  2004 02 39 PM   est  A Temperat   wet    Notes  UM                                        HCVquan V17    HCVquan 0    Temperat     Melt   raw       CVa    HCVquan K3  HCVquan B1 J  HCVquan B2            r   1    700 01 02 03 0 4 0 5 06 07 08 9 8 10    Degrees C             With this euge you can mre a eT curve vy TET emm    Sacace    HCV Real TM Quant April 16  2011 10          Smart Cycler  User Tees Setup Tools Help      4 Run   Status    Run    T       27 2  Run Name     Resutts Table Type Seg StdiRes Btd Res                     HCVquan 1612041  ass Settings      Protocols     f Temperature          FAM 2  orn           User  Default User    Started  Dec 16  2004 12 53 PM  Finished  Dec 16  2004 02 39 PM                                                                                      f Standartcy3 JA HCVquan   25081 426 78 677271 0519 50    Status  Done   Jaz  HcVquan AT6 UNKN 06 1 421696572021     ND  HCVquan HCV1000      UNKN
4.                l Abbreviations   UNKN   Samples  STD   Standards  PKO1  PKO2   HCV Positive Controls  OKO     Negative Control  Sample  K1  K2  K3     HCV Quantitative Standards  B1  B2  B3     IC Quantitative Standards    Sacace    HCV Real TM Quant April 16  2011 11    RotorGene    3000 6000 Q  Corbett Reasearch  Qiagen     1  Select New Run and Dual Labeled Probe  Click New     New Run   x           Quick Start Advanced    i    Perform Last Run  d   SYBR Green Fi    3 Labeled Probe          Quenched FRET  E P DMA  Concentration Measurement  E   Hybridization  Melt    m      d Rotor Gene Demo Kit    Open  amp  Template In Another Folder    Cancel          Help          Show this Screen when Rotor Gene Opens          2  Program Rotor Gene 3000 6000 Q as follows    lt  Select Rotor Type 36 Well Rotor and No Domed 0 2 ml Tubes  Nepos Rum Wizard     Welcome to the Rotor Gene Run Wizard      Rolo Type                         Tawel Rotor  Genebisc  7299 8               gt  Reaction Volume  ul   25    gt  Temperature Profile   Hold  50  C     30 min  Denature  95  C     15 min  Cycling   95  C     20 sec  60  C     40 sec    Cycle Repeats     42 times  Fluorescence is measured at 60  C on FAM  Green  and JOE  Yellow  channels     Sacace    HCV Real TM Quant April 16  2011 12    VR Edit Profile             e      New pen Save 35   Heb      mea m Se LE diedas         0    The run wil take aperasmately 131 mintelz  to complete  The graph below represents the run to be performed      
5.               Click the menu Sample Information  S    lt  Sample Data    S   Insert Sample name  Property  U for unknown  samples and negative PCR control  S for Standards  and insert standard concentration reported in Data Card                  Click   or click File   gt  Run Programs  R  to Run the PCR program   Sacace    HCV Real TM Quant April 16  2011 23    Results Analysis   FAM  In the Quantitative Analysis tab  select from the menu Dyes D    gt  FAM F   Select standard Concentration boxes     left mouse click CTRL  for Internal Control  see picture below    Group   Sample  1 Neg extr  1 POS1  1 1  1 12  1 2  1 051 HCV  5  1 QS2 HCV 5 46e 003  1 QS3 HCV 1    1 8  1 88e 005  1 asc  1 QS3IC 8 20e 002    1 NTC       Concentration            Then do in this order   1  Under Analysis Method select Fit Points   2  Under Analysis Step Click Zero Adjust  select Auto and then click OK  Then click Baseline    3  Select 2 points and set the base line  drag and drop the red baseline  as low as possible but above the noise  of each sample  usually 6 8    4  Under Analysis Step click Analysis  Calculated copies reaction of Internal Control and Ct Results should  appear in the Ct and the    Calculated Concentration    column respectively        Analysis Method Analysis Step Points Selected Baseline and Threshold   Display Mode   EJ2nd Driv  Max  Zero Adjust Baseline Logarithmic  ElBaseline Full Scale   Fit Points ElAnalyss Show Points Threshold Smooth low         Results Analysis   J
6.        Smart Cycler      8 x   User Logs Setup Tools Help    1    Create Run Check Status Stop Run                                                                                                                      4  Site  Protocol   Sample ID   Eee Status FAM Cy3   Cy3Ct    Run Name                Std Res  2 JA1 HCVquan W17     UNKN   OK POS 4 POS 1950 NEG        p HCVquan 476 UNKN   OK POS 2681  Pos 2021                 HCVquan 00 1000   UNKN OK POS 2710 POS   3270          yp TE ONES      Temperature  HCVquan 6 1000     UNKN      OK Pos 2702  POS 3239 NEG    Started  Dec 16 2004 12 53 pM  FA   HCVquan HCV500   UNKN OK POS 2694 POS 3531   NEG      OT  B  HCVquan HCV500   UNKN OK POS 26 88 POS     3537 0  ere minit Standart Cy3            HCVquan PKO1 UNKN OK POS 2713 POS 3267 NEG  4 las HCVquan PKO2 UNKN OK POS 26 76 POS   2572 NEG  Pia  Done 49 HCVquan OKO UNKN OK Pos 2729 NEG 0 00 NEG  Notes  jA10 HCvauan 1 UNKN OK  NEO 0 00 POS   1867 NEG  jaM HcVquan       UNNN   20K Neo 000     Pos 2562 NEG  jA12 HCVquan 3   UNKN OK NEG 0 00 POS 3425 NEG    jA13 HCVquan 4     UNKN    OK Pos 22268   NEG X 000 NEG di  A14 HCVquan 5 POS   0 00 NEG  HCVquan 6 EEE NEG             POS      Site ID Protocol         HCVquan V17  Hein A uan 8    er  jStandatCy3        Number of Sites  16    SE in the new TEES Tm experiment ent with me standard curves curves   click OK     C smart Cycer IE Ereto  la x   User Logs Setup Tools LLL ain    LT Rum Check Static ton Ear    Ca Select a Run            
7.    OK  0  05 555 3270 ND  Noise  CVquan HCV1000    UNKN OK 2 129 718 3239 ND    HCVquan HCV500    UNKN   OK  22478 3626 94 18716 35 31 ND  6 HCVquan HCV500      UNKN   OK  23429 6626 88  7 987 13537 NO  JA  HCVquan PKO1 UNKN   OK 3 80792 13267 ND  Jas  HCVquan PK02 UNKN   OK     25428 7626 76  10856 606025 72     ND  A9 HcVquan OKO      UNKN   Ok 17713802728 ND 000     ND  CVquan K1 UNKN   OK ND 0 00 1116980311867 ND  CVquan K2 UNKN OK ND  0 00  11608 05725 62     ND  CVquan  K3 UNKN   OK ND  0 00 37714  3428 ND  HCVquan B1 UNKN 6 421168 122 08 0 00 ND  Dye Set  FCTC25 HCVquan B2 OK  148938 7125 81 0 00  0 pa OK 49429 foo             Protocols  2 1           Lot Number      HCVquan                   tja  HCVquan                     HCVquan o9   uv     70   5 Pom e    Number of Sites  22 8   HCVquan  econ   0 00 00 0 00 20 0 00 40 0 01 00 0 01 20 00140   HCVquan s        81204 1 m   Ja   5 HCVquan   Sunt 3 gp                                                                     r En NN    In any case  if the calibrators were inserted with the clinical samples in the same experiment or after the importation  of standard curves from another experiment  in the table of results  in the column FAM Std Res for IC HCV and in  the column Cy3 Std Res for cDNA HCV should appear the calculated values       oc nim as   or 5s 22108  OK 19856                    x p  a0     pe           Hevaan s     Bm   ox Yoon om  em aaps        OK Wo000 226   o00 OJ  OK  X0000   25         OK a             
8.   Applied Biosystems   is trademarks of Applera Corporation     SmartCycler   is a registered trademark of Cepheid     LightCycler   is a registered trademark of Roche     LineGene K   is a registered trademark of Bioer     Eco PCR Real Time System   is a registered trademark of Illumina    Sacace Biotechnologies Srl  via Scalabrini  44     22100     Como     Italy Tel  390314892927 Fax  390314892926  mail  info sacace com web  www sacace com    SISTEMA       SISTEMA  DI GESTIONE  CERTIFICATO    9001   2008    DI GESTIONE  CERTIFICATO  13485 2004        PCR  The Polymerase Chain Reaction  PCR  process is covered by patents owned by Hoffmann La Roche and applicable in certain countries  Sacace does not encourage or support the  unauthorized or unlicensed use of the PCR process  Use of this kit is recommended for persons that either have a license to perform PCR or are not required to obtain a license    32    
9.  FAM channel for Internal   Control      amp  Smart Cycler    User Logs Setup Tools Help      1            A                                    i       Maintenance                                                                Create Run Check Status Stop Run  M Views   SitelD   Protocol  Sample ID  Run Name    Results Table     HCVquan  HCVquan 141204 1 f Analysis Settings    Sf Standard FAM    User  Default User      Standard cya 8   3  Started  Dec 14 2004 05 01 PM   c    8  Finished  Dec 14  2004 06 46 PM   5    5             Status  Done     elt 2   FAM    Notes  2  Views    Resutts Table  2  Analysis Settings  Dye Set  FCTC25 j 8  A z  2 g  Protocols         Protocol Lot Number      5  HCVquan      Number of Sites  16 j Cycles  a HC             perm EC EET     Select ens   View Another Run   Delete Run s    Update Analysis   me E    Sacace    HCV Real TM Quant April 16  2011 9    Results Analysis   Choose in the menu    Analysis settings    the value 20 for the channels Fam and Cy3  In the table of results  Results  Table  appear the values of Ct  Threshold cycle  for Fam and Cy3 channels    The calculation of HCV RNA concentration in the clinical specimens sample and standards can be performed in the  same experiment  but with the SmartCycler software it is possible to calculate the samples concentration by  importing the experiment with Standard Curve in the experiment with clinical samples  You can import curves from    another experiment clicking on Import Std Curve
10.  Prepare quickly the Reaction mix     10  Specimens may be infectious  Use Universal Precautions when performing the assay    11  Specimens and controls should be prepared in a laminar flow hood    12  Handle all materials containing specimens or controls according to Good Laboratory Practices in order to  prevent cross contamination of specimens or controls    13  Clean and disinfect all spills of specimens or reagents using a disinfectant such as 0 5  sodium hypochlorite  or  other suitable disinfectant  Follow by wiping down the surface with 70  ethanol    14  Avoid contact of specimens and reagents with the skin  eyes and mucous membranes  If these solutions come  into contact  rinse immediately with water and seek medical advice immediately    15  Material Safety Data Sheets  MSDS  are available on request    16  Use of this product should be limited to personnel trained in the techniques of amplification    17  Workflow in the laboratory must proceed in a uni directional manner  beginning in the Extraction Area and  moving to the Amplification Area  Do not return samples  equipment and reagents in the area where you  performed previous step  Personnel should be using proper anti contamination safeguards when moving  between areas     8  STORAGE INSTRUCTIONS   Part N   1        Controls    must be stored at 2 8  C    Part N  2    HCV Real TM Quant  must be stored at  20 C    The kit can be shipped at 2 8  C for 3 4 days but should be stored at 2 8  C and  20  C immediate
11. 2021 222324 25 26 27 28 29 30 31 32 3334 35 36 37 38 39 40 41 42 12345 6 7 8 91011121314151617 18192021 222324 25 26 27 28 283031 32 3334 35 36 37 38 39 40 41 42  Cycles Cycles  Fig  1 FAM   530 channel  Internal Control  Fig  2 Joe   560 channel  HCV   Standard Curve  t 28   amp  26  Error  0 0104 g  Efficiency  1 964 LES  5 22  3 4 5  Log Concentration  Fig  3 Standard Curve  At first usage the operator must perform a color compensation experiment asking Sacace for the detailed  instructions     Sacace    HCV Real TM Quant April 16  2011 22    LineGene K    Bioer   Open LineGeneK Software  Click File   gt  New  In the Setup Programs tab  insert User Name and Test Name  Click  Dyes button and select FAM  JOE  Insert the correct liquid quantity  25 ul    User nane  Testnane      Liquid Quant    25   Click the button 9 7  3 times and program the 3 thermalcycling steps  At the end it must look like this   ed Cycles  1 Cydes  1 Cycles  42         95 0   95 00       On the second phase  60 0   C  of step 3 select    Single    on the Sample Mode column to collect fluorescence data  as  shown in this picture         Target                95 0 00 00 20 4 0         None Y  i  None      Single    Click the menu System Parameters  P    gt  Gain Setup  G   Select Auto Gain box and click OK  see picture below for  Gain Setup details                 File E  Sample Information S    riter   Groups G  Dyes D  T      Raw Data R        0L Fao  Grad Temp Calculation C       25    25   2   25   
12. 6    BIOTECHNOLOGIES          HCV Real TM Quant  Handbook    Real Time Kit for the Quantitative detection of  Hepatitis C Virus in human plasma    for use with RotorGene    3000 6000 Q  Corbett Research  Qiagen    SmartCycler    Cepheid     1Q 1Cycler    and 1Q5     Biorad   MX3000P   and  MX3005P    Stratagene   Applied Biosystems   7300 7500 StepOne Real Time  PCR Systems  Applera   LightCycler 2 069  Roche   LineGene K    Bioer    Eco Real Time PCR System    Illumina     V1 100 2FRT  16 04 2011    NV 0    Sacace    HCV Real TM Quant April 16  2011    TABLE OF CONTENTS    v Ge              E  RADI      Roo ER   gt  07 N mM       15   16   17   18   19     EI M                                                                                             PUY OCU CHOM M                                                                                E B c                                               R  Principle 0I XS9dV oeesses vetas E FEE E UP bx cR TQUE ME MEX DERI EMI MK EECU DENEN   Materials Provided e                                             Materials Required but Not Provided    eere eee e rese ue S ev roe ERE ERR E E EN PRA ER SEP FEES UE TERI es  Warnings and Precautions c  sievev vastos duo upEvan UE RREFENUERULUM FI icQuIM EVI S NOT PUn MES CAVE FER UNDE MEE  Virg            1010 TR                                                                                 wICIDILAT E E EU Et      Specimen collection  Storage and Transport           ccccccccccccccccccccccccccc
13. 6 53 7 27 10  IEEE    Analytical sensitivity   The kit HCV Real TM Quant allows to detect HCV RNA in 100  of the tests with a sensitivity not less than 20   U ml  value obtained using the    Magno Virus    extraction kit Sacace REF K 2 16 and RotorGene 6000   The  detection was carried out on the control standard and its dilutions by negative plasma     Human immunodeficiency virus 1 Sa       Linearity   The linearity of the HCV Real TM Quant assay was tested with the HCV RNA Standard and it   s dilution using  HCV negative human plasma  Each dilution was analysed three times and the mean HCV RNA titer of each sample  was determined    HCV Real TM Quant is linear from 2 5 x 10   to 5 x 10    IU mL  Test results greater than 50 000 000 IU mL are  above the upper limit of quantitation of the test and should be reported as    greater than 50 000 000 IU mL     If  quantitation results are desired for such samples  the specimen should be diluted 1 10 with negative serum and  retested  Test results less than 250  U mL are below the lower limit of quantitation of the test and should be reported  as    less than 250 IU mL        28    Table 1  Comparison among the results of HCV RNA viral load detection using    HCV Real TM Quant     Sacace  kit    and reference manufacturer   s kit     Sample  ae xl cdi  pie   wm    sm   e   sw   99   ww   ww            9       sm omm   w   ome   o     99   mw   wem            7    sem   mee   os     mee   e    3         sew   nem   aor   ome o   99
14. C    ResWs                           Area to analyze     Amplification plots   C Plate sample values  C Standard curve   C Initial template quantity  C Dual color scatter plot  C Text report    Last cycle  ls      Fluorescence     dA        Threshold fluorescence     v JOE  500 0002  Reset        FAM  1000 00  Reset       NVA     Reset                      Fluorescence  dR     Select amplification plots to display       B2 JOE Unknown 22 18  B2 FAM Unknown 5  B3 JOE Unknown 25 38  B3 FAM Unknown 24 07  B4 JOE Unknown 28 71  B4 FAM Unknown 25 25  B5 JOE Unknown 34 32  B5 FAM Unknown 8  BB JOE Unknown 32 58  BE FAM Unknown 25 05  Cycles B  JOE Unknown 35 32  B  FAM Unknown 8      Select All   Disable Dyes    Dyes shown Well types shown  JOE FAM     Unknown Standard      Results  Door Closed   Lamp Off  F          Sacace    HCV Real TM Quant April 16  2011 18    Applied Biosystems   7300 7500 StepOne Real Time PCR Systems  Applera     1  Select in the main menu the option    New Experiment     Advanced Setup   insert experiment details  Select your  Instrument Type  7500   7500 fast StepOne StepOne Plus   Quantitation     Standard curve  Taqman Reagents   Standard ramp speed  DO NOT select Fast RampSpeed      Which instrument are you using to run the experiment              Which reagents do you want to use to detect the target sequence        Set up  run  and analyze an experiment using a 4  or 5 color  96 well system   The PCR reacticns cottain primers des gned to amalify the 
15. Cal E 2D ul   suggest to select 20 ul and use 30 ul polycarbonate capillaries     2  Program the thermalcycling as follows     ied Real TM Quant    one None   Stage   Profile       NENNEN          yu Hi      39c 309b        cycles     None 3    95  C   15 00                                 T 95  C     0 15  Progra   None 60  C   0 40    a a a                 Program 42 E  Quantification    Program Temperature Targets            Acquisition mode   Single                                  Target    C      Hold  hh mm ss    Ramp Rate    C s    Sec J   Step Size    C    Step Delay Acquisition Mode   cycles   5    00   g Overview     100  O eee  d 0 00 00 0 30 02 0 45 09 0 57 27 1 10 05 1 22 23 1 34 46    Time  h mm ss        3  Click and insert sample names  then select Absolute Quantification from Analysis Type and select 530   560 in Selected Channels section           Sample data            1 Analysis Type       Reset Samples    Impe   P Absolute Quantification   Color Compensation   Genotyping   4 X Nucleic Acid Quantification  Qualitative Detection   Relative Quantification   Dual Color        Relative Quantification   Monocolor     Tm Calling       Selected Channels  530  560 61 o  64       4  Insert target Name  Sample Type  Standard for standards  Unknown for all other samples   and insert sample  concentration values as reported in the Data Card                                                                                         Pos   Sample Name Channel   Target Name 
16. HCV  Reporter HEX  and the other one for Internal Control  Reporter  Fam   select Quencher     Non fluorescent for both targets  Click    Assay Mame Color Reporter Quencher  Internal Control C FAM NonFluorescent  HBV 6 HEX NanFlucrescent    4  Click     and enter name for all samples  controls and standards  Click   5  Associate samples and standards with the two previously designed assays following this procedure  select the  well  select Unknown for samples and Standard for standards and to assign the corresponding assay click the  white circle of the HCV and Internal control assays  the circle will become colored   for each well both Assays  must be assigned     6  Click   m and in the Set Up Standards box insert each value reported in the DataCard  click Enter after  inserting each value  both for HCV assay and Internal control assay  then associate concentration values to the  corresponding well  In the end it should look like above       Set Up Standards 1 2 3    Sampei ENENEEMNEEEEENNM         Sample3      vas  zu   9086       2                   ee Apply Standards  68350    Sampe4           SampeS ample ple  8220  gos  go 0    Ww w an D  760 B wy  e  uj   s 68350  s8220  s 760 w    0      ermal Profile s        7  Click on   mE   and program the instrument as follows  reaction volume is 20 ul1        ILL    HCV Real TM Quant    Reverse Polymerase PCR Cycling    Transcription Activation    3 95  C     0 15  42  60  C     1 00         Add 10 ul of Reaction Mix into  eac
17. Sample Type Concentration  8  Sample 8 530  IC Unknown  560  HCV Unknown  3  Sample 3 530 JIC Unknown  560  HCV Unknown  10  Cneg EXTR 530 JIC Unknown  560  HCV Unknown  11  POS 1 530  IC Unknown  560  HCY Unknown  12  POS 2 530 JIC Unknown  560  HCV Unknown  13  Neg PCR 530 JIC Unknown  560  HCV Unknown  14  Q51 HCV 530  IC Unknown  560  HCY Standard 8 72E5  15  Q52 HCV 530  IC Unknown  560  HCV Standard 5 46E3  16  Q53 HCY 530  IC Unknown  560  HCV Standard 3 10E1  17  Q81 IC 530  t Standard 1 88E5  560  HCV Unknown  18  QS2IC 530  IC Standard 1 88E4  560  HCV Unknown  19  QS31C 530  IC Standard 8 20E2  560  HCV Unknown                         Click   totus   to start PCR reaction     6  After Run is completed click  Analysis    button and select Absolute quantification  Select Method  Auto     Standard Curve  In Run   Ct values and calculated results will appear  for the selected channel  under the CP  and Concentration column in the Absolute Quantification table        Ese ee Quantification  HCV Quant FAM JOE    Channel  560    Color Compensation  On    Program   Samples Results    n     vi Sample 1  2 nple 2 32 92  2 40E 1           8 nple 3  1 29E0   Sample 4 31 98 1  Sample 5 31 17 9 27E1   6 Sample 6 31 72 6 06E1  Sample 7 31 59 6 67E1    8 Sample 8 31 8   5 53E1    v    7  Calculate results manually using the usual formula as indicated in the User Manual of the kit     Amplification Curves Amplification Curves          8       z z  12345 6 7 8 9 1011121314151617 1819
18. cccccccsccssssssssssssssecs   2 RNA COLO                                                           es esiersts   Bhuodgcenig c                                                     Rear PEeDAIEOUOIL ececaciccioe access neues OR EEE e RU PEU Dong eiF Usa cta D  sio Vota rU ERES RIT FOE      Protocol and Data Analysis      SmariCycler   6 inm                                    e  RotorGene M 3000 6000  Corbett Research              ccccccccccccccccccccccccccccccccccsccccesccscces   6 jw IWC ycler    and 105    BiOrad  sicicccccsasensevssadeccscsaetver venues UI URS EEN DIR DEDI NAAR Ua FA EEUU     MX3000P and MX3005P    Stratagene          ccccccccccccccccccccccccccccccsssssssssssssssssssseees  e Applied Biosystems   7300 7500 Real Time PCR Systems  Applera             ccccccscccccccceees  e    gt            yder 2 08   0076  e LimeGene KO   1067  6 Eco Real Time PCR System  Tlumima                ccccccccccccccsscccccccccccsssccsssccccccsscsssseees  Results              tes  Performance 6          06001  0 exe eeeressenss  0  Explanation of 005  10 tine even eonadsetareudsaacoesserdisecieiess    1  NAME  HCV Real TM Quant    2  INTRODUCTION   Hepatitis C virus  HCV  is a single stranded RNA virus that belongs to the family Flaviviridae  In 1974 HCV was  initially recognized as non A  non B hepatitis virus  NANBH  until 1989 when the etiologic agent was cloned  It is  estimated by the World Health Organization that worldwide there are 170 million HCV infected persons    The prima
19. ch the amplification will be carried out   Make sure that the following settings for dynamicwf are selected   1 cycle 95   C    8  1 cycle 95   C     30 s    fluorescence measurement on the second step  1  Select in the main menu    Define Protocols  and click    Create a new protocol   Set the following parameters     30 0  2 1    9s    2         B   00 20     p w dw    pm       PCR   Melk Data  Acquisition        1 30 00 50 0     1 1500 95 0       1 0 20 95 0  gt    2 1 00 60 0  i Real Time        2  Select    Edit Plate  for iQ5 or  View Save Data  for iQiCycler to create the    plate    for samples and controls   Enter the concentrations of the Quantitative Standards  reported on the HCV TM 1Q Quant Data Card  in the   Standard Quantity  box  QS1 HCV  QS2 HCV  QS3 HCV as Standard in the HEX channel and Unknown  in the Fam  QS1 IC  QS2 IC  QS3 IC as Standard in the Fam channel and Unknown in the HEX channel    Use icon  Unknown  for all samples    3  Choose the FAM and HEX channels for all samples    4  For the iQ5 instrument enter the reaction volume    Sample volume  25 yl  Select caps type    Seal Type  and  tubes type    Vessel Type   Save plate setup    5  Begin the amplification    For the 1Q iCycler instrument activate the key    Run with selected plate  and choose in the window of the  opened menu  the reaction volume 25 ul  select    PCR Quantification Melt Curve  and    Experimental  Plate   Click the button    Begin Run  and save the experiment settings    For t
20. entration in the Quantity field as indicated in the Data Card     Target HCY RNA v  Task Standard v   Quantity 1000 0  Target Internal Control    Task Unknown v Quantity       Sample      Q81 HCV vi                   6  Click on id oes  insert Reaction volume  25 ul      Number of Cycles  42 2    HCV Real TM Quant    See  Pete  a       3 95  C     0 15  42  60  C     1 00      Legend    Data Collection On     Data Collection Off          7  Click to begin PCR reaction     8  After the run is complete  select all samples and standards and click    Analysis Settings       D      Real TM Quant April 16  2011 19    NONCAIM UC  V  Sacace    HCV    9     10   11     D    13     900 000    850 000    800 000    750 000 1    700 000    650 000    600 000      550 000      600 000    c 450 000      AR    350 000      300 000    250 000    200 000    150 000  gt     100 000  60 000  o     50 000      NS    400 000      Deselect Use default settings  Automatic threshold and automatic baseline  suggested values are      Cr Settings for HCV Cr Settings for Internal Control  CT Settings to Use      Use Default Settings Cr Settings to Use      Use Default Settings  E  Automatic Threshold    Automatic Threshold  Threshold    30 000 Threshold    35 000  E  Automatic Baseline    Automatic Baseline  Baseline Start Cycle    3 5 End Cycle    4  Baseline Start Cycle    3     End Cycle    15 7      for StepOne instrument recommended Threshold values are  Joe  HCV    1 000  Fam  IC    2 000   NOTE  
21. f patients with chronic hepatitis C Virus Infection       HIGH VIRAL    LOAD O Treatment Started    LOW VIRAL  LOAD        Virological Response  me   D  Nonresponse     E  Slow Response     F  Early Response   G  Rapid Response     H  Relapse          D Sustained Response    Log g HCV RNA    HCV RNA Testing  4 Quantitative  4 Qualitative or  Highly Sensitive  Stop Treatment  x All Genotypes  69 Genotype 2 or 3      Genotype 1 or 4          6 12 18 24 30 3  42 48 54 60 66 72  A A A    Time From Treatment Onset  wks       4  PRINCIPLE OF ASSAY   Kit HCV Real TM Quant is a Real Time test for the Quantitative detection of Hepatitis C Virus in human plasma   HCV RNA is extracted from plasma  amplified using real time amplification and detected using fluorescent reporter  dye probes specific for HCV or HCV IC    Internal Control  IC  serves as an extraction and an amplification control for each individually processed specimen  and to identify possible inhibition  IC is detected in a channel other than the HCV RNA  Monitoring the  fluorescence intensities during Real Time allows the detection and quantification of the accumulating product  without having to re open the reaction tube after the real time amplification     5  MATERIALS PROVIDED  Part N   1      Controls     RNA isolation controls   Part N   2      HCV Real TM Quant   Real Time kit     Part N  2      HCV Real TM Quant       RT PCR mix 1 TM     Standards    and controls    concentrations are specific for every lot     mu
22. h tube    Add 10 ul of extracted RNA sample  or quantitative standards to the  appropriate tube with Reaction Mix                   Number of Cycles  42             8  Click Start Run to begin PCR reaction     E    9  After the run is complete click     then click  o  tab to interpret the results     25    10  Click v to Export the samples  in the Export Options select Results Table    6 Export Options    Experiment File Name  HCV Quant exp 2 results run    Export Location  CN Browse     CSV PDF Report     Excel  Export Options  CO Component Data O Plate Layout  O Component Melt Data  C  Thermal Profile  O Amplification Data E Results T able  CO Melt Data        11  Using the exported file  calculate results pasting data from the Sample Name and the Quantity columns in the       provided HCV Quant Result Calculation sheet              FAM   Internal Control       Color Slope Y Intercept Efficiency    2      10  103 10  105 10  107 108 Bn  3 376 39 324 97 783  0 999 10  103 10      105 105    Standard Curve Standard Curve       26    15  RESULTS INTERPRETATION  The Internal Control  IC  is detected on the FAM channel and HCV RNA on the Joe HEX Cy3 channel     For each control and patient specimen  calculate the concentration of HCV RNA using the following formula     HCV DNA copies specimen  Joe HEX Cy3 channel       x coefficient    IU HCV mL  IC DNA copies specimen  FAM channel    coefficient is specific for each lot and reported in the HCV TM RG Quant Data Card provided in 
23. he 1Q5 instrument activate the button    Run    and choose in the window of the opened menu    Collect Well  Factors from Experimental Plate   Click the button  Begin Run  and save the experiment settings     15    Data Analysis   The results are interpreted with the software of    IQ iCycler  or  1Q5  through the presence of crossing of  fluorescence curve with the threshold line  Internal Control  IC  is detected on the FAM channel and HCV RNA on  the HEX channel   Activate the button    PCR Quant    for the results analysis    Results analysis for Fam channel  IC   Activate the button    Log View   Put the threshold line  with the left button of the mouse  at such level where curves of  fluorescence are linear  see figure        Results analysis for HEX channel  HCV RNA  Activate the button    Log View     Put the threshold line  with the left button of the mouse  at such level where the curves of  fluorescence are linear  see figure     ue      1 1  77       i                 t 15 5 40          16    MX3000P   and MX3005P    Stratagene     1  Open the program  select    Quantitative POR pe Standarts     and click    OK      New Experiment Options E   xj              CA ble    1         Feal  ane  i Molocuar Haacan bise Dome  Plate read    T Quantitative Plate Read     C Plate Read   Allele Discrimination           Meo Tumdempenderwemrug  f Don t show this again    At the top left of the window choose    Plate Setup      In the window    Well type    set    Unknown    for 
24. into the tube with DTT  300 ul of RT PCR mix 1  200 ul of RT   PCR mix 2 and vortex for at least 5 10 seconds  This mix is stable for 1 month at  20  C  Add for each sample  N   in the new sterile tube 12 5 N ul of mix  0 5  N ul of TaqF Polymerase and 0 25 N ul of M MLV  For the second  run it s recommended to import the standard curve generated in the first run   provided that at least one HCV and  IC standards is used    4  Add 12 5 ul of Reaction Mix into each tube   5  Add 12 5 ul of extracted RNA sample to the appropriate tube with Reaction Mix and mix by pipetting   If the Ribo Sorb isolation kit is used as a RNA extraction kit  re centrifuge all the tubes with extracted RNA for 2 min at  maximum speed  12000 16000 g  and take carefully supernatant  N B  don t disturb the pellet  sorbent inhibit reaction   6  Prepare for each run 6 standards and 1 negative control     e add 12 5 ul of Quantitation Standards HCV  QS1 HCV  QS2 HCV  QS3 HCV  into 3 labelled tubes   e add 12 5 ul of Quantitation Standards IC  QS1 IC  QS2 IC  QS3 IC  into 3 labelled tubes   e add 12 5 ul of TE buffer to the tube labelled Negative Control   Close the tubes and transfer them into the Real Time PCR instrument     14  PROTOCOL DATA ANALYSIS   SmartCycler    Cepheid     1  Select in the main menu    Define Protocols    and press the button    New Protocol     Give a name to the protocol  and set the following parameters     1  Stagel 50  C     1800 sec  2  Stage2 95  C     900 sec    2 Temperatu
25. ion of clinical specimens must comply with country  federal  state and local regulations for the  transport of etiologic agents     DOES UN    11  RNA ISOLATION  The following isolation kits are recommended     gt  Ribo Virus 100  spin column extraction kit  Sacace  K 2 C 100        Ribo Sorb 100  Sacace   Please carry out the RNA extraction according to the manufacturer s instructions  Add 5 ul of Internal Control  during the RNA isolation procedure directly to the sample lysis mixture  see 12  Internal Control     12  INTERNAL CONTROL  HCV Rec IC    HCV Rec IC is a quantitative Internal Control  concentration reported in Data Card  and represents recombinant  RNA containing structure which carried through all steps of analysis from nucleic acid extraction to PCR  amplification detection  The presence of quantitative HCV Rec IC allows not only to monitor the extraction  procedure and to check possible PCR inhibition but also to verify possible losses of the RNA during extraction  procedure thus enabling to calculate precisely the HCV viral load     13  REAGENT PREPARATION  Note  Reaction Mix volume   25 ul  1  Thaw one set of reagents  vortex and centrifuge briefly the tubes   2  Prepare reaction tubes or PCR plate   3  Prepare Reaction Mix  add into the tube with DTT  300 ul of RT PCR mix 1  200 ul of RT PCR mix 2  20  ul of Hot Start Taq Polymerase and 10 ul of M MLV Revertase  Vortex thoroughly and centrifuge briefly    If it is necessary to test less than 25 samples add 
26. l 16  2011 17       if Mx3000P  Standalone    Quantitative PCR    New Experiment mxp       5  x   File Edit Instrument Tools Options Section View Window Help    Plate Setup Thermal Profile Setup    Thermal Profile  gt        Deas     Add        Normal Segment    Plateau with Ramp    Data collection marker by dragging   Q  All points  Enders    Selection    Edit Properties         Delete      Temperature           a    Thermal profile comments     Segment 1 Segment 2 Segment 3  1 Cycle 1 Cycle 45 Cycles       Full Screen Profile   Start Hun      To edit  select plateaus  ramps or segment boundaries   Door Closed   Lamp Off  zm   Data Analysis   1  Soon after the amplification is over  choose the button    Analysis    at the top left of the window     2  Choose the button    Results      3  At the right angle of the window    Area to analyze  select    Amplification plots     4  Set    Threshold fluorescence    500 for the Joe channel and 1000 for the Fam channel    5  Inthe window    Text report  appear for each sample the values of Ct and experimental values of copies cDNA    HCV and cDNA IC   6  Take care that the value of RSq  correlation coefficient  in the window    Standard curve    is not lower than 0 9  for both channels      f Mx3000P  Standalone    Quantitative PCR       Server  Documents 3a    es Banepa Real  Stratagene Quantitative PCR  08 18 2005  17Hr 42Min mxp     E la  x   File Edit Instrument Tools Options Section view Window Help    Analysis Selection Setup C
27. ly on receipt     9  STABILITY   HCV Real TM Quant Test is stable up to the expiration date indicated on the kit label  The product will maintain  performance through the control date printed on the label  Exposure to light  heat or humidity may affect the shelf  life of some of the kit components and should be avoided  Repeated thawing and freezing of these reagents should  be avoided  as this may reduce the sensitivity  Components stored under conditions other than those stated on the  labels may not perform properly and may adversely affect the assay results     10  SAMPLE COLLECTION  STORAGE AND TRANSPORT  Note  Handle all specimens as if they are potentially infectious agents     1  EDTA tubes may be used with the HCV Real  TM Quant  Follow sample tube manufacturer s instructions    2  Whole blood collected in EDTA should be separated into plasma and cellular components by centrifugation at   800 1600 x g for 20 min within six hours  The isolated plasma has to be transferred into a sterile polypropylene   tube  Plasma may be stored at 2 8  C for an additional 3 days  Alternatively  plasma may be stored at  18  C for   up to one month or 1 year when stored at  70 C    Do not freeze whole blood    Specimens anti coagulated with heparin are unsuitable for this test    Thaw frozen specimens at room temperature before using    Whole blood must be transported at 2 25 C and processed within 6 hours of collection  Plasma may be   transported at 2 8  C or frozen    7  Transportat
28. nnel reported  in the protocol   e The IC was not added to the sample during the pipetting of reagents       Make attention during the RNA extraction procedure   Weak  Ct    40  signal on the Cy3 channel  retesting of the sample is required   The correlation coefficient R  is less than 0 9  retesting of all samples is required   The calculated concentrations of HCV Rec Pos  and or HCV Rec Pos2 are different from given control  concentrations  reported in the Data Sheet  retesting of all samples is required   Any signal on the Joe HEX Cy3 channel with Negative Control of extraction   e Contamination during RNA extraction procedure  All samples results are invalid     gt  Decontaminate all surfaces and instruments with sodium hypochlorite and ethanol     gt  Use only filter tips during the extraction procedure  Change tips among tubes     Repeat the RNA extraction with the new set of reagents   Any signal with Negative PCR Control   e Contamination during PCR preparation procedure  All samples results are invalid     gt  Decontaminate all surfaces and instruments with sodium hypochlorite and ethanol or  special DNA decontamination reagents        Pipette the Positive controls at the end       Repeat the PCR preparation with the new set of reagents     18  REFERENCES    l     3     11     L2     13   14     15     16     World Health Organization  Hepatitis C    Global prevalence  update   Wkly Epidemiol Rec  1999  74  425   427    Alter  H  1999 Discovery of non A  non B hepati
29. oe  In the Quantitative Analysis tab  select from the menu Dyes D    gt  Joe F   Select standard Concentration boxes  left  mouse click CTRL  for HCV   Status C   Sample   Concentration    P             Neg extr         POS1       m POS1         POS2      m POS2     Ej QS1 HCV 8 72e 005    Bou 5 46e 003     E Q53 HCV 310e  001        QS1 IC 1 88e 005    E QS2 IC 1 88e 004         QS3 IC 8 20e 002        NTC    Then do in this order   1  Under Analysis Method select Fit Points   2  Under Analysis Step Click Zero Adjust  select Auto and then click OK  Then click Baseline  3  Select 2 points and set the base line  drag and drop the red baseline  as low as possible but above the noise  of each sample  usually 15 20   4  Under Analysis Step clic Analysis  Calculated copies reaction of HCV and Ct Results should appear in the  Ct and the    Calculated Concentration    column respectively     Analysis Method Analysis Step   Points Selected Baseline and Threshold Display Mode               Driv  Max  Elzero Adjust Baseline  Logarithmic  ElBaseline ElFull Scale   MIFit Points Analysis Show Points Threshold   1 00 Smooth hi    24    Eco Real Time PCR System  Illumina          Quantification    Hydrolysis Probe    1  Open Eco software  click experiment and under Experiment Type select l   select RNA  as starting material for HCV Real TM Quant     2 U     o Standard Curve        nder Quantification Method select   and insert experiment name  Click    3  Click Q set up two assays  one for 
30. patients who have recovered from infection may remain anti HCV positive for many years   Conversely  during seroconversion  antibody tests may be negative   The direct molecular biology detection of HCV RNA by reverse transcriptase polymerase chain reaction  RT PCR   is considered the gold standard for the diagnosis of HCV infection     3  INTENDED USE   Kit HCV Real TM Quant is a Real Time Amplification test for the Quantitative detection of Hepatitis C Virus in  human plasma and the simultaneous detection of a HCV specific Internal Control  IC   by dual color detection   Quantitative HCV RNA testing provides prognostic information regarding likelihood of treatment response and it  plays an important role in monitoring the antiviral response to treatment  Sustained virological response is defined  as testing negative for HCV RNA  6 months after cessation of therapy  Recent studies suggest that the rate of  response to therapy is also important  For example  conversion to an HCV RNA negative test result after 4 weeks of  therapy constitutes a rapid virological response and is a strong predictor of treatment success  Patients who have not  had an early virological response  defined as at least a 2 log decline in HCV RNA  after 12 weeks of therapy are  unlikely to respond with an additional 36 weeks of therapy and should stop therapy    The same type of quantitative HCV RNA test should be used throughout a patient   s treatment course     Fig  2  Monitoring treatment response o
31. please slightly adapt suggested values according to your result curves      Apply Analysis Settings  Click S    Select all samples and standards  click View well table and then Click File   gt  Export  select Results only     Custornise Export       Sample Setup  v  Results  E  Raw Data  E  Multicomponent Data    Amplification Data    Export Properties          1  Select data to export     Click Customize Export and sort samples by reporter  as indicated in the above image  Then click CS J    export results in a file   Using the exported file  calculate results pasting data from the Sample and the Quantity columns in the provided  HCV Quant Result Calculation sheet     Amplification Plot    Amplification Plot    650 000      600 000    550 000    500 000 1  450 000    400 000    350 000      ARn    300 000    250 000 1    200 000      150 000    100 000 1    50 000                                                     28 x 32 3 35 38 40 2 4 2 4 6 a 10 12 14 16 18 20 2 24 28 28       2 E 3 3a 4n 42    2 24 2  Cycle Cycle    FAM   Internal Control Joe     HC    20    LightCycler 2 0    Roche     1  Select  New    LightCycler Experiment   Click E and insert as follows           Sety    Default Channel  530    e Max  Seek Pos  indicates the number of samples used in the  experiment    e Choose Instrument Type according to your instrument         Seek Temperature    0          Max  Seek Pos  13    Instrument Type   E Ch       Choose capillary Size according to your capillaries  we  
32. re 95  C   20 sec  Cycle 60  C   40 sec   Repeat     42 times        optics ON  2  Choose  Save Protocol   3  Click the  Create Run  button in the main menu  then the button  Dye Set  and select FCTC25      amp  Smart Cycler    18  x     User Logs Setup Tools Help              Check Status Stop Run                                                        sitein   Protocol Sample         Sample Type Notes FAM Std      3             TXR Std   Cy5 Std   Run Name  2     E 2    Selections   Site   Protocol   Notes    atl     AddRemow Sites     Dye Set   FCTC25 v   ch  Dye  Usage  Bkand        Name Sub  Mir      11 FAM Assay  ON 5  Protocols            Protocol        toNumer     3 Assay ON  5      213 TXR Assay  ON 5     Hn   N            Cy5 Assay  O 5 4  Graphs  HNA   EPEA  Standard FAM 2 Select All Sites  Standard Cy3  E d Cance Ol  FAM E   FR SST St SEE HIER         Rox       emperature m   secre err TE ERE EE EE     Start an    CancetRun Setup    Report Run Setup    SelectGranhe    ConyRun Setup     4  Choose    Add Remove Sites    and select in the new windows the protocol and the sites for analysis  Click      OR       Choose    Start Run  and give a name to the experiment    6  Insert in the column Sample Type UNKN for samples  Enter the concentrations of the Quantitative Standards   reported on the HCV TM SC Quant Data Card  in the columns Fam Std and Cy3 Std in order to generate  Standard curves    7  Fluorescence is observed in Real Time on the Cy3 channel for HCV RNA and
33. region  c100 3  as well as a part of c100 3  named 5 1 1  Third  generation anti HCV ELISA was introduced in Europe in 1993 and in the USA in 1996  In addition to the antigens  of ELISA 2  third generation anti HCV ELISA uses a NS5 region antigen of the viral genome  However  synthetic  peptide antigens  c22 and c 100  replaced recombinant antigens of ELISA 2  Other manufactures  for example  Abbott Diagnostics  used recombinant antigens derived from the same regions of HCV genome  Despite the  increased sensitivity and specificity with each generation of ELISA  false positive antibody results continue to be  observed  particularly among low risk blood donors  Thus  supplemental or confirmatory assays were developed in  parallel with ELISA  The recombinant immunoblot assay  RIBA  was used extensively to confirm the presence or  the absence of antibody to HCV epitopes  In RIBA  recombinant or peptide HCV antigens are blotted as separate  bands onto a nitrocellulose strip flanked by a weak positive  Level I  and a moderately positive  Level II  strip  control     Fig  1 Genome organization of HCV and antigens licensed for diagnostic use               Structural proteins           gt  gt   4 amp                      Nonstructural proteins                                                    EN EN a   5 1 1 NS5       c33c c100 3       c22 3 c200    Since ELISA and RIBA are antibody tests  the positivity of either one or both does not necessarily indicate current  HCV infection  as 
34. ribavirin compared with interferon alfa 2b plus ribavirin for initial treatment of  chronic hepatitis C  a randomized trial  Lancet  2001  358  958 965   Weiland O  Braconier JH  Fryden A  Norkrans G  Reichard O  Uhnoo I  Influence of pretreatment factors on outcome of interferon alpha treatment of patients  with chronic hepatitis C  Scand J Infect Dis  1999  31  115 118    Fried MW  Shiffman ML  Reddy KR et al  Peginterferon alfa 2a plus ribavirin for chronic hepatitis C virus infection N Engl J Med  2002  347 975 982    Jensen DM  Morgan TR  Marcellin P  et al  Early identification of HCV genotype 1 patients responding to 24weeks peginterferon alpha 2a  40kd  ribavirin  therapy  Hepatology  2006  43 954 960   Davis GL  Wong JB  McHutchison JG  Manns MP  Harvey J  Albrecht J  Early virologic response to treatment with peginterferonalfa 2b plus ribavirin in  patients with chronic hepatitis C  Hepatology  2003  38 645 652   Terrault NA  Pawlotsky JM  McHutchison J et al  Clinical utility of viral load measurements in individuals with chronic hepatitis C infection on antiviral  therapy  J Viral Hepat  2005  12 465 472    31    19  EXPLANATION OF SYMBOLS    Catalogue Number  RUO For Research Use Only    LOT  Lot Number    b Expiration Date    Contains reagents    Temperature limitation      Cycler    and iQ5    are trademarks of Bio Rad Laboratories     Rotor Gene    Technology is a registered trademark of Corbett Research   MX3000P   and MX3005P   are trademarks of Stratagene  
35. ry mode of HCV transmission is the exposure to infected human blood via intravenous  drug use or  unscreened transfusions  Nosocomial HCV transmission during dialysis  colonoscopy  and surgery has also been  reported  Perinatal and sexual transmission of the virus is inefficient  but occurs more frequently if the HCV   infected mother or sexual partner is also infected with human immunodeficiency virus type 1    Most people with acute HCV infection are asymptomatic or have mild symptoms  fatigue  nausea  jaundice  but  they are unable to clear the virus and in approximately 80  of cases this leads to chronic infection  In 15 to 20  of  patients chronic HCV infection progresses at a variable rate to cirrhosis  with a 1 to 4  annual risk of developing  hepatocellular carcinoma    The discovery of HCV genome in 1989 by Choo et al  paved the way for the development of serological and  molecular assays for viral hepatitis C  In the first generation of an enzyme linked immunosorbent assay  ELISA    wells of microtitre plates were coated with purified recombinant antigen c100 3 which was derived from the non   structural 4  NS4  region of the HCV genome  However  ELISA 1 was associated with a high percentage  50  to  70   of false positive results among low risk blood donors and in the presence of hyperglobulinemia  Thus  second   generation anti HCV ELISAs were developed  ELISA 2 by Ortho Diagnostics contained recombinant antigens from  the core  c22 3   NS3 region  c33c   and NS4 
36. ss Analysis and then select Quantitation Cycling A  Fam  Cycling A Green   Show   Turn off the automatic option Threshold    Press the buttons Dynamic Tube  Slope Correct   In the table of results  Quantitation Analysis  select More settings and set NTC Threshold to 10    Select Threshold  0 03   In the table of results  Quantitation Results  appear the values of Ct  Threshold cycle  which should be  lt  30    In the menu window Quantitation Results  column Calculation concentration  appear values of IC cDNA  copy specimen     Quantitation Analysis   Cycling AFAM Sybr  Page t             aS c ps                 ee   9       ABVRBLSRL SRS 5          HCV amplification analysis  Cycling A Joe  Yellow     1  Press Analysis and then select Quantitation Cycling A Joe  Cycling A  Yellow      Show   2  Turn off the automatic option Threshold    3  Press the buttons Dynamic Tube  Slope Correct   4  Inthe table of results  Quantitation Analysis  select More settings and set NTC Threshold to 6    5  Select Threshold  0 03   6  Inthe table of results  Quantitation Analysis  appear the values of Ct  Threshold cycle     7  Inthe new window Quantitation Results appear values of HCV cDNA copy specimen          amp  Quantitation Analysis   Cycling A JOE  Page 1              Sacace    HCV Real TM Quant April 16  2011 14    iQ iCycler    and iQ5     Biorad     Make sure that the 1Q5 instrument is calibrated for working with 50 ul reaction mix  Perform the calibration with the  same tubes in whi
37. st be used during the sample preparation procedure  see RNA isolation       must be used during the sample preparation procedure  add 100 ul of C     Negative Control  to labeled Cneg  add 90 ul of C     Negative  Control  and 10 ul of HCV Rec Pos controls to the tubes labeled Cposl and Cpos2       6  MATERIALS REQUIRED BUT NOT PROVIDED  RNA isolation kit  see 11  RNA isolation    Desktop microcentrifuge for    eppendorf    type tubes  Vortex mixer   Disposable gloves  powderless   Biohazard waste container   Refrigerator  Freezer   Real Time Thermal cycler   Workstation   Pipettes  adjustable    Sterile pipette tips with filters   Tube racks    Sacace    HCV Real TM Quant April 16  201 1 6    7  WARNINGS AND PRECAUTIONS  1  Wear disposable gloves  laboratory coats and eye protection when handling specimens and reagents   Thoroughly wash hands afterward     2  Use routine laboratory precautions  Do not eat  drink  smoke  apply cosmetics  or handle contact lenses in  laboratory work areas  Do not pipette by mouth    3  Do not use a kit after its expiration date    4  Do not mix reagents from different kits    5  Dispose all specimens and unused reagents in accordance with local regulations    6  Heparin has been shown to inhibit reaction  The use of heparinized specimens is not recommended    7  Avoid repeated thawing and freezing of the reagents  this may reduce the sensitivity of the test    8  Once the reagents have been thawed  vortex and centrifuge briefly the tubes    9 
38. targat sequence and a TaqMan   probe t                What type of experiment do you want to set up        Which ramp speed do you want to use in the instrument run     cus              For optimal results with the standard ramp speed  Applied Biosystems recommerds using standart                                  Use standards to determine the absolute quantity oftargel nucleic acid sequence in samples              Note  when using StepOne software  be sure to click New Experiment     Advanced Setup    5  StepOne    Software v2 1    File Edit Instrument Analysis   Wdvenced Satun           New Experiment      Advanced setup       Design Wizard          From Template       2  Click EM add two targets  one for HCV  Reporter Joe  and the other one for Internal Control   Reporter Fam   select Quencher     None for both targets   3  Click Add new samples multiple times and enter name for all samples  controls and standards    4  Click Assign target and Samples tab  associate samples  m h and standards  ES with the two targets and  select None as passive reference dye as shown in the following pictures     3         1   2              Assign target s  to the selected wells            Q 1 ic 52 IC 53 Ic       Task               Assign Target Neg Control Sample 1 Sample 2 nev RNA            OT   a                              Select the dye to       5  Double click on each standard well  a new small window will appear  and enter both for HCV and Internal  control the corresponding conc
39. the kit   Results may by calculated also using    HCV Quant Result Calculation sheet  provided with the kit    To obtain the results in copies mL multiply the IU HCV ml value by 4   IU RNA HCV mL x 4   copies RNA HCV mL    21    16  PERFORMANCE CHARACTERISTICS   The sensitivity  the specificity  the performance and the linearity of the kit HCV Real TM Quant were determined   with the following samples    1  The HCV RNA standard with the concentration of 2x10    U ml and its dilution    2  The samples of blood plasma obtained from 215 patients with IgG antibodies to HCV  The results were  compared to those ones obtained by the reference manufacturer   s kit    3  80 samples of blood plasma obtained from donors      Analytical specificity   The analytical specificity of the primers and the probes was validated with 80 negative samples  They did not generate  any signal with the specific HCV primers and probes  The specificity of the kit HCV Real TM Quant was 100    The potential cross reactivity of the kit HCV Real TM Quant was tested also against the group control listed in the  following table  It was not observed any cross reactivity with these pathogens     Table 1  Testing the specificity of the kit with other pathogens     Adenovirus type         o              Adenovirus type3                       Herpes simplex virus 1  Herpes simplex virus 2    Human herpes virus 6    Human herpes virus 8    HPV groups 6  y  u  1 3 4 5 8 37 38 65 20 24 49 50 15     HPV group a  6 11 2
40. the samples and    Standard    to identify calibrators    In the window    Collect fluorescence data  select for all samples the channels Fam and Joe    Enter the concentrations of the Quantitative Standards  reported on the HCV TM MX Quant Data Card  in the   Standard Quantity  box     yf Mx3000P  Standalone    Quantitative PCR    New Experiment mxp   E 2  18  x   File Edit Instrument Tools Options Section View Window Help    De  B amp B oo 8 RE    Thermal Profile Setup      bm 9              Import      Defaults   bg    Buick Setup    Welltype   Standard  gt      Collect fluorescence data   Iv JOE   HEX     Rox       FAM  Reference dye    lt none gt     All wells  Standard quantity     lt all gt  2     O0e 000    Auto Increment  10x   m    Standard units   copies      Identify replicates  Replicate symbol    lt none gt  x   Auto Increment    Clear Selected Wells      Plate setup comments     Full Screen Plate   Next  gt            FAM   Pe    Assign Well Type and Dyes to selected wells  The Ctrl key selects multiple wells   Door Closed   LampOff  Sm  6  At the top left of the window select button    Thermal Profile Setup   7  Set the following parameters of amplification    1  50  C     30 min   2  95  C     15 min   3  05 C     20 sec   60  C     60 sec   45 Cycles  8  Fluorescence is measured at 60 C  To do this  set on the Thermal Profile graph the  Endpoints  marker   9  Click  Run  button  enter a name for the experiment and save it     Sacace    HCV Real TM Quant Apri
41. tis and identification of its etiology  Am  J  Med 107  Supp1 6B  16S   20S    Bendinelli  M   M  Pistello  F  Maggi  and M  Vatteroni  2000  Blood borne hepatitis viruses  hepatitis B  C  D  and G viruses and TT virus  p 306   337  In  S Specter  R L  Hodinka  and S A  Young  ed    Clinical virology manual  3  ed  ASM Press  Washington  D C    Choo  Q  L   G  Kuo  A J  Weiner  L R  Overby  D W  Bradley  and M  Houghton  1989  Isolation of a cDNA clone derived from a blood borne non A  non B  viral hepatitis genome  Science 244  359 362    Lauer  G M   and B D  Walker  2001  Hepatitis C virus infection  N  Engl  J  Med  345  41   52   De Medina M  Schiff ER  Hepatitis C  diagnostic assays  Sem Liver Dis  1995  1  33 40   Kao  J  H   M  Y  Lai  Y  T  Hwang  P  M  Yang  P  J  Chen  J  C  Sheu  T  H  Wang  H  C  Hsu  and D  S  Chen  1996  Chronic hepatitisC without anti   hepatitis C antibodies by second generation assay  a clinicopathologic study and demonstration of the usefulness of a third generation assay  Dig  Dis  Sci  41   161 165    Schiff ER  de Medina M  Kahn RS  New perspectives in the diagnosis of hepatitis C  Sem Liver Dis  1999  19  3 15    Lok ASF  Gunaratnam NT  Diagnosis of hepatitis C  Hepatology  1997  26 Suppl 1   48 565    Pawlotsky J M   Bouvier Alias M  Hezode C  Darthuy F  Remire J  Dhumeaux D  Standardization of hepatitis C virus RNA quantification  Hepatology  2000   32  654 659    Manns MP  McHutchison JG  Gordon SC  et al  Peginterferon alfa 2b plus 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Western Digital XE 300GB  Sh 30 - Grundig Business Systems GmbH  LQD04B Bx FITDESIGN  FDL-PCI  EN_Access_Manager_User_Manual-DC1-0057A-Rev  Scania press release  BIO 300 - Broyeurs de végétaux  Procédures d`installation de mise/remise à niveau pour série 300 et  User Guide Manual - Service Module For Branches - Xplor  SEG DPP 1135-090    Copyright © All rights reserved. 
   Failed to retrieve file