Home
R&S®FSV Signal Analyzer Specifications
Contents
1. Frequency Frequency range R amp S FSV 3 DC coupled 9 kHz to 3 6 GHz AC coupled 1 MHz to 3 6 GHz R amp S FSV 7 DC coupled 9 kHz to 7 GHz AC coupled 1 MHz to 7 GHz with R amp S FSV B29 option DC coupled 20 Hz to fmax Frequency resolution 0 01 Hz Reference frequency internal Accuracy time since last adjustment x aging rate temperature drift calibration accuracy Aging per year standard 1x 10 with R amp S FSV B4 OCXO reference 1x 107 frequency option Temperature drift 5 C to 45 C standard 1x 10 with R amp S FSV B4 OCXO reference 1x 107 frequency option Max initial calibration accuracy standard 5x10 with R amp S FSV B4 OCXO reference 5x 10 frequency option Frequency readout Marker resolution 1 Hz Uncertainty marker frequency x reference uncertainty 10 x resolution bandwidth Ye span sweep points 1 1Hz Number of sweep trace points default value 691 range 101 to 32001 Marker tuning frequency step size marker step size sweep points span Sweep points 1 marker step size standard span default sweep points 1 Frequency counter resolution 0 001 Hz Count accuracy frequency x reference uncertainty Ya last digit Display range for frequency axis 0 Hz 10 Hz to max frequency Resolution 0 1 Hz Max span deviation 0 1 Spectral purity SSB phase noise frequenc
2. Output level center frequency gt 10 MHz span 0 Hz 1 V open circuit nominal signal at reference level and center frequency Trigger out Connector BNC female Output TTL compatible 0 V 5 V USB interface 2 ports type A plug version 2 0 Version 01 00 June 2008 R amp S FSV Signal Analyzer 11 Ordering information Designation Type Order No Signal Analyzer R amp S FSV3 1307 9002 03 Signal Analyzer R amp S FSV7 1307 9002 07 Accessories supplied Power cable quick start guide and CD ROM with operating manual and service manual Options Designation Type Order No Retrofittable Remarks Ruggedized Housing R amp S FSV B1 1310 9500 02 no Audio Demodulator R amp S FSV B3 1310 9516 02 yes retrofit in service center OCXO Reference Frequency R amp S FSV B4 1310 9522 02 yes user retrofittable Additional Interfaces R amp S FSV B5 1310 9539 02 yes IF out video out AUX port trigger out 2 x USB Spare Hard drive R amp S FSV B19 1310 9574 02 yes user retrofittable RF Preamplifier 9 KHz to 7 GHz R amp S FSV B22 1310 9600 02 yes user retrofittable Electronic Attenuator 1 dB steps R amp S FSV B25 1310 9622 02 yes user retrofittable Frequency Range Extension 20 Hz R amp S FSV B29 1310 9639 02 yes retrofit in service center 40 MHz Analysis Bandwidth R amp S FSV B70 1310 9645 02 yes user retrofittable Firmware softwar
3. lt 110 dBm typ 120 dBm 1 kHz lt 120 dBm typ 130 dBm Spurious responses spurious at fn 4205 MHz Image response fin 2 x 8410 MHz 1st IF typ lt 80 dBc fin 2 x 730 MHz 2nd IF lt 80 dBc fin 2 x 90 MHz 3rd IF lt 80 dBc Intermediate frequency response 1st IF 8410 MHz typ lt 70 dBc 2nd IF 730 MHz lt 80 dBc 3rd IF 90 MHz lt 80 dBc Residual spurious response 0 dB RF attenuation f lt 1MHz lt 90 dBm f gt 1 MHz lt 103 dBm Local oscillator related spurious 1 kHz lt offset from carrier lt 10 MHz lt 70 dBc offset from carrier gt 10 MHz lt 80 dBc Other interfering signals Subharmonic of 1st LO spurious at 8410 MHz 2 x fin lt 70 dBc Harmonic of 1st LO mixer level lt 25 dBm lt 70 dBc Level display Logarithmic level axis 1 dB to 200 dB in steps of 1 2 5 Linear level axis 10 of reference level per level division 10 divisions or logarithmic scaling Number of traces 6 Trace detector Max Peak Min Peak Auto Peak Normal Sample RMS Average EMI detectors Quasi Peak Trace functions Clear Write Max Hold Min Hold Average View Setting range of reference level 130 dBm to 10 dBm RF attenuation RF preamplifier gain in steps of 0 01 dB Units of level axis logarithmic level display dBm dBuV dBmV dBA dBpW linear level disp
4. nominal Shape factor 60 dB 3 dB lt 5 nominal Channel filters Bandwidths 3 dB standard RRC root raised cosine 100 Hz 200 Hz 300 Hz 500 Hz 1 1 5 2 2 4 2 7 3 3 4 4 4 5 5 6 8 5 9 10 12 5 14 15 16 18 RRC 20 21 24 3 RRC 25 30 50 100 150 192 200 300 500 kHz 1 1 228 1 28 RRC 1 5 2 3 3 84 RRC 4 096 RRC 5 10 20 28 MHz with R amp S FSV B70 option 40 MHz additionally Bandwidth accuracy lt 2 nominal Shape factor 60 dB 3 dB lt 2 nominal EMI filters Bandwidths 6 dB 200 Hz 9 kHz 120 kHz 1 MHz Bandwidth uncertainty lt 3 nominal Shape factor 60 dB 3 dB lt 6 nominal 1 Net sweep time without additional hardware settling time Time for data acquisition for FFT calculation 4 R amp S FSV Signal Analyzer Version 01 00 June 2008 Video bandwidths 1 Hz to 10 MHz in 1 2 3 5 sequence 20 MHz 28 MHz with R amp S FSV B70 option 40 MHz additionally Signal analysis bandwidth standard 28 MHz nominal with R amp S FSV B70 option 40 MHz nominal Level Display range displayed noise floor up to 30 dBm Maximum input level DC voltage AC coupled 50 V DC coupled OV CW RF power RF attenuation 0 dB RF preamplifier OFF 20 dBm 0 1 W with R amp S FSV B22 option RF preamplifier ON 13
5. dB Linear level display S N gt 16 dB 0 dB to 70 dB 5 of reference level Total measurement uncertainty signal level 0 dB to 70 dB below reference type sweep RF attenuation 10 dB 20 dB level S N gt 20 dB sweep time auto sweep 30 dB 40 dB RF preamplifier OFF span RBW lt 100 95 confidence level 20 C to 30 C 9 kHz lt f lt 10 MHz 0 39 dB 10 MHz lt f lt 3 6 GHz 0 28 dB 3 6 GHz lt f lt 7 GHz 0 39 dB Measurement speed Local measurement and display update rate 2 ms 500 s Remote measurement 1000 sweep averages 1 ms 1000 s Remote measurement and LAN transfer 3 ms 333 s Marker peak search 1 5 ms Center frequency tune and transfer 15 ms 3 Therefore the reference level setting causes no additional uncertainty in measurement results Version 01 00 June 2008 The setting of the reference level affects only the graphical representation of the measurement result on the display not the measurement itself R amp S FSV Signal Analyzer 7 Trigger functions Trigger Trigger source free run video external IF power Trigger offset span 2 10 Hz 31 25 ns to 30 s resolution 31 25 ns min or 1 of offset span 0 Hz sweep time to 30 s resolution 31 25 ns min or 1 of offset Max deviation of trigger offset 7 8125 ns 0 1 x trigger offset IF power trigge
6. lt f lt 7 GHz typ 2 Setting range of attenuator standard 0 dB to 75 dB in 5 dB steps with R amp S FSV B25 option 0 dB to 75 dB in 1 dB steps RF preamplifier gain with R amp S FSV B22 option 20 dB nominal Probe power supply Supply voltages 15 V DC 12 6 V DC and ground max 150 mA nominal Noise source drive Connector BNC female Output voltage 0 V 28 V max 100 mA switchable nominal Power sensor Connector 6 pin LEMOSA female for supported R amp S NRP Zxx power sensors USB interface 2 ports type A plug version 2 0 Reference output Connector BNC female Impedance 50 Q Output frequency internal reference 10 MHz external reference same as reference input signal Level gt 0 dBm nominal Reference input Connector BNC female Impedance 50 O Input frequency range Required level 1 MHz lt fin lt 20 MHz in 1 MHz steps gt 0 dBm from 50 Q External trigger gate input Connector BNC female Trigger voltage 0 5Vto3 5V Input impedance 10 kQ IEC IEEE bus control interface in line with IEC 625 2 IEEE 488 2 Command set SCPI 1997 0 Connector 24 pin Amphenol female Interface functions SH1 AH1 T6 L4 SR1 RL1 PP1 DC1 DT1 CO External monitor Connector VGA compatible 15 pin mini D Sub Version 01 00 J
7. 9 22 Version 01 00 June 2008 R amp S FSV Data without tolerance limits is not binding Subject to change 0 14 min within German wireline network rates may vary in other networks wireline and mobile and countries
8. R amp S FSV signal Analyzer Specifications Si ooosee eae Ceara ma j MII Years of ya nae a Qu DE amp SCHWARZ Test amp Measurement CONTENTS Seca en 3 FIOQUONCY A ONO 3 NS 4 EA E O NANA 4 Ll ii A A da AA AAA AA AAA RNGA A Ad 5 MGaSUTEMENE SPO e A davecu ceva div ustesnduitusvotedsy cvs sueduecabacy adi da baeudsevh coun coettestensades 7 NO 8 VO datas sti A AAA et ee ie a it a ee en 8 INPUTS ANG OUTPUTS aiii 9 General AAN 10 OPTIONS ests ts Becca cetes T OO 11 R amp S ESV283 audio demodulator iia tata 11 REUS ESVEB aia A ance do 11 Ordering information iii ii 12 Opt Sia aa A AAA A A ed si anes eis 12 dll AAA O NN 13 Power sensors supported by RESTFSV K9 Option c cscsscsssssssssessesssesussnssssesscssssussusssssssessessssnssnssssessessssussnsarsaseeteesscsussusareasecsecareeseeees 13 2 R amp S FSV Signal Analyzer Version 01 00 June 2008 Specifications Specifications apply under the following conditions 30 minutes warm up time at ambient temperature specified environmental conditions met calibration cycle adhered to and all internal automatic adjustments performed Data without tolerances typical values only Data designated nominal applies to design parameters and is not tested Rohde amp Schwarz equipment is designed for reliable operation up to an altitude of 3000 m above sea level and for transport up to an altitude of 4500 m above sea level
9. average noise level 0 dB RF attenuation termination 50 Q log scaling normalized to 1 Hz RBW RBW 1 kHz VBW 3 kHz zero span sweep time 50 ms sample detector trace average sweep count 20 mean marker RF preamplifier OFF 9 kHz lt f lt 100 kHz lt 130 dBm typ 140 dBm 100 kHz lt f lt 1 MHz lt 145 dBm typ 150 dBm 1 MHz lt f lt 1 GHz lt 152 dBm typ 155 dBm 1 GHz lt f lt 3 6 GHz lt 150 dBm typ 153 dBm 3 6 GHz lt f lt 6 GHz lt 148 dBm typ 151 dBm 6 GHz lt f lt 7 GHz lt 146 dBm typ 149 dBm with R amp S FSV B22 option RF preamplifier ON 0 dB RF attenuation termination 50 Q log scaling normalized to 1 Hz RBW RBW 1 kHz VBW 3 kHz zero span sweep time 50 ms sample detector trace average sweep count 20 mean marker 100 kHz lt f lt 1 MHz lt 150 dBm typ 155 dBm 1 MHz sf lt 1 GHz lt 162 dBm typ 165 dBm 1 GHz lt f lt 3 6 GHz lt 160 dBm typ 163 dBm 3 6 GHz lt f lt 6 GHz lt 158 dBm typ 161 dBm 6 GHz lt f lt 7 GHz lt 156 dBm typ 159 dBm with R amp S FSV B29 option RF preamplifier OFF 0 dB RF attenuation termination 50 Q log scaling normalized to 1 Hz RBW RBW 5 Hz VBW 5 Hz zero span sweep time 500 ms sample detector trace average sweep count 20 mean marker 20 Hz lt 100 dBm typ 110 dBm 100 Hz
10. dBm 0 02 W RF attenuation 210 dB RF preamplifier OFF 30 dBm 1 W with R amp S FSV B22 option RF preamplifier ON 23 dBm 0 2 W Pulse spectral density RF attenuation 0 dB RF preamplifier OFF 97 dB uV MHz Max pulse voltage RF attenuation 10 dB 150 V Max pulse energy RF attenuation 2 10 dB 10 us 1 mWs Intermodulation 1 dB compression of input mixer RF attenuation 0 dB RF preamplifier OFF 3 dBm nominal with R amp S FSV B22 option RF preamplifier ON 12 dBm nominal Third order intercept point TOI level 2 x 10 dBm Af gt 5 x RBW or 10 kHz whichever is larger RF preamplifier OFF 10 MHz lt fin lt 100 MHz gt 12 dBm typ 15 dBm 300 MHz lt fin lt 3 6 GHz gt 13 dBm typ 16 dBm 3 6 GHz lt fin lt 7 GHz gt 15 dBm typ 18 dBm with R amp S FSV B22 option RF preamplifier ON level 2 x 10 dBm Af gt 5 x RBW or 10 kHz whichever is larger 10 MHz lt fin lt 100 MHz 3 dBm nominal 300 MHz lt fi lt 3 6 GHz 2 dBm nominal 3 6 GHz lt fin lt 7 GHz 0 dBm nominal Second harmonic intercept SHI 100 MHz lt fin 3 5 GHz RF preamplifier OFF typ 40 dBm with R amp S FSV B22 option RF preamplifier ON 25 dBm nominal Version 01 00 June 2008 R amp S FSV Signal Analyzer 5 Displayed
11. e Analog Modulation Analysis AM FM M R amp S FSV K7 1310 8103 02 Power Sensor Support R amp S FSV K9 1310 8203 02 supports R amp S NRP power sensors Analysis GSM EDGE Analysis R amp S FSV K10 1310 8055 02 3GPP FDD BS Analysis R amp S FSV K72 1310 8503 02 3GPP FDD UE Analysis R amp S FSV K73 1310 8555 02 WLAN IEEE 802 1 1a b g j Analysis R amp S FSV K91 1310 8903 02 WLAN IEEE 802 11n Analysis R amp S FSV K9In 1310 9468 02 requires R amp S FSV B70 WiMAX IEEE 802 16 OFDM OFDMA R amp S FSV K93 1310 9416 02 EUTRA LTE BS Analysis R amp S FSV K100 1310 9051 02 EUTRA LTE UE Analysis R amp S FSV K101 1310 9100 02 12 R amp S FSV Signal Analyzer Version 01 00 June 2008 Recommended extras Designation Type Order No Headphones 0708 9010 00 IEC IEEE Bus Cable 1 m R amp S PCK 0292 2013 10 IEC IEEE Bus Cable 2 m R amp S PCK 0292 2013 20 19 Rack Adapter not for R amp S FSV B1 R amp S ZZA 478 1096 3248 00 Soft Carrying Case grey R amp S ZZT 473 1109 5048 00 Matching pads 50 75 Q L Section matching at both ends R amp S RAM 0358 5414 02 Series Resistor 25 Q matching at one R amp S RAZ 0358 5714 02 end taken into account in instrument function RF INPUT 75 Q SWR bridges 50 Q SWR Bridge 5 MHz to 3 GHz R amp S ZRB2 0373 9017 5X SWR Bridge 40 kHz to 4 GHz R amp S ZRC 1039 9492 5X High power att
12. enuators 100 W 3 6 10 20 30 dB 1 GHz R amp S RBU100 1073 8495 XX XX 03 06 10 20 30 50 W 3 6 10 20 30 dB 2 GHz R amp S RBU50 1073 8695 XX XX 03 06 10 20 30 50 W 20 dB 6 GHz R amp S RDLS50 1035 1700 52 Connectors and cables Probe power connector 3 pin 1065 9480 00 DC blocks DC Block 10 kHz to 18 GHz type N R amp S FSE Z4 1084 7443 02 Power sensors supported by R amp S FSV K9 option Order designation Type Order No Average Power Sensor R amp S NRP Z11 1138 3004 02 10 MHz to 8 GHz 200 mW Average Power Sensor R amp S NRP Z21 1137 6000 02 10 MHz to 18 GHz 200 mW Average Power Sensor R amp S NRP Z22 1137 7506 02 10 MHz to 18 GHz 2 W Average Power Sensor R amp S NRP Z23 1137 8002 02 10 MHz to 18 GHz 15 W Average Power Sensor R amp S NRP Z24 1137 8502 02 10 MHz to 18 GHz 30 W Power Sensor Module with Power Splitter R amp S NRP Z27 1169 4102 02 DC to 18 GHz 500 mw Power Sensor Module with Power Splitter R amp S NRP Z37 1169 3206 02 DC to 26 5 GHz 500 mw Average Power Sensor R amp S NRP Z91 1168 8004 02 9 kHz to 6 GHz 200 mW Thermal Power Sensor R amp S NRP Z51 1138 0005 02 0 Hz to 18 GHz 100 mW Thermal Power Sensor R amp S NRP Z55 1138 2008 02 0 Hz to 40 GHz 100 mW Wideband Power Sensor R amp S NRP Z81 1137 9009 02 50 MHz to 18 GHz 100 mW Version 01 00 June 2008 R amp S FSV Signal Analyzer 13 Service
13. ing use of instrument The instrument complies with the emission requirements stipulated by EN 55011 class A This means that the instrument is suitable for use in industrial environments According to EN 61000 6 4 operation is not covered in residential commercial and business areas nor in small size companies Thus the instrument must not be operated in residential commercial and business areas nor in small size companies unless additional measures are taken to ensure that EN 61000 6 3 is met 10 R amp S FSV Signal Analyzer Version 01 00 June 2008 Options R amp S FSV B3 audio demodulator Demodulation AF demodulation types AM and FM Audio output loudspeaker and phone jack Marker stop time in spectrum mode 100 ms to 60 s AF output Connector 3 5 mm mini jack Output impedance 100 Open circuit voltage up to 1 5 V adjustable R amp S FSV B5 additional interfaces User port Connector 9 pin D Sub male Output TTL compatible 0 V 5 V max 15 mA Input TTL compatible max 5 V IF video demod out span 0 Hz Connector BNC female 50 Q IF out Bandwidth equal to RBW setting IF frequency 32 MHz Output level gain versus RF input RF attenuation 0 dB RF preamplifier OFF 0 dB nominal Video out Bandwidth equal to VBW setting Output scaling display scale log log scaling display scale lin lin scaling
14. lay uV mV HA mA pW nW 6 R amp S FSV Signal Analyzer Version 01 00 June 2008 Level measurement uncertainty Absolute level uncertainty at 64 MHz RBW 10 kHz level 10 dBm reference level 10 dBm RF attenuation 10 dB 20 C to 30 C lt 0 2 dB a 0 07 dB 5 C to 40 C lt 0 35 dB o 0 12 dB Frequency response referenced to 64 MHz DC coupling RF attenuation 10 dB 20 dB 3 RF preamplifier OFF 20 C to 30 C 0 dB 40 dB 9 kHz lt f lt 10 MHz lt 0 5 dB o 0 17 dB 10 MHz lt f lt 3 6 GHz lt 0 3 dB a 0 1 dB 3 6 GHz lt f lt 7 GHz lt 0 5 dB a 0 17 dB any setting of RF attenuation or RF preampl ifier 5 C to 40 C 9 kHz lt f lt 3 6 GHz lt 1 dB o 0 33 dB 3 6 GHz lt f lt 7 GHz lt 1 5 dB o 0 5 dB with R amp S FSV B29 option DC coupling RF preamplifier OFF 5 C to 40 C 20 Hz lt f lt 9 kHz lt 1 dB o 0 33 dB Attenuator switching uncertainty f 64 MHz 0 dB to 70 dB referenced to 10 dB attenuation lt 0 2 dB o 0 07 dB Uncertainty of reference level setting 0 dB Bandwidth switching uncertainty referenced to RBW 10 kHz sweep filters lt 0 1 dB a 0 04 dB FFT filters lt 0 2 dB o 0 07 dB Display nonlinearity Logarithmic level display S N gt 16 dB 0 dB to 70 dB lt 0 1 dB a 0 04
15. r Sensitivity minimum signal power 60 dBm RF attenuation RF preamplifier gain maximum signal power 10 dBm RF attenuation RF preamplifier gain IF power trigger bandwidth RBW gt 500 kHz swept RBW gt 20 kHz FFT 40 MHz nominal RBW lt 500 kHz swept RBW lt 20 kHz FFT 6 MHz nominal Gated sweep Gate source video external IF power Gate delay 31 25 ns to 30 s min resolution 31 25 ns or 1 of delay Gate length 31 25 ns to 30 s min resolution 31 25 ns or 1 of gate length Max deviation of gate length 7 8125 ns 0 1 x gate length 1 Q data Interface GPIB or LAN interface Memory length max 200 Msamples and Q Word length of I Q samples sampling rate gt 64 MHz or 18 bit number of samples gt 100 Msample otherwise 24 bit Sample rate standard 100 Hz to 45 MHz with R amp S FSV B70 option 100 Hz to 128 MHz Max signal bandwidth equalized standard 28 MHz with R amp S FSV B70 option 40 MHz Amplitude flatness 0 3 dB nominal Deviation from linear phase 1 nominal 8 R amp S FSV Signal Analyzer Version 01 00 June 2008 Inputs and outputs RF input Impedance 50 Q Connector N female VSWR RF attenuation 210 dB 10 MHz lt f lt 3 6 GHz lt 1 5 3 6 GHz lt f lt 7 GHz lt 2 RF attenuation lt 10 dB DC coupled 10 MHz
16. une 2008 R amp S FSV Signal Analyzer 9 General data Data storage Internal hard disk gt 40 Gbyte nominal External supports USB 2 0 compatible memory devices Temperature Temperature operating temperature range 5 C to 40 C permissible temperature range 0 C to 50 C storage temperature range 40 C to 70 C Climatic loading 40 C at 90 rel humidity in line with EN 60068 2 30 Mechanical resistance Vibration sinusoidal 5 Hz to 150 Hz max 2 g at 55 Hz 0 5 g from 55 Hz to 150 Hz in line with EN 60068 2 6 random EN 60068 2 64 10 Hz to 130 Hz acceleration 1 2 g rms Shock 40 g shock spectrum in line with MIL T 28800F classes 3 and 5 EMC EMC Directive 2004 108 EC including EN 61326 class B emission CISPR 11 EN 55011 group 1 class A emission EN 61326 table A 1 immunity industrial Recommended calibration interval 1 year Power supply AC supply 100 V to 240 V 3A to 1 25 A 50 Hz to 400 Hz class of protection to VDE 411 Power consumption typ 90 W max 175 W with all options Safety in line with EN 61010 1 UL 3111 1 CSA C22 2 No 1010 1 IEC 1010 1 Test mark VDE GS CSA CSA NRTL Weight and dimensions Dimensions WxHxD 412 mm x 197 mm x 417 mm 16 22 in x 7 76 in x 16 42 in Net weight without options nominal R amp S FSV3 R amp S FSV7 9 5 kg 20 94 Ib 4 Note regard
17. y 500 MHz carrier offset 100 Hz lt 84 dBc 1 Hz 1 kHz lt 101 dBc 1 Hz 10 kHz lt 106 dBc 1 Hz 100 kHz lt 115 dBc 1 Hz 1 MHz lt 134 dBc 1 Hz 10 MHz typ 150 dBc 1 Hz Residual FM frequency 500 MHz RBW 1 kHz lt 3 Hz nominal sweep time 100 ms Version 01 00 June 2008 R amp S FSV Signal Analyzer 3 60 70 80 90 100 110 120 130 SSB phase noise dBc 1Hz 140 150 160 100 Hz 1 kHz Sweep time 10 kHz 100 kHz 1 MHz 10 MHz e center frequency lt 20 MHz e center frequency 200 MHz a center frequency 500 MHz a center frequency 1 GHz x center frequency 3 GHz x center frequency 6 GHz Typical phase noise at different center frequencies Range span 0 Hz 1 us to 16000 s span gt 10 Hz swept 1 ms to 16000 s span gt 10 Hz FFT 7 us to 16000 s Sweep time accuracy span 0 Hz 0 1 nominal span gt 10 Hz swept 3 nominal Resolution bandwidths Sweep filters and FFT filters Resolution bandwidths 3 dB span gt 10 Hz sweep filters 1 Hz to 10 MHz in 1 2 3 5 sequence span gt 10 Hz FFT filters 1 Hz to 300 kHz in 1 2 3 5 sequence span 0 Hz 20 MHz 28 MHz additionally with R amp S FSV B70 option span 0 Hz 40 MHz additionally Bandwidth uncertainty lt 3
18. you can rely on 1 In 70 countries 1 Person to person Customized and flexible 1 Quality with a warranty 1 No hidden terms About Rohde amp Schwarz Rohde amp Schwarz is an independent group of companies specializing in electronics It is a leading supplier of solu tions in the fields of test and measurement broadcasting radiomonitoring and radiolocation as well as secure com munications Established 75 years ago Rohde amp Schwarz has a global presence and a dedicated service network in over 70 countries Company headquarters are in Munich Germany Regional contact Europe Africa Middle East 49 1805 12 42 42 or 49 89 4129 137 74 customersupport rohde schwarz com North America 1 888 TEST RSA 1 888 837 8772 customer support rsa rohde schwarz com Latin America 1 410 910 7988 customersupport la rohde schwarz com Asia Pacific 65 65 13 04 88 customersupport asia rohde schwarz com Certified Quality System Certified Environmental System ISO 9001 ISO 14001 DOS REG NO 1954 QM DOS REG NO 1954 UM For product brochure see PD 5214 0499 12 and www rohde schwarz com search term FSV Rohde amp Schwarz GmbH 8 Co KG MuhldorfstraBe 15 81671 Munchen Phone 4989 41 290 Fax 49894129121 64 www rohde schwarz com R amp S is a registered trademark of Rohde 8 Schwarz GmbH amp Co KG Trade names are trademarks of the owners Printed in Germany as PD 5214 049