Service Manual TOUL 400 TOUL 200

image

Contents

1. Loose connection Broken cable None of the above Due for yearly maintenance A connector is loose The control panel is broken The sterile shield is mounted incorrectly Bar code reader is not working The bar code reader is dirty 36 Try to restart the unit Contact Toul Meditech AB or your distributor Check the service interval and if necessary carry out service according to instructions in this manual Replace the fan Contact Toul Meditech AB or your distributor Check the connectors at the controller module and monitor module plate Contact Toul Meditech AB or your distributor Check that the cables at the power transformer are connected Check the cable at the control panel Contact Toul Meditech AB or your distributor Check the service interval and perform service if necessary Check the connectors on the controller module Replace the entire control panel Check that the sterile shield has the bar code towards the reader Check that the bar code reader gives out a red light when the Sterile button is pressed Clean the bar code reader with isopropyl alcohol on a cotton applicator TOUL 400 200 10 EMC declaration of TOUL 400 200 Guidance and manufacturer s declaration electromagnetic emissions The Toul 400 200 system is intended for use in the electromagnetic environment specified below The customer or the user of the Toul 400 200 system should assure
2. Hyper terminal if the unit has been used for a long time If you want to save log data you can copy the text and paste it into a word document Change date Enter time and date in the format and time yymmddhhmmss and press s for setting the system clock with the new time and date Enter all characters in a sequence without any blanks For instance 14January 2013 3 pm is 130114150000 t Ye on EES ooon time hours speed speed Factory setting Don not use this setting sets the fan speed to speed low Changing the fan speed alters the performance of the unit and the air flow capacity Do not change the fan speed to low Normal fan speed should be set to Medium n m Exiting from service mode The unit cannot be operated as normal while you are in Service mode 1 Exit the service mode by pressing the space bar or the enter key Service manual TOUL 200 400 17 TOUL 200 400 5 Contaminated unit If the unit has been contaminated during a procedure or if spilled fluid has entered the unit you must take special precautions Always wear rubber gloves when such cleaning should be performed The Operator s manual contains information on how the unit should be decontaminated For internal cleaning follow the cleaning instructions below 10 11 12 Service manual TOUL 200 400 Turn off the unit disconnect the power Remove the single use sterile shield and dispose of it according to your local routines
3. Service manual TOUL 200 400 3 3 TOUL 400 200 NOTE As for the lower holder make sure that the front of the sterile shield holder comes into contact with the elevation on the inside on the hood The red arrow points out the small elevation where the front end of the sterile shield holder needs to come into close contact with NOTE This view shows the small elevation and sterile shield holder front from underneath of the inside the hood 14 Sweep your fingers across the holder with pressure making sure the holder sticks 15 Check that the sterile shield holders are mounted correctly by inserting an old sterile shield cardboard strip in the holders Pull the paper strip in both directions through the plastic holders It shall run smoothly but with some resistance without getting stuck If the resistance is too big insert your fingers between the plastic sterile holder and bend carefully to widen the gap 16 Check the lower sterile shield holder in the same manner as the upper 17 Put back the HEPA filter Service manual TOUL 200 400 3 A Trouble Shooting List TOUL 400 200 The table below lists possible problems and gives general advice on what actions to take when problems are met Always contact Toul Meditech AB for advice if you come across problems that cannot be solved Problem The unit does not start The fan does not start Service manual TOUL 200 400 Possible Cause The pow
4. for managing infectious refuse Remove the filter according to previous instructions Moisten a soft cloth with a cleaning solution of low level disinfectant such as isopropyl alcohol or similar and wipe all accessible inner and outer surfaces gently Clean the inside of the fan box with alcohol based detergent The frame where the filter gasket 1s fitted must be thoroughly cleaned with alcohol Remove any debris or dirt If liquids have entered the unit leave the unit to dry for at least 24 hours It should be disassembled and turned off and be placed in room temperature Inspect all connectors inlets and openings and make sure that no stains soil or dust is present Inspect the filter thoroughly before you put it back If the filter has been soiled spilled on it should be replaced with a new one Assemble the unit according to earlier instructions and make sure put everything back correctly Attach a new sterile shield Plug the unit in and turn it on Leave the unit running for at least 30 minutes Turn the unit off and remove the sterile shield and dispose of it according to your local routines for managing infectious refuse 18 TOUL 400 200 6 Parts overview TOUL 200 mount HEPA T200 400 01 HEPA filter absolute 300178 1 032 300178 3 0018 Lower sterile shield holder 300178 3 0012 Air inlet grid Service manual TOUL 200 400 19 TOUL 400 200 Cable tie foot n 40 290 88 Sing
5. the fuses need to be replaced follow these steps 1 Disconnect power and open the lid of the compartment by unlocking the tab 2 Replace the fuses with 3 15 A for 230 V or 5 A for 115 V 3 Close the compartment lid Fuse compartment TOUL 400 and TOUL 200 TOUL 200 has no power inlet Removing the fan The fan may need service if any of the following conditions apply e The fan will not start e The fan seems to be stuck or does not run smoothly e The fan has a strange noise Follow these steps to remove the fan 1 Remove the filter according to previous instructions on Replacing HEPA filter 2 Remove the screws holding the fan on the stainless steel plate 3 Disconnect the fan connector and lift out the fan 4 Perform necessary service on the fan or replace it if it cannot be repaired 5 Connect and mount back the fan and filter in reverse order Service manual TOUL 200 400 27 TOUL 400 200 9 Possible problems All TOUL units have built in self diagnostics for detection of faulty conditions Detected problems need of service or an internal equipment failure are presented on the control panel Other specific problems that you may encounter are mentioned here Bar code reader The bar code reader is situated at the bottom of the air hood and is directed towards the sterile shield bar code The unit uses an algorithm to keep track on which sterile shield that is being used If the same sterile shield i
6. the position will be fixed 6 Remove the seal from the adhesive tape from the new lower sterile shield holder TOUL 400 200 Service manual TOUL 200 400 52 7 Now place the sterile shield holder with the adhesive tape side pointing down and place the front end precisely behind the small elevation as indicated by the arrow and as you did in step 5 Pull forward towards the elevation NOTE Make sure that the front of the adhesive is in contact with the elevation on the hood See arrow on the picture 8 Push down with pressure until it sticks 9 Sweep your fingers across the holder with pressure making sure the holder sticks TOUL 400 200 10 Follow this step to assure where to place the upper sterile shield holder Without removing the seal of the adhesive tape try out the position of the upper sterile shield holder Move it forwards until it comes into contact with the small elevation in the front of the hood Keep the back of the sterile shield holder tilted so it doesn t come into contact with the inner hood surface ee 11 Once in position and in contact with the small elevation flip the rear end upwards so it comes into contact with the inner hood surface 12 Remove the seal of the adhesive tape from the upper sterile shield holder 13 Repeat the previous steps but with the seal removed and adhesive tape pointing up Push it towards the small elevation in the hood
7. UL 400 200 System as recommended below according to the maximum output power of the communications equipment Rated Separation distance according to frequency of transmitter maximum m output Auer of 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz transmitter For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 40 TOUL 400 200 11 Manufacturer How to get in contact with Toul Meditech AB Toul Meditech AB Tunbytorpsgatan 31 721 37 V ster s Phone int 46 21 13 50 00 Fax int 46 21 13 86 45 Internet www toulmeditech com E mail support toulmeditech com Your distributor TOUL 400 200 TOUL 400 200 TOUL 400 200 toul medite c h 43488 omen pa Intertek Toul Meditech AB Tunbytorpsgatan 31 SE 721 37 V ster s Sweden
8. acecccasencvcsesacvuceuuavsseusecvecsusonuseaeseeesceaseussveusecveceasonvenseeesvaces 8 V Carly Mamtenant tssecrssnsonsrsraenes vs tacescsuecscessevaseuseease seh EEEE Ara EAA Ea OEE ESER 9 3 Replacing HEPA Bref simsereteineeholssdandberseaneverdandebnnadhondaodans a T Il Ah SCT VICO MOU Oriini a A E A E A T A A aaehshasd senaten 13 USB CONnECIOL en nennen ree T T A T A R 13 Installation Of driver assnret ea 006000 a aea Eae E oE aC Esai 13 setuns up TNE CONECO 23530553 000045 500050 arn E A EA E cee abeunadunnbenkies 13 Activating the service mode sssssssosoossssssssseocors vso0o000000000s000000000000000000000000000000000000000000000000000000 16 Resetting the counti onan sa S 16 Sery Ce COMMANDS scccsedi sicsivciecescasscsssedeiucscaidasdechesinceduecoeiidedsbecsisidaasechastncessusibbededscoesisissetacesssececsuns 17 Exitine front Service Modessa e E ee Nasse ene se bss nan eeen 17 3 Contamm ted nbs aneh aes roksan aoe obd anda 18 0 F rts OVErVIeWestsesdrrewsdrnirendendenerearninendondebesdrndeiheendhnarsatninendendetensdrndenbeendhearsnininendandebedsndeeln 19 TOUL 200 cereias aaae cc sccaes ca dan von ass Ken na E ENE ESEE E AEEA A OEA REEE EERST 19 POUR A senin sa ceca ede dasa Daca pa ee otis e cease ata Sa lana NRA SR 21 7 ENCCITICGL CONNCCHIONS sista niin etende aandeed 23 8 Circuit boards and ACCESSING parts sececcccccccccccccccccccsssssssssssscsccccccccccccccccccsssssssssssssssees 25 COTO COOLS wavncescacanscenscus
9. aintenance Ensuring Safety To ensure safety and to keep your equipment in good working order it is necessary to carry out recommended maintenance and inspections A qualified service technician should carry out technical maintenance and inspection Contact your local distributor for information on whom you can consult for the yearly service of your unit or contact Toul Meditech AB directly if you do not have access to a qualified service technician in your hospital Address and phone numbers are listed at the end of this manual Service indicators When the unit needs to be serviced a warning lamp will show up on the control panel as follows Symbol Lamp Meaning 1 Scheduled service will be needed within 50 operational hours Start planning for service The HEPA filter shall be replaced 1 Service is required immediately A technical problem has occurred or the scheduled service is due A Do not modify Toul 400 200 without authorization of the manufacturer Warning HEPA Filter The filter needs to be replaced if any of the following conditions applies Yearly Maintenance procedure is due The filter has been contaminated The service symbol on the control panel is lit The efficiency of the filter is suspected to be low The fan has a strange noise Service manual TOUL 200 400 8 TOUL 200 400 Yearly Maintenance Once a year or more frequently if necessary you should replace the filter and perform an inspe
10. ar code reader S 9 Monitor module S Service manual TOUL 200 400 2 3 TOUL 400 200 BTN board relay control buttons PWR board relay module Power supply 12 V Fan Raise lowering control unit 24 V supply TOUL 400 only d4 Fa o O Service manual TOUL 200 400 9 4 TOUL 400 200 8 Circuit boards and accessing parts This section describes the circuit boards and specific parts and how to access them in a safe and proper way aa The circuit boards are situated behind the control panel Remove the 10 screws around the control panel and gently fold down the control panel Disconnect all connectors and remove the circuit boards completely ShenZhen Ghaik Tech co Ltd Httpillwvw ghaik com Model No GHK 060P NO000 Material No O AOPG06526V8 0 G E Ne I GHK 20121031 OPNO126 pen m T ym 4 2 Eas aaa 1 EE dnek Eee gt Service manual TOUL 200 400 2 5 TOUL 400 200 Connectors Power supply LCD display LCD display signal Signal from camera distance sensor and airflow External USB Video slave monitor Power supply to PCB Height adjustment TOUL 400 Power supply fan 230 V for fan and up down Toul 400 only Earth cable Power transformer The power transformer is situated under the circuit boards control panel Service manual TOUL 200 400 26 TOUL 400 200 Replacing fuses The 2 fuses are placed in the compartment just above the power switch If
11. ce Outlook X r 2 Microsoft Office Word 2007 2 Min musik i as Den har datorn Adobe Reader 9 Mina natverksplatser Microsoft Office Excel 2007 SSS gt B rn HyperTerminal Click on HyperTerminal Mobile Connect Q QuickTime Player Alla program 19 0 10 54 Arkiv Redigera Visa DS Rina upp verf r Select 5 small letters Anslutnings Ny anslutning Skriv ett namn och v lj en ikon f r ansl ne Namn Toul Toul AKELE Click OK Automatisk identil Rula CAPS NuUM Infangst Utskriftseko Identifiera ritningar 1612 Start drivrutin kom Ey a jing Hyper Toul HyperTerminal Arkiv Redigera Visa Ring upp verf r Hj lp Select the highest port Anslut till Bou Ange information om telefonnumret du vill inga Land Sverige 46 Riktnummer 021 Telefonnummer Anslut med Agere Systems HDA Modem Agere Systems HDA Modegs COMS TCP IP Winsock Automatisk identil Rula gt ritningar 1612 Start g 4 drivrutin kom Toul HyperTerminal Service manual TOUL 200 400 14 TOUL 200 400 Click OK Frnkopplad Identifiera Automatisk identil Rulla APS M Inf ngst Utskriftsek value 115200 value 8 value None value 1 value None Click OK Wait
12. ction to make sure all functions and control indicators work satisfactorily It is also necessary to reset the counter for operational hours and to carry out a leakage test in order ensure accurate performance Also carry out this procedure if the service indicator is lit indicating need for service Follow these steps for yearly maintenance l 2 Turn off the unit and disconnect the power Replace the filter as described under Replacing HEPA filter see next section Clean the cover and inner fan box with alcohol based detergent Put everything back correctly according to the instructions under Replacing HEPA filter Connect the power Start the unit and check that all buttons and controls function properly Follow the instructions given under Functional test in the General section in the Operators manual Perform a safety and leakage test using a safety tester Set up the program according to Class I and B Record the measurements on a protocol and save it so comparisons can be made with future tests 1 USB connector RCA video connector Do not connect earth or the safety tester to the RCA video connector ZN since this could damage the unit Caution Service manual TOUL 200 400 7 TOUL 200 400 With the fan turned on use a flow meter and measure the airflow in the direction of the airflow up to 1 2 meter from the sterile shield The air that comes out from the sterile shield sho
13. d or a table that can be lowered and raised since this will facilitate the mounting A Ceiling plate and plastic cover B Flange tube 7 C Locking button D D Suspension adjustment screw E 4 x MC6S M8x25 bolts 7 Pull out the locking button C and simultaneously push down the pendant arm by force This step can be DANGEROUS and it requires TWO PEOPLE Shift the arm towards the TOUL 200 unit and connect the pendant cable and plug into the corresponding connector on the TOUL 200 unit 8 Arrange the surplus cable length inside the arm and make sure it will not be squeezed 9 Push down the final bit on the pendant arm so it connects in its tube seating on the TOUL 200 unit Fixate the connection by tightening the 4 x M8 bolts E Service manual TOUL 200 400 6 TOUL 200 400 10 The suspension of the arm can be adjusted with the suspension adjustment screw using a hex key D Try to find the proper balance and adjust the screw so that the arm is correctly balanced The arm shall stay in position holding up the sterile shield and at the same time being able to move freely up and down Also refer to the OndaSpace instruction 11 Start the TOUL 200 unit Refer to the TOUL 200 Operator s manual A N Adjusting the suspension might be necessary to repeat once in a while The suspension shall be checked on the yearly procedure Note Service manual TOUL 200 400 7 TOUL 200 400 2 Recommended M
14. er cable has been disconnected Rear cables disconnected broken Power transformer not connected working A fuse has blown None of the above The sterile shield has not been replaced The fan has been turned off longer than 30 minutes The unit needs service Power is disconnected to the fan 35 Remedy Check that the cable is plugged in at the wall outlet and the unit properly Check the rear cables Check the power transformer and the connections located underneath the control panel The fuses are located near the power switch For 230 V replace with fuse 3 15 A slow 5 x 20 mm For 115 V replace with fuse SA Contact Toul Meditech AB or your distributor Make sure to use a new sterile shield for every procedure If the 30 minute s interval is exceeded a new sterile shield must be mounted Check if the unit is due for service after 2000 hours Check that the fan gets power Remove the filter and fan Disconnect the fan cable and measure the power between the grey and blue line It shall be 230 or 115 V If the fan seems to be faulty replace it for a new one The fan has strange noise No picture from the camera Cannot raise lower the hood TOUL 400 The red service indicator is lit No response from control panel Cannot read recognise the sterile shield Service manual TOUL 200 400 TOUL 400 200 Other The yearly service interval is due Loose connection
15. erivadinsscsevsarsdiacavsdensdunsdcusd nssadsusenvadwdatadensduseddend suasenudendexwsarudessdanniientievdeseteest 26 Power tFANSEOEMEE osan E SER GU SodR Ros hostnet E 26 Replacme FASES srsanneretnnseere hansen enten 0 RAD RR RAA be 27 Removine the BA ieeccascscvccecceruscecsereawusuavaceaeoesveceasesuaccwnesyaneuavaaeavoeevecdocesvacceeouacueuesussavocvuecdssenvacsoeens 27 9 Fossible proble S creci ense ren se teten sars iran bei E E E NAR 28 Bar code reader asunnon E 28 Pally OF TOUL AO snusa a a desen innemen dende INR BN tested innemende 29 Sterleshield holders orinn E e EN 30 Service manual TOUL 200 400 3 Trouble Shooting Listessi sernai iE EE E EEEE E Eaa EEEE ETEEN 10 EMC declaration of TOUL 400 200 ennn vvooevenennnnnnnsoorvvnssnnnnnnnnssrveven Service manual TOUL 200 400 4 TOUL 200 400 1 Installation instruction for TOUL 200 The following installation instructions regarding mounting the unit and pendant arm into the ceiling only apply to TOUL 200 OndaSpace Pendant The pendant that supports TOUL 200 is manufactured by Ondal GmbH Upon delivery TOUL 200 is accompanied by a separate instruction for the OndaSpace pendant Refer to this instruction before you start mounting the pendant and select the appropriate mounting option that corresponds to the requirements and the construction of the ceiling Electrical installation Electrical installation is required and the ceiling space where the pendant arm shall be mounted
16. for 5 seconds for the text RTC DateAndTime Get 07 06 29 13 22 35 to be displayed on the screen Date and time on the TOUL unit shall be the same as on your PC If this is not the case you can set date time from the service mode see Service commands The PC is now connected to the TOUL unit and will receive data from the unit continuously To activate the service mode see below If the driver doesn t work properly it may be a compatibility problem with Note your PC or operating system The driver should work properly with XP and Windows 7 Service manual TOUL 200 400 15 TOUL 200 400 Activating the service mode Start the service mode by pressing the two middle buttons and keep them pressed down for at least 3 seconds tiel TOUL 400 above the same blank buttons are used on TOUL 200 After 3 seconds the two buttons must be released at the same time If the buttons are released before that time the system will go into standby mode Note The Toul unit cannot be started when you are in the service mode The following text will appear in the Hyper terminal FOR RTC DateAndTime Get 11 02 04 11 43 03 RTC DatefindTime Get 11 02 04 11 43 08 ADMIN MODE SELECTED STATE IDLE STATE MODE PRIORITY EXIT STATE SERVICE STATE MODE ENTER Press n turn off bar code reader beep Press r to read logg Type yymmddhhmmss and press s to set time t to get run time Shift 2 to
17. le pole safety connector SEB 2610 F4 8 300178 3 0115 Overlay Toul 200 Service manual TOUL 200 400 20 TOUL 400 200 TOUL 400 300178 3 0025 Hood T400 TM ECM TM ECM 1 3 300178 3 0008 4 300178 3 0009 Fan distance ring Service manual TOUL 200 400 21 TOUL 400 200 300178 3 0014 6 i D2E146HS9721 300178 3 0019 Upper sterile shield holder HEPA T200 400 01 HEPA filter absolute 9 2 300178 3 0012 Air inlet grid 10 300178 1 011 300178 3 0018 Lower sterile shield holder 300178 3 0015 Camera front Glass 300178 3 0016 EA1050F 120 AC DC converter 300178 3 0018 Midpart distance T400 300178 3 0112 Control panel T200 T400 300144 3 0018 Camera IR smst 300178 3 0023 Midpart topdisk AA 01066 MD Rev G p AA 01066 MD Rev Gh Trafo holder 932 325 100 CA 6 GD socket KD14 1 101 151 IEC Appliance Connector Cable tie foot Cable tie foot 12 T11539 Telescopic frame Nut protector M8 Nut protector M8 2 2 x 00770382 Wheels without breaks 1 58 000 013 30 MECANO z 29 Leveller pole Leveller pole 30 1 300178 3 0039 5 5 Top backer 300178 3 0038 Top bracket 300178 3 0040 Bearing plate T400 37 I 40 290 88 Single pole safety connector SEB 2610 F4 8 300178 3 0078 3 Scanner mounting plate 300178 3 0077 2 Scanner adjustment plate 300178 3 0079 Cover glass for barcode reader Service manual TOUL 200 400 22 TOUL 400 200 7 Electrical connections 6 Camera S 8 B
18. must be prepared with electrical wiring and a socket The electrical installation shall be connected to a single fuse and a main power switch needs to be mounted on the wall All electrical installation steps must be carried out by a certified electrician Consult the maintenance department at the Hospital or your workplace since they most likely have qualified staff or can give advice who to contact A N When mounting the OndaSpace pendant it is necessary to be at least two persons The device is heavy and it requires extra hands to hold some Warning items while mounting the pendant Mounting TOUL 200 For mounting the TOUL 200 and the pendant arm follow the steps below and see the illustration on the next page 1 Measure all parts and adjust the lengths of the fixating items It might be necessary with some slight adjustments 2 Start by mounting the ceiling set A into the ceiling Connect the earth wire to the earth pin on the ceiling plate to be connected by a certified electrician 3 Mount the flange tube B on the ceiling plate The flange tube B can be sawed into the correct length Service manual TOUL 200 400 5 TOUL 200 400 4 Mount the pendant arm on the flange tube and place the plastic cover over the ceiling plate A 5 Pull the cable through the pendant arm and leave some extra cable length in the upper connection 6 Place the TOUL 200 unit on a table or a support stand If possible use a Stan
19. o line s 2 kV line s to earth lt 5 Ut gt 95 dip in Ur for 0 5 cycle 40 UT 60 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles lt 5 Ut gt 95 dip in Ur for 5 sec 3 A m 6 kV contact 8 kV air 2 kV for power supply lines 1 kV line s to line s 2 kV line s to earth lt 5 Ut gt 95 dip in Ur for 0 5 cycle 40 Ut 60 dip in Ur for 5 cycles 70 Ur 30 dip in Ur for 25 cycles lt 5 Yo Ur gt 95 dip in Ur for 5 sec 1000 m Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 The customer PC used for the system provides TOUL 400 200 Amplifier with power The PC itself is not part of this declaration The mains power quality used for the PC should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the TOUL 400 200 System requires continued operation during power mains interruptions it is recommended that the TOUL 400 200 System be powered from an uninterruptible power Supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical location in a typical comme
20. onnect the connector underneath the hood Service manual TOUL 200 400 29 TOUL 400 200 2 Turn the hood sideways from the plate 3 Unscrew the four bolts and carefully lift the hood Sterile shield holders If the sterile shield holders become damaged or bent the sterile shield will not fit tightly and air leakage may appear To replace the sterile shield holders follow these steps 1 Disconnect the power 2 Remove the HEPA filter see previous instructions on removing the HEPA filter 3 Remove the old sterile holders by pulling them off Make sure all adhesive comes off 4 Clean the upper lower surfaces thoroughly with alcohol based detergent NOTE Make sure to clean thoroughly in order for the adhesives on the new sterile shield plastic holders to stick properly to the surface If the surfaces aren t clean the holders risk becoming loose Also make sure the surfaces are dry before mounting the holders Service manual TOUL 200 400 30 TOUL 400 200 Service manual TOUL 200 400 31 5 Do this step to assure where to place the lower sterile shield holder Without removing the seal of adhesive tape try out the position of the lower sterile shield holder by placing the front end where the adhesive sits precisely behind the small elevation as indicated by the arrow Pull forward towards the elevation It should fit perfectly Once you remove the seal the adhesive will stick and
21. quency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the EQUIPMENT or SYSTEM is used exceeds the applicable RF compliance level above the EQUIPMENT or SYSTEM should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the TOUL 400 200 System P Over the frequency range 150 kHz to 80 MHz field strengths should be less than 10 V m 39 TOUL 400 200 Recommended separation distances between portable and mobile RF communications equipment and the TOUL 400 200 System The TOUL 400 200 System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the TOUL 400 200 System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the TO
22. rcial or hospital environment NOTE Uris the a c mains voltage prior to application of the test level 38 TOUL 400 200 The TOUL 400 200 System is intended for use in the electromagnetic environment specified Guidance and manufacturer s declaration electromagnetic immunity below The customer or the user of the TOUL 400 200 System should assure that it is used in such an environment Immunity test IEC 60601 test Complianc Electromagnetic environment level e level guidance Portable and mobile RF communications equipment should be used no closer to any part of the TOUL 400 200 System including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 10 Vrms de 0 4 P IEC 61000 4 6 150 kHz to 80 MHz d 0 3VP 80 MHz to 800 MHz Radiated RF 3 V m IEC 61000 4 3 80 MHz to 2 5 GHz d 0 5VP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufac turer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol o NOTE 1 At 80 MHz and 800 MHz the higher fre
23. s mounted twice the unit will not accept it and the fan cannot be started Possible problems that may arise with the sterile shield is either that an old sterile shield is used or that the time limit has expired for how long the fan or the unit can be turned off Keep in mind the following time limits e The unit can be in standby mode for as long as needed with a new sterile shield registered before the fan is started e The fan can be turned off for 30 minutes with the same sterile shield mounted e The unit can be powered off for 15 minutes with the same sterile shield mounted If there are problems with recognising the bar code the glass may be dirty or scratched The bar code reader glass can be cleaned with isopropyl alcohol on a cotton applicator if it 1s dirty If the glass is scratched it needs to be replaced unscrew the protective glass Remove the glass and mount a new glass Make sure that the glass fit tight against the hood otherwise unfiltered air may be distributed Service manual TOUL 200 400 28 TOUL 400 200 E Bar code reader glass 4 gt Connector on pillar of TOUL 400 Pillar of TOUL 400 The TOUL 400 unit has the pillar steering cable underneath the control panel This cable must not be bent or destroyed in any way Check that this cable has free passage and is a good condition if you come across problems lowering or raising the hood To separate the pillar from the hood 1 Disc
24. set Model to 200 Shift 3 to set Model to 300 Shift 4 to set Model to 400 h to set Fan speed high m to set Fan speed medium to set Fan speed low Resetting the counter The unit has a built in counter in the controller module for keeping track of operational hours When the pre set value of 1950 hours is reached the Service indicator will become yellow indicating need for service After 50 more hours the indicator will alert in red The counter needs to be reset manually after the filter has been replaced 1 Press fin the Service mode to clear and reset the operational counter to Zero Service manual TOUL 200 400 16 TOUL 200 400 Service Commands A Warning There are several service commands that can be used for controlling the TOUL unit They are accessed from the keyboard on the PC when it is connected to the unit The commands are listed in the Hyper terminal Below is a detailed explanation for each command i Beep on Press b to activate a beep sound each time the bar code reader reads a new Sterile shield f Reset Press f to clear and reset the operational operational counter to zero This should be done each time counter the HEPA filter is replaced Beep off Press n to turn off the beep sound each time the bar code reader reads a new Sterile shield r Log file Press r for showing the log file of the unit When this function is executed a large amount of data will show up in the
25. that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance The TOUL 400 200 uses RF energy only for its RF emissions internal function Therefore its RF emissions are CISPR 11 very low and are not likely to cause any interference in nearby electronic equipment RF emissions The TOUL 400 200 System is suitable for use in all Class B establishments other than domestic and those directly connected to the public low voltage power Supply network that supplies buildings used for H ee domestic purposes armonic emissions N tsbbilesble IEC 61000 3 2 ET Voltage fluctuations flicker emissions Not applicable IEC 61000 3 3 CISPR 11 37 TOUL 400 200 Guidance and manufacturer s declaration electromagnetic immunity The TOUL 400 200 System is intended for use in the electromagnetic environment specified below The customer or the user of the TOUL 400 200 System should assure that it is used in such an environment IEC 60601 Immunity test test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 Electrical fast transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV line s t
26. toul meditcech E Service Manual TOUL 400 TOUL 200 Toul is registered by TOUL Meditech AB TOUL 200 and TOUL 400 bears the C marking and comply with the provisions of the Medical Device Directive 93 42 EEC Toul Meditech AB takes no responsibility for use of the product other than for those uses specified herein and disclaims any liability resulting from other uses Observe all warnings and cautions This manual covers software version 1 06 Toul 400 200 Type A2 Manual revision 3 01 February 2013 Toul Meditech AB Copyright Toul Meditech AB 2012 All rights reserved Unauthorized use copying or distribution of this manual is prohibited without the written permission from Toul Meditech AB Content POULE ZOO rr RRE SINA nadoen EE ae bakske dens edna sans anderen RARE NINI UU ESAIAS SITER RARE 5 I Installation instruction for TOUL 200 snvnisrstererersertennbererbinharesertenrweeederkrdern kin hoassdenes 5 OndaS pace Pendant vessesissctssiestslcisiisrnudsrslseiseesieseiisnuculininidcuidsiandlsniinideede 5 Electrical mstalla amp f msstsrosrrterveerstneddaerdenkendkusodoersnaweerdn ARN RAR SERA RAR KARA FN RASAR 5 Mountne FOUL 200 smste reeet ARS NN NRA 5 POULE 400 200 ossen san ates NR RR SOT de Saeed 8 2 Recommended Maintenant assesoires aAA a aE aA ATENA REEE aR 8 Ensurme Safety sneren eenen sek deter EEE eae 8 SEE VICE TINCICACORS res deden eden denon beleren hinde este 8 HEPA FIOT someone venne reet ceseai
27. uld have a constant speed of 0 4 0 5 meter per second If the air speed is not within this range contact Toul Meditech for assistance On TOUL 400 check that the lock levers on the castors function properly by locking and releasing them When in the released position check that all the castors can rotate freely If a castor is defective it has to be replaced Also visually check that there are no cracks dents or other signs of damage on any of the castors On TOUL 200 check that the pendant arm is correctly balanced can move freely and stays in position If the suspension of the arm needs to be adjusted refer to installation instruction of TOUL 200 Be sure to put back all parts in the correct order after finishing service Warning and inspection The filter may not be fitted incorrectly Check especially that there are no remaining items outside or inside the unit Service manual TOUL 200 400 10 TOUL 200 400 3 Replacing HEPA Filter 1 Turn off the unit and disconnect the power 2 Remove the sterile shield and dispose of it Uo Untighten the bolts on the small filter brackets that fixate the HEPA filter 4 Do not remove the brackets Let them hang loose pointing down as shown in the picture Service manual TOUL 200 400 1 1 TOUL 200 400 Remove the HEPA filter carefully by pulling out both sides simultaneously Don t use any tools to remove the filter Wipe the inside of the hood
28. with alcohol Mount a new filter with the rubber gasket facing towards hood Dispose of the old filter according to your local waste management regulations Put back the brackets that holds the filter and tighten the bolts 9 The operational counter has to be reset after replacing the filter Refer to Resetting the counter in the Service Mode section Service manual TOUL 200 400 12 TOUL 200 400 4 Service Mode USB connector To start the service mode you first need to connect a PC at the USB port 1 USB connector connector Em RCA video connector 1 Connect the USB cable from the PC into the USB socket TOUL 400 and TOUL 200 have converters for USB to RS 232 which emulates a COM port For installation of a suitable driver see Install driver in the next section The RCA video connector can be used for Video to connect a slave monitor Installation of driver To communicate with the unit you need to have the correct driver installed The driver is supplied separately on a USB stick or it can be emailed upon request if you contact Toul Meditech Setting up the connection 1 Start the TOUL unit 2 Start Windows 3 Connect a USB cable between the PC and the TOUL unit see picture above Service manual TOUL 200 400 1 3 TOUL 200 400 Click on START Lars Johansson e Internet D Mina dokument Internet Explorer Senast anv nda gt E post dokument Microsoft Offi

Download Pdf Manuals

image

Related Search

Related Contents

スマートアグリシステム  König HC-PS101N personal scale  Philips Softone Energy saving bulb 8718291682738  Contrat de Recherche Associé  Gérer son équilibre nutritionnel après les fêtes de fin d`année 10.01  Mode d`emploi METTLER TOLEDO MultiRange Terminal de pesée    Spire CXP8000 Color Server für das Xerox DocuColor 8000 Digital  AMPL CPLEX 9.0 User Guide  Battery and Electrical Systems - Les Entreprises Marc Latulippe  