kENDALL GENIUS™ 2 - Tiger Medical, Inc
Contents
1. ooooooooooooo eee nes 100 240V 0 8 0 4A 47 63Hz Fuse Requirements 0 ccc ccc cece te ene eens Schurter FSF 5 x 20mm Fast Blow Leg iheed O BON ee eeu Tears aa eee ee Low Breaking Capacity 250 V 3 15A Recommended Checker Calibrator Calibration Interval One Year ESD Immunity gah k tenia dd arribar aca ibid Meets IEC 61000 4 2 Non continuous Criteria C Pollution AA Installation Category Mowe Kd Soke ORM SNES TRE KEK SHEE SEEK CURES NESSES GAO 88 Post recalibration accuracy using the GENIUS 2 Checker Calibrator may not necessarily be equivalent to factory calibration Section IX Cleaning Instructions CAUTION DO NOT IMMERSE THE Checker Calibrator CABLES OR POWER CORD IN WATER OR OTHER CLEANING SOLUTION CLEAN USING A DAMP NOT WET CLOTH FAILURE TO FOLLOW CLEANING PROCEDURES DESCRIBED HEREIN COULD RESULT IN HAZARDS TO USERS As with any AC powered electrical device care must be taken to prevent liquid from entering the Checker Calibrator to avoid electrical shock hazard fire hazard or damage to electrical components GENIUS 2 Checker Calibrator 9 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate General Cleaning Instructions Cleaning of the GENIUS 2 Checker Calibrator may be performed as follows FIRST UNPLUG THE GENIUS 2 Checker Calibrator FROM POWER SOURCE NEVER CLEAN THE
2. DEVICE WHILE CONNECTED TO THE MAINS SUPPLY e A mild detergent should be used for general cleaning of outside surfaces Also a cloth damped with isopropyl alcohol can be used to clean the unit DO NOT USE abrasives or solvents Avoid excess moisture around the target assemblies and the electrical connectors e Do not attempt to clean the target below the rubber flaps e Do not clean the target nests or flaps with alcohol If there is any doubt about the compatibility of a cleaning agent with part of this equipment or its materials contact Tyco Healthcare Kendall customer service See Section XI Section X Troubleshooting Some of the common problems are as follows GENIUS 2 Checker Calibrator will GENIUS 2 Checker Calibrator not receiving AC power Check AC power connections not power up Test report does not write to USB USB flash drive not formatted Use PC or laptop to format flash drive flash drive USB flash drive write protected Use PC or laptop to turn off write protection USB flash drive does not contain enough free memory space Use PC or laptop to delete unnecessary items or use different USB flash drive USB flash drive incompatible with GENIUS 2 Checker Calibrator Use different USB flash drive GENIUS 2 Checker Calibrator has not had enough time to detect Retry write process the newly inserted USB flash drive Multiple GENIUS 2 thermometers Probe cover stuck in one of the blackbody targets Remove probe cover from blackbo
3. a report is desired insert a flash drive into the front panel USB port Note allow about 30 seconds after insertion for the USB flash drive to initialize Use the encoder knob to scroll to either Yes or No and press the encoder knob to make a selection If No is selected the screen returns to the Main Menu If Yes is selected a report is written to the USB flash drive using the thermometer serial number for filename and TXT as the file extension The file contains the test results as well as other information such as serial number date time user name and organization This file is a simple ASCII text file and is only available in English This file can be read by MicrosoftTM Windows programs such as Microsoft Notepad and sent to a printer After the file is written the display will show Push knob to continue Press the encoder knob to return to the Main Menu Step 10 After completing a calibration check recalibration remove the probe connector from the battery compartment using a rocking motion to free the connector Then reinstall the 3 AAA batteries observing the proper polarity orientation Finally reinstall the battery compartment cover and tighten the screw Step 11 Remove USB flash drive before closing the lid of the case The lid should be closed when the GENIUS 2 Checker Calibrator is not in use Section VII Error Messages The error messages will have a number associated with them such as Error 1 etc The followin
4. blown fuse e This device should not be used in the presence of flammable anesthetics There is risk of an explosion in the presence of these anesthetics e There are no user serviceable components in this device Do not remove the front panel Refer servicing to qualified service personnel e Use only the supplied power module with this device e When prompted to insert the GENIUS 2 Thermometer into one of the blackbody targets there is a 60 second window of time to accomplish the insertion If the time delay exceeds this period the procedure will be terminated e Prior to checking calibrating a GENIUS 2 Thermometer ensure that the thermometer lens is clean Effective cleaning will ensure that the calibration tolerances are correctly verified Refer to the cleaning instructions in the Genius 2 Thermometer operating manual e When installing a GENIUS 2 probe cover on to the tip of the GENIUS 2 Thermometer use only covers that are in a cassette and not released Try to avoid all hand contact with the probe covers in order to prevent inaccurate readings Do not reuse probe covers Use only GENIUS 2 probe covers Let the GENIUS 2 Checker Calibrator warm up after power up for at least 15 minutes before use Make sure that the GENIUS 2 Thermometers probe covers and the GENIUS 2 Checker Calibrator have had enough time to equilibrate to room temperature before proceeding Allow a longer warm up period if the device or thermometers were at the extreme
5. special grades of thermoplastic materials are also tightly controlled and are specifically chosen to deliver consistent temperature measurements with GENIUS 2 Thermometers The factory calibration and final determination of an acceptable GENIUS 2 Thermometer is dependent on the use of acceptable Tyco Healthcare Kendall GENIUS 2 Thermometer ACCUSYSTEMy probe covers The use of generic probe covers or other probe covers not produced by Tyco Healthcare Kendall is not supported or sanctioned by Tyco Healthcare Kendall The use of unauthorized GENIUS 2 Thermometer probe covers could jeopardize the accuracy of the GENIUS 2 Thermometer Check your probe cover carton for the ACCUSYSTEMy logo to be certain that your thermometer will deliver an accurate temperature every time it is utilized GENIUS 2 Checker Calibrator 3 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section IV Required Equipment NOW BW DN GENIUS 2 Checker Calibrator Power module with AC adapter Probe cable Operators Manual GENIUS 2 Thermometers to be tested GENIUS 2 disposable probe covers Small Phillips screwdriver Section V System Setup T ONO BW NY he following actions should precede use of the GENIUS 2 Checker Calibrator Connect the correct AC plug adapter for your country to the power module Connect the power module cable to the power jac
6. 2 Checker Calibrator to be free of defects in materials and workmanship under normal use and service for a period of 1 year from the date of delivery by Tyco Healthcare Kendall to the first purchaser Liability hereunder is limited to repair or replacement at Tyco Healthcare Kendall s option at Tyco Healthcare Kendall service facility for any product which shall under normal use and service appear to Tyco Healthcare Kendall to have been defective in material or workmanship This warranty shall not apply to and Tyco Healthcare Kendall shall not be responsible for any loss arising in connection with the purchase of any product that has been repaired by anyone other than Tyco Healthcare Kendall or its authorized representative or which has been subject to misuse neglect or accident or which has been used otherwise than in accordance with the instructions furnished by Tyco Healthcare Kendall Tyco Healthcare Kendall neither assumes nor authorizes any representative or other person to assume for it any liability other than as expressly set forth herein Tyco Healthcare Kendall DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION OTHER THAN AS EXPRESSLY SET FORTH IN THE PRODUCT LABELING IN NO EVENT WILL Tyco Healthcare Kendall BE LIABLE FOR ANY INCIDENTAL INDIRECT OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE PURCHASE OR USE OF THIS PRODUCT GENIU
7. 888 e www tycohealthcare com ECJREP TYCO HEALTHCARE U K LTD GOSPORT P013 OAS U K 182563 01 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Table of Contents Engish Section l Functional DO iaa 1 Section II Precautions and Warnings cccssssssssscssssssssecssssscsssccsssccssuccsssecssusecessccssnscsssscssnsccsssecsnsccsssecsssecessesssscessuessseceesees 2 Section Il Icon den ON coi 3 Section IV Required Equipment ccccccssssssssssssssssssssescssecssssssssnccssscsssuscssucessuscssscessnscsssscssnscssssecsnscsessecsssccssscsssscessuesssscessees 4 Section V System CU sicario 4 Section VI Operational Procedure ara id ic 5 O A TE A IEEE AE AI EE 5 A 5 A aa cacteaiinaeed Olt E eM aH aici AoE Meal 5 A O teasing eee 5 AE AAA A A eee Pate 5 EA A A EEEE 5 A a OAOE S enna de AMHR 5 SA IM anida 6 MERINO did 6 Section Vil EOS OS oi AA AAA A ARA 8 Section VIII Checker Calibrator SpecificatiONS o nonnonininodinnninmsrrerrrreererrrer errar 9 Section IX Cleaning Instructions cosari niosioci nncinn crean ti cnc cen cincel 9 Section X Tr USO MO ina da 10 Section Al CUSTO MESE airis 11 Section Xil Wa Many oca iia 11 Section XIII Electromagnetic Conformity Declaration cccsssssssssssssecssnssssssecssessssescssscceseecsssccessecsssccesuesssseessuecesseess 12 This product contains softwar
8. Calibrator High Temperature Target High Temperature Target LED LCD Display Encoder Knob USB port Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section Il Precautions and Warnings The GENIUS 2 Checker Calibrator is designed to be operated as specified in this manual Protection provided by the safety devices in the GENIUS 2 Checker Calibrator may be impaired if the device is not used in accordance with the directions contained in this manual e The GENIUS 2 Checker Calibrator enclosure is not waterproof Avoid spilling liquid of any kind onto the front panel of the device Likewise the GENIUS 2 Checker Calibrator is powered from an AC power source Avoid the potential for contact between liquid and any external surface of the device while connected to the mains source The surface quality of the interior of the reference blackbody targets is critical to their operation as efficient infrared radiators Care should be exercised when working near the entrance to the blackbody targets to ensure that no contact is made with the surface inside the rubber flaps e In the event of a blown fuse replace the fuse with a fuse that exactly matches the rated voltage current and form factor as specified in the GENIUS 2 Checker Calibrator Specifications section of this manual Make sure the device is disconnected from the AC power source before changing a
9. Electrical and Electronic Waste a High Temperature Target F C LABORATORY EQUIPMENT 46EA Low Temperature Target F C C US ULListing Information 97 3 m i Caution Indoor Use Only 2 GENIUS 2 Thermometer Connection The GENIUS 2 Infrared Tympanic Thermometer is a reliable and accurate temperature taking device The reason for the accuracy of the GENIUS 2 Thermometer is the design controlled calibration methods and stringent manufacturing controls One of the most critical functional parts of a thermometer is the ACCUSYSTEMy probe cover produced by Tyco Healthcare Kendall The GENIUS 2 Thermometer ACCUSYSTEMy probe cover when placed on a GENIUS 2 Thermometer serves as an infection control barrier between the patient and the device and the medium for heat transmission from the patient to the thermometer The functionality of the GENIUS 2 Thermometer ACCUSYSTEMY probe cover is extremely important for preventing device contamination and also for allowing accurate patient temperature measurements The reason the GENIUS 2 Infrared Tympanic Thermometer and GENIUS 2 Thermometer ACCUSYSTEMy probe covers have satisfied our customers expectations is due to the integration of these two parts during the manufacturing process The GENIUS 2 Thermometer ACCUSYSTEMy probe cover molding process parameters are tightly controlled at Tyco Healthcare Kendall to minimize variation and produce consistent and reliable products The
10. S 2 Checker Calibrator 11 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section XIII Electromagnetic Conformity Declaration The Genius 2 Checker Calibrator has been built and tested according to IEC EN61010 1 and EN61326 1 standards The Checker Calibrator is intended for use in the electromagnetic environment specified below The user of the equipment should assure that it is used in such an environment Electromagnetic Environment Guidance Conducted and Radiated Group 1 Class B The Checker Calibrator emissions are very low and are not likely Emissions EN61326 1 to cause interference in nearby electronic equipment Harmonic Current Class A EN61000 3 2 Voltage Fluctuation and Flicker Complies EN61000 3 3 Radiated Disturbance Immunity Complies EN61000 4 3 Conducted Disturbance Complies Immunity EN61000 4 6 Power Frequency Magnetic Field Complies Immunity EN61000 4 8 Voltage Dips and Interrupts Complies Immunity EN61000 4 11 Electrical Fast Transient Bursts Complies Immunity EN61000 4 4 Electrostatic Discharge Meets Criteria C EN61000 4 2 Non Continuous GENIUS 2 Checker Calibrator 12 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245
11. Status Approved Doc AW10001726 Effective Date nulldate TYCO Healthcare KENDALL Checker Calibrator Checker Calibrator Edeyxtmc Baguov unonc Operation and Service Manual Eyxetpidto Aettoupyiac kat o pBic Appareil de v rification Calibrateur Ov rovac Kalibrator Manuel d utilisation et d entretien U ivatelsk a servisn p ru ka Pr fger t Kalibrator Ellen rz kalibr l k sz l k Benutzer und Service Handbuch M k dtet si s karbantart si k zik nyv Sistema di controllo della KOHTpOnbHO 13MepuTenbHoe yctponctBo Kanu6patop Pyk0BOACTBO no 3KCMIMNyYaTau un Y TeEXHNYECKOMY o6cnyKNBaHulo calibrazione calibratore Manuale operativo e di assistenza Verificador de calibracion ap Kalibrator Manual de operaci n y servicio Instrukcja obs ugi i serwisowania Kalib reringsenhet Kontrolor Kalibrator Anvandar och servicemanual Kullanim ve Servis Kilavuzu Validator Kontroll og kalibreringsapparat Gebruiks en onderhoudshandleiding i n Bruker og servicehandbok Verificador Calibrador Manual de utiliza o e assist ncia Kalibrointilaite K ytto ja huolto opas Kontrol og kalibreringsapparat Betjenings og servicehandbog CE La Trademark of Tyco Healthcare Group LP or its affiliates Mal TYCO HEALTHCARE GROUP LP 15 HAMPSHIRE ST MANSFIELD MA 02048 MADE IN U S A 2008 TYCO HEALTHCARE GROUP LP ALL RIGHTS RESERVED 090909 PRODUCT INFO IN U S 1 800 962 9
12. brator Specifications Temperature Set Points Low larget 5554006 to oe ocr ee eee hn eee hoe Renee ena es 32 2 C 0 3 C 90 0 F 0 5 F HQ TANCE oe a s 3 ach ea aoe haha ee ae Oe CA a ROS 40 6 C 0 3 C 105 0 F 0 5 F Target Temperature Accuracy 0 0 0 0 ooo ooo Initial Limit 0 03 C 0 05 F o do tet Post year 1 0 06 C 0 10 F GENIUS 2 Thermometer Accuracy After Recalibration Readings between 36 0 C and 39 0 C 96 8 F and 102 2F 0 2 C 0 4 F Readings less than 36 0 C 96 8 F or greater than 39 0 C 102 2 F 0 3 C 0 5 F Warm up Time Before Use o o ooooooooooooooo 15 minutes from non extreme ambient Ambient Temperature Operating Range o oo oooooocoooco o 21 1 26 7 C 70 80 F Relative Humidity OperatingRange 0 0 cece ee ee eee 50 20 non condensing Altitude Range cx ohn os Creda chee tear ess Fans Aa ead Up to 2000 meters 6563 feet Storage Temperature Range 0000 25 C 55 C 13 F 131 F 85 RH non condensing Approximate Size Length oe a o II dA ASN ate wt gale ee 31 1 cm 12 25 o A NEES SEARO REA POE DE EOS OURS OSS ede TRO eA SS 15 9 cm 6 25 A ae ee Bice rok ek te ce ed dc pe ds aa Ah ora dlue Sue aes a og 26 0 cm 10 25 A e ee ee eee ee ee ee ee ee ee eee 3 3 kg 7 2 lbs Power Requirements
13. dy target using small needle nose pliers or fail recalibration hemostats Do not scratch the interior surface of the target Dirty probe cover Use a new probe cover for each reading Do not re use a probe cover Probe covers should never be touched by hand Dirty thermometer lens Clean the thermometer lens per the instructions in the GENIUS 2 operating manual GENIUS 2 thermometer not seated correctly Ensure that the GENIUS 2 thermometer is completely seated in the target and at the proper angle GENIUS 2 thermometers probe covers and or checker calibrator Ensure that the GENIUS 2 thermometers probe covers and or checker not at room temperature calibrator have had enough time to equilibrate to room temperature before proceeding Error 6 repeatedly appears while Potential power consumption issue with GENIUS 2 thermometer Repeat calibration check recalibration with a different GENIUS 2 testing the same GENIUS 2 thermometer and return problem thermometer for factory servicing thermometer GENIUS 2 thermometer seems Probe cover reused during calibration check recalibration Repeat calibration check recalibration using a new probe cover prior to to read high or low following a every target insertion during the process successful recalibration GENIUS 2 Checker Calibrator 10 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section XI Cust
14. e solely owned by Tyco Healthcare Group LP Tyco Healthcare Group LP grants the user a non exclusive limited license to use the software according to the operating instructions A copy of the license can be obtained from Tyco Healthcare Group LP 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section Functional Description This manual describes the operation of the GENIUS 2 Checker Calibrator This device has been developed to check the accuracy of the GENIUS 2 Tympanic Thermometer and automatically recalibrate the thermometer if necessary In addition the GENIUS 2 Checker Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB flash drive The GENIUS 2 Checker Calibrator will work with all GENIUS 2 Thermometer software revisions The GENIUS 2 Checker Calibrator contains two independently controlled infrared calibration targets that are similar to factory calibration targets These blackbody targets are designed to have efficient radiative heat transfer and produce infrared radiation that approximates the human ear at temperatures of 32 2 C 90 0 F and 40 6 C 105 0 F Checker Calibrator LL fe Figure 1 Front Panel Layout Legend 1 Power Connector 7 2 Power Switch 8 3 Fuse Holder 9 4 Low Temperature Target LED 10 5 Low Temperature Target 11 6 GENIUS 2 Thermometer Connector GENIUS 2 Checker
15. each target There is a prompt to press the encoder knob to continue to the next screen If either target cannot reach the correct temperature an Error message screen will be displayed If this happens check to be sure the GENIUS 2 Checker Calibrator is within the correct ambient room temperature range and has been given at least 15 minutes of warm up time since power was switched on Step 2 Inspect the GENIUS 2 Thermometer that will be tested Ensure that the thermometer lens is not scratched and is free from ear wax and or other contaminants If the thermometer lens is dirty clean the lens according to the instructions in the GENIUS 2 Thermometer operating manual Once the thermometer lens is clean connect the GENIUS 2 Thermometer to the Checker Calibrator by plugging the round end of the thermometer interface cable into the round connector socket on the GENIUS 2 Checker Calibrator front panel refer to Figure 3 Open the battery door of the thermometer by using a small Phillips screwdriver Remove the batteries and set aside Connect the other end of the thermometer probe cable into the battery compartment making sure the connector snaps into place securely Refer to Figure 2 in Section V Note A new probe cover must be used prior to every target insertion Finger oils on previously used probe covers or re used probe covers that have been heated by GENIUS 2 Checker Calibrator target can cause erroneous readings which may result
16. een will be displayed showing the thermometer serial number along with an indication that the thermometer passed calibration check If the GENIUS 2 thermometer is found to be outside the specified accuracy limits the display will indicate to the user that the GENIUS 2 Checker Calibrator is entering into a calibration sequence that requires the user to repeat Steps 3 through 6 three more times in order to gather recalibration data Following the last pass through Steps 3 through 6 the GENIUS 2 Checker Calibrator will attempt to recalibrate the GENIUS 2 thermometer If the recalibration is successful a Results screen will be displayed showing the thermometer serial number along with an indication that the thermometer passed calibration If the recalibration is unsuccessful or the thermometer is unable to be calibrated a Results screen will be displayed showing the thermometer serial number along with an indication that the thermometer failed calibration GENIUS 2 Checker Calibrator 7 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Step 8 After the pass fail Results screen has been displayed press the encoder knob to display a second Results screen containing the target temperatures the GENIUS 2 readings and the target versus thermometer variances Step 9 The next screen will display Write report to USB flash drive and display Yes and No below If
17. eter while it is inserted into the Low temperature target The GENIUS 2 Thermometer will output 3 short beeps as it takes the temperature of the blackbody target After the temperature is taken the next screen will display Step 5 Withdraw the thermometer from the Low temperature target and eject the probe cover The screen display will change to Install a new probe cover and insert probe into the High target Again use a new GENIUS 2 probe cover from the cassette and be careful not to point the probe tip at any object including hands fingers or LCD display Inspect the probe covers to make sure it is fully seated no space between the cover and the tip base and no holes tears or wrinkles are present in the plastic film When the cover is on insert the probe tip into the High temperature target which will have a blinking LED indicator next to it Make sure that the probe tip is fully inserted into the target well Step 6 The insertion of the probe tip into the High target will automatically be detected and the screen display will change to Press GENIUS 2 Thermometer scan button Press the scan button on the GENIUS 2 Thermometer while it is inserted into the High temperature target The GENIUS 2 Thermometer will output 3 short beeps as it takes the temperature of the blackbody target After the temperature is taken the next screen will display Step 7 If the GENIUS 2 thermometer is within specified accuracy limits a Results scr
18. g is an explanation of the errors Error 1 Internal Error This error is an indication of a problem within the GENIUS 2 Checker Calibrator If this error persists after cycling power off and back on return the unit for factory servicing Error 2 Unable to communicate with GENIUS 2 Thermometer This error is an indication of either a lack of communication or a miscommunication between the GENIUS 2 Thermometer and the GENIUS 2 Checker Calibrator Check both ends of the thermometer probe cable especially the end that connects inside the battery compartment The pins in the thermometer probe cable can lose spring action if they are bent or damaged After inspecting the probe cable re attempt the calibration check If the problem persists try a different GENIUS 2 Thermometer if available If this does not resolve the problem return the unit for factory servicing along with the thermometer cable and the GENIUS 2 Thermometer that indicated this error Error 3 Timeout This error is an indication that the operator took too long to perform an operation This error might be seen when connecting to the GENIUS 2 Thermometer or while installing a probe cover on the thermometer The timeout is set for one minute Try repeating the operation that gave this error Error 4 Target Out of Range This error is an indication that one of the blackbody targets is not within the range of temperature allowed Check that the unit is being operated within the allowed a
19. in a failed calibration check or recalibration Therefore it is extremely important that a new probe cover is used each time Step 3 Press the encoder knob to continue then install a new GENIUS 2 probe cover onto the GENIUS 2 Thermometer Make sure to use a cover that is held in the cassette Once the cover is on the thermometer do not point the probe tip at any object including hands fingers or LCD display as this will cause an inaccurate temperature reading Inspect the probe cover to make sure it is fully seated no space between the cover and the tip base and no holes tears or wrinkles are present in the plastic film When the cover is on insert the probe tip into the Low temperature target which will have a blinking LED indicator next to it Make sure that the probe tip is fully inserted into the target well Refer to Figure 3 for proper insertion In order to avoid a time out error insert the covered probe tip into the target within 60 seconds from the time the encoder knob is pressed GENIUS 2 Checker Calibrator 6 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Figure 3 Proper Insertion of GENIUS 2 Probe Tip into Target Step 4 The insertion of the probe tip into the target will automatically be detected and the screen display will change to Press GENIUS 2 Thermometer scan button Press the scan button on the GENIUS 2 Thermom
20. k on the front panel Connect the power module to an AC power outlet Turn the unit on via the switch on the front panel Let the GENIUS 2 Checker Calibrator unit warm up for at least 15 minutes Set aside a cassette with new GENIUS 2 probe covers Let the GENIUS 2 Thermometers equilibrate to the room temperature Prepare the GENIUS 2 Thermometers by thoroughly cleaning the lenses prior to use Refer to the GENIUS 2 Thermometer operating manual for cleaning instructions 9 Remove the battery compartment access panel by using a small Phillips screwdriver Remove all three AAA batteries 10 Connect the probe cable round connector to the front panel of the GENIUS 2 Checker Calibrator 11 Connect the other end of the probe connector into the battery compartment The connector will snap into place where the batteries were Refer to Figure 2 Figure 2 Insertion of Probe Connector into GENIUS 2 Battery Compartment GENIUS 2 Checker Calibrator 4 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section VI Operational Procedure The following procedure will describe the steps required to operate the GENIUS 2 Checker Calibrator In order to facilitate familiarity with the product it would be advisable to use this section of the manual during the calibration check of the first few thermometers Installation of USB Flash Drive Optional In
21. limits of allowed ambient temperature The portability of the GENIUS 2 Checker Calibrator will allow for relocation of the device to areas that meet the ambient temperature requirements e To maintain Electromagnetic Compatibility EMC conformance use a USB flash drive that is CE marked and meets EN55022 standard e Do not put or rest the GENIUS 2 Thermometer on the surface of the GENIUS 2 Checker Calibrator at any time The thermal sensors inside the thermometer probe tip will detect the heat transferred from the blackbody targets through the front panel surface This will cause instability of the thermometer sensors which will produce erroneous temperature readings The GENIUS 2 Checker Calibrator may only be used indoors in an area free from drafts and wide ambient temperature swings e The USB port on the front panel is intended to be used with only a USB Flash memory drive No other USB devices should be inserted into the USB port The USB flash memory drive should be installed prior to device power up In the event that the GENIUS 2 Checker Calibrator is dropped return the unit to the factory for re calibration GENIUS 2 Checker Calibrator 2 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Section IlI Icon Identification S N Serial Number al Manufacturer AN e ad ata am Date of Manufacture sandy Pover El eS pm CE USB Port Dispose of as
22. mbient temperature operating range Refer to Section VIII If operating within the specified ambient conditions does not eliminate the error the blackbody target s may be damaged Return the unit for factory servicing Error 5 USB Flash Drive Write Error This error is an indication of a problem when writing the report file to a USB flash drive Check that the write protect switch on the USB drive is in the off position Make sure that there is sufficient free space on the drive and that security or encryption features are not in use Error 6 Ambient Temperature Out of Range This error is an indication that the ambient temperature is not within the specified range for performing an accurate recalibration Check that the unit is being operated within the allowed ambient temperature operating range Refer to Section VIII GENIUS 2 Checker Calibrator 8 Operation and Service Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Error 7 Wrong Target This error is an indication that the GENIUS 2 Thermometer has been inserted into wrong target or was inserted into the target prematurely Please take care to follow the instructions on the screen and not perform steps prior to the instructions being displayed The Error screen will be displayed for 60 seconds before returning to the Main Menu To exit the Error screen sooner press the encoder knob Section VIII Checker Cali
23. n use the encoder knob to scroll over and highlight the desired menu option then press the encoder knob Preferences The Preferences menu is used to setup the language option Press the encoder knob to enter the Language Selection screen Language Selection The Language Selection screen lists the available languages for the LCD screen text To select a language scroll over the selection and it will be highlighted Press the encoder knob to complete the selection and return to the Main Menu screen The selected language will be stored in non volatile memory and will be used next time the unit is powered on Set User Site Information The Set User Site Information screen is used to enter the operator s name and the organization and or site information This information will be written to the test report file To enter the information use the encoder knob to scroll through the alphabet for a given letter position in the operator name or site information When the correct letter is shown press the encoder knob to select the letter and advance to the next letter position To enter a space just press the encoder knob when a space or blank letter is shown in that letter position The user name can be up to 20 letters long Use the encoder knob to scroll to the organization field and enter a name for the site in a similar manner To exit the menu scroll to the DONE text and press the encoder knob GENIUS 2 Checker Calibrator 5 Operation and Se
24. omer Service The GENIUS 2 Checker Calibrator requires factory calibration and inspection once a year Notify and return the device to Tyco Healthcare Kendall on or before the calibration date for the Checker Calibrator The Checker Calibrator will display a message one month prior to the calibration date as a reminder CAUTION THERE ARE NO SERVICEABLE PARTS INSIDE RETURN TO FACTORY FOR SERVICE In the event that it is necessary to return a unit for repair please observe the following 1 Contact Tyco Healthcare Kendall as shown below or your local customer service representative for an Authorized Return Number and shipping instructions 2 Ship insured parcel to your local service contact or the appropriate location below United States Canada Outside US amp Canada Tyco Healthcare Group LP Tyco Healthcare Canada Tyco Healthcare Group LP 98 6 Faichney Drive 7300 Trans Canada Highway Unit 2 Talisman Business Center London Road Watertown NY 13601 Pointe Claire QC Bicester 0X26 6HR UK 1 800 448 0190 or HOR 1C7 44 1869 328065 315 788 5246 1 877 664 8926 or 514 695 1220 Parts Listing Please contact your local customer service center or sales representative for the parts listed below Description GENUS 2 CNE CU CMIAON sg peon eeRRASOOED depara erre ta bin 303097 Refurbished GENIUS 2 Checker Calibrator sce caren ean dea racer eee ap eae esac ee ee aha chen 303096 Section XII Warranty Tyco Healthcare Kendall warrants the GENIUS
25. rvice Manual 10 23 2008 Artwork s WN 182563 01 Job 16245 Status Approved Doc AW10001726 Effective Date nulldate Set Date and Time The Set Date and Time screen menu option is used to setup the date and time that is used on the test report The time and date are internally stored and automatically updated when the unit is powered up To enter time use the encoder knob to select the hours and minutes Note the hours is setup using a 24 hour clock format where 1PM is entered as 13 00 etc When the value is correct push the encoder knob to move to the next field of the time and then to the date field The date field is setup using a MMDDYYYY format After the last field of the date is entered the next encoder push will exit this menu and return to the Main Menu Note the time and date are shown at the bottom of the Main Menu Check Calibration This menu option performs the calibration check of a GENIUS 2 Thermometer and if required automatically recalibrates the thermometer using a series of step by step screens Before entering this menu option make sure that the required equipment listed in Section IV is available This includes a cassette with unused new probe covers Step 1 Warm up of infrared blackbody targets This screen is displayed while the targets are checked for proper temperature Both the Low Target and the High Target are checked and when they are at the correct temperature the screen will change to OK for
26. stall a USB flash drive into the USB port located on the front panel if a report file written to the USB flash drive is desired If the USB flash drive is installed after performing a calibration check an error message may occur when attempting to write the report file if the operator does not wait for the USB flash drive to complete initialization If the GENIUS 2 Checker Calibrator is powered off all internal memory not yet saved to the USB flash drive will be lost Apply Power Make sure the AC adapter end of the power module is plugged into a grounded outlet Connect the power module to the front panel power connector Apply power to the GENIUS 2 Checker Calibrator by using the power switch on the front panel Start up Screen The first screen on the LCD display is the Initialization Screen which displays the software revision and date of the GENIUS 2 Checker Calibrator If there is an internal problem during initialization an Error screen will be displayed See Section VII Reminder Screen The next screen on the LCD display is a reminder to let the GENIUS 2 Checker Calibrator warm up for at least 15 minutes after applying power This will let the infrared blackbody targets warm up to the proper temperatures To exit this screen press the encoder knob Main Menu The Main Menu screen is displayed next This screen has four menu options Check Calibration Preferences Set User Site Information Set Time Date To enter any menu optio