Home

Troubleshooting Procedure for ArtisanLink Special Staining

image

Contents

1. iene Shooting Troubleshooting Procedure for ArtisanLink Special Staining System utomation in histology is here to stay With the shortage of experienced histotechnologists an increased workload and demands for faster turnaround times laboratories are looking for ways to produce consistent quality stained slides with less staffing in the most efficient time The arrival of automated instruments has forced histotechnologists to familiarize themselves with hardware software and chemistry in order to adequately troubleshoot automated platforms Understanding the differences and the similarities between manual and automated special staining is half the battle when it comes to troubleshooting Back to Basics Every histotechnologist understands that to achieve quality staining results performed manually the following basic factors should be in place in the laboratory 1 Proper grossing infiltration of tissue with fixative and dehydration steps 216 Debra Cobb HT MBA Dako North America Research and Development for Special Stains Carpinteria CA USA 2 A certain tissue thickness recommended for a particular procedure For instance it is recommended to cut tissue for staining with Congo Red into a thickness of 8 um and tissue for staining with Jones Basement Membrane Stain is recommended to be cut into a thickness of 2 um 3 Drying slides at the right temperature because drying slides at temperatures a
2. Dispenser showing build up of chemical deposits Staining Artifacts Problem 1 Tissue slides stored for long periods of time months to years 2 Dust particles settling on slides especially those that are charged 3 Dispensing of reagents from dirty bulk liquid containers Fig 1 Solution 1 For slides that may be dusty it is recommended to try the following procedure after the deparaffinization and hydration steps m Place the slides in running tap water for 2 3 minutes sm Rinse in deionized water m Place the slides in Artisan Wash Solution for 5 minutes before loading them on ArtisanLink 2 For slides that have dispensing artifacts it is recommended to use clean bulk liquid containers Reagent packs that are placed on the instrument need to be checked for any type of build up of chemical deposits that can be dispensed on the slide during the staining run It is recommended to check the dispenser tip and to mix and prime the packs before daily use Fig 2 It is also important to store the reagent packs tip down to prevent damage to the dispenser s internal parts Staining artifacts may also arise from reagent dyes Dyes such as Nuclear Fast Red or Crystal Violet are known to cause artifacts For instance Nuclear Fast Red is sensitive to cold If the reagent is stored in the refrigerator 218 or left out on the loading dock in freezing weather there is a risk that the reagent w
3. bove 62 C can cause damage to the tissue leading to a suboptimal staining The same basic factors are required for staining slides on an automated instrument such as ArtisanLink However two more factors should be taken into consideration a The user must follow the procedures and instructions as provided by the instrument manufacturer b The user must follow a daily maintenance procedure in addition to the biannual or annual preventive maintenance procedure Troubleshooting Issues When troubleshooting ArtisanLink Special Staining System the histotechnologist will not only be required to rely on his her knowledge of histology procedures and special stains chemistry but also on his her computer skills in order to optimize staining protocols using the ArtisanLink software Because the instrument applies reagents as well as heats and incubates according to the instructions given by the user it is important that the user programs the instrument properly taking into consideration that all other processes such as grossing sectioning fixation etc are optimized before the sample is placed on the instrument In other words how to optimize individual slides based on prestaining processes can only be defined by a user The most common staining issues on ArtisanLink are m Staining artifacts m Background staining m Inconsistent staining Figure 1 Artifacts on slides caused by dirty bulk liquid containers Figure 2
4. ill start to precipitate Once this reaction starts it will continue regardless of correcting the storage temperature This precipitate will then be dispensed onto the slide Fig 3a Crystal Violet is another example of a dye that has a tendency to leave crystal like shards on the tissue Fig 3b Reagents used on ArtisanLink have been validated and optimized for staining quality as well as for correct storage and usage temperatures It is important that the user follows the recommended temperature for storage and usage for each of the staining kits All ArtisanLink staining kits have been validated to be used at room temperature Deviating from the recommendations may Cause an uneven staining shorten the shelf life and possibly cause a false negative staining Reagent storage and usage temperatures can be located on the reagent packs and in the package inserts Background Staining Background staining is a common problem when using charged slides or when there are additives in the water bath or by heat When using a methenamine silver solution a mirroring also know as grey haze can occur on the slide Mirroring is defined as metallic silver deposited non specifically over large areas of the front and back of the glass slide The silver solution deposits a fine brown black granular precipitate over the section and the glass portion of the slide Fig 4 Microscopically this type of precipitate is usually not found
5. nd the automated procedures is very similar The table above shows the comparison ArtisanLink Special Staining System provides the histotechnologist with a tool to produce the same diagnostic quality stains as found in a manual staining technique However unlike a manual staining ArtisanLink allows the histotechnologist to walk away from the special staining bench to pursue other tasks Reference 1 Brown RW Ed 2009 Histologic Preparations Common Problems and Their Solutions College of American Pathologists Northfield IL 2009 221
6. on the same plane as the tissue 1 To prevent mirroring or grey haze it is important to check the temperature of the slide heater and or the incubation time set in the software Using non charged slides except unlike a manual staining ArtisanLink allows the histotechnologist to walk away from the special staining bench to pursue other tasks for decalcified tissue can also help lessen the amount of background staining when performing stains with silver nitrate solutions Inconsistent Staining This may be caused due to m Inconsistent tissue thicknesses from slide to slide For instance thicker sections will show a strong staining whereas thinner sections will show a weak staining Fig 5 It is important to keep to the correct thickness throughout the microtomy sectioning to obtain a consistent staining m Cold reagent packs not brought to room temperature before use All reagent packs used on ArtisanLink Special Staining System must be brought to room temperature 22 25 C Allow 45 minutes for the reagent packs to warm up before placing them on ArtisanLink Using a cold or cool reagent pack may prevent the chemical reaction from taking place and will give the impression that the instrument is producing inconsistent staining results within the same run m Reagent packs not primed before use This can cause either a non dispense of reagent or an incomplete dispense of reagent It is important to alwa
7. se Use clean slides Clean bulk liquid containers Check dispenser tip for debri build up Use the default protocol for silver stains Do not add adhesives to the water bath Employ proper fixation dehydration and infiltration Cut all sections into a thickness of 4 um except for Jones Basement Membrane Stain for kidney which should be cut into a thickness of 2 um and Congo Red Stain which should be cut into a thickness of 8 um soak the slides in Artisan Wash Solution for 5 minutes before placing them on the instrument m Uneven staining This can be caused by the reagent not spreading evenly across the tissue section or sections This is particularly problematic with reagents containing acids such as acetic acid or periodic acid The solution to this problem is to soak the slides in a Artisan Wash solution for 5 minutes before loading the slides onto the slide carousel This procedure conditions the slides for spreading the reagents Fig 6 Another reason for an uneven staining may be the placement of the tissue sections on the slide close to the slide clip This can be rectified by making sure that the tissue sections are centered on the slide Placing the sections close to the slide clip causes a wicking effect which prevents complete aspiration of the reagents causing an uneven staining Conclusion The solution to staining artifacts background staining and inconsistent staining for both the manual a
8. ys prime the packs before daily use to ensure proper dispensing of reagents Figure 3a Nuclear Fast Red precipitate u AN or bi ait fi _ Ink k d a 4 by eh 4 un u a im Figure 3b Crystal Violet showing shards Connection 2010 219 Figure 5 Inconsistent tissue thicknesses from the same run Warthin Starry Stain Figure 4 Examples of mirroring grey haze The cause of this effect is overheating of the methenamine solution i Figure 6 Periodic Acid Schiff Diastase PAS D staining of liver tissue The uneven staining is due to amylase not spreading SA evenly causing part of the liver section to wn stain undigested glycogen 220 Connection 2010 Issue Manual Procedure ArtisanLink Procedure staining Artifact Background Staining Inconsistent Staining Store reagents at correct temperature See package insert for specification Filter reagents before use Use clean slides Clean glassware in soap and hot water Rinse in deionized water and air dry Use acid clean glassware Do not use metal forceps Employ proper fixation dehydration and infiltration Increase or decrease staining time by looking at individual slides under the microscope For an uneven staining adjust the volume of reagent used to ensure proper spreading Store reagent packs at correct temperature See package insert for specification Mix and prime reagents before u

Download Pdf Manuals

image

Related Search

Related Contents

Manual de Usuario User manual Mode d´emploi  ptx Radio IC 16  Guide méthodologique destiné aux gestionnaires financiers  T'nB CBFLAT1BL mobile phone cable  HQ Power PAR56  manual do usuário  Manual - Innovative Technology  EZ CE Panel Getting Started  SC-323PU Bedienungsanleitung iOS 7  WERKPLAATSPERS PRESSE D`ATELIER  

Copyright © All rights reserved.
Failed to retrieve file