         Contents
1.           TekLink                3  i PE  AO O                  ne OUT               EXTERNAL TRIG OUT                               EXTERNAL TRIG IN        2771 002                                                                            y     l                                              Figure 2  Accessories connections    TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Additional Accessory Connect additional accessories as needed   See Table 3    Connection Information    Table 3  Additional accessory connection information    Item Description       Monitor There are two display ports on the logic analyzer  VGA  and DVI I  digital analog out   Simulscan  mode displays the same information on the internal display and the external monitor connected to  either Primary or Secondary     To change the display settings  right click the mouse on the desktop  and select Properties  gt  Settings  gt   Advanced  gt  Intel Graphics  Note that some of the display settings may not function until you connect an  external monitor     To connect a VGA monitor to the DVI I connector  use a VGA to DVI adapter  Part numbers for  accessories and options are listed at the end of this manual   See page 27  Accessories and Options    If you use a nonstandard monitor  you may need to change the Windows display settings to obtain  the proper resolution        LAN Connect the logic analyzer to your network through the Gbit LAN connectors  Remotely control the logic  a
2.          LL 13   Change the BIOS Settings oceni iei srn ea Ria 14  Reinstalling the TLA Application Software                 0 15  Installing the TLA Application Software on a PC                   een ence eee ee eee eneeneeeaeeeneenaes 15  Calibrate the Touchscreen econ end aio 17  Install Other Softwares  dao 18  Upgrading or Restoring Firmware                   non n nn cnn nan n eee tne narran nana 18  Appendix A  User Service Procedures                     L cnn nn ron rr nr nr ron nr EE enn EEE EES 21  AA O 21    TLA6000 Series Logic Analyzers Installation Manual 1    Table of Contents    Warranty Repair Service y orale prices bei 21  Calibration and Repair Service                 0 21  General Ci aa Mod nt al e e a nana 21  Preventive  Mamtenance    irc diet 22  Cleaning the Flat Panel Display             0 22  Exterior Surfaces    iui an a ada 22   In  Case of Problems    a Ad van A lid ai  s 23  Diagnostiese a is 23  Software Problems 0 A A AAA aero i 23  Hardware Problems aii A aiar 24  Network Connection Problems                   L cnc rrrnnnannnnnos 24   Check for Common Problems            0 0 00 ce cece cece nen eee nnn ron nnn EEE EEE Ene EEE ea 24  Repacking for Shipment                   0 25  Appendix B  Accessories and Options              0 cece 0 27  O ON 27  OPIS A AA AA 27  Appendix C  Network Installation Site SUrvey               ui 29    Index    11 TLA6000 Series Logic Analyzers Installation Manual    List of Figures    Figure i  TLA6000 series log
3.      If the Connection dialog box appears  select the instrument that you want to  upgrade  typically the Local instrument  and click Connect     Click the Load button in the Instrument Module Firmware section  bottom  part of the dialog box      If the firmware version is less than the FW Required version  then select Load  Firmware from the Execute menu     Select the proper firmware file  TLA620X lod    Click OK  You will be prompted to confirm your action  click Yes     When the process has completed  exit the firmware loader program and power  off the logic analyzer  You must power off the logic analyzer to allow the  TLA application to start up properly     Locate the big label on the instrument     Record the firmware version that is printed on the label  You will need this  information to see that the firmware version matches the label     Power on the logic analyzer     After the logic analyzer completes the power on diagnostics  select System  Properties from the System menu     Click the LAI tab     Verify that the firmware version matches the version on the label that you  recorded in step 20     If the firmware versions do not match  power off the instrument  and update  the label     TLA6000 Series Logic Analyzers Installation Manual 19    Restoring and Reinstalling Software    20 TLA6000 Series Logic Analyzers Installation Manual    Appendix A  User Service Procedures    This appendix describes high level service information and procedures for the  Tektronix lo
4.   with 8 GHz MagniVu timing  235 MHz  state clock  2 Mb record length      TLA6204 136 channel logic analyzer  with 8 GHz MagniVu timing  235 MHz  state clock  2 Mb record length       The logic analyzers are built on the Microsoft Windows operating system  which  allows you to install PC compatible  third party hardware and software on the  instrument     The user interface operates under the Microsoft Windows operating system   Microsoft recommends the following to make sure your instrument is protected     m Use an internet firewall     Install operating system updates regularly       Use up to date antivirus software     TLA6000 Series Logic Analyzers Installation Manual xi    Preface    Documentation    Related Documentation    ltem                                                 2771 001    Figure i  TLA6000 series logic analyzer    The following table lists related documentation available for your logic analyzer   The documentation is available on the TLA Documentation CD and on the  Tektronix Web site  www tektronix com manuals      For documentation not specified in the table  contact your local Tektronix  representative     Purpose Location       TLA Quick Start User Manuals    High level operational overview ES      S  gt  LD       Online Help          In depth operation and UI help                      xii    TLA6000 Series Logic Analyzers Installation Manual    Related Documentation  cont      ltem    Purpose    Location    Preface       Installation Quick Refere
5.  Analyzers Installation Manual 13    Restoring and Reinstalling Software    Restoring the Operating  System from the Restore  Discs    Change the BIOS Settings    You can only use the restore discs on the instrument that created them     NOTE  This procedure requires a mouse and that the DVD drive is set as the first  startup device  this is the default setting  If necessary  you can change the startup  setting   See page 14  Change the BIOS Settings         1  Insert restore disc 1 into the instrument DVD drive     2  Restart the instrument  The restore software opens automatically if the DVD  drive is the first startup device  If the DVD drive is not the first startup device   you need to enable it as the first startup device before performing a restore  from the restore discs     3  Click Restore     4  Inthe Confirmation dialog box  click Yes to restore the operating system and  instrument application  or No to exit the restore process     5  When the restore process is completed  remove the restore disc and restart the  instrument     Continue with any restore discs when prompted   When prompted  read and accept the end user license agreement  EULA      Click Next and then configure your regional settings  as needed     Se A    Click Next   The instrument will apply the settings and then restart     10  After the instrument restarts  reinstall the TLA Application software   See  page 15  Reinstalling the TLA Application Software      Before reinstalling the operating sys
6.  DMAX P6960 logic analyzer probes  45 Increase speed to 450 MHz state clock       TLA6000 Series Logic Analyzers Installation Manual 27    Appendix B  Accessories and Options    Option Description                                              Manuals   LO Add English manuals  TLA6000 Installation Manual and TLA Documentation CD    L5 Add Japanese manuals  TLA6000 Installation Manual and TLA Documentation  CD    L10 Add Russian manuals  TLA Documentation CD    L99 No manuals   Service offerings   CA1 Single calibration event or coverage   C3 Calibration services extended to cover three years   C5 Calibration services extended to cover five years   D1 Calibration data report   D3 Test Data for calibration services in Opt  C3   D5 Test Data for calibration services in Opt  C5   R3 Repair warranty extended to cover three years   R5 Repair warranty extended to cover five years       Table 7  Power cords                                  Option Description Part number  AO North America  IEC320 C13 161 0104 00  Al Universal European  IEC320 C13 161 0104 06  A2 United Kingdom  IEC320 C13 161 0104 07  A3 Australia  IEC320 C13 161 0104 14  A4 240 V North American  IEC320 C13 161 0104 08  A5 Switzerland  IEC320 C13 161 0167 00  A6 Japan  IEC320 C13 161 A005 00  A10 China  IEC320 C13 161 0306 00  A11 India South Africa  IEC320 C13 161 0324 00  A99 No power cord       28 TLA6000 Series Logic Analyzers Installation Manual    Appendix C  Network Installation Site Survey    Customer     Photoc
7.  NOTE  This procedure creates a set of restore discs for the Microsoft Windows  operating system  After restoring the operating system  use the TLA Application  disc to reinstall the TLA application software   See page 15  Reinstalling the TLA  Application Software      6 TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Complete the following steps to create a set of restore discs  You will need blank  discs  one for each backup file      1     enn APA       10     11     Insert a blank disc in the instrument DVD drive    Click Start  gt  All Programs  gt  Nero 8  gt  Nero Express Essentials    Click Image  gt  Project  gt  Copy    Click Disk Image or Saved Project    Navigate to C  backup    Select file backup  and click Open    Click the Verify data on disk after burning check box to enable this function     Click the Burn button  The application writes the backup file to the disc and  then verifies that the data matches the source file     When the application reports that the write process has completed  successfully  remove the disc and label it appropriately  include backup file  name  instrument name  instrument serial number  and date      Copy the disk image files    iso  located in the C  backup directory to a  network location  a separate hard disk  or optical media for backup purposes     Store the backup discs as defined by your company policy        NOTE  You can only use restore discs on the instrument that created them        Perform
8.  appear on the product     A A QQ     QO      CAUTION WARNING Protective GroundEarth Terminal i Mains Disconnected Mains Connected  Refer to Manual High Voltage  Earth  Terminal Chassis Ground OFF  Power  ON  Power   Standby    TLA6000 Series Logic Analyzers Installation Manual v    Service Safety Summary    Service Safety Summary    vi    Only qualified personnel should perform service procedures  Read this Service  Safety Summary and the General Safety Summary before performing any service  procedures     Do Not Service Alone  Do not perform internal service or adjustments of this  product unless another person capable of rendering first aid and resuscitation is  present     Disconnect Power  To avoid electric shock  switch off the instrument power  then  disconnect the power cord from the mains power     Use Care When Servicing With Power On  Dangerous voltages or currents may  exist in this product  Disconnect power  remove battery  if applicable   and  disconnect test leads before removing protective panels  soldering  or replacing  components     To avoid electric shock  do not touch exposed connections     TLA6000 Series Logic Analyzers Installation Manual    Compliance Information    This section lists the EMC  electromagnetic compliance   safety  and  environmental standards with which the instrument complies     EMC Compliance    EC Declaration of Meets intent of Directive 2004 108 EC for Electromagnetic Compatibility   Conformity   EMC Compliance was demonstrated t
9. A Application  software  run the TLA Mainframe Diagnostics located under the Windows Start  menu using the following path  Start  gt  All Programs  gt  Tektronix Logic Analyzer   gt  TLA Mainframe Diagnostics  Exit the TLA Application before running the  external diagnostics     Back up your user files on a regular basis  Use the Windows backup tools or copy  the files to another media  Always keep a backup copy of files that you access  on a regular basis     Removing the Replaceable Hard Disk Drives    CAUTION  To avoid permanently damaging the replaceable hard disk drive  do  not remove it while the instrument is powered on  Always power off the instrument  before removing the hard disk drive        1  Verify that the instrument is powered off     2  Pull on the replaceable hard disk drive cartridge to remove it from the chassis     TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Hard disk a  AI AF    ZS    AM                                  2771 04    Figure 5  Accessing the replaceable hard disk drive    Connecting Probes to the Target System    The logic analyzer connects to the target system through the probes  The logic  analyzer probes allow you to connect to the target system in several different  ways  For probe specific connection details  refer to the appropriate probe  instruction manual or browse through the Tektronix Web site     TLA6000 Series Logic Analyzers Installation Manual 9    Basic Installation    10 TLA6000 Series Logic A
10. External connectors  11    F    Firmware  upgrading  18  Front panel controls  11    G    General maintenance  21  Ground  chassis  2    H    Hard disk  drive  removal  8  Hardware problems  24    Incoming inspection  7  Installing  software  TLA application  software  15    L    LA module See Logic analyzer  Logic analyzer  chassis ground connections  2  connecting probes  5  installing software  13  reinstalling software  13  shipping list  1  site considerations  1  TLA6000 Series  description  xi    Mainframe  diagnostics  23    N    Network problems  24  NI GPIB USB software  16    O    On Standby switch  5  Operating system  restoring  13    TLA6000 Series Logic Analyzers Installation Manual    Operating system restore discs  creating  6   Optional accessories  27   OS restore discs  creating  6    P    Power cords   part numbers  28  Power on diagnostics  23  Power on procedure  5  Preventive maintenance  22   cleaning  22   LCD display  22  Probes   connecting probes  5  9  Problems   hardware  24   network connection  24   software  23  Product   accessories  27    R    Reinstalling software   TLA application software  15  Removing the hard disk drive  8  Repacking for shipment  25  Repair service  21  Replaceable hard disk  8  Requirements   site consideration  1  Restore discs   creating  6  Restoring the operating   system  13    S    Safety Summary  1v    31    Index    Service  calibration and repair  service  21  mainframe diagnostics  23  offerings  21  preven
11. TLA6000 Series  Logic Analyzers    Installation Manual       AURA MA VAT AKA UA ON Tektronix  071 2771 00 Y    TLA6000 Series  Logic Analyzers    Installation Manual    www tektronix com  071 2771 00    Tektronix  7    Copyright O Tektronix  All rights reserved  Licensed software products are owned by Tektronix or its subsidiaries  or suppliers  and are protected by national copyright laws and international treaty provisions     Tektronix products are covered by U S  and foreign patents  issued and pending  Information in this publication  supersedes that in all previously published material  Specifications and price change privileges reserved     TEKTRONIX and TEK are registered trademarks of Tektronix  Inc   MagniVu and TekLink are trademarks of Tektronix  Ic     Contacting Tektronix    Tektronix  Inc    14200 SW Karl Braun Drive  P O  Box 500   Beaverton  OR 97077   USA    For product information  sales  service  and technical support     In North America  call 1 800 833 9200     Worldwide  visit www tektronix com to find contacts in your area     Warranty    Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one  1   year from the date of shipment  If any such product proves defective during this warranty period  Tektronix  at its  option  either will repair the defective product without charge for parts and labor  or will provide a replacement   in exchange for the defective product  Parts  modules and replacement 
12. al  Injury    iv    Review the following safety precautions to avoid injury and prevent damage to  this product or any products connected to it     To avoid potential hazards  use this product only as specified   Only qualified personnel should perform service procedures     While using this product  you may need to access other parts of a larger system   Read the safety sections of the other component manuals for warnings and  cautions related to operating the system     Use proper power cord  Use only the power cord specified for this product and  certified for the country of use     Connect and disconnect properly  Do not connect or disconnect probes or test  leads while they are connected to a voltage source     Ground the product  This product is grounded through the grounding conductor  of the power cord  To avoid electric shock  the grounding conductor must be  connected to earth ground  Before making connections to the input or output  terminals of the product  ensure that the product is properly grounded     Observe all terminal ratings  To avoid fire or shock hazard  observe all ratings  and markings on the product  Consult the product manual for further ratings  information before making connections to the product     The inputs are not rated for connection to mains or Category II  III  or IV circuits   Connect the probe reference lead to earth ground only     Power disconnect  The power cord disconnects the product from the power source   Do not block the power co
13. ave an older version of the software on the hard disk  the installation  program will detect it and ask to remove it  Follow the on screen instructions to  remove the software  answering  Yes  to any prompts  Restart the instrument  when prompted and run the Setup exe program again     TLA6000 Series Logic Analyzers Installation Manual    Restoring and Reinstalling Software    Start the software by double clicking on the TLA Application icon  The TLA  Startup dialog box displays     1     Select an instrument in the TLA Connection dialog box  and then click the  Connect button  If your instrument is connected to a network  you can connect  to any instrument on your local network that has a TLA server running   enabled  and is not being used by another user     Click Offline to start an offline version of the TLA application software   The TLA Offline software allows you to run the TLA application without  connecting to a real instrument  You can analyze previously acquired data  from a logic analyzer  create or modify reference memories  or perform  system tests without being connected to a real instrument     Calibrate the Touchscreen    Perform the following steps to calibrate the touchscreen     1     Go to the Windows desktop and the double click the Touch screen cofigurator   the pointing Device icon or small mouse graphic   The Touch Screen  Properties window appears     Click Calibrate and follow the on screen instructions to perform the alignment  routine  Use a stylu
14. b xii  Basic  Installation viii ta aa 1  Check the Shipping List  allagata rr 1  Site  COnSIderations ici aaa lato 1  Chassis Ground Connections                   scene ence ne ene ence eens eee e ee ene ene eeaeeenneneeeaeeneenennaeees 2  Connecting Accessories iiss  sie a iS A ade 3  Additional Accessory Connection Information                   0 4  Connecting  Probes vi asus ehebes Sense ceatadaaeteae nd TOTE 4  First Time Operation a A RR 5  Turning the Logic Analyzer On 2 0 0    0    cece cece ene enn nnn nora n cr nora nrr nana ne tne ea 5  Turning off the Logic Analyzer               Li 6  Creating Operating System Restore DISCS            0    ccc cece eee ce teen eee eee ee ene ene en eden nee e eae nne ena 6  Performing the Incoming Inspection                      0 7  Checking the Logic Analyzer Probes  Optional                        8  Backing Up User FS ini A cas 8  Removing the Replaceable Hard Disk Drives                  0c ccc ccc cece eee ence cnet eee e ene e ene eneeenaes 8  Connecting Probes to the Target System                 0 9  PIOAUCLOVETVIEW   Luria caos 11  Front Panel Controls ia il LEA iaia ila 11  External COnneGtors   s scandire 11  Restoring and Reinstalling Software 2 0 0 0    0    ccc cece ence eee non nn con rr nn cnn rr ne cnn rr nee eee tne eta ES 13  Restoring the Instrument Operating System                 ccc cece cece cece ee eee ee nn ran ron nr narra ninas 13  Restoring the Operating System from the Instrument Hard Disk           
15. ccessories that you ordered    Read this section before installing the logic analyzer  This section describes  operating considerations and power requirements for your logic analyzer  The  environmental considerations apply to all TLA6000 series products     Table 1  Environmental considerations    Feature Description       Temperature Operating  5   C to  45   C    Nonoperating  20   C to  60   C       Humidity Operating  lt 30   C  80  relative humidity  29   C maximum wet bulb  20  to 80  temperature     Nonoperating 8  to 80   29   C maximum wet bulb temperature        Altitude Operating To 3000 m  9843 ft      and  Nonoperating          NM CAUTION  Allow a 15 3 cm  6 in  clearance at the top  back  and sides of the   instrument to allow proper cooling  Avoid blocking any exhaust fans or vents  when using the instrument on a cart or in an instrument rack  Inadequate  clearances can cause the instrument to overheat and shut down        TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Use the logic analyzer on a bench  in an instrument rack  or on a cart in the normal  position  on the bottom feet   The front feet extend to give a better view of the  instrument display     Table 2  Instrument power considerations             Feature Description   Voltage range and 90 VAC to 250 VAC at 45 Hz to 66 Hz  frequency 100 VAC to 132 VAC at 360 Hz to 440 Hz  Input current 7 A maximum at 90 VAC  70 A surge   Power consumption 750 W maximum       Chassis Gro
16. der dialog box appears  You have the choice of loading  either Mainframe and TekLink Firmware or Instrument Module Firmware   Click the Load button corresponding to Mainframe and TekLink Firmware   top part of the dialog box      5  A list of all the currently installed Mainframe and TekLink firmware displays   The list shows the version currently running and the version available for  installation  Select the items that you want to load  If all firmware versions  are up to date  click Close and go to step 14     6  Click the Update Firmware button  A dialog box displays a list of available  firmware images     7  Select the appropriate file  Click Open to open the selected firmware image   You must do this for each firmware image selected     8  The Firmware Load Progress dialog displays  Click the Start Flash Operation  button to start the flash operation for the selected flash images  Once the flash  operation is started 1t cannot be canceled     9  Astheflash operation runs  it identifies what steps are being performed  When  complete  it will inform you to restart the instrument     10  Turn off and restart the instrument after the flash operation is complete     TLA6000 Series Logic Analyzers Installation Manual    11   12     13     14     15     16   17   18     19     20     21   22     23   24     25     Restoring and Reinstalling Software    Exit the logic analyzer application     Click Start  gt  All Programs  gt  Tektronix Logic Analyzer  gt  TLA Firmware  Loader
17. g system from either a file on the  instrument hard disk drive or from the set of instrument restore discs   See page 6   Creating Operating System Restore Discs   The preferred restore method is to  use the hard disk restore file        NM CAUTION  Restoring the operating system reformats the hard disk drive  All   saved data including the files that allow you to create the restore discs will be lost   Create the restore discs and then save important files to external media before  restoring the operating system        Restoring the Operating The instrument contains an operating system restore file on a separate partition  System from the of the hard drive     Instrument Hard Disk 1  Restart the instrument  During the startup process  you will see the following    message at the top of the screen     Starting Acronis Loader press F5 for Acronis Startup  Recovery Manager    2  Repeatedly press the F5 key until the Acronis True Image Tool opens  There  is a short delay from the time the message appears until the instrument  proceeds with the normal instrument startup  If the instrument does not  open the Acronis application  power off the instrument  and then power on  the instrument and try again     3  Click Restore     4  In the Confirmation dialog box  click Yes to restore the instrument  operating system  or No to exit the restore process  The restore process  takes approximately 30 minutes  the actual time depends on the instrument  configuration     TLA6000 Series Logic
18. gic analyzers     Service Offerings    Tektronix provides service to cover repair under warranty and other services that  are designed to meet your specific service needs     Whether providing warranty repair service or any of the other services listed  below  Tektronix service technicians are well equipped to service the logic  analyzers  Services are provided at Tektronix Service Centers and on site at your  facility  depending on your location     Warranty Repair Service Tektronix warrants this product as described in the warranty statements at the  front of this manual  Tektronix technicians provide warranty service at most  Tektronix service locations worldwide  The Tektronix product catalog lists all  service locations worldwide     Calibration and Repair In addition to warranty repair  Tektronix Service offers calibration and other  Service Services that provide cost effective solutions to your service needs and quality   standards compliance requirements  Our instruments are supported worldwide  by the leading edge design  manufacturing  and service resources of Tektronix to  provide the best possible service     General Care    Protect the instrument from adverse weather conditions  The instrument is not  waterproof  Do not store or leave the instrument where the LCD display will be  exposed to direct sunlight for long periods of time        CAUTION  To avoid damage to the instrument  do not expose it to sprays  liquids   or solvents        TLA6000 Series Logic Analy
19. ic analyzer               i xii  Figure 1  Location of the ground connection on the logic analyzer                  0 2  Figure 2  Accessories connections    0ooooocoocconnconcconnoonnonn nro nr r rn ron EEE EE ERE EEE EEE EE EEE EEE EERE EEE 3  Figure 3  Connecting probes to the logic analyZer            0    cece cece cece eee cent cnet ene ene e snes eneeenaes 5  Figure 4  On Standby switch location                  cece eee ene eens ene n ee ene ee rr ene e ene e nee ena es 6  Figure 5  Accessing the replaceable hard disk drive               ce cece cece ence cece eens eee e ene eenaeeneeenaes 9  Figure 6  Logic analyzer front panel           0 0    ccc cece cence eee eee nen ne cor rr ene eee nn ete bE eta ES 11  Figure 7  External COMECtors ii a sae 12  List of Tables  Table 1  Environmental considerations                      0 1  Table 2  Instrument power considerations              0c cece eee e eect eee eee eee cnn ron nn rr rar e ene ene e eats 2  Table 3  Additional accessory connection information                      0 4  Table 4  Failure symptoms and possible causes                   ence eee e eee eee e ene eee e eee nneeeaeenne eens 24  Table 5  Standard accessories csi ia ire 27  Table 6  Optional accessories                      een cnn rr rr EEE EEE EEE rn EEE EEE EET EES 27  Table 7  POWer COTOS estriado ld riada dida 28    TLA6000 Series Logic Analyzers Installation Manual 111    General Safety Summary    General Safety Summary    To Avoid Fire or Person
20. ing the Incoming Inspection    Incoming inspection consists of verifying the basic operation of the logic analyzer   The power on diagnostics check the basic functionality  The diagnostics run  every time you turn on the logic analyzer     You can also verify more detailed functionality by running the self calibration and  extended diagnostics        NOTE  Allow the logic analyzer to warm up for 30 minutes before running the  self calibration        TLA6000 Series Logic Analyzers Installation Manual 7    Basic Installation    Checking the Logic  Analyzer Probes  Optional     Backing Up User Files    To run self calibration and diagnostics  perform the following   1  Disconnect any probes attached to the instrument   2  Select the System menu and click Calibration and Diagnostics     3  Run the self calibration  and then run the extended diagnostics by selecting  the proper tab  Results of the tests display on the individual property page        NOTE  The time required to run the self calibration on the logic analyzer depends  on the number of acquisition channels  Logic analyzers with a large number of  channels may take several minutes to run the self calibration        Connect the logic analyzer probes to a signal source and  if necessary  adjust the  threshold levels for the probes to acquire the signals  Use the threshold and activity  indicators in the LA Setup window to verify signal activity at the probe tips     To check the mainframe diagnostics not covered by the TL
21. mity   EMC    Safety Compliance    EC Declaration of  Conformity   Low Voltage    U S  Nationally Recognized  Testing Laboratory Listing    Canadian Certification    Additional Compliances    Equipment Type  Safety Class    Pollution Degree  Description    viii    Complies with the EMC provision of the Radiocommunications Act per the  following standard  in accordance with ACMA     m CISPR 11 2003  Radiated and Conducted Emissions  Group 1  Class A  in  accordance with EN 61326 1 2006 and EN 61326 2 1 2006     Compliance was demonstrated to the following specification as listed in the  Official Journal of the European Communities     Low Voltage Directive 2006 95 EC        EN 61010 1  2001  Safety requirements for electrical equipment for  measurement control and laboratory use       UL 61010 1 2004  254 Edition  Standard for electrical measuring and test  equipment     m CAN CSA C22 2 No  61010 1 2004  Safety requirements for electrical  equipment for measurement  control  and laboratory use  Part 1     m JEC 61010 1  2001  Safety requirements for electrical equipment for  measurement  control  and laboratory use     Test and measuring equipment     Class 1     grounded product     A measure of the contaminants that could occur in the environment around  and within a product  Typically the internal environment inside a product is  considered to be the same as the external  Products should be used only in the  environment for which they are rated       Pollution Degree 1  No 
22. nalyzer through a LAN switch using a PC  loaded with the TLA application software         Printer The instruments send printer information to the USB ports  Use any of the four USB ports for your printer        Connecting Probes    After you have connected all the accessories  connect the probes to the logic  analyzer     For additional information on the individual probes  refer to the instructions that  accompanied your probes   See Figure 3         AN CAUTION  When attaching the probe to the logic analyzer  you must use care to   evenly tighten probe screws until they are snug  First slightly tighten screws  then  snug each screw to 4 in lbs  maximum   Under tightening the probe screws can  result in intermittence  Over tightening can result in stripped screws        4 TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Wo    S    Ane          QO                                     Match color coded labels    VIN                                  SAU  S    2774 0068    Figure 3  Connecting probes to the logic analyzer    First Time Operation    Use the information in this section to turn on the logic analyzer for the first time        CAUTION  Connecting the accessories after turning on the logic analyzer may  damage the accessories  Connect accessories before applying power to the logic  analyzer        Turning the Logic Analyzer Follow these steps to turn the logic analyzer on   On    1  Connect the power cord to the logic analyzer     2  If you ha
23. nalyzers Installation Manual    Product Overview    This chapter describes the product controls and connectors of the logic analyzers   Refer to the TLA Quick Start User Manual for operating information on the logic  analyzer     Front Panel Controls    Use the front panel controls to operate most of the logic analyzer   See Figure 6      Run Stop    Co  tt  Moves the display pointer       ae    and selects objects Se C2                                  O  x  Mew        _ Zo Mark  Navigation controls       dI MagniVu  Search      CTD  Scrolls data vertically         gt          Scrolls data horizontally            Changes Time Div in    selected waveform the Waveform window     Waveform window only  ya         Off Ctrl Shift    O  Changes the height of             cy    DO          Figure 6  Logic analyzer front panel    External Connectors    The following mainframe external connections are available   See Figure 7 on  page 12      m System Trigger In and System Trigger Out  used to receive or send a trigger  from to an external source    m External Signal In and External Signal Out  used to receive or send a signal  from to an external source      TekLink  used to coordinate trigger signals  input output signals  and time  references between mainframes    m Accessory connections  such as USB  LAN  and audio video outputs    TLA6000 Series Logic Analyzers Installation Manual 11    Product Overview    DVI I LAN    Lo 7     LAN ON  10 100 1000 Mbps LINE IN  Dl            IDI o
24. nce Cards    Installation Manuals    High level installation information    Detailed first time installation information         ra   gt                       XYZs of Logic Analyzers    Logic analyzer basics       www Tektronix com       Declassification and Securities  instructions    Data security concerns specific to  sanitizing or removing memory devices  from Tektronix products       Application notes    Product Specifications  amp  Performance  Verification Procedures    Collection of logic analyzer application  specific notes    TLA Product specifications and  performance verification procedures          TPI NET Documentation    Detailed information for controlling the  logic analyzer using  NET       Field upgrade kits    Upgrade information for your logic  analyzer                 Optional Service Manuals    Self service documentation for modules  and mainframes    TLA6000 Series Logic Analyzers Installation Manual    OQ         xiii    Preface    xiv TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    This chapter describes the steps needed to install your Tektronix logic analyzer  and related accessories     Check the Shipping List    Verify that you have received all of the parts of your logic analyzer using the  shipping list and the accessories list   See page 27  Accessories   Also check  for the following     Site Considerations    Power cords are correct for your geographical area  Correct probes  Standard accessories    All optional a
25. nnected to the low voltage installation        m Measurement Category I  For measurements performed on circuits not  directly connected to MAINS     Overvoltage Category  Overvoltage Category II  as defined in IEC 61010 1     Environmental Considerations    This section provides information about the environmental impact of the product     Product End of Life Observe the following guidelines when recycling an instrument or component   Handling   Equipment recycling  Production of this equipment required the extraction and  use of natural resources  The equipment may contain substances that could be  harmful to the environment or human health if improperly handled at the product   s  end of life  To avoid release of such substances into the environment and to  reduce the use of natural resources  we encourage you to recycle this product in  an appropriate system that will ensure that most of the materials are reused or  recycled appropriately     This symbol indicates that this product complies with the applicable European  Union requirements according to Directives 2002 96 EC and 2006 66 EC   on waste electrical and electronic equipment  WEEE  and batteries  For  information about recycling options  check the Support Service section of the  Tektronix Web site  www tektronix com      TLA6000 Series Logic Analyzers Installation Manual ix    Compliance Information    Mercury notification  This product uses an LCD backlight lamp that contains  mercury  Disposal may be regulated d
26. ns power on diagnostics every time you start the TLA  application if the box is checked in the Configuration dialog box  View the  results of the diagnostics by selecting Calibration and Diagnostics from the  System menu  If desired  run more detailed diagnostics by selecting Extended  diagnostics  Here you can run all tests  loop on one or more tests  or loop on a test  until a failure occurs     Before running the extended diagnostics  disconnect any attached probes     TLA Mainframe Diagnostics  The TLA mainframe diagnostics program is a  stand alone application  These diagnostics check the operation of the mainframe  beyond the basic PC circuitry  These diagnostics also check the front panel knobs  of the portable mainframe     Checklt Utilities  The CheckIt Utilities is a separate application located in the  Windows Start Programs menu  Start  gt  Programs  gt  CheckIt Utilities   The  diagnostics check the basic PC operations of the controller     Software Problems Your logic analyzer comes with most software installed  Before running any of  the diagnostics  check the online release notes to verify that the logic analyzer  software is compatible with the module firmware     Many software problems are due to corrupted or missing software files  In most  cases the easiest way to solve software problems is to reinstall the software and  follow the on screen instructions   See page 13  Restoring and Reinstalling  Software      TLA6000 Series Logic Analyzers Installation Ma
27. ntegration increases the time or difficulty  of servicing the product     THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY  OTHER WARRANTIES  EXPRESS OR IMPLIED  TEKTRONIX AND ITS VENDORS DISCLAIM ANY  IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE   TEKTRONIX    RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE   AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY   TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT  SPECIAL  INCIDENTAL   OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS  ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES      W2     15AUG04     Warranty    Tektronix warrants that the media on which this software product is furnished and the encoding of the programs on  the media will be free from defects in materials and workmanship for a period of three  3  months from the date of  shipment  If any such medium or encoding proves defective during the warranty period  Tektronix will provide   a replacement in exchange for the defective medium  Except as to the media on which this software product is  furnished  this software product is provided    as is    without warranty of any kind  either express or implied   Tektronix does not warrant that the functions contained in this software product will meet Customer   s requirements  or that the operation of the programs will be uninterrupted or error free     In order to obtain ser
28. nual 23    Appendix A  User Service Procedures    Hardware Problems    Network Connection  Problems    Check for Common  Problems    If you are certain that you have installed the logic analyzer correctly  run the  extended diagnostics  located under the System menu  to identify any problems   If your logic analyzer powers on so that you have access to the desktop  run the  CheckIt Utilities software to identify possible controller hardware problems     You can also run the external TLA mainframe diagnostics to identify problems  not covered by other diagnostics  The TLA mainframe diagnostics are located  under the Start menu under the Tektronix Logic Analyzer programs     Version 5 6 or higher of the TLA application software allows you to connect the  logic analyzer to a network where you can control the instrument from a PC   Verify that the PC and the logic analyzer are connected to the network  The TLA  application software must be installed on the PC and the logic analyzer  The TLA  server must be running on the logic analyzer     Complete the following steps to start the TLA server     1  On the logic analyzer  click Start  gt  Programs  gt  Tektronix Logic Analyzer  gt   TLA Server  An icon with a red circle appears in the toolbar     2  Right click the TLA Server icon and select Start TLA Server  The red circle  disappears indicating the server is running     Check with your system administrator to address other network problems     Use the following table to isolate 
29. o the following specifications as listed in the  Official Journal of the European Communities     EN 61326 1 2006  EN 61326 2 1 2006  EMC requirements for electrical equipment  for measurement  control  and laboratory use    23    CISPR 11 2003  Radiated and conducted emissions  Group 1  Class A  IEC 61000 4 2 2001  Electrostatic discharge immunity   IEC 61000 4 3 2002  RF electromagnetic field immunity   IEC 61000 4 4 2004  Electrical fast transient burst immunity   IEC 61000 4 5 2001  Power line surge immunity   IEC 61000 4 6 2003  Conducted RF immunity   IEC 61000 4 11 2004  Voltage dips and interruptions immunity 4    EN 61000 3 2 2006  AC power line harmonic emissions    EN 61000 3 3 1995  Voltage changes  fluctuations  and flicker    European contact     Tektronix UK  Ltd   Western Peninsula  Western Road  Bracknell  RG12 1RF  United Kingdom    This product is intended for use in nonresidential areas only  Use in residential areas may cause electromagnetic  interference     Emissions which exceed the levels required by this standard may occur when this equipment is connected to a  test object     For compliance with the EMC standards listed here  high quality shielded interface cables should be used     Performance Criterion C applied at the 70  25 cycle Voltage Dip and the 0  250 cycle Voltage Interruption test  levels  IEC 61000 4 11      TLA6000 Series Logic Analyzers Installation Manual vii    Compliance Information    Australia   New Zealand  Declaration of  Confor
30. om    Possible causes and recommended action       Instrument turns on but does not  complete the power on sequence    Make sure that the instrument boots from the hard disk drive  Remove any discs from the  floppy disk drive     Possible software failure or corrupted hard disk drive   See page 23  Software Problems         Power on diagnostics fail    Instrument failure  contact your local Tektronix service center        Instrument does not recognize  accessories such as monitor   printer  or keyboard    Check that accessories are properly connected or installed  Try connecting other standard  PC accessories or contact your local Tektronix service center        Windows comes up but the TLA  application does not    Instrument not set up to start the TLA application at power on  Start application from the  desktop  by double clicking the TLA icon located on your desktop     Faulty or corrupt software  Reinstall the application software        Windows comes up in Safe mode    Exit the Safe mode and restart the instrument     Incompatible hardware and hardware driver software  Either install the hardware driver or  remove the incompatible hardware        Repacking for Shipment    If the logic analyzer is to be shipped to a Tektronix service center for repair  attach  a tag to the mainframe or module showing the owner   s name and address  the  serial number  and a description of the problem s  encountered and or service  required  Always include the probes so that the entire uni
31. opy this sheet and use it for your system installation        Installer        Date        Networking information  Share with customer IT department                    Item Description Value   1 IP address default DHCP   2 Operating System Service Pack   3 Computer name   4 Admin password   5 Firewall Microsoft  enabled   6 Virus Scanner None   7 Network Port Information TLA Application     port 111    TPI NET     port 9000  Windows XP Remote Desktop     port 3389       Network installation information  Provided by IT department     Fi  3    Description    Value       User Name       Default password       Domain name       DNS suffix       IP address  Fixed or DHCP        Security requirements       Computer naming convention  DNS address       O O  NOD  a  I wl rm        Proxy Server address          o    RJ45 outlet location activation       Other notes                          TLA6000 Series Logic Analyzers Installation Manual    29    Appendix C  Network Installation Site Survey    30 TLA6000 Series Logic Analyzers Installation Manual    Index    A    Accessories  27  connecting  3  optional  27  standard  27   Activity indicators  8    B    BIOS settings  changing  14    C    Calibration   service  21   Touchscreen  17  Chassis ground location  2  Checking the shipping list  1  Cleaning  22   LCD display  22  Connecting   accessories  3   probes  5  9  Connectors   external  11    D    Diagnostics  23  mainframe  23  documentation  xii    E    Extended diagnostics  23  
32. pollution or only dry  nonconductive pollution occurs   Products in this category are generally encapsulated  hermetically sealed  or  located in clean rooms     m Pollution Degree 2  Normally only dry  nonconductive pollution occurs   Occasionally a temporary conductivity that is caused by condensation must  be expected  This location is a typical office home environment  Temporary  condensation occurs only when the product is out of service     TLA6000 Series Logic Analyzers Installation Manual    Compliance Information      Pollution Degree 3  Conductive pollution  or dry  nonconductive pollution  that becomes conductive due to condensation  These are sheltered locations  where neither temperature nor humidity is controlled  The area is protected  from direct sunshine  rain  or direct wind       Pollution Degree 4  Pollution that generates persistent conductivity through  conductive dust  rain  or snow  Typical outdoor locations     Pollution Degree Pollution Degree 2  as defined in IEC 61010 1   Note  Rated for indoor use only     Installation  Overvoltage  Terminals on this product may have different installation  overvoltage  category  Category Descriptions designations  The installation categories are       Measurement Category IV  For measurements performed at the source of  low voltage installation     m Measurement Category III  For measurements performed in the building  installation       Measurement Category II  For measurements performed on circuits directly  co
33. problems  This list is not exhaustive  but it may  help you eliminate problems that are easy to fix  such as an open fuse     Table 4  Failure symptoms and possible causes    Symptom    Possible causes and recommended action       Instrument does not turn on    Verify that the power cord is connected to the instrument and to the power source     Check that the instrument receives power when you press the On Standby switch  Check that  fans start and that front panel indicators light     Check that power is available at the power source  Instrument failure  contact your local  Tektronix service center        External monitor does not turn on    Check the monitor power cord connection    Check for failed fuse  if one exists    Check external display connection    Monitor failure  contact the vendor of your monitor for corrective action        Monitor display is blank    Check that the monitor is connected to the mainframe  replace the cable if necessary     If instrument display is blank  try connecting an external monitor  if both displays are blank   contact your local Tektronix service center     External monitor controls turned down  adjust monitor controls for brightness and contrast   Check the controller BIOS setups for the monitor   Faulty monitor  contact the vendor of your monitor for corrective action        24    TLA6000 Series Logic Analyzers Installation Manual    Appendix A  User Service Procedures    Table 4  Failure symptoms and possible causes  cont      Sympt
34. products used by Tektronix for warranty  work may be new or reconditioned to like new performance  All replaced parts  modules and products become  the property of Tektronix     In order to obtain service under this warranty  Customer must notify Tektronix of the defect before the expiration of  the warranty period and make suitable arrangements for the performance of service  Customer shall be responsible  for packaging and shipping the defective product to the service center designated by Tektronix  with shipping  charges prepaid  Tektronix shall pay for the return of the product to Customer if the shipment is to a location within  the country in which the Tektronix service center is located  Customer shall be responsible for paying all shipping  charges  duties  taxes  and any other charges for products returned to any other locations     This warranty shall not apply to any defect  failure or damage caused by improper use or improper or inadequate  maintenance and care  Tektronix shall not be obligated to furnish service under this warranty a  to repair damage  resulting from attempts by personnel other than Tektronix representatives to install  repair or service the product   b  to repair damage resulting from improper use or connection to incompatible equipment  c  to repair any damage  or malfunction caused by the use of non Tektronix supplies  or d  to service a product that has been modified or  integrated with other products when the effect of such modification or i
35. rd  it must remain accessible to the user at all times     Do not operate without covers  Do not operate this product with covers or panels  removed     Do not operate with suspected failures  If you suspect that there is damage to this  product  have it inspected by qualified service personnel     Avoid exposed circuitry  Do not touch exposed connections and components when  power is present     Use proper fuse  Use only the fuse type and rating specified for this product     TLA6000 Series Logic Analyzers Installation Manual    General Safety Summary    Do not operate in wet damp conditions   Do not operate in an explosive atmosphere   Keep product surfaces clean and dry     Provide proper ventilation  Refer to the manual   s installation instructions for  details on installing the product so it has proper ventilation     Terms in This Manual These terms may appear in this manual        WARNING  Warning statements identify conditions or practices that could result  in injury or loss of life           CAUTION  Caution statements identify conditions or practices that could result in  damage to this product or other property        Symbols and Terms on the These terms may appear on the product   Product      DANGER indicates an injury hazard immediately accessible as you read  the marking       WARNING indicates an injury hazard not immediately accessible as you  read the marking       CAUTION indicates a hazard to property including the product     The following symbol s  may
36. s for the best accuracy     You can also customize some of the functions of your touchscreen in the Touch  Screen Properties window  Right click the selection text to view a description of  the function     TLA6000 Series Logic Analyzers Installation Manual 17    Restoring and Reinstalling Software    Install Other Software    If you purchased any microprocessor support packages  you need to reinstall  them after you reinstall the operating system and TLA application  If you do not  have a copy of the microprocessor support software  contract your Tektronix  Account Manager to order a replacement copy  If you cannot contact the account  manager  contact the Tektronix Support Center  refer to Contacting Tektronix at  the beginning of this document      Upgrading or Restoring Firmware    You may have to upgrade the firmware on the logic analyzer if the firmware  version is not compatible with the TLA application software version  This will  be indicated by one or more messages when you start the logic analyzer  The  most current firmware for the TLA application software resides in a file on the  hard disk of the logic analyzer     Use the following procedure to upgrade the firmware   1  Exit the logic analyzer application     2  Click Start  gt  All Programs  gt  Tektronix Logic Analyzer  gt  TLA Firmware  Loader     3  If the Connection dialog box appears  select the instrument that you want to  upgrade  typically the Local instrument  and click Connect     4  The Firmware Loa
37. successfully     1   2   3     Log on to the instrument as Administrator and quit any applications   Install the TLA Application Software CD in the CD drive of the logic analyzer     If the installation software does not start automatically  click Run in   the Windows Start menu to display the Run dialog box  Enter DATLA  Application SW Setup exe in the Run dialog box  if your CD drive is not the  D drive  enter the appropriate letter for your drive      Click OK to perform the installation     If you have an older version of the software on the hard disk  the installation  program will detect it and ask to remove it  Follow the on screen instructions  to remove the software  answering  Yes  to any prompts     After the software is successfully installed  restart the instrument     Installing the TLA Application Software on a PC    Install the TLA application software on a PC for the following purposes       Run the TLA application software in the Offline mode       Control a logic analyzer with the same TLA Application Software over a    network     TLA6000 Series Logic Analyzers Installation Manual 15    Restoring and Reinstalling Software    Third Party Software    Install the  NI GPIB Software    Install the Snaglt Software    Install the TLA Application    Software on a PC    To use all of the features of the TLA application software on your PC  you must  install additional third party software  The following third party software is  available on the TLA Application Soft
38. t can be tested     When packing an instrument for shipment  use the original packaging  If it is  unavailable or not fit for use  contact your Tektronix representative to obtain  new packaging     TLA6000 Series Logic Analyzers Installation Manual 25    Appendix A  User Service Procedures    26 TLA6000 Series Logic Analyzers Installation Manual    Appendix B  Accessories and Options                                                                            Accessories  Table 5  Standard accessories  Quantity Accessory Part number  1 ea Power cord clip 343 1213 00  1 ea Mini keyboard 119 7275 00  1 ea USB optical wheeled mouse 119 7054 00  1ea Front panel cover 200 4939 00  Product CD  TLA application software V5 7  Nero CD and license  NIST  2540 1  and ISO9000 Calibration Certificate  Literature package  Table 6  Optional accessories  Quantity Accessory Part number  1ea Removable HDD assembly  80 GB  w o SW 650 4805 xx  1ea Wheeled transit case 016 1522 00  1ea Rackmount kit 020 2664 00  Options  Option Description  18 Add touchscreen  1C Add iView External Oscilloscope Cable kit  2C Add USB iView External Oscilloscope Cable kit  1K Add LACART Instrument Cart  AM Enable Analog Mux  PO Add accessory pouch  TL Add TekLink cable  1S Increase to 8 Mb record length  235 MHz state speed  28 Increase to 32 Mb record length  235 MHz state speed  3S Increase to 128 Mb record length  235 MHz state speed  1P Add general purpose P6810 logic analyzer probes  2P Add Mictor probes  3P Add
39. tem from the restore discs  you might need  to change the BIOS settings to enable starting the logic analyzer from the DVD  drive  You might also need to change the BIOS settings if you want to change the  video mode to exclude the external display     1  Restart the logic analyzer and then press the Delete key repeatedly to enter  the BIOS setup     2  Inthe BIOS setup  go to the Boot menu     3  Use the on screen instructions to change the BIOS settings  set the CD DVD  as the first boot device     4  Save the settings by pressing function key F10 and confirm that you want to  save the new settings     TLA6000 Series Logic Analyzers Installation Manual    Restoring and Reinstalling Software    Reinstalling the TLA Application Software    Complete the following steps to reinstall the latest version of the TLA application  software  Use these steps as a first resort to recovering from application software  problems  If you still experience problems you may have to restore the hard disk  image   See page 13  Restoring the Instrument Operating System      While using this procedure you will be asked to log on as Administrator  The  logic analyzer is initially set up to automatically log on as Administrator  with no  password  so you may not see the login prompt  If the network setups have been  changed on your instrument  make sure that you log on as Administrator or as a  user who has administrator privileges  Failure to do so can prevent the software  upgrade from completing 
40. tive maintenance  22  warranty repair service  21  Shipping  25  Shipping list  checking  1  Site considerations  1  Snaglt software  16    32    Software   firewall  xi   installing  13   installing support   packages  18   mainframe diagnostics  23   NI GPIB USB  16   problems  23   Snaglt  16   third party  16   TLA application software  15  Standard accessories  27  Support packages   installing  18    y    Third party software  16   Threshold indicators  8   TLA application software  installing  15    TLA6000 Series   description  xi  Touchscreen   calibrating  17  Trigger   system  external signal  11  Troubleshooting  23   common problems  24    U    Upgrading firmware  18   User files  backups  8   User service  general maintenance  21  preventive maintenance  22    W    Warranty repair service  21    TLA6000 Series Logic Analyzers Installation Manual    
41. ue to environmental considerations  Please  contact your local authorities or  within the United States  refer to the E cycling  Central Web page  www eiae org  for disposal or recycling information     Perchlorate materials  This product contains one or more type CR lithium  batteries  According to the state of California  CR lithium batteries are  classified as perchlorate materials and require special handling  See  www dtsc ca gov hazardouswaste perchlorate for additional information     Restriction of Hazardous This product is classified as Monitoring and Control equipment  and is outside the  Substances Scope of the 2002 95 EC RoHS Directive     x TLA6000 Series Logic Analyzers Installation Manual    Preface    This manual contains information needed to install your Tektronix logic analyzer  and related accessories     To prevent personal injury or damage  consider the following requirements before  starting service     m  The procedures in this manual should be performed only by qualified service  personnel     E Read the General Safety Summary and Service Safety Summary found at  the beginning of this manual     Be sure to follow all warnings  cautions  and notes in this manual     TLA6000 Series Logic Analyzers    The TLA6000 Series Logic Analyzers are stand alone logic analyzers consisting  of the following instruments     Mm TLA6202 68 channel logic analyzer  with 8 GHz MagniVu timing  235 MHz  state clock  2 Mb record length    m TLA6203 102 channel logic analyzer
42. und Connections    Use the chassis ground connections to connect the grounds of the target system   system under test  to the logic analyzer to verify a common ground connection  between instruments   See Figure 1         CAUTION  To reduce the risk of ground loop noise  ground all of the instruments  in the system to the logic analyzer using the ground connections shown                                                                                                                                                                                                                                                                                                                         Owes CE  e        Xe o       o  OJ        o  TekLink         ol    Chassis ground   2771005    Figure 1  Location of the ground connection on the logic analyzer    2 TLA6000 Series Logic Analyzers Installation Manual    Basic Installation    Connecting Accessories    After installing the logic analyzer in the desired location  connect the accessories  such as an external monitor  keyboard  and printer   See Figure 2                                                                                                                           DVI I LAN  O   O    aa   UNE IN    ce   Ton  P  ie  X      _    KO O  a e      oA               VGA   USB   TL   LU  O  EXTERNAL SIG OUT             EXTERNAL   E  SIGNAL IN   TL                  EXTERNAL SIG IN                          lt           Activity light         
43. ve an external monitor  connect the power cord and turn on the    monitor   3  Press the On Standby switch to turn on the logic analyzer   See Figure 4    4  Wait for the logic analyzer to complete the power on self tests     TLA6000 Series Logic Analyzers Installation Manual 5    Basic Installation                                                                                     RISE             2771 06    Figure 4  On Standby switch location    Turning off the Logic To power off the logic analyzer  press the On Standby switch  The logic analyzer  Analyzer has a built in soft power off function that safely powers off the logic analyzer  when you press the On Standby switch        WARNING  To avoid the chance of electrical shock when opening the instrument  or accessing any location that may have a voltage applied  always unplug the  instrument to disconnect it from the mains power  Do not unplug the instrument  when it is powered on otherwise you can damage the instrument     Creating Operating System Restore Discs    The instrument does not ship with an operating system restore disc  Use the  following procedure to create a set of discs that enable you to restore the operating  system if the need arises        CAUTION  Restoring the operating system reformats the hard disk drive  All  saved data  including the operating system restore file on the hard disk will be  lost  Create the operating system restore discs before attempting to restore the  operating system          
44. vice under this warranty  Customer must notify Tektronix of the defect before the expiration  of the warranty period  If Tektronix is unable to provide a replacement that is free from defects in materials and  workmanship within a reasonable time thereafter  Customer may terminate the license for this software product  and return this software product and any associated materials for credit or refund     THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY  OTHER WARRANTIES  EXPRESS OR IMPLIED  TEKTRONIX AND ITS VENDORS DISCLAIM ANY  IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE   TEKTRONIX    RESPONSIBILITY TO REPLACE DEFECTIVE MEDIA OR REFUND CUSTOMER   S  PAYMENT IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH  OF THIS WARRANTY  TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT   SPECIAL  INCIDENTAL  OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX  OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES      W9b     15AUG04     Table of Contents    General Safety Summary rari RO n iv  Service Safety SUMMALY rr A RR RI vi  Compliance InfOrmMati Mii ea eee vii  EMC    ComplianGe    O RON iia vii  Safety Compliance               Li viii  Environmental Considerations                     0 bi ix  Preface AAA E ETO EE ARA EEA e xi  TLA6000 Series Logic Analyzers 1 0 0    0 0    cece cece ence een eee en nee ne cnn narran rr een nee eaten bettas xi  Documentation visiti i
45. ware CD V5 6  or higher     m  The NI GPIB USB software allows you to use the   View software with your  PC        The Snaglt software is useful for copying screen shots of logic analyzer data  for use with other applications        NOTE  Several third party software applications are included to use with your  logic analyzer  These applications may include software license agreements  Be  sure to abide by those license agreements        Complete the following steps to install the third party software on your PC   1  Install the first TLA Application Software CD in the CD drive   2  Browse to the NI GPIB USB folder on the CD and run the Setup exe program   3  Follow the on screen instructions and note the items below     When prompted  select the Typical installation option     When the Add GPIB wizard appears  select GPIB USB B       After restarting the instrument  the NI 488 2 Getting Started wizard  displays  This is not needed  select Do not show at Windows startup       When you first connect the 1View cable  the instrument will detect the new  hardware  Select Install the software automatically  Recommended      Complete the following steps to install the SnagIt software on your PC   1  Browse to the SnagIt folder on the CD and run the Setup exe program     2  Follow the on screen instructions     Browse to the TLA Application software folder on the CD and run the Setup exe  program  Follow the on screen instructions to install the TLA application  software     If you h
46. wt  FREE                                                                                                                eta     O   aasaga 815  O O  SERIAL 2 IDI vea    OJA O  D Ai fee     oF     I VGA   USB  cerro      sm   TL   I     EXTERNAL SIG OUT                    BIGNAN            lt     EXTERNAL SIG IN    OL Activity light                                     TekLink                            LL EXTERNAL TRIG OUT       lt  lt  EXTERNAL TRIG IN    4 2771 002                                                                                            af                                                                   Figure 7  External connectors    TLA6000 Series Logic Analyzers Installation Manual    Restoring and Reinstalling Software    Most of the software is installed when your instrument is shipped from the factory   Refer to this section if you need to reinstall your software  These instructions  refer only to reinstalling the TLA application software and operating system  To  upgrade to a newer TLA application software version  contact your local Tektronix  representative  This section also provides information on installing related logic  analyzer software on a PC for remote operation or for offline applications        NOTE  Ifyou install or reinstall software on a remote PC  make sure that the  software version matches that of the main application on the logic analyzer        Restoring the Instrument Operating System    You can restore the instrument operatin
47. y Wipes   05701  available from  Kimberly Clark Corporation      If the display is very dirty  moisten the wipe with distilled water or a 75   isopropyl alcohol solution and gently rub the display surface  Avoid using excess  force or you may damage the plastic display surface     Clean the exterior surfaces with a dry  lint free cloth or a soft bristle brush  If dirt  remains  use a cloth or swab dampened with a 75  isopropyl alcohol solution  A  swab is useful for cleaning in narrow spaces around the controls and connectors   Do not use abrasive compounds on any part of the instrument        CAUTION  To avoid damaging the instrument follow these precautions        TLA6000 Series Logic Analyzers Installation Manual    Appendix A  User Service Procedures    Avoid getting moisture inside the instrument during external cleaning and use  only enough solution to dampen the cloth or swab     Do not wash the front panel On Standby switch  Cover the switch while washing  the instrument     Use only deionized water when cleaning  Use a 75  isopropyl alcohol solution as  a cleanser and rinse with deionized water     Do not use chemical cleaning agents  they may damage the instrument  Avoid  chemicals that contain benzene  toluene  xylene  acetone  or similar solvents     In Case of Problems    Review the following sections if you encounter any problems with your  instrument  For additional help  contact your local Tektronix Service  Representative     Diagnostics The logic analyzer ru
48. zers Installation Manual 21    Appendix A  User Service Procedures    Preventive Maintenance    22    Cleaning the Flat Panel  Display    A    Exterior Surfaces    A    Once a year the electrical performance should be checked and the instrument  accuracy should be certified  calibrated   This service must be performed by  a qualified service technician using the procedures outlined in the appropriate  service manual for the Tektronix Logic Analyzer product     Preventive maintenance mainly consists of periodic cleaning  Periodic cleaning  reduces instrument breakdown and increases reliability  Clean the instrument as  needed  based on the operating environment  Dirty conditions may require more  frequent cleaning than computer room conditions     Refer to your logic analyzer probe instructions for specific cleaning procedures  for the probes and probe connectors     The LCD flat panel is a soft plastic display and must be treated with care during  cleaning        CAUTION  Improper cleaning agents or methods can damage the flat panel  display        Do not use abrasive cleaners or commercial glass cleaners to clean the display  surface     Do not spray liquids directly on the display surface   Do not scrub the display with excessive force     Avoid getting moisture inside the instrument while cleaning the display  use only  enough solution to dampen the wipe     Clean the flat panel display surface by gently rubbing the display with a  cleanroom wipe  such as Wypall Medium Dut
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Sony WM-FX561 User's Manual  Ug Manuel d`utilisation  mANeTTe  AEG 54710 Dishwasher User Manual  Dossier Que faire avec un doctorat en chimie  Revolution Z-7789  OneTouch™ AT - Iris Networks  取扱説明書  Solaris 9 4/03 Release Notes Supplement for Sun Hardware  User Manual Milli-Q Direct 8/16 System    Copyright © All rights reserved. 