1. Resetting Turn the button in the direction of the arrows until the button pops out Emergency lowering Manual emergency lowering For manual emergency turn the round plastic knob on the actuator clockwise Electrical emergency lowering For electrical emergency use the down button on the control box Trouble shooting If the lift is not working properly up or down check the following The emergency stop device is not pressed in All cables are properly connected The battery charges via the adaptor If the lift is not working properly contact RoMedic The lift can be raised but not lowered or vice versa or the base width adjustment only works in one direction Check the following That all cables are properly and securely connected Pull out the contact and plug it in again firmly If the lift is not working properly contact RoMedic If the lift makes unusual noises Try to locate the source of the sound take the lift out of operation and contact RoMedic Eva450EE SystemRoMedic Safe working load Different products on the assembled lift system ie lift slingbar sling scale and other accesories may have different max loads It is always the lowest maximum allowable load indicated for the respective products on the system that applies for the entire system Always check the markings on all products in the unit and contact RoMedic with any questions Other maintenance details
2. The lift must undergo thorough inspection at least once per year This must be done by authorized RoMedic service personnel and in accordance with RoMedic s service manual Repairs and maintenance may only be done by authorized personnel using original spare parts X Old batteries are to be left at the nearest recycling station or given to personnel authorized by RoMedic Cleaning disinfection When necessary clean the lift with warm water or rubbing alcohol and ensure that the castors are free of dirt and hair Do not use cleaning agents containing phenol or chlorine as this may damage aluminum and polyamide materials Storage If the lift is not to be used for some time or e g when transporting we recommend the emergency stop button to be pushed in Store the lift at a temperature exceeding freezing and at normal relative humidity about 60 Service agreements RoMedic offers the possibility of service agreements for maintenance and regular testing of your Eva lift Contact your local RoMedic representative 10 Eva450EE SystemRoMedic Accessories Slings The RoMedic Slings are available in different fabrics and several sling models The choice of model and material is determined by the patient s functional capacity and the situations in which the product will be used For slings and other lifting accessories see RoMedic s user manual for slings Wy Slingbars SlingBar S 13 8 inch 350mm SlingBar M 17 7 inch 450m
3. mast Secure the mast with the locking handle Place the battery in the mount in the control box Mount handlebar with two wrenches Connect the cables The cable for the handset in outlet HS the cable for the hoist motor in outlet M1 and the cable for the base motor in outlet M2 Release the emergency stop and perform a final inspection see final inspection HS Mt M2 Eva450EE SystemRoMedic Final inspection Inspect the lift for signs of wear and damage Check all four castors and castor brakes Check all connections and fixtures including screws and bolts Check the packaging to ensure that there are no loose parts Check emergency stop function by depressing the emergency stop and then pressing either the up or down button If nothing happens when the up or down buttons are pressed the emergency stop is functioning properly Grasp the hand control press the up button and run the lift arm all the way up Then press the down button and run the lift all the way down Test base width adjustment function by pressing the button for base width adjustment Widen the base to max width and then press the other button to narrow it again Test lift function by lifting a person not a user using an approved sling At the same time check the emergency lowering function with someone in the lift see section on Emergency lowering If the equipment is satisfactory upon inspection connec
4. CE Manual Oe Eva450EE SystemRoMedic Boom Mast Handlebar Battery pack Emergency stop Control box Motor for base width adjustment Rear castors with brakes SlingBar with safety latch 10 Front castors 11 Emergency lowering 12 Motor actuator for boom 13 Mast height adjustment Sas oS eS oN Max load 205 kg 450lbs Mobile lift Eva450EE has been developed to meet most lifting needs in combination with the right accessories It is a lift that can lift patients in a seated or supine position The Eva 450EE is made largely of aluminium which makes it relatively light considering the weight it can lift RoMedic s lifts and slings are part of the SystemRoMedic concept which takes a holistic approach to patient transfers SystemRoMedic has four categories transfer repositioning Support and lifting e Functional inspection Visual inspection Inspect lift functions regularly Check to ensure that material is free from damage Ci Always read the instructions Read instructions for all assistive devices used in connection with transfers Keep the manual where it is accessible to users of the product Do not leave the patient unattended during a lifting situation Before use Make certain the lift is properly assembled Check Slingbar connection and safety latch function Check lifting function and base width adjustment Under no circumstances may the lift be used by persons who have not received instru
5. ction in the operation of the lift E Manual nr 00789En Ver 1 091201 RoMedic A company in the Handicare Group Eva450EE SystemRoMedic Table of contents Technical information Important information Assembly Final inspection Dimensions Charging batteries Handset Emergency stop emergency lowering Trouble shooting Safe working load Other maintenance details Accessories 6 7 10 10 11 Eva450EE SystemRoMedic Technical information Lifting speed Batteries Charger Motor mast Motor base Sound level Material Emergency lowering Castors Weight IP class Expected lifetime Operating forces controls 34 mm s without load Two 12V 2 9 Ah valve regulated sealed lead accumulator gel type batteries Max 400mA DC 24 V 10 A IP X4 Operationtime 10 at maximum continous running of 2 minutes maximum 5 switching cycles per minute Push 10 OOON 24 V 3A IP X4 Operationtime 10 at maximum continous operation of 2 minutes maximum 5 cycles per minute Push SOOON With load upwards 74 7 dB A downwards 52 6 dB A Operations force for the handset 4 N Aluminum Mechanical and electrical Front 4 100 mm back 4 100mm 75 Ibs 34 kg IP X4 10 years Buttons on hand control 4 N The lift have been tested by an accredited testing institute and complies with the requirements of Council Directive 93 42 EEC of 14 June 1993 concerning me
6. dical devices A The device is intended for indoor use Type B according to the degree of protection against electric shock A Important Information e The lift must be assembled according to the assembly instructions provided by RoMedic e The lift may only be used on a level floor e Lifting accessories must be properly trial fitted and tested in relation to the user s needs and functional ability e Under no circumstances may the safe working load be exceeded See section on Technical information and the prod uct label on the lift e Never move the lift by pulling on the actuator Do not push e The lift must not be exposed to running water e Batteries control unit and actuators on the lift must not come in direct contact with water e Eva400EE must not be charged in a wet room or used in connection with showering and bathing e To ensure optimal function the lift must be inspected regularly See section on other maintenance details e Warranty applies only if repairs or alterations are made to the product by personnel who are authorized by RoMedic Eva450EE SystemRoMedic Assembly Check to ensure that all components are included Mast with boom lift motor carry bar control box and battery pack Base with motor and locking handle Handset and cord Instructions charging cable and power supply cord for charging Place the mast in the base There are three alternatives for adjusting the height of the
7. m G y SlingBar L 23 6 inch 600mm StretcherBar and StretcherSling for lifting in the horizontal position AmbulationArm armrest for gait training Scale for weighing patients who use patient lifts SwiftHook for SlingBar Extra batteries Positioning aids TopSheet SafeHandlingSheet RoMedicClips 11 SystemRoMedic We love easy transfers RoMedic s lifts and lifting accessories are part of SystemRoMedic This means they are part of a holistic approach to lift ing and transferring SystemRoMedic includes four transfer segments transfer repositioning Support and lifting
8. ngbar at maximum reach legs spread U is minimum distance from wall to slingbar at minimum height legs spread Eva450EE SystemRoMedic Charging batteries Charge the lift after use to ensure that the battery is always fully charged Lock the castor wheels when charging the battery 1 Connect the charging cable to a power outlet 2 Check to ensure that the lamps on the control box light up The green LED lamp indicates that the charger is receiving power and the yellow LED lamp indicates that the battery is charging 3 Charging stops automatically when the battery is fully charged Wall mounted charger 1 Remove the battery pack from the hoist and place it in the wall mounted charger 2 Check to ensure that the LED lamp on the front of the charger lights up NOTE Before the lift is used for the first time it must be charged for at least 4 hours For maximum battery life charge batteries regularly We recommend daily charging when the lift is used daily The emergency stop button must be pulled out during charging Handset Manoeuvring Raising lower the boom Arrows indicate direction of travel Motion stops as soon as the button is released Widening narrowing the base Markings on the buttons indicate function Motion stops as soon as the buttons are released Eva450EE SystemRoMedic Emergency stop To activate emergency stop Depress the red emergency stop button on the control box
9. t the charger to the lift and to a power outlet and then check to ensure that the charging lamp on the control box lights up NOTE Before the lift is used for the first time it must be charged for at least 4 hours See section on Charging batteries Keep the manual where it is accessible to users of the product Eva450EE SystemRoMedic Dimensions Eva450EE Pat 58 186 GAS D A2 63 191 A3 68 196 B1 50 169 B2 54 174 G B3 59 179 i C 130 B D 8 E 11 5 F 6 Gt 493 192 G2 138 197 G3 143 202 H 55 Column A measurement is in cm and weight is in kg Column B measurement is in inches and weight is in Ibs B 22 85 73 28 24 82 75 25 26 79 77 22 19 70 66 59 21 28 68 56 23 25 70 53 51 22 34 67 4 53 2 36 52 4 75 65 54 37 77 62 56 34 79 59 21 67 Dimensions Eva450EE a Column A measurement is in cm and weight is in kg Column B measurement is in inches and weight is in Ibs A 44 45 134 34 14 5 27 41 292 45 Calal to ee Tr S M is turning diameter N is total weight of lift 58 5 88 68 5 98 Eva450EE SystemRoMedic B 17 34 17 73 23 05 34 67 26 99 38 61 54 74 9 31 9 10 64 16 1 8 7 172 0 is the weight of the heaviest component P is movement in forward direction R is referencemeasure 70 cm with max legspreading S is minimum distance from wall to slingbar at maximum height legs spread T is minimum distance from wall to sli

Download Pdf Manuals

image

Related Search

Related Contents

Global Direct 21147 Installation Guide    VOLTAGE DANGEREUX - Insect-O  Mouvement d`un solide, exploitation d`un document vidéo  Digital Media Adapter DMA-100  住まいるサポ[E型]2:2  Channel Vision CVT-15WB TV signal amplifier  KitchenAid KDRP463LSS0 User's Manual  Mapelastic HPG Mapelastic HPG  LP Gas Grill OPERATION INSTRUCTIONS  