AirAssure Indoor PM2.5 Particle Counter
Contents
1. Configuration Static IP Address Static IP Mask Static IP Gateway Use DHCP vf ESCH IP Address 10 1 na p 114 i Server IP Port 8080 EE Factor 1 00 Calibration Coefficients 1670 0 441 0 21 000 y 0 888 Apply Changes Network Ethernet Setup 7 15 7 16 9 To reset the Server IP address the monitor has to be power cycled Click the Apply Changes button in the webpage to reboot the microprocessor CN ES http 10 1 3 117 locComm cai P BCx a Location Comm Configuration Static IP Address Static IP Mask Static IP Gateway Use DHCP F Server IP Address 10 1 11 114 Server IP Port 8080 Correction Factor 1 00 Calibration Coefficients 1670 0 441 0 21 000 0 888 Submit Changes Apply Changes There is a Reset push button on the circuit board of the unit to the right of the display This can be used to reset the unit without removing it from the wall in cases where there is no access to the web page or to AirAssure Configuration Manager 10 Repeat steps 1 to 9 above for each of the monitors configured AirAssure Indoor PM2 5 Mass Concentration Monitor 11 Allow five 5 minutes for each sensor to show up if it has just been configured Re open the AirAssure Web Application or refresh it if it is already open Each sensor should now appear in the Instrument column of the AirAssure Web page TSI UNDERSTANDING ACCELERATED LOGIN 7 V ifejjufa W Instruments Instruments 25 50
2. Setup will install Apache Tomcat in the following folder To install in a different folder dick Browse and select another folder Click Install to start the installation C Program Files Apache Software Foundation Tomcat 7 0 Space required 7 3MB Space available 98 466 Nullsoft Install System v2 46 Network Ethernet Setup 1 1 9 Finalize setup by selecting Run Apache Tomcat and click Finish Completing the Apache Tomcat Setup Wizard Apache Tomcat has been installed on your computer Click Finish te dose this wizard Run Apache Tomcat Show Readme http tomcat apache org Apache Tomcat 7 cancel Service Manager is attempting to start the following service Apache Tomcat 7 0 Tomcat 11 In the toolbar double click on the Tomcat server icon to open the configuration dialog box Alternately you can type configure tomcat in the search box and run the configuration 4 26 PM 12 7 2014 z P AirAssure Indoor PM2 5 Mass Concentration Monitor 12 Change the startup type from manual to automatic so that Tomcat will start up automatically if the computer is power cycled Click OK General Log On Legging Java _ Startup Shutdown Service Name Tomcat Display name Apache Tomcat 7 0 Tomcat Description Apache Tomcat 7 0 52 Server http tomcat apache Path to executable C Program Files Apache Software Foundation
3. Ifthe unit is reporting an error then the color code will be gray and the reading will indicate the error code which appears as a negative number Instruments 25 50 All PM 2 5 ug m3 Correction Factor Reading Date SN11 CONTROL LZ 1 000 Feb 13 2015 4 23 10 PM Beijing Office S N 15 1 000 Feb 13 2015 4 22 26 PM Beijing Office S N 04 Pgo 1 000 Feb 13 2015 4 23 03 PM Breakroom EE Gel 1 000 Feb 13 2015 4 18 13 PM IPM251503030 a ee 1 000 Feb 13 2015 4 19 38 PM IPM2515N3N39 SS es 4 nnn Eeh 13 2015 4 27 27 PM See the Troubleshooting section of this manual for details and suggested actions regarding specific error conditions Single Unit Configuration Administrator An administrator can set up individual units from the single unit configuration screen Configuration changes can include changing the device name or location correction factor and calibration coefficients Follow these directions to configure individual devices 1 Login as the administrator using the administrator username and password defined during setup of Apache Tomcat Chapter 7 LOGIN KH b UNDERSTANDING ACCELERATED T m Ai em A l A a_i l V A i f d Instruments Username Password LOGIN Copyright 2015 by TSI Inc All rights reserved 8 6 AirAssure Indoor PM2 5 Mass Concentration Monitor 2 From the main page click on a specific instrument location to display monitor details including location IP address
4. These ae SPM251447123 E coefficients must be 1582 811 28 5 0 888 programmed into the monitor after powering back on The monitor is ready for reinstallation Re connect all cable connections Slide unit back onto wall Tighten mounting screws AirAssure Indoor PM2 5 Mass Concentration Monitor 12 Place cover on by placing cover over the display and snapping the bottom of the cover into the base 13 After powering on confirm a successful installation Listen for three 3 chirps of the pump and verify that no error codes are shown on the display within 30 seconds of power up 14 Verify that the IP address displays on the monitor s screen An IP address will not appear on the screen of the monitor if an active Ethernet cable is not connected to the device Please refer to instructions in Chapter 6 Analog Output for configuring a monitor without Ethernet Maintenance and Troubleshooting 9 5 15 On a computer connected to the same network open a web browser and enter the IP address of the monitor as it appears on the display screen of the device 16 A password screen will open like the one below Enter the serial number as the password from the label on the back panel or table e g IPM251501002 and click the Login button AirAssure x d Login e D 192 168 11 2 login cgi Password 17 From step 9 enter the four calibration coefficients recorded f
5. Tomcat 7 0 bin Tomcat Startup type Service Status Started stop Apache Tomcat Application Manager Interface The web application for the AirAssure monitor is packaged as a Web Application Archive war file Apache Tomcat has an application manager interface to upload and control web applications The installation medium folder contains the web application named pmweb war To Install the Web Application 1 Open a web browser and navigate to hitp localhost 8080 manager or http IPADDRESS 8080 manager where IPADDRESS is the IP address of the server 2 Enter the username and password recorded in step 6 above from installing Tomcat and click OK Windows Security The server localhost at Tomeat Manager Application requires a username and password Warning This server requesting that your username and password be sent in an insecure manner basic authentication without a secure connection B admin EI Remember my credentials Remember my credentials Network Ethernet Setup 7 9 7 10 3 Under WAR file to deploy click Browse and select pmweb war from the installation medium 4 Click Deploy a a CEA Ee P BEX E manager nki v 3 Search gt Hi Share More gt gt Signin A Context Path required XML Configuration file URL WAR or Directory URL Apache Tomcat 7 0 52 1 8 0_20 b26 Oracle Corporation
6. correction factor value and most recent readings Instruments Configuration Instruments 25 50 All Executive Suite Jan 26 2015 3 27 21 PM Lower Level inactive Weg GE Jan 23 2015 2 53 32 PM Instruments Configuration Serial 07 Status active Executive Suite 10 1 3 119 1 100 1506 0 696 0 25 000 0 888 Details 1 100 Jan 26 2015 3 39 37 PM E 1 100 Jan 26 2015 3 39 06 PM aS 1 100 Jan 26 2015 3 38 35 PM EE 1 100 Jan 26 2015 3 38 05 PM ae 1 100 Jan 26 2015 3 37 34 PM 3 Admin users can configure 4 individual devices on AirAssure Web monitor Shown below is the individual configuration screen Instruments Configuration Edit Instrument 07 Executive Suite 10 1 3 119 255 25 1 100 1506 0 696 0 25 Update Cancel Copyright 2014 2015 by TSI Inc All rights reserved Using the Web Application 8 7 Serial Serial number of the monitor Not editable Instrument Enter device location This will appear as the Location instrument name on the main page IP Address IP Address of the sensor Editable from the unit webpage but not editable from AirAssure Web Correction Factor Used to adjust the reading to agree witha known reference device For example if the meter is reading 100 g m and the actual concentration is known to be 110 ug m the correction factor can be changed from 1 000 to 1 100 to multiply all future readings by 1 1 Note that the correction fact
7. e Wall mount to drywall Type 86 gang box American single gang box Local service available Sensor and some key components are field serviceable A 1 A 2 Service Kits Ethernet connection Password protection capability Data export capability Supports up to 99 sensors Compatible with mobile devices Located on local server Display measurement on remote screen Output coordinated to Air Quality Index color rating Supports any operating system that can run Java Runtime 8 and above IPM2 5 Sens Calibrated replacement sensor element IPM2 5 Kit Replacement pump fan and filter AirAssure Indoor PM2 5 Mass Concentration Monitor TSE UNDERSTANDING ACCELERATED TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 874 2811 India Tel 91 80 67877200 UK Tel 44 149 4 459200 China Tel 86 10 8219 7688 France Tel 33 1 41 19 21 99 Singapore Tel 65 6595 6388 Germany Tel 49 241 523030 P N 6008418 Rev C US 2015 TSI Incorporated Printed in U S A
8. please see the Troubleshooting section at the end of Chapter 9 AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 8 Using the Web Application The AirAssure monitor sends data readings to the web application once every five 5 minutes The application can support up to 99 devices at any given time The web application displays up to 24 hours of data points Historical data can be downloaded and imported into a spreadsheet program such as Microsoft Excel spreadsheet program for offline analysis AirAssure Web Main Page Functions The main page of AirAssure Web is located at http IRADDRESS 8080 pmweb where IPADDRESS is the IP address of the server The main page of AirAssure Web shows all AirAssure units set up on the network q p UNDERSTANDING ACCELERATED LOGIN ph i j ji if e V y CaF Ai Ss Instruments Instruments 25 50 All Instrument PM 2 5 ug m3 JF Home SE a Beijing Office SS Ss Shoreview MN Outdoor NACH pear Been Machine Shop SE GR Receiving Dock AE Lunchroom E esch 8 2 Language The webpage can be displayed using either English or Chinese LOGIN simplified Mandarin The language selection is made by clicking on the respective flag icons in the upper TS right hand corner of the web _ application Shown below is a web page in Chinese Note that Chinese characters can also be entered in the Instrument Location field in the single unit configuration view al
9. All Instrument PM 2 5 ugim3 JF Home eh ee Beijing Office ESS Shoreview MN Outdoor LSS SSS Front Lobby e Machine Shop SS Receiving Dock NL Lunchroom EE Sa 12 By logging in as administrator you can click on each instrument listed and enter a more meaningful description location based on your notes from the table in Chapter 3 More information on the AirAssure Web application is in the following chapter Network Ethernet Setup 7 17 13 If a phone or tablet is on the same network it can also access the web app with the same address TSE UNDERSTANDING ACCELERATED AirAssure Wlan Instruments Instruments Instrument JF Home Beijing Office Shoreview MN Outdoor Front Lobby Machine Shop Receiving Dock Lunchroom Sales Marketing Assembly Floor Engineering West Engineering East Arden Hills MN Outdoor kalen A 1 3080 pmweb LOGIN 25 50 All PM 2 5 ug m3 kel A ES pmweb tsi cc O Mo q gt UNDERSTANDING ACCELERATED LOGIN e AirAssure Watt ka Instruments Serial 4 Status active Instrument Location IP Address Correction Factor Beijing Office 10 120 1 110 1 000 Details PM 2 5 ug m3 Correction Factor Reading Date EE i000 Aug 6 2015 3 41 32 PM WE 1 000 Aug 6 2015 3 36 28 PM EE 1000 Aug 6 2015 3 31 25 PM MC 1 000 Aug 6 2015 3 26 22 PM eC 1000S Aug 6 2015 3 21 18 PM E l 1 NNN Auen R 20D1E 216 16 14 If the monitors do not show up on the web page
10. Incorrect polarity may damage unit Hardware Installation 3 5 3 6 8 Use a flathead screwdriver to separate the cover from the base The slot is located on the back of the unit along the bottom seam between the unit cover and base 10 Connect Ethernet if digital connection is desired 11 Connect 4 20 mA if analog connection is desired Wires are held in place by depressing spring clips inserting wires and releasing the spring clips AirAssure Indoor PM2 5 Mass Concentration Monitor 12 Slide the unit base downward on to the wall plate to fasten to wall plate hooks 13 Tighten the lower right mounting screw to secure the unit to the wall plate 14 When the unit is secure place the cover over the mounted display and snap the bottom of the cover to the base 15 The monitor is ready for use and powers on when the 24 VAC VDC connection is turned on 16 Self test When the unit is first powered up it performs a quick self check by turning the pump on to give three 3 quick chirps If an error occurs it will display on the front panel as a 1 2 or 3 within 30 seconds of powering up See the table in Chapter 9 under Troubleshooting if one of these codes appears Hardware Installation 3 7 Table 3 1 Reference Table for Sensor Configurations Serial number IP Address Description e g IPM251501002 e g 192 168 11 5 e g Accounting Floor 5 Northeast 3 8 AirAssure I
11. Instruments Configuration Serial 4 Status active Instrument Location IP Address Correction Factor Calibration Coefficients Firmware Version Action Beijing Office 10 120 1 110 1 000 1485 0 788 0 19 800 0 888 1 0 S Details PM 2 5 ug m3 Correction Factor Reading Date ee es 1 000 Aug 6 2015 3 51 39 PM NET 1 000 Aug 6 2015 3 46 35 PM eege Sn 1 000 Aug 6 2015 3 41 32 PM ee ical 1 000 Aug 6 2015 3 36 28 PM ERT Tee 1 000 Aug 6 2015 3 31 25 PM Eege nd 1 000 Aug 6 2015 3 26 22 PM Using the Web Application 8 9 3 Here global changes can be made to all sensors on the network and AirAssure Web data can be exported and cleared Global Correction Use this feature to set all units on AirAssure Factor Web to acommon correction factor Sensor Records Use this feature to set the number of sensor on Page records displayed on all individual unit pages Up to 24 hours of data 288 records can be displayed Update AirAssure Web with updates to the global correction factor and or sensor records on page Export AirAssure When Export AirAssure Web Data as csv is Web Data as csv clicked all AirAssure Web data will be exported in a csv format In this format data can be opened in a spreadsheet for offline analysis The file size may be large depending on the length of time AirAssure Web has been running and the number of units on your network Clear AirAssure To keep file sizes manageable it is Web Data recommended that
12. Reading Date Executive Suite o 1 100 Jan 26 2015 3 27 21 PM Lower Leva inactive 1 000 Jan 23 2015 2 53 32 PM Front kees D ae 1 100 Jan 26 2015 3 27 25 PM Engineering Lab 8 Sass 1 000 Jan 25 2015 3 28 33 PM Beijing Office S N 04 DA 1 100 Jan 26 2015 3 22 27 PM Beijing Office S N 09 DM 1 100 Jan 26 2015 3 27 19 PM Breakroom oo el 1 000 Jan 26 2015 11 25 10 AM Copyright 2014 2015 by TSI Inc All rights reserved 3 Click on a specific instrument location to display monitor details including location IP address correction factor value and most recent readings D Username admin gt UNDERSTANDING ACCELERATED LOGOUT Wane E Al w Al If y ar Instruments Configuration Serial 4 Status active instrument Location IP Address Correction Factor Calibration Coefficients Firmware Version Action Beijing Office 10 120 1 110 1 000 1485 0 788 0 19 800 0 888 1 0 f XR Details PM 2 5 ugim3 Correction Factor Reading Date 1 000 Aug 6 2015 3 51 39 PM 1 000 Aug 6 2015 3 46 35 PM 1 000 Aug 6 2015 3 41 32 PM 1 000 Aug 6 2015 3 36 28 PM 1 000 Aug 6 2015 3 31 25 PM 1 000 Aug 6 2015 3 26 22 PM Using the Web Application 8 5 AirAssure Web monitor is only updated with the most recent data when the webpage is reloaded Therefore to always view the most recent data the browser must be refreshed For automated display of real time data third party applications for auto refreshing web pages can be utilized 4
13. TSI Incorporated 500 Cardigan Road Shoreview MN 55126 USA 651 490 3824 For country specific terms and conditions outside of the USA please visit www tsi com Seller warrants the goods sold hereunder excluding software under normal use and service as described in the operator s manual to be free from defects in workmanship and material for twenty four 24 months or if less the length of time specified in the operator s manual from the date of shipment to the customer This warranty period is inclusive of any statutory warranty This limited warranty is subject to the following exclusions and exceptions a Hot wire or hot film sensors used with research anemometers and certain other components when indicated in specifications are warranted for 90 days from the date of shipment b Pumps are warranted for one year or 3000 hours whichever comes first c Parts repaired or replaced as a result of repair services are warranted to be free from defects in workmanship and material under normal use for 90 days from the date of shipment d Seller does not provide any warranty on finished goods manufactured by others or on any fuses batteries or other consumable materials Only the original manufacturer s warranty applies e This warranty does not cover calibration requirements and seller warrants only that the instrument or product is properly calibrated at the time of its manufacture Instruments returned for calibration are not
14. enable the Java installation in your browser Network Ethernet Setup 7 3 Install Apache Tomcat Application 7 4 Apache Tomcat Application Server Servlet Container is an open source software implementation of the Java Servlet and Java Server Pages technologies Apache Tomcat is commonly used for both development and production and is part of the reference platform for the AirAssure Web Application Installing Apache Tomcat Application It is recommended that all open applications are closed before starting setup to eliminate a computer reboot 1 From the installation package select Tomcat 7 0 xx right click and select Run as Administrator e Run as administrator BUDleshoot COMpatibility J Scan with Microsoft Forefront Endpoint Protection Express Zip Share with Pin to Taskbar Pin to Start Menu Restore previous versions D Send to Cut Copy Se ei gt Install Medium WY Create shortcut Organize v Open Share w Delete a N Rename Ce Favorites E E Desktop E apaci Properties mi Downloads E Recent Places m ow Libraries Documents a Music Pictures FE Videos Ap 641 Installing_CPM 2_5_Web_Application Sy jre 8u20 windows i586 Sy jre 8u20 windows x64 _ pmweb war 4 m Size 8 86 MB sade apache tomcat 7 0 52 Date modified 9 24 2014 1 33 PM VS Application A 10 6 2014 5 29 PM 9 24 2014 2 23 PM 9 24 2014 2 24 PM 10 10
15. or representative or call TSI at 1 800 874 2811 USA or 651 490 2811 TSI and the TSI logo are registered trademarks of TSI Incorporated AirAssure is a trademark of TSI Incorporated Microsoft Internet Explorer Excel and Windows are registered trademarks of Microsoft Corporation Java is a registered trademark of Oracle and or its affiliates Apache Tomcat and Tomcat are trademarks of the Apache Software Foundation Safety This chapter provides instructions to promote safe handling and correct operation of the AirAssure monitor All repair and maintenance of the AirAssure monitor should be done by qualified trained technicians All maintenance and repair information in this manual is included for use by a qualified trained technician The AirAssure monitor must be used in the manner described in this manual Failure to follow all of the procedures described in this manual can result in serious injury to you or can cause irrevocable damage to the instrument Description of Safety Labels This information explains the advisory and identification labels used on the instrument and in this manual to reinforce the safety features built into the instrument Caution Symbols Caution means be careful If you do not follow the procedures described in this manual you may damage the instrument Caution also indicates important information about the operation and maintenance of this instrument Warning Symbols Warning me
16. that Know are consistently much too low or too high compared to a reference device Monitor periodically makes a hissing noise Cause Display shows lt 5 or gt 300 Display shows 1 2 or 3 Display shows 2 frequently Monitor seems too high or low compared to a reference device Sensor has degraded over time Reading is too low Zero module pump has failed or filter is completely clogged Reading is too low Internal fan has failed Response time is very slow Reading is much too high Calibration is incorrect Reading is too high or low Normal operation of the zero module Maintenance and Troubleshooting Suggested Action This is normal lt 5 indicates a concentration of less than 5 ug m due to the limited resolution of the monitor Likewise gt 300 indicates a concentration greater than 300 g m This indicates an error condition 1 indicates a sensor element error Check for a damaged sensor or disconnected cable Replace if needed 2 indicates an auto zero error This could be caused by the pump not turning on filter not inserted or a leak around the filter If you have replaced the pump filter check that they are installed correctly If not replace the pump filter and cable assembly 3 indicates a fan error The fan may be faulty or disconnected Check or replace the fan If the 2 error occurs more than once per year it might indicate that the installation en
17. the setup file SSS 38 AirAssure Configuration Manager InstallShield Wizard Welcome to the InstallShield Wizard for _AirAssure Configuration Manager The InstallShield R Wizard will install AirAssure Configuration aseene your Computer To continue dick Next WARMING This program is protected by copyright law and international treaties Follow these steps to connect to monitor 1 Before opening AirAssure Configuration Manager you must connect USB To access the USB connector remove the cover of the monitor by pressing fingers firmly on bottom of device cover squeeze and pull outwards Analog Output 6 3 6 4 2 Connect the unit to a personal computer using a Micro USB cable not provided The USB connector is on the left hand side of the device Use caution when plugging in cable as excessive perpendicular force may damage the connection PM2 5 pg m 3 Open up AirAssure Configuration Manager on your personal computer AirAssure Indoor PM2 5 Mass Concentration Monitor 4 After the splash screen the AirAssure Configuration Manager main page will appear vi AirAssure Configuration Manager Configuration Serial Port Static IP Address Static IP Mask Static IP Gateway Use DHCP Server IP Address Server IP Port 6080 Correction Factor 10 Calibration Coefficients 100005 Message 5 Click the Serial Port drop down arrow and select the
18. 2014 1 58 PM AirAssure Indoor PM2 5 Mass Concentration Monitor 2 Atthe Apache Tomcat Setup window click Next a2 Apache Tomcat Setup Welcome to the Apache Tomcat Setup Wizard This wizard will guide you through the installation of Apache Tomcat Itis recommended that you dose all other applications before starting Setup This will make it possible to update relevant system files without having to reboot your Click Next to continue Le L H w EE U a GC CG z o Ka E a e d Apache Tomcat 3 Click Agree button to accept license and proceed Lal Apache Tomcat Setup gy on wel x Please review the license terms before installing Apache Tomcat Press Page Down to see the rest of the agreement Version 2 0 January 2004 http www apache orgJicenses TERMS AND CONDITIONS FOR USE REPRODUCTION AND DISTRIBUTION 1 Definitions License shall mean the terms and conditions for use reproduction and distribution as defined by Sections 1 through 9 of this document If you accept the terms of the agreement dick I Agree to continue You must accept the agreement to install Apache Tomcat Nullsoft Install System v2 46 4 Atthe Choose Components window select the type of components to install a Tomcat b Start Menu Items c Manager Network Ethernet Setup 7 5 5 Click Next to proceed a Apache Tomcat Setup mg mg mp we a zm e Choose Componen
19. 4 3 CHAPTER 5 Setup Overview sssnnnnnnnnnnnnnnnonnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnn 5 1 CHAPTER 6 Analog Output ccccseeeeeeeeeeeeseeeeeceeessseeeeseeeeeeeeeeees 6 1 A OIA Eege Ee 6 1 AlrASSUre Configuration Manager 6 2 CHAPTER 7 Network Ethernet Setup cscceeecccssssseeeeeeeeeeeeeeeees 7 1 Install the Web Applcatpon 7 1 JEE 7 2 Install Apache Tomcat Application cccccccccsssceceeseeseeeeeseeeeenees 7 4 Installing Apache Tomcat Application nnssnesnnnnnnnnnnnsnennnnnnni 7 4 Apache Tomcat Application Manager Interface ceeeeeee 7 9 AirAssure Monitor IP CGontguraton 7 12 CHAPTER 8 Using the Web Application cccccccsssssessseeeeeeeeeeeenees 8 1 AirAssure Web Main Page Functions ssssesensesensennrensrnnrresnrnrrrenenn 8 1 Languages EE 8 2 LOGOO Hl 8 3 Ghanging the Password seen 8 4 Vew RE EE 8 5 Single Unit Configuration Administrator cccseeseeeeeeeeeeeeeeeeeeeeees 8 6 Multi Unit Configuration Administrator 8 9 vii viii CHAPTER 9 Maintenance and Troubleshooting cccccccssseseeseeeeees 9 1 Device Maintenance eelere 9 1 Sensor Element Hepl cement 9 1 Pump Filter Fan Hepnlacement 9 7 Bisieteelletebelgo VE 9 15 Bellen ee ee EE 9 16 heChiniGal GOMtACIS 32sec E E 9 19 APPENDIX A Specifications eeeesgeggeegeee eege geesde ege A 1 AirAssure Indoor PM2 5 Mass Concentration Monitor
20. AIRASSURE INDOOR PM2 5 MASS CONCENTRATION MONITOR MODEL IPM2 5 AD OPERATION AND SERVICE MANUAL P N 6008418 REVISION C AUGUST 2015 START SEEING THE Semel tle L REGISTERING TODAY Thank you for your TSI instrument purchase Occasionally TSI releases information on software updates product enhancements and new products By registering your instrument TSI will be able to send this important information to you http register tsi com As part of the registration process you will be asked for your comments on TSI products and services TSI s customer feedback program gives customers like you a way to tell us how we are doing TSE UNDERSTANDING ACCELERATED TSI Incorporated Visit our website www tsi com for more information USA Tel 1 800 874 2811 India Tel 91 80 67877200 UK Tel 44 149 4 459200 China Tel 86 10 8219 7688 France Tel 33 141192199 Singapore Tel 656595 6388 Germany Tel 49 241 523030 2015 TSI Incorporated Printed in U S A Manual History The following is a history of the AirAssure Indoor PM2 5 Mass Concentration Monitor Model IPM2 5 AD Operation and Service Manual Revision Date A December 2014 B February 2015 C August 2015 Part Number Copyright Address Fax No Limitation of Warranty and Liability effective April 2014 Service Policy Trademarks Warranty 6005887 Revision C August 2015 TSI Incorporated 2014 2015 All rights reserved
21. AirAssure Web data be archived and cleared at least once per year To clear AirAssure Web Data click Clear AirAssure Web Data Data records for all sensors will be cleared and will no longer be available for viewing or export q J UNDERSTANDING ACCELERATED Username admin LOGOUT res TU d jae Instruments Configuration Configuration Sensor configuration Global Correction Factor 1 0 Sensor records on page 288 Update Export AirAssure Web Data as csv Clear AirAssure Web Data Delete AirAssure Web Data Application Version 1 0 4 AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 9 Maintenance and Troubleshooting This chapter describes recommended maintenance procedures for the AirAssure monitor and is intended for use by a trained service technician The frequency of routine maintenance procedures depends upon the frequency of use of the instrument and the operating conditions If contacting TSI for assistance please have the AirAssure monitor close to the telephone when discussing the problem with a TSI technician Table 9 1 AirAssure Monitor Replacement Part Kits Qty Part Number Description IPM2 5 Sens Calibrated replacement sensor element IPM2 5 Kit Replacement pump fan and filter kit Service and Maintenance procedures described in this section should only be performed by a qualified trained service technician Device Maintenance Sensor Element Replacement Replacement of
22. CHAPTER 1 Product Overview This chapter contains an introduction to the Model IPM2 5 AD AirAssure Monitor and provides a brief explanation of instrument operation Product Description The AirAssure monitor is a mass concentration measuring device that mounts to a wall and continuously measures indoor PM2 5 concentrations displaying and controlling particulate pollution in indoor environments 34 The AirAssure monitor is typically used to monitor air quality in Grade A commercial gel offices government buildings hospitals private schools and high end residences that want to provide controlled regulated indoor air quality for employees and residents Features e 4 3 color LCD display that is easy to read from across the room Figure 1 1 e Languages Simplified Mandarin English A Moore Mase e Range of measurement 5 300 ueim Concentration Monitor e Calibrated with professional instruments for accuracy e Large easy to read display with readings that change colors to signify changes in air quality e Easy to install e Low maintenance with no batteries to replace e Analog output to communicate with building control system that updates every second e Ethernet digital output and web application used to extract and store data for viewing on a monitor using a web application that updates every 5 minutes e Web application controls up to 99 instruments from 1 server identifies problem areas in building a
23. COM port connected to the IPM2 5 AD AirAssure monitor If there are multiple COM ports listed the highest COM port is usually the most recently added AirAssure monitor AirAssure Configuration Manager Configuration Serial Port COM comi7_ Static IP Address 6 Click the Open button to connect to the monitor te AirAssure Configuration Manager Configuration Serial Port ein Static IP Address Analog Output 6 5 6 6 7 You are now connected to the IPM2 5 AD AirAssure monitor From this screen you can configure IP settings update the correction factor or modify the calibration coefficients when installing a new sensor element AirAssure Configuration Manager Configuration Serial Port COM17 Static IP Address 192 8 200 Static IP Mask 255 255 255 Static IP Gateway IZ 8 200 Use DHCP Server IP Address Server IP Port Correction Factor Calibration Coefficients Please refer to Chapter 7 for details about how to configure IP settings and refer to Chapter 9 for details about how to install a new sensor element and update calibration coefficients AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 7 Network Ethernet setup Up to 99 Model IPM2 5 AD AirAssure monitors can be configured on a local area network with the TSI AirAssure Web application running on a computer on the same network The custom application allows monitoring and configuration
24. Ethernet Setup 7 11 9 From step 6 above reference the IP address of the Tomcat server and replace the test localhost with the IP you recorded such as 192 168 11 22 The address should now read something like http 192 168 11 22 8080 omweb Once the monitors have been configured properly this is the address to use for users to view readings from the monitors i i I SS eg os Ti i P b A l JY es d S I Sidd Al re EI se LOGIN q b UNDERSTANDING ACCELERATED instruments No instruments found Copyright 2015 by TSI Inc All rights reserved The next step is to configure each of the AirAssure monitors with the IP address of the Tomcat server so they know where to send data to the server Follow the steps under AirAssure Monitor IP Configuration below AirAssure Monitor IP Configuration 7 12 Each AirAssure monitor must be configured to communicate with the AirAssure Web application before data will be available on the web pages After the web application service is installed the monitor is ready for connection to a network Each monitor is configured from the factory to get an IP address from a DHCP server on the network This is how most networks are configured lf this is not the case the monitor must be connected using the internal USB connector and AirAssure Configuration Manager software see Chapter 6 to override DHCP with a fixed IP address Please contact your IT department for he
25. SES INJURIES OR DAMAGES CONCERNING THE GOODS INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE TORT STRICT LIABILITY OR OTHERWISE SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE OR AT THE OPTION OF SELLER THE REPAIR OR REPLACEMENT OF THE GOODS IN THE CASE OF SOFTWARE SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS BUSINESS INTERRUPTION OR ANY SPECIAL DIRECT CONSEQUENTIAL OR INCIDENTAL DAMAGES SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION DISMANTLING OR REINSTALLATION COSTS OR CHARGES No Action regardless of form may be brought against Seller more than 12 months after a cause of action has accrued The goods returned under warranty to Seller s factory shall be at Buyer s risk of loss and will be returned if at all at Seller s risk of loss Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give prompt attention to any problems If any malfunction is discovered please contact your nearest sales office
26. Windows7 6 1 192 168 11 4 Copyright 1999 2014 Apache Software Foundation 5 From the lower right corner of the web page record the Tomcat server IP address You will need it later to configure each of the sensors Tomcat Server IP Address 6 The web application is now installed and ready for use The web page should refresh after a few seconds and add the omweb entry under applications as shown below AirAssure Indoor PM2 5 Mass Concentration Monitor 7 Click on the pmweb link in the application manager to navigate to this application a I 7 EE O R O xf El manager M 3 Search Hi Share More gt gt Signin Tomcat Web Application Manager List Applications HTML Manager Help Manager Help Server Status Start Stop Reload Undeploy I None specified Welcome to Tomcat true 0 spire sessions with idte 30 minutes Stat Stop Reload Undeploy 2 aa o Ere sessions wihide2 0 minutes pa pmweb None pecified 0 l with idle gt 30 minutes 8 When you first open the web page there will be no instruments listed unless you have already configured each of the monitors manager None specified Tomcat Manager Application sh AirAssure LOGIN AN Le A ee Fa W EC d AIASSUIS ver Instruments No instruments found Copyright 2015 by TSI Inc All rights reserved Network
27. ans that unsafe use of the instrument could result in serious injury to you or cause irrevocable damage to the instrument Follow the procedures prescribed in this manual to use the instrument safely Un insulated voltage within the instrument may have sufficient magnitude to cause electric shock It is dangerous to make any contact with any part inside the instrument yi Labels Advisory labels and identification labels are attached to the interior and exterior of the AirAssure monitor Labels are described below Back Panel Label displayed on the back panel IS IPM2 5 AD AirAssure PM2 5 Monitor Serial Number IPM251501001 TSI Incorporated www tsi com 24VAC DC 10 W Sensor Element Label internal inside the front cover AirAssure Indoor PM2 5 Mass Concentration Monitor Contents Mal Wall HIStory E lil Waraniyi A AE iv EE v Description of Safety Labele cansi N V Caution Gvmbole V Warning OYMDOIS EE V Eer vi CHAPTER 1 Product Overview ccccccssssssseeeeeeeseceeeesneeeeeeeeeoeensees 1 1 PrOOUCL bedeite EE 1 1 SEI 1 1 FIOW IE ei 1 2 CHAPTER 2 Unpacking tail iouiitii neice a 2 1 PACKING BC EE 2 1 Bletz et ue ME 2 1 CHAPTER 3 Hardware Installation ssccseeeeesseessssssseeeeeseeeeeeeees 3 1 YW All WOU aes EE 3 1 aang Box MOUN EE 3 5 CHAPTER A AirAssure Monitor Operation ccccccccesesseesseeeeeeeeeeeeeeeeees 4 1 Air Quality Index Reading EE 4 2 Bdderg elulal Wie Te CN
28. ation running on a computer on the same network The custom application allows monitoring and configuration of the sensors remotely AirAssure Web data is archived and can be downloaded and imported into a spreadsheet program such as Microsoft Excel spreadsheet program for offline analysis AirAssure Indoor PM2 5 Mass Concentration Monitor A IPM2 5 AD p Atte PMS 25 Mente ra ant IN ecard en ve td low d owa Mace om z D E gt Figure 1 2 AirAssure Monitor Back Panel Ports Product Overview 24 VDC or VAC power Ethernet port 1 3 1 4 This page intentionally left blank AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 2 Unpacking Packing List The packing list described in Table 2 1 lists the components shipped with the AirAssure monitor Table 2 2 lists the AirAssure service kits available from TSI The service kits are sold separately Table 2 1 AirAssure Monitor Packing List Qty Description 1 IPM2 5 AD AirAssure PM2 5 Indoor Particle Monitor and 24 V connector Optional wall plate Mounting hardware 3 drywall screws 3 drywall plugs 2 gang box screws one 6 mm screw NIST Statement of conformance Quick Start Guide Table 2 2 AirAssure Monitor Service Kits Qty Part Number Description IPM2 5 Sens Calibrated sensor element IPM2 5 Kit Replacement pump fan and filter Unpacking Use the Packing List in Table 2 1 to
29. covered by this warranty f This warranty is VOID if the instrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g This warranty is VOID if the product has been misused neglected subjected to accidental or intentional damage or is not properly installed maintained or cleaned according to the requirements of the manual Unless specifically authorized in a separate writing by Seller Seller makes no warranty with respect to and shall have no liability in connection with goods which are incorporated into other products or equipment or which are modified by any person other than Seller The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE WITH RESPECT TO SELLER S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS BUYER S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER S OPTION REPLACEMENT OF THE GOODS WITH NON INFRINGING GOODS TO THE EXTENT PERMITTED BY LAW THE EXCLUSIVE REMEDY OF THE USER OR BUYER AND THE LIMIT OF SELLER S LIABILITY FOR ANY AND ALL LOS
30. e Shop 1 000 Aug 6 2015 3 46 48 PM Receiving Dock 1 000 Aug 6 2015 3 47 10 PM Using the Web Application 8 3 8 4 Changing the Password To change password of the user currently logged into AirAssure Web monitor follow these steps 1 Inthe top right corner of the web application click on admin 2 At the Change Password window enter the Current Password enter a New Password and re enter to confirm the new password 3 Click the Change button to accept new password UNDERSTANDING ACCELERATED spe eat admin oe i kal gem om quest i UAG feat Instruments Configuration Change Password Current Password New Password Confirm Password Change Cancel Copyright 2015 by TSI Inc All rights reserved AirAssure Indoor PM2 5 Mass Concentration Monitor View Data Data received from AirAssure monitors on the network is available for viewing by all users It is displayed on the main page of the AirAssure Web application 1 A monitor is identified by its location in the Instruments column e An active status indicates the monitor is actively communicating with the web application e An inactive status indicates application has not received data from the unit in the past 5 minutes 2 The concentration in ug m measured by the monitor is displayed along with its correlating AQI color code Instruments Configuration Instruments 25 50 All Instrument PM 2 5 ug m3 Correction Factor
31. een the plug and the screw head AirAssure Indoor PM2 5 Mass Concentration Monitor 10 Pull cables through hole in drywall 11 Remove 24 V screw terminal connector from the back of the unit Verify that polarity matches the polarity shown on the back of the unit Incorrect polarity may damage unit The power input must be wired to 24 VAC or 24 VDC Wiring the unit to 120 230 VAC will cause serious damage to the unit and void the warranty 12 Use a flathead screwdriver to separate the cover from the base The slot is located on the back of the unit along the bottom seam between the unit cover and base 13 Insert stripped leads from building 24 V power cable into screw terminal connector and tighten screws on terminal 14 Insert connector into the unit Hardware Installation 3 3 19 18 19 20 21 Connect 4 20 mA if analog Align monitor flush against the three 3 Connect Ethernet if digital connection is desired connection is desired Wire is held in place by depressing spring clip inserting wire and releasing spring clip mounting screws and push downward onto screw heads Tighten mounting screws Place cover over the mounted display and snap the bottom of the cover to the base The monitor is ready for use and powers on when the 24 VAC VDC connection is turned on Self test When the unit is first powered up it performs a quick self check by turning the pu
32. egative number appears on the display this indicates an error condition PM2 5 ug m3 The error codes are as follows Error Code Description Cause Sensor element is damaged or needs 1 Sensor Error replacement here is a leak 2 Autozero Error Pump is not turning on or there is a lea around the filter Fan is damaged or needs replacement Refer to the Troubleshooting section of the manual for other causes and recommended actions AirAssure Monitor Operation 4 3 4 4 This page intentionally left blank AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 5 Setup Overview The AirAssure monitor provides data in three ways 1 Display only only the power connection is required for the unit to display readings on the display for use by anyone walking nearby The display is large and colorful making it easy to read from a long distance Analog output for use with building automation systems that accept 4 20 mA input The 4 20 mA output can be connected to a pre existing building automation system and configured to show from lt 5 to gt 300 g m output over the 4 20 mA output range Ethernet output for use with AirAssure Web a custom TSI web application A standard Ethernet output can connect up to 99 devices on a local area network to a software application running on a computer on the same network The custom application allows monitoring and configuration of the sensors remo
33. follow the directions below 1 Remove unit from wall unplug power and remove sensor element See previous section for details 2 Locate the pump filter and fan near the bottom of the device The pump rests on top of the filter held in place with tabs Fan Cable Guide Pump Filter behind pump 3 Use a small flathead screwdriver and insert behind the pump to carefully snap out 4 Move pump away from the area to expose the filter underneath 9 8 AirAssure Indoor PM2 5 Mass Concentration Monitor 5 Place the screwdriver behind the filter and push upward to remove Discard filter 6 Remove cables from the cable guide Maintenance and Troubleshooting 9 9 9 10 Disconnect the pump cable from the main board by pushing on the connector tab Discard pump and cable assembly Remove the fan by pressing down on the retainer tab and pulling the fan out of the fan slot A flathead screwdriver can be used from the opposite side bottom of device to help push fan out of the fan slot Remove the fan connector from the circuit board by pushing on the connector tab Discard fan AirAssure Indoor PM2 5 Mass Concentration Monitor 10 Connect new fan into the circuit board 4 pin connector 11 To place fan in correct position the arrow on the fan housing should point downward away from the display The fan should be oriented such that arrows are at the bottom
34. log output or Ethernet with a building automation system or with the TSI AirAssure Web software it will be helpful to record the serial number physical location and IP address for each sensor in Table 3 1 at the end of this chapter The serial number is located on the back of the unit below the bar code 3 1 3 2 1 Locate the wall mount template located in the Quick Start guide Remove enn 2 Use center punch or sharp toot to the template from the Quick mart hve cabo aa 3 Remove tempiste Start guide 2 Tape template to wall in the desired location using a level to straighten Recommended mounting height is 1 4 to 1 6 meters from the floor 3 Locate the three 3 mounting hole marks on the template and the four 4 corners of the rectangular cutout 4 Ateach location use a pencil or sharp tool to poke through the template and mark desired position on the wall 5 Remove template 6 At each mounting location drill three 3 4 5 mm holes through the drywall 7 Cuta rectangular hole in 4 drywall using the wall cutout corner marks on the template 8 Insert the provided drywall Cutout rectangle using plugs into the three 3 4 corner marks Insert 3 mounting locations Ensure v drywall plugs are fully seated by l plugs tapping with a hammer 9 Fasten the provided 25 mm WW screws into drywall plug and screw into drywall leaving approximately 3 mm space betw
35. lp in this case AirAssure Indoor PM2 5 Mass Concentration Monitor Connect power and an Ethernet cable into the monitor s ports as shown below The computer with the Tomcat server must be on the same network as the AirAssure monitors See the sections above for configuring Java and the Tomcat server 24V DC or AC power Ethernet 6 ae taral Warder TO IN ze perst d Figure 7 1 AirAssure Back Panel 2 After a valid connection is located from a DHCP server an assigned IP address displays on the monitor s screen Network Ethernet Setup 7 13 3 Record the IP address in the table at the end of Chapter 3 for each monitor 4 On a computer connected to the same network open a web browser and enter the IP address of the monitor as it appears on the display screen of the device D A password screen will open like the one below Please refer to the table at the end of Chapter 3 for the serial number of each instrument that you recorded during installation Enter the serial number as the password from the label on the back panel or table e g IPM251501002 and click the Login button EI AirAssure x Login Eg 192 168 11 2 login cg Password 6 The configuration page should now be displayed as shown below If the screen does not appear please contact local IT support to see if your computer has access to the network d A http 10 1 3 117 locComm cgi A Location omr Configura
36. mp on to give three 3 quick chirps If an error occurs it will display on the front panel as a 1 2 or 3 within 30 seconds of powering up See the table in Chapter 9 under Troubleshooting if one of these codes appears AirAssure Indoor PM2 5 Mass Concentration Monitor Gang Box Mount Tools required for installation into a Gang Box e Wire stripper e Flat head screw driver e Phillips head screw driver e Level In cases where there is a gap between the gang box and wall plate it is recommended that spacers be used to provide support for the wall plate Gang box spacers can be obtained from a local hardware supplier 1 Insert the short 6 mm self tapping screw into the plate The back of the screw should be flush with the back of the plate 2 Place optional wall plate up to the gang box 3 Pull cables through the wall plate opening 4 Fasten the wall plate to the gang box using the two 2 provided 20 mm M4 screws 5 Further secure the plate into the drywall using the two 2 provided 25 mm screws and drywall anchors If mounting to masonry or other hard surface install screws appropriate for that surface 6 Remove green connector labeled 24V IN from the back of the device 7 Insert stripped leads from building 24 V power cable into screw terminal connector and tighten screws on terminal Verify that 24 V power polarity matches the polarity on the back of the unit
37. nd updates monitor configuration e Built in autozero mechanism periodically zeroes instrument for sustained accuracy over time e Mounts to drywall type 86 gang box or American single gang box e Field serviceable components 1 1 How it Works 1 2 The AirAssure monitor incorporates a photometer with light scattering technology to closely estimate particulate mass concentrations The monitor measures particles less than roughly 2 5 micrometers in diameter PM2 5 Regular air quality readings provide employees and building occupants assurance that the indoor environment is at an acceptable level for human health The monitor can be either mounted flush into drywall or to a gang box using the included adapter plate There are three ways that the AirAssure monitor can provide data 1 Display only only the power connection is required for the unit to display readings on the display for use by anyone walking nearby The display is large and colorful making it easy to read from a long distance 2 Analog output for use with building automation systems that accept 4 20 mA input The 4 20 mA output can be connected to a pre existing building automation system and configured to show from lt 5 to 300 g m output over the 4 20 mA output range 3 Ethernet output for use with AirAssure Web a custom TSI web application A standard Ethernet output can connect up to 99 devices on a local area network to a software applic
38. ndoor PM2 5 Mass Concentration Monitor CHAPTER 4 AirAssure Monitor Operation At power up a splash screen displays with the TSI logo model number serial number and firmware version Serial Number 1PM251501001 Firmware Rev 1 0 Color bar indicates range and progression through the color codes 4 bars per color except for maroon which has 1 bar The monitor s display indicates PM2 5 concentration in units of ug m This value is a 5 minute trailing average and updates once per second PM2 5 ug m The air quality index AQI color changes depending on the PM2 5 concentration When the monitor is connected to a network the IP address will be shown in the lower left hand corner of the display 4 1 Air Quality Index Reading As air quality fluctuates beyond the good limit range the color shown on the display changes to reflect the severity of the change The AirAssure monitor uses the China air quality index AQI standard which is based on the recommendations of the World Health Organization WHO from 2005 Six AQI categories and PM 2 5 value ranges can display on the monitor Green Range 5 35 Good Red Range 116 150 Unhealthy Yellow Range 36 75 Moderate Purple Range 151 250 Very Unhealthy Orange Range 76 115 Unhealthy for Sensitive Groups Maroon Range 251 gt 300 Hazardous 4 2 AirAssure Indoor PM2 5 Mass Concentration Monitor Error Conditions If a n
39. of the fan slot 12 Snap fan into place until the tab is securing the fan 13 Place filter into the cavity Do not forget to install the filter If the pump Is installed without the filter the unit will not make accurate measurements Maintenance and Troubleshooting 9 11 14 16 Place the pump into place Secure cables into the cable Connect the new pump into the 10 pin connector on the circuit board resting on filter by sliding under the top tab and pushing down on the bottom tab until it snaps into place Avoid pressing on the center of the pump Position the ferrites gray beads between cable guide as shown There should be five 5 ferrites above the fan and five 5 ferrites below the fan guide to avoid pinching cables by the cover AirAssure Indoor PM2 5 Mass Concentration Monitor 18 19 20 21 Maintenance and Troubleshooting Slide the sensor into position until it snaps into place secured by the white tab Attach 6 pin cable to sensor To avoid pinching the cable when installing the cover verify that there are five 5 ferrite beads above the fan and five 5 below the fan as shown Reposition if necessary Place sensor cable underneath cable guide 9 13 22 The monitor is ready for reinstallation Re connect all cable connections 23 Slide unit back onto the wall Tighten mounting screws 24 Place cover over the moun
40. of the sensors remotely AirAssure Web data is archived and can be downloaded and imported into a spreadsheet program such as Microsoft Excel for offline analysis Five steps are required to establish an Ethernet connection 1 Install the Web Application 2 Obtain the IP address for the server computer that the web application will reside Configure each monitor with the IP address of the server computer 4 Open the AirAssure Web Application to view and control the monitors Recommended browsers include Internet Explorer Version 10 and higher Chrome Safari and Android Recommended operating systems are Windows 7 and Windows 8 1 Use of other browsers or operating systems may limit functionality or prevent access to the webpage or web application Install the Web Application There are seven steps to configuring the system for Ethernet output for use with the TSI custom web application Follow the outline below There are more details for each step below 1 Download the installation package from http www tsi com airassure get started 2 Install Java 8 on a computer server connected to an Ethernet network using jre 8u20 windows x64 exe for 64bit computers or jre 8u20 windows i586 exe for 32 bit computers To prevent problems with changing IP address it is strongly recommended that the computer server has a fixed address Install Apache Tomcat on the computer server Load the web application file from T
41. omcat omweb war Physically install the sensors on the Ethernet network 7 1 6 Open the web page for each sensor and configure the web page on each sensor to send data to the web application 7 Open the web application and configure the data and sensor labels Install Java 8 All Java application servers require a Java virtual machine The vendor and version of Java required depends on the application server chosen Java 8 or higher is recommended The reference platform and instructions below use Oracle Java 8u20 64 bit running on Windows 7 64 bit Professional Other platforms are compatible provided they have a Java 7 or higher environment and Java application server available Ubuntu 12 04 64 bit with Oracle Java 7 and Apache Tomcat 7 has been tested for compatibility In addition to the installation package Java 8 is downloadable directly from www java com It is recommended before proceeding with online installation to disable your Internet firewall In some cases the default firewall settings are set to reject all automatic or online installations such as the Java online installation If the firewall is not configured appropriately it may stall the download install operation of Java under certain conditions Refer to your specific Internet firewall manual for instructions on how to disable your Internet Firewall The installation medium contains Oracle Java 8 JRE 32 and 64 bit for Windows R
42. or can also be changed on the unit webpage Calibration These are only modified when a new sensor Coefficients element is installed see Chapter 9 Although typically set up on the device webpage Chapter 7 they can also be entered here Note that the four coefficients must be separated by a forward slash when entered 4 Admin users can delete Ka individual devices on AirAssure Web This might be necessary if a monitor has been removed from the network If a unit is deleted Ka from AirAssure Web then the data records of the unit are deleted also If you wish to keep data records from the unit use the export data feature on the Configuration screen to export all AirAssure Web data before deleting the unit 8 8 AirAssure Indoor PM2 5 Mass Concentration Monitor Multi Unit Configuration Administrator An administrator has access to the Configuration screen which has configuration options for the overall network Follow these directions to access the configuration screen 1 Log in as the administrator using the administrator username and password defined during setup of Apache Tomcat Chapter 7 LOGIN q p UNDERSTANDING ACCELERATED l WW e EF Se Instruments Username Password LOGIN Copyright 2015 by TSI Inc All rights reserved WW 2 From the main page click on the Configuration tab Username admin D UNDERSTANDING ACCELERATED LOGOUT N Pe E bel P Af ASUI S yY y f
43. out the center hole with some clean air If the high reading persists consider replacing the sensor element See Device Maintenance section If the sensor has been replaced recently check that the calibration parameters for the sensor element have been entered into the configuration web page for the monitor Incorrect calibration parameters can cause incorrect readings The zero module has a pump that can make a hissing noise for a few minutes This is normal When the monitor is first powered on it makes a quick three chirps or hisses to indicate the zero module is operating properly In normal operation the pump runs 30 minutes after power up and then every 8 hours afterward Problem Suggested Action The AirAssure Web Java installation Tomcat The installation of the components required for web based application does not show installation AirAssure Web monitoring are somewhat complex If this is a first time the monitors I ve installed installation or IP installation carefully read the instructions and review all the configurations incorrect steps Install only from the install package from TSI that has been tested Review the system requirements and compatibility requirements to make sure your hardware operating system and software are compatible IP configuration of the On the monitor webpage make sure that Server IP Address monitors is incorrect has the IP address of the server There is a windo
44. re from the Software and Firmware Wizard see previous steps 5 Launch the application It may be necessary to select the COM port connected to the IPM2 5 AD AirAssure monitor Click Update Firmware Allow process to complete following onscreen prompts The duration of the update is 5 to 7 minutes Troubleshooting The table below provides basic information about instrument problems and suggestions for corrective action Table 9 1 Troubleshooting Problem Cause Suggested Action Monitor does not power No power to the sensor Check with a voltmeter that the connector is providing either up blank screen 24 VAC or VDC power Incorrect wiring Verify that the 24 VAC or VDC is wired into the 24V connector and not into the 4 20 mA connector Power connector not seated Make sure the green power connector is seated firmly in the properly socket in the monitor back panel Screen or circuit board If the power is supplied to the unit when the cover is damaged removed there should be a green pulsing LED on the upper left of the circuit board If not then there may still be a problem with the power or the circuit board may be damaged If the LED is pulsing but the screen is blank the screen may be damaged In either case the monitor must be returned to the factory or replaced 9 16 AirAssure Indoor PM2 5 Mass Concentration Monitor Problem Display shows unusual numbers Display shows numbers
45. rom the label on the replacement sensor element and click Submit Changes me SPM251447123 W I SEE ETET 219212010 Configuration Stabe IP Address Stabe Ip Mask Static IP Gateway 1562 fT J255 0 0p AirAssure Indoor PM2 5 Mass Concentration Monitor Alternatively the sensor calibration coefficients can be updated using AirAssure Configuration Manager Chapter 6 or TSI AirAssure Web application Chapter 8 18 Allow the unit to warm up and run for a couple of hours and then check that the readings look reasonable If not consult the Troubleshooting section at the end of this chapter Error code 2 may appear during for the first 8 hours of operation after changing the sensor element This is not a problem as it indicates that the monitor has detected a shift in calibration parameters However if the 2 error code does not go away after 8 hours check the troubleshooting section at the end of this chapter for possible solutions Pump Filter Fan Replacement Periodic replacement of pump filter and fan may be required for optimum performance Review the Troubleshooting section at the end of this chapter for situations when the pump filter assembly or fan should be replaced Ras Filter Fan Pump Maintenance and Troubleshooting 9 7 Service kit IPM2 5 KIT includes a replacement pump filter and fan and is available for purchase from TSI To service these components
46. ted display and snap the bottom of the cover to the base 25 After powering on confirm a successful installation Listen for three 3 chirps of the pump and verify that no error codes are shown on the display within 30 seconds of power up 9 14 AirAssure Indoor PM2 5 Mass Concentration Monitor Upgrading Firmware Check the TSI website periodically for upgrades to IPM2 5 AD firmware Firmware upgrades will be available for download To determine whether a IPM2 5 AD firmware upgrade is available follow these instructions 1 Open the TSI website at www tsi com 2 From the Support tab gt TSI Software and Firmware gt Software and Firmware Wizard 3 Follow the onscreen instructions to locate firmware for the AirAssure IPM2 5 AD monitor If firmware is found online by the Software and Firmware Wizard and you want to install it onto your monitor follow these instructions 1 Before running the firmware installer you must connect USB To access the USB connector remove the cover of the monitor by pressing fingers firmly on bottom of device cover squeeze and pull outwards 2 Connect the unit to your personal computer using a Micro USB cable not provided The USB connector is on the left hand side of the device Use caution when plugging in cable as excessive perpendicular force may damage the connection Maintenance and Troubleshooting 9 15 pm2 s pg m 4 Download the firmwa
47. tely AirAssure Web data is archived and can be downloaded and imported into a spreadsheet program such as Microsoft Excel spreadsheet program for offline analysis 5 1 4 2 This page intentionally left blank AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 6 Analog Output The 4 20 mA current loop is a very robust sensor signaling standard used commonly in building automations systems Current loops are ideal for data transmission because of their inherent insensitivity to electrical noise The current provided by the analog output from 4 20 mA is linearly proportional to the reading on the front of the display 4 20 mA Setup Refer to Chapter 3 for hardware installation instructions The electrical connections required for use of the analog output are e 24 VAC VDC power e 4 20 mA dual conductor connection to building automation system Note that the connectors are different for these two connections ff aw be 1 zm 24 VDC or VAC power A 1PM2 5 AD gt Menat PMS Men Seral Marber TL WW te Figure 6 1 AirAssure Monitor Back Panel Power and Analog Output Connections Please consult the documentation for your particular building automation system for how to connect a sensor to the system and any special requirements for that system Also refer to the building automation system documentation and the table and formula below to configure the system for 6 1 this sensor
48. the internal sensor element is recommended annually in order to avoid degradation effects Calibrated sensor element service kit IPM2 5 SENS is available for purchase from TSI Incorporated 9 1 9 2 1 To replace the sensor element first remove the cover of the monitor Remove the cover by pressing fingers firmly on the bottom of device cover squeeze and pull outwards 2 Power down the unit either by powering down the building circuit or by removing the unit from the wall and unplugging power as shown If you remove the unit from the wall it may be necessary to loosen the mounting SCrews 3 To remove the sensor element grip the 6 pin cable between thumb and forefinger underneath connector and pull to disconnect AirAssure Indoor PM2 5 Mass Concentration Monitor Press on the white tab to the left of the sensor and slide sensor outward to remove Discard sensor Slide the new sensor into position until it snaps into place secured by the white tab Attach 6 pin cable to sensor To prevent the cable from getting pinched by the cover position the ferrites gray beads between cable guides as shown There should be five 5 ferrites above the fan and five 5 below the fan Maintenance and Troubleshooting 9 3 8 10 11 Place sensor cable underneath the cable guide Take note of the new sensor label that identifies the four emm v i calibration coefficients
49. though shown in English below H gt UNDERSTANDING ACCELERATED ANIFASSULS Ep eZ os bed D IJ de 29 50 All RE PM 2 5 ug m3 Executive Suite E Lower Level RRA E Front Lobby TE Engineering Lab 8 E eee Beijing Office S N 04 ee sy Beijing Office S N 09 SSS Sq Breakroom ss RRELAR A 2014 2015 TIAI RBMA RAR a al AirAssure Indoor PM2 5 Mass Concentration Monitor Logging In The administrator can log in to access single unit and multi unit configuration options Follow these steps to log in as the administrator 1 Click LOGIN in the upper right hand corner of the web application 2 Enter the administrator username and password defined during set up of Apache Tomcat Chapter 7 TSE UNDERSTANDING ACCELERATED LOGIN ka AirAssure Wer Instruments Password LOGIN Copyright 2015 by TSI Inc All rights reserved 3 Press LOGIN to proceed Once you are logged in you return to the AirAssure Web main screen Now the Configuration tab is visible along with the correction factor and latest reading timestamp for each device on the network UNDERSTANDING ACCELERATED ane admin A T ii ge da ry AE si z kg AifrAssUure Wee Instruments Configuration Instruments 25 50 All 1 000 Aug 6 2015 3 46 37 PM JF Home Beijing Office 1 000 Aug 6 2015 3 46 35 PM Shoreview MN Outdoor 1 000 Aug 6 2015 3 47 23 PM Front Lobby 1 000 Aug 6 2015 3 46 43 PM Machin
50. tion Static IP Address Static IP Mask Static IP Gateway Use DHCP Server IP Address 10 1 AN _ 114 Server IP Port 6080 Correction Factor 1 00 Calibration Coefficients 1670 0 441 0 21 000 0 886 submit Changes Apply Changes AirAssure Indoor PM2 5 Mass Concentration Monitor 7 Generally the only information that has to change at this point is the server IP address for the Tomcat server configured earlier in the chapter and entered in step 6 under Apache Tomcat Application Manager Interface above The top section of the configuration web page is only used when overriding DHCP and entering a static IP address If this is the case please contact your IT department for help If you turn off DHCP and enter a static IP address and click Submit changes the monitor cannot be reconfigured without connecting to a network that will provide the given static IP address If the address is inadvertently changed you can use AirAssure Configuration Manager to re enable DHCP over USB See Chapter 6 for details on using AirAssure Configuration Manager The bottom section of the configuration web page provides information on calibration coefficients for a replacement internal sensor This is only edited during service or when the calibration factor is updated by an administrator 8 Enter the Tomcat Server IP address and click Submit Changes CH http 10 1 3 117 locCormm cgi H e GQ X a Location Comm
51. ts Choose which features of Apache Tomcat you want to install Check the components you want to install and uncheck the components you don t want to install Click Next to continue Or select the optional e RB Tomcat le leslie ar ad Jl Start Menu Items Beien Manager l Fi Host Manager Examples Descrint Space required 7 3MB 6 Enter application manager details at the Configuration window and click Next Check the Create shortcuts for all users box The user name and password used at this screen will be required again when installing Application Manager Interface Record your username and password below for reference admin and admin are common but not secure Username Password Configuration Tomcat basic configuration Server Shutdown Port HTTP 1 1 Connector Port AJP 1 3 Connector Port 8009 Windows Service Name Tomcat Create shortcuts for all users Fi Tomcat Administrator Login User Name admin optional Password TEIT Roles manager gui 7 6 AirAssure Indoor PM2 5 Mass Concentration Monitor 7 Select the path of a Java JRE installed and click Next a Apache Tomcat Setup Java Virtual Machine path selection 7 i Java Virtual Machine Java Virtual Machine path selection Please select the path of a Java SE 6 0 or later JRE installed on your system C Program Files Java jre1 8 0_20 i Nullsoft Install System v2 46 Choose the folder in which to install Apache Tomcat
52. type A maximum cable length of 300 meters is recommended to avoid excessive loss of voltage Typically a system is configured within the building automation software by a technician to convert the signal into a meaningful measurement If your building automation system accepts only a voltage input then a load resistor will be needed Recommended load resistor specifications are as follows e 0 5V gt 0 25W 2500 0 1 lt 25 PPM temp coefficient e 0 10V gt 0 33W 5000 0 1 lt 25 PPM temp coefficient Table 6 1 provides some examples of conversions based on the formula below Table 6 1 Conversion from 4 20 mA Signal to PM2 5 Measurement mA Lig m 4 LU 10 ats 15 20 300 or more The conversion formula is as follows Measurement 300 5 20 4 Signal 4 AirAssure Configuration Manager Ethernet is typically used to adjust calibration parameters or configure IP settings However if you need to configure the monitor but are not using Ethernet AirAssure Configuration Manager can be used to configure the device over USB The USB connection is located inside of the device Procedures described in this section should only be performed by a qualified trained service technician 6 2 AirAssure Indoor PM2 5 Mass Concentration Monitor Follow these steps to download AirAssure Configuration Manager 1 Download AirAssure Configuration Manager from http www tsi com airassure get started 2 Run
53. un the installer following the installation prompts 1 On the download page click the Download button to start the installation process The File Download dialog box appears prompting to run or save the download file a Click Run to run the installer b Click Save to save the file for later installation 2 Choose the folder location and save the file to your local system Double click on the saved file to start the installation process The installation process starts 7 2 AirAssure Indoor PM2 5 Mass Concentration Monitor 4 Atthe Java Setup Welcome screen click Install ORACLE Welcome to Java Java provides safe and secure access to the world of amazing Java content From business solutions to helpful utilities and entertainment Java makes your internet experience come to life Note No personal information is gathered as part of our install process Click here for more information on what we do collect Click Install to accept the license agreement and install Java now Change destination folder ae 5 Click Close at the final installation window ORACLE W You have successfully installed Java You will be prompted when Java updates are available Always install updates to get the latest performance and security improvements More about update settings The security level has been set to a minimum value of High More information Note A restart close and re open of the browser may be required to
54. uter Make sure Tomcat server is set to start not display in browser computer reboot automatically after reboot during web application installation 9 18 AirAssure Indoor PM2 5 Mass Concentration Monitor Technical Contacts If you have technical or application questions contact an applications engineer at TSI Incorporated If the AirAssure monitor fails contact TSI at Technical Support TSI Instrument Beijing Co Ltd Unit 1201 Pan Pacific Plaza No 12 A Zhongguancun South Avenue Haidian District Beijing 100181 CHINA Telephone 86 10 8219 7688 Fax 86 10 8219 7699 E mail tsibeijing tsi com Maintenance and Troubleshooting 9 19 This page intentionally left blank 9 20 AirAssure Indoor PM2 5 Mass Concentration Monitor APPENDIX A Specifications This appendix provides specifications for the AirAssure Indoor PM2 5 Mass Concentration Monitor Model IPM2 5 AD which are subject to change without notice Light scattering photometer 5 to 300 ug m PM2 5 mass concentration and color change to represent Air Quality Index 1 g m increments 10 ug m Less than 65 RH 5 minute trailing average screen updates every 1 sec 0 2 kg 0 5 Ib Packaged 0 5 kg 1 0 Ib Physical Size Dimensions HWD 16 2 cm x 8 5 cm x 3 3 cm 6 4 in x 3 4 in x 1 3 in 3 8 cm 1 5 in e Ethernet e Analog 4 20 mA 24 volt AC or DC 5W max Statement of Conformance e Simplified Mandarin
55. verify that there are no missing components If anything is missing or appears to be damaged contact a TSI representative or TSI Customer Service using one of the following methods Telephone 86 10 8219 7688 China 1 800 874 2811 within the US 001 651 490 2811 outside the US and China E mail technicalservice tsi com 2 1 2 2 This page intentionally left blank AirAssure Indoor PM2 5 Mass Concentration Monitor CHAPTER 3 Hardware Installation The AirAssure monitor can be mounted to either drywall or to a switch box Type 86 gang box or American single gang box When mounting to a switch box or when covering up a larger than needed hole the wall plate accessory can be used The wall plate accessory is shipped with the monitor Prior to installation the cable connections should be prepared and accessible at the desired wall mount location ee EMPPORTANT e Before selecting a mounting option check local codes and regulations to verify if a gang box is required e The AirAssure monitor requires 24 VAC or VDC If the installation location has only 220 VAC then a transformer is needed This transformer can be purchased from a local supplier Drywall Mount Tools required for installing into Drywall e Wire stripper e Flat head screw driver e Phillips head screw driver e Level e Drill e Pencil e Wallboard saw e Notepad to record serial numbers Before Starting If you plan to use the ana
56. vironment has sources of random high EMF electromagnetic field which is affecting sensor accuracy Depending on the instrument the readings over a short period may be significantly different However over an averaged time period of 24 hours for a week the monitors should compare well against a high quality reference instrument such as a TSI DustTrak Aerosol Monitor Also sources can be very local Make sure the measurements are located very close together The correction factor can be adjusted to weight the reading as necessary to agree with a known reference device Refer to Chapter 8 for using the correction factor feature If the sensor element in the monitor is over 1 year old and the reading is low the sensor element may have to be replaced See Device Maintenance section If the zero pump has failed or the filter is clogged the zero offset will be incorrectly high since it is zeroing with dirty air When this large number is subtracted the reading can be incorrectly low The zero pump and filter element and filter should be replaced See Device Maintenance section There is a small mixing fan that pulls air from the room into the sensor If it fails the response could be slow or erratic to changing air quality levels Consider replacing the fan See Device Maintenance section There could be a foreign element such as large dust or a hair in the sensor viewing volume Open the cover and inspect the sensor element Blow
57. ws or Contact your IT department and have them unblock port 8080 another firewall blocking on the firewall for the Tomcat server computer Contact your HTTP traffic on port 8080 TSI representative for an application note on this topic Tomcat server IP address Each monitor has to be power cycled after updating the has been entered on the Tomcat server IP address If you haven t please remove and monitor web page and restore the power to the monitor submitted but the monitor isn t sending the data Tomcat server and monitors The server and monitor have to be on the same network If are not on the same you are using a combination of wired and wireless networks network make sure they are on the same network Often wireless networks give guest access that blocks access to the wired network Contact your IT department for assistance On AirAssure Web Unit is inactive Calibration coefficients and correction factor may be blank if individual unit calibration the unit is not currently posting data to AirAssure Web See Unit is in the process of Refresh your browser posting data to AirAssure Web When troubleshooting a Depending on the operating Use Ethernet if you need to monitor multiple units device over internal USB system of your PC it may cannot connect multiple not be possible to connect units to my personal more than one unit toa computer single computer over USB Tomcat application does Java install requires Reboot comp