Home
SP-500-user manual(Ver.1)
Contents
1. Tip amp Handpiece Tip amp Handpiece can be sterilized in autoclave at 135 C for 3 minutes without protective package or for 30 minutes in a protective package Caution Tip amp handpiece can be damaged by dropping on the floor Do not clean the handpiece in an ultrasonic cleaner Do not use chemical solvents abrasive cleaner Tips must be sterilized in autoclave before using Tip Holder The torque wrench can be cleaned with a cloth soaked in any of the common non abrasive mild cleaning liquids containing alcohol Torque Wrench The torque wrench can be cleaned with a cloth soaked in the common non abrasive mild cleaning liquids containing alcohol Transformer The casing of transformer can be cleaned with a cloth soaked in any of the common non abrasive mild cleaning liquids containing alcohol External Casing The casing of the Turboson can be cleaned with a cloth soaked in any of the common non abrasive mild cleaning liquids containing alcohol 6 Instruction for Use When the connections are completed as indicated in Installation you may use the device adhering to the following instruction 1 Tip shall be sterilized before use or reuse after autoclave 2 Use to prevent infection or injured especially transporting the tips 3 Select the suitable tip for using Caution Hand and finger can be injured by the sharp edge of the tip 4 By pressing the POWER button on the unit it will
2. again completely 2 Tip is distorted Answer Replace with a new one 3 Tip is worn out Answer Replace with a new one The handpiece whistle 1 Tip is not properly fastened onto handpiece Answer Unscrew the tip and screw again completely Water is not coming out of the tip 1 Water control situated in front of the unit is closed Answer Turn the WATER knob located on the unit 2 The water supply hose is not connected or not completely connected to the coupling at the back of the unit Answer Connect the water supply hose securely 3 Tip has not been fastened correctly onto handpiece Answer Unscrew the tip and screw again completely 4 Tip is distorted Answer Replace with a new one 8 5 Tip is worn out Answer Replace with a new one 6 Handpiece is not connected to its cord Answer Connect the handpiece to the cord securely 7 Break in handpiece wire and current leakage through patient Answer Switch OFF the power switch and disconnect transformer from the AC power source wall outlet Contact the local dealer or our company for service 8 Breakage of handpiece Answer Replace with a new one Technical specifications Dimension 18 L x 14 5 W x 6 5 H cm Weight 607 g Water supply pressure 1 5 kg cm 1 5 bar 15 72 psi Frequency 28 k Hz 3 k Hz SWITCHING POWER ADAPTER Input AC 100 240V 50 60Hz 0 5A Output DC 24V 1A Weight 175g COMPLY SAFETY STANDARDS UL CUL PSE E
3. switch the unit ON OFF as needed 5 Select the level of power which is suitable to the desired treatment by turning the VOLUME knob located on the unit 6 Regulate the flow of water which is suitable to the desired treatment by turning the WATER knob located on the unit with the footswitch pressing Caution In case of using without water supply handpiece and tip will get overheated To preventing bio contamination Tips amp Handpiece must be sterilized in autoclave before using 7 The handpiece does not have heating effect A fine spray is sufficient to cool the point of friction between the tip and the tooth Troubleshooting Guide The unit does not switch ON when the POWER switch is pressed 1 The power cord is not completely connected to power source or wall electric socket or else to the socket at the rear of the main unit Answer Connect the power cord securely 2 Apparatus is supplied with a different voltage power to the one prescribed Answer Check the value of the applied voltage power supply corresponds to that indicated on the identification label on the bottom of the unit The unit is switch ON but does not work 1 The plug of footswitch cord is not connected or not completely connected to its socket at the rear of the unit Answer Connect the footswitch cord securely Insufficient power vibration 1 Tip is not properly fastened onto handpiece Answer Unscrew the tip and screw
4. 241618 JQE10968 GS EN60950 EMI CE FCC Conditions for use 10 C 40 C 30 75 relative humidity Storage and transport conditions 10 C 40 C 10 95 relative humidity 500 hpa 1060 hpa NMAONITEX NDUSTRIAL CO LTD 6F 70 GUANG FU ROAD SEC 1 SAN CHONG CITY TAIPEI TAIWAN TEL 886 2 29993869 FAX 886 2 29993647 E mail monitex monitex com tw http www monitex com tw Copyright 2010 MONITEX Industrial CO LTD All Rights Reserved 2010 12 Ver 1
5. MONITEX Turboson SP 500 Ultrasonic Scaler USER MANUAL Contents Preface de Gate orne S ee M ee aie a Or en lo re Art en cb AR E AR CR Le DS Cia ne ee nes Components and Accessories Installation eo eee es Cleaning amp Maintenance n Instruction for Use NT ae eS SOs Sb es Troubleshooting Guide Technical specifications OT OT ED TO ET CPE ERP Preface Thank you for purchasing Turboson piezomatic ultrasonic scaler Turboson is designed for high efficiency for perio debridement with its easy charging capability Turboson can remove the stubborn old calculus from coronal as well as the sulcular and pocket epithelium in root epithelial or debris scaling with maximum frequency Prior to using this equipment please read this USER MANUAL carefully and follow the maintenance to enable the function to be fully implemented Please store this USER MANUAL in a safe place after reading No further advice will be issued shall any of the further modifications be made in the future The Turboson is for dental use only Turboson may only be used by licenced doctor Dentist with proper awareness of the standard operating procedure in accordance to this manual Please Contact your local agent distributor for furthur assist of trouble shooting when problems occured Any attempt of repairing this product by unauthorized people is not allowed Disclaimer We shall not assume any responsibility if any malfunctio
6. n damage or accident of personal body injury is caused by the improper removal maintenance modification or repair conducted by unauthorized dealer We shall not assume any responsibility if any malfunction damage or accident of personal body injury is caused by the use of the product not provided by us We shall not assume any responsibility if any malfunction damage or accident of personal body injury is caused by the improper removal modification maintenance or repair using the parts not provided by us We shall not assume any responsibility if any malfunction damage or accident of personal body injury is caused by the improper operation incompliant with this manual We shall not assume any responsibility if any malfunction damage or accident of personal injure is caused by the surrounding environment incompliant with the installation environment and operating conditions specified on this manual We shall not assume any responsibility if any malfunction damage or accident of personal injure is caused by fire disaster earthquake flood struck by lightning or any unpredicable natural calamity Components and Accessories Control for power adjustment Control for water flow adjustment LEDs for operating indicator A O N On Off power switch Components and Accessories fr Handpiece 1 O9 oO p Torque Wrench Tip Holder Handpiece Cord Wate
7. r supply hose coupling Power adapter connection Footswitch cord connection T Footswitch Power Adapter Water Hose T Connecter Installation In order to install Turboson the following connection is needed An electrical supply of DC 24V Water supply with a source pressure between 0 1 0 2 Mpa Connect to water supply Connect the water hose to a cold water supply using T shape connector Connect the coupler of the hose to the socket at the back of the unit Important If the unit is connect to the public water directly please check the public pressure of water supply before installation The internal device of this unit might be damaged if the water pressure exceed the required specification of water supply pressure Please see page 9 Connect to foot switch Connect the round plug of the foot switch to the socket at the back of the unit Connection of the handpiece Screw the tip onto the handpiece Tighten the tip by using the torque wrench Connect the handpiece on its cord and place it on its holder Connection to the electricity supply WARRING Always use the correct voltage of power supply supply of DC 24V is available Cleaning amp Maintenance Handpiece Cord The cord can be cleaned with a cloth soaked in the common non abrasive mild cleaning liquids containing alcohol Dry the handpiece connector with compressed air blow before connecting the handpiece
Download Pdf Manuals
Related Search
Related Contents
manual de instruções da estação de retrabalho smd Mode d`emploi - Rice Lake Weighing Systems Palm Pilot 1553 User Manual, UM11009 Kompernass KH 2370 Operating Instructions Bedienungsanleitung Aritco 7000er Serie ISDN Basic Rate Interface Feature Fundamentals User Manual - Compare Cellular Dell SP2309W User's Manual Manuel d`utilisation DCO7 Copyright © All rights reserved.